Browse wiki

Jump to navigation Jump to search
The following is a summary of information The following is a summary of information that was received via email:</br>At 1141 EST, on 12/19/20, an authorized user (AU) for the licensee determined that one of their portable nuclear gauges was damaged. The AU felt the index rod loosen and then break off when he went to retrieve the gauge from its transport case. The gauge had been used the day before at a jobsite and did not experience any accidents. The gauge had not been dropped and was in the possession of the AU at all times during this event. The source remained in the shielded position within the gauge throughout this event. </br>The AU contacted the radiation safety officer (RSO) and the gauge was transported to an authorized storage location. The RSO took measurements of the outer transport box surface (1 mR/hr) and the gauge surface (5 mR/hr). The gauge was placed into an authorized storage shed and readings at the surface of the shed were non-distinguishable from background. </br>The gauge was a Troxler 3440 (s/n: 16938) with an 8 mCi Cs-137 source and a 40 mCi Am-241/Be source. </br>NC event number: NC200023-241/Be source. NC event number: NC200023  +
16:41:00, 19 December 2020  +
55,042  +
16:27:00, 21 December 2020  +
16:41:00, 19 December 2020  +
The following is a summary of information The following is a summary of information that was received via email:</br>At 1141 EST, on 12/19/20, an authorized user (AU) for the licensee determined that one of their portable nuclear gauges was damaged. The AU felt the index rod loosen and then break off when he went to retrieve the gauge from its transport case. The gauge had been used the day before at a jobsite and did not experience any accidents. The gauge had not been dropped and was in the possession of the AU at all times during this event. The source remained in the shielded position within the gauge throughout this event. </br>The AU contacted the radiation safety officer (RSO) and the gauge was transported to an authorized storage location. The RSO took measurements of the outer transport box surface (1 mR/hr) and the gauge surface (5 mR/hr). The gauge was placed into an authorized storage shed and readings at the surface of the shed were non-distinguishable from background. </br>The gauge was a Troxler 3440 (s/n: 16938) with an 8 mCi Cs-137 source and a 40 mCi Am-241/Be source. </br>NC event number: NC200023-241/Be source. NC event number: NC200023  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 21 December 2020  +
060-0353-4  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
16:05:37, 15 January 2021  +
16:27:00, 21 December 2020  +
1.99 d (47.77 hours, 0.284 weeks, 0.0654 months)  +
16:41:00, 19 December 2020  +
Damaged Nuclear Gauge  +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.