Browse wiki

Jump to navigation Jump to search
Event type: A medical event (ME) involvingEvent type: A medical event (ME) involving the administration of Yittrium-90 SIR-Spheres which is reportable under 10 CFR 35.3045(a)(1)(i).</br>Notifications: On March 21,2012, the Department's (Pennsylvania Department of Environmental Protection, Bureau of Radiation Protection) Southeast Regional Office received notification via a phone call about this ME.</br>Event Description: The patient was being treated for disease of the liver and received 77% of the intended dose.</br>CAUSE OF THE EVENT: Currently under investigation and unknown at this time.</br>ACTIONS: No harm to the patient is expected. The treating physician notified the patient. </br>The Department (Pennsylvania Department of Environmental Protection, Bureau of Radiation Protection) plans to do a reactive inspection.</br>PA Event Report ID: PA120009.</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.</br>* * * RETRACTION FROM JOSEPH MELNIC TO VINCE KLCO ON 3/26/12 AT 0937 EDT VIA FAX * * * </br>Notifications: On March 21, 2012, the Department's Southeast Regional Office received notification via a phone call about this ME. After the HOO notification was made the Department was informed that the measurement for actual dose delivered was repeated two times, both indicating the administration was within acceptable limits.</br>Event Description: The patient was being treated for disease of the liver and was initially thought to have received 77% of the intended dose, however after additional measurements were made the dosage delivered was determined to be 81.2% and 82.2%.</br>ACTIONS: Medical Event is being retracted.</br>Notified the R1DO (Newport) and FSME EO (McIntosh).the R1DO (Newport) and FSME EO (McIntosh).  +
04:00:00, 20 March 2012  +
47,763  +
20:26:00, 22 March 2012  +
04:00:00, 20 March 2012  +
Event type: A medical event (ME) involvingEvent type: A medical event (ME) involving the administration of Yittrium-90 SIR-Spheres which is reportable under 10 CFR 35.3045(a)(1)(i).</br>Notifications: On March 21,2012, the Department's (Pennsylvania Department of Environmental Protection, Bureau of Radiation Protection) Southeast Regional Office received notification via a phone call about this ME.</br>Event Description: The patient was being treated for disease of the liver and received 77% of the intended dose.</br>CAUSE OF THE EVENT: Currently under investigation and unknown at this time.</br>ACTIONS: No harm to the patient is expected. The treating physician notified the patient. </br>The Department (Pennsylvania Department of Environmental Protection, Bureau of Radiation Protection) plans to do a reactive inspection.</br>PA Event Report ID: PA120009.</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.</br>* * * RETRACTION FROM JOSEPH MELNIC TO VINCE KLCO ON 3/26/12 AT 0937 EDT VIA FAX * * * </br>Notifications: On March 21, 2012, the Department's Southeast Regional Office received notification via a phone call about this ME. After the HOO notification was made the Department was informed that the measurement for actual dose delivered was repeated two times, both indicating the administration was within acceptable limits.</br>Event Description: The patient was being treated for disease of the liver and was initially thought to have received 77% of the intended dose, however after additional measurements were made the dosage delivered was determined to be 81.2% and 82.2%.</br>ACTIONS: Medical Event is being retracted.</br>Notified the R1DO (Newport) and FSME EO (McIntosh).the R1DO (Newport) and FSME EO (McIntosh).  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 26 March 2012  +
PA-0131  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
23:17:09, 24 November 2018  +
20:26:00, 22 March 2012  +
2.685 d (64.43 hours, 0.384 weeks, 0.0883 months)  +
true  +
04:00:00, 20 March 2012  +
Agreement State Report - Underdose During Liver Disease Treatment  +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.