Browse wiki

Jump to navigation Jump to search
The following was received via e-mail: OnThe following was received via e-mail: </br>On March 14, 2019, the Texas Department of Health Services was notified by the licensee's radiation safety officer (RSO) that a therapy event had occurred. A patient was scheduled for three fractions using a high dose rate after loader containing a 6.8 curie iridium-192 source. Multiple catheters were to be used. On March 14, 2019, prior to the third fraction, the licensee determined that one of the catheters was shorter than the length required and the patient received 350 centigrays (50 percent) of the prescribed dose to the thigh resulting in the target tissue receiving 50 percent of the prescribed dose (350 centigrays). The third fraction was not administered. The RSO stated they intended to create a new treatment plan to correct the error and insure the intended area receives the correct dose. Additional information will be provided as it is received in accordance with SA-300.</br>Texas Incident Number: 9665</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
05:00:00, 7 March 2019  +
53,932  +
16:05:00, 14 March 2019  +
05:00:00, 7 March 2019  +
The following was received via e-mail: OnThe following was received via e-mail: </br>On March 14, 2019, the Texas Department of Health Services was notified by the licensee's radiation safety officer (RSO) that a therapy event had occurred. A patient was scheduled for three fractions using a high dose rate after loader containing a 6.8 curie iridium-192 source. Multiple catheters were to be used. On March 14, 2019, prior to the third fraction, the licensee determined that one of the catheters was shorter than the length required and the patient received 350 centigrays (50 percent) of the prescribed dose to the thigh resulting in the target tissue receiving 50 percent of the prescribed dose (350 centigrays). The third fraction was not administered. The RSO stated they intended to create a new treatment plan to correct the error and insure the intended area receives the correct dose. Additional information will be provided as it is received in accordance with SA-300.</br>Texas Incident Number: 9665</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 14 March 2019  +
L05108  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
11:30:10, 25 March 2019  +
16:05:00, 14 March 2019  +
7.503 d (180.08 hours, 1.072 weeks, 0.247 months)  +
05:00:00, 7 March 2019  +
Agreement State Report - Misadministration of Dose to Patient  +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.