Browse wiki

Jump to navigation Jump to search
During a liver radio embolization procedurDuring a liver radio embolization procedure with Y-90 SIRTEX Sir-Spheres on January 24, 2014 an unanticipated shunting occurred and was identified during the procedure. The procedure for the right liver lobe was stopped when the shunting was identified. The licensee runs contrast between vial doses to verify proper delivery of the microspheres. After imaging of the patient, the licensee determined on January 27, 2014 that the unintended dose to the gastric fundus was 11 Gray. The licensee reported the medical event to ODH BRP (Ohio Department of Health - Bureau of Radiation Protection) by telephone on February 11, 2014 at 3:30pm (EST).</br>The referring physician and patient were notified on January 24, 2014. The patient is asymptomatic (no adverse effects) and is currently receiving medical monitoring.</br>Ohio Report #: OH140002</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
05:00:00, 24 January 2014  +
49,815  +
17:37:00, 11 February 2014  +
05:00:00, 24 January 2014  +
During a liver radio embolization procedurDuring a liver radio embolization procedure with Y-90 SIRTEX Sir-Spheres on January 24, 2014 an unanticipated shunting occurred and was identified during the procedure. The procedure for the right liver lobe was stopped when the shunting was identified. The licensee runs contrast between vial doses to verify proper delivery of the microspheres. After imaging of the patient, the licensee determined on January 27, 2014 that the unintended dose to the gastric fundus was 11 Gray. The licensee reported the medical event to ODH BRP (Ohio Department of Health - Bureau of Radiation Protection) by telephone on February 11, 2014 at 3:30pm (EST).</br>The referring physician and patient were notified on January 24, 2014. The patient is asymptomatic (no adverse effects) and is currently receiving medical monitoring.</br>Ohio Report #: OH140002</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 11 February 2014  +
02110180013  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
22:20:44, 24 September 2017  +
17:37:00, 11 February 2014  +
18.526 d (444.62 hours, 2.647 weeks, 0.609 months)  +
05:00:00, 24 January 2014  +
Agreement State Report - Medical Event Involving Unintended Dose  +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.