Browse wiki
Jump to navigation
Jump to search
The following report was received by the C … The following report was received by the California Radiation Control Program via email: </br>On April 15, 2025, the radiation safety officer at Cedars Sinai Medical Center contacted Los Angeles County Public Health Radiation Management (part of the California Radiation Control Program), first by telephone at 0955 PDT and then by email at 1055 PDT, to report a medical event that occurred at Cedars Sinai Medical Center. During the administration of Y-90 TheraSpheres to a patient, the Y-90 TheraSpheres became obstructed within the intravenous tubing such that the authorized user was unable to administer the full dosage. The activity in the syringe was approximately 35 mCi, but only 12.14 mCi was delivered to the target organ. This was 35 percent of the prescribed activity. Additionally, the prescribed dose to the liver (the target organ) was 78.8 gray (7,880 rads). The liver received a dose of 27.6 gray (2,760 rads), a difference of 51.2 gray (5,120 rads).</br>A site visit by the principal radiation protection specialist, Los Angeles County Public Health Radiation Management is scheduled for April 17, 2025." </br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.</br>California 5010 Number: 041525he patient.
California 5010 Number: 041525 +
07:00:00, 14 April 2025 +
57,668 +
13:57:00, 16 April 2025 +
07:00:00, 14 April 2025 +
The following report was received by the C … The following report was received by the California Radiation Control Program via email: </br>On April 15, 2025, the radiation safety officer at Cedars Sinai Medical Center contacted Los Angeles County Public Health Radiation Management (part of the California Radiation Control Program), first by telephone at 0955 PDT and then by email at 1055 PDT, to report a medical event that occurred at Cedars Sinai Medical Center. During the administration of Y-90 TheraSpheres to a patient, the Y-90 TheraSpheres became obstructed within the intravenous tubing such that the authorized user was unable to administer the full dosage. The activity in the syringe was approximately 35 mCi, but only 12.14 mCi was delivered to the target organ. This was 35 percent of the prescribed activity. Additionally, the prescribed dose to the liver (the target organ) was 78.8 gray (7,880 rads). The liver received a dose of 27.6 gray (2,760 rads), a difference of 51.2 gray (5,120 rads).</br>A site visit by the principal radiation protection specialist, Los Angeles County Public Health Radiation Management is scheduled for April 17, 2025." </br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.</br>California 5010 Number: 041525he patient.
California 5010 Number: 041525 +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 16 April 2025 +
0404-19 +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
11:20:05, 24 April 2025 +
13:57:00, 16 April 2025 +
2.29 d (54.95 hours, 0.327 weeks, 0.0753 months) +
07:00:00, 14 April 2025 +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.