The following report was received from theThe following report was received from the New York State Department of Health, Bureau of Environmental Radiation Protection via facsimile:</br>Radiation safety staff from a hospital called on 5/24/2016 to report a medical event that meets the reportability criterion on the borderline. 79.5 percent of intended dose is believed to have been delivered to the left lobe of the liver of a patient receiving Y90 treatment. The patient had received treatment to the right lobe a month ago. The left lobe was smaller with small vessels. The waste measured after the procedure showed that only about 79 .5 percent had been delivered. The licensee usually is able to administer 95 percent of intended dose in this procedure. We were informed that since the amount retained in the device is about the same, smaller doses tend to show a greater percent in retained in the device. They will send a report within 7 days.</br>NYS NRC Event Report ID No.: NYDOH-16-02</br>NY State Incident No.: 1146</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
04:00:00, 24 May 2016  +
51,970  +
15:17:00, 1 June 2016  +
04:00:00, 24 May 2016  +
The following report was received from theThe following report was received from the New York State Department of Health, Bureau of Environmental Radiation Protection via facsimile:</br>Radiation safety staff from a hospital called on 5/24/2016 to report a medical event that meets the reportability criterion on the borderline. 79.5 percent of intended dose is believed to have been delivered to the left lobe of the liver of a patient receiving Y90 treatment. The patient had received treatment to the right lobe a month ago. The left lobe was smaller with small vessels. The waste measured after the procedure showed that only about 79 .5 percent had been delivered. The licensee usually is able to administer 95 percent of intended dose in this procedure. We were informed that since the amount retained in the device is about the same, smaller doses tend to show a greater percent in retained in the device. They will send a report within 7 days.</br>NYS NRC Event Report ID No.: NYDOH-16-02</br>NY State Incident No.: 1146</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 1 June 2016  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
01:44:57, 2 March 2018  +
15:17:00, 1 June 2016  +
8.47 d (203.28 hours, 1.21 weeks, 0.278 months)  +
04:00:00, 24 May 2016  +
Agreement State Report - Medical Underdose to the Left Lobe of Liver  +