The State reported that a licensee, KooneyThe State reported that a licensee, Kooney X-Ray Inc., shipped a radiography camera (Model #150; S/N 197) to QSA Global in Baton Rouge, LA, for replacement of the iridium-192 camera source. When the camera arrived and was inspected at QSA Global, low-level radioactive contamination was found on the camera. A wipe test (swipe area not specified) revealed a contamination level of 0.0059 microcuries. The acceptance limit for QSA Global is 0.0050 microcuries. QSA Global believes that the contamination is from depleted uranium shielding in the camera S-tube and not from the currently installed iridium-192 source. Kooney X-ray has requested that QSA Global return the camera. Kooney plans to have the camera S-tube repaired as necessary. The State reported that the contamination was confined to the camera and this event did not result in any personnel or uncontrolled contamination on any other items.</br>Texas Report Number 8278 any other items. Texas Report Number 8278  +
18:30:00, 23 November 2005  +
42,167  +
14:39:00, 23 November 2005  +
18:30:00, 23 November 2005  +
The State reported that a licensee, KooneyThe State reported that a licensee, Kooney X-Ray Inc., shipped a radiography camera (Model #150; S/N 197) to QSA Global in Baton Rouge, LA, for replacement of the iridium-192 camera source. When the camera arrived and was inspected at QSA Global, low-level radioactive contamination was found on the camera. A wipe test (swipe area not specified) revealed a contamination level of 0.0059 microcuries. The acceptance limit for QSA Global is 0.0050 microcuries. QSA Global believes that the contamination is from depleted uranium shielding in the camera S-tube and not from the currently installed iridium-192 source. Kooney X-ray has requested that QSA Global return the camera. Kooney plans to have the camera S-tube repaired as necessary. The State reported that the contamination was confined to the camera and this event did not result in any personnel or uncontrolled contamination on any other items.</br>Texas Report Number 8278 any other items. Texas Report Number 8278  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 23 November 2005  +
L01074  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
02:19:46, 2 March 2018  +
14:39:00, 23 November 2005  +
-0.16 d (-3.85 hours, -0.0229 weeks, -0.00527 months)  +
18:30:00, 23 November 2005  +
Agreement State Report - Radioactive Contamination on Radiography Camera  +