The following is excerpted from an email:
… The following is excerpted from an email:</br>On January 20, 2017 the licensee informed the Department (Pennsylvania Department of Environmental Protection Bureau of Radiation Protection) of an under dose incident involving yttrium-90 (Y-90) TheraSpheres. It is reportable as per 10 CFR 35.3045(a)(1)(i).</br>On January 19, 2017 a patient underwent a Y-90 TheraSphere treatment. The staff reported the procedure went as planned; however, upon surveying the waste and performing a dose calculation it was found that the patient received approximately 29% of the prescribed dose. The patient was notified of this under dose on January 20, 2017. The physician utilized contrast medium to view the flow prior to the procedure and no issues were seen. No increased resistance was noted and the physician was able to flush the line post administration. The manufacturer was notified and will be conducting a joint investigation with the hospital staff. No harm is expected to the patient.</br>A reactive inspection is planned by the Department. More information will be provided upon receipt.</br>PA Event Report ID No: PA 170001</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient. +
05:00:00, 19 January 2017 +
52,504 +
08:53:00, 23 January 2017 +
05:00:00, 19 January 2017 +
The following is excerpted from an email:
… The following is excerpted from an email:</br>On January 20, 2017 the licensee informed the Department (Pennsylvania Department of Environmental Protection Bureau of Radiation Protection) of an under dose incident involving yttrium-90 (Y-90) TheraSpheres. It is reportable as per 10 CFR 35.3045(a)(1)(i).</br>On January 19, 2017 a patient underwent a Y-90 TheraSphere treatment. The staff reported the procedure went as planned; however, upon surveying the waste and performing a dose calculation it was found that the patient received approximately 29% of the prescribed dose. The patient was notified of this under dose on January 20, 2017. The physician utilized contrast medium to view the flow prior to the procedure and no issues were seen. No increased resistance was noted and the physician was able to flush the line post administration. The manufacturer was notified and will be conducting a joint investigation with the hospital staff. No harm is expected to the patient.</br>A reactive inspection is planned by the Department. More information will be provided upon receipt.</br>PA Event Report ID No: PA 170001</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient. +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
Yttrium-90 + and Y-90 +
00:00:00, 23 January 2017 +
PA-0131 +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
22:50:48, 24 November 2018 +
08:53:00, 23 January 2017 +
4.162 d (99.88 hours, 0.595 weeks, 0.137 months) +
05:00:00, 19 January 2017 +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.