Browse wiki
Jump to navigation
Jump to search
The following was received from the State … The following was received from the State of Pennsylvania via email:</br>Notifications: The (Pennsylvania Bureau of Radiation Protection) Department's Central Office was informed of this event on March 19, 2014. This event is reportable within 24-hours per 10 CFR 30.50(b)(2).</br>Event Description: During a routine monthly inspection of the gauge, (the licensee) discovered that the collimator block had broken off of the shutter assembly. The collimator block was found lying in the quarter inch steel box that the gauge is housed in to protect it from the operating conditions on the battery. The cause of this event is still being investigated. Radiation surveys performed indicated no employees were exposed as a result of this event.</br>Manufacturer: Thermo Measure Tech</br>Model: 5204</br>Serial Number: B392</br>Isotope: Cs-137</br>Activity: 4 Ci </br>CAUSE OF THE EVENT: Unknown at this time.</br>ACTIONS: The broken shutter mechanism was removed and replaced by a licensed service provider. The shutter mechanism was tested and confirmed as operating properly. The (State of PA) Department plans a reactive inspection.</br>Media attention: None at this time.</br>PA Event Report ID No: PA140009this time.
PA Event Report ID No: PA140009 +
04:00:00, 13 February 2014 +
49,936 +
12:35:00, 20 March 2014 +
04:00:00, 13 February 2014 +
The following was received from the State … The following was received from the State of Pennsylvania via email:</br>Notifications: The (Pennsylvania Bureau of Radiation Protection) Department's Central Office was informed of this event on March 19, 2014. This event is reportable within 24-hours per 10 CFR 30.50(b)(2).</br>Event Description: During a routine monthly inspection of the gauge, (the licensee) discovered that the collimator block had broken off of the shutter assembly. The collimator block was found lying in the quarter inch steel box that the gauge is housed in to protect it from the operating conditions on the battery. The cause of this event is still being investigated. Radiation surveys performed indicated no employees were exposed as a result of this event.</br>Manufacturer: Thermo Measure Tech</br>Model: 5204</br>Serial Number: B392</br>Isotope: Cs-137</br>Activity: 4 Ci </br>CAUSE OF THE EVENT: Unknown at this time.</br>ACTIONS: The broken shutter mechanism was removed and replaced by a licensed service provider. The shutter mechanism was tested and confirmed as operating properly. The (State of PA) Department plans a reactive inspection.</br>Media attention: None at this time.</br>PA Event Report ID No: PA140009this time.
PA Event Report ID No: PA140009 +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 20 March 2014 +
PA-1280 +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
23:09:33, 24 November 2018 +
12:35:00, 20 March 2014 +
35.399 d (849.58 hours, 5.057 weeks, 1.164 months) +
04:00:00, 13 February 2014 +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.