The following information was obtained froThe following information was obtained from the Commonwealth of Pennsylvania via facsimile:</br>EVENT DESCRIPTION: Failure of the shutter mechanism on one of its generally licensed thickness gauges was identified immediately to operators through a computer based warning message and visually by a red status light adjacent to the gauge location. Adjustment of the air supply allowed the shutter to close but still not operate properly. Routine maintenance activities were not successful in allowing the shutter to return to operational status and Applied Health Physics were notified to provide radiological support. No radiation exposure to personnel ensued during this event. The device is identified as: Manufacturer (IRMS), Model (TG-2), Serial# (00M0397-I5), Isotope (Americium-241), Activity (3 Ci), Location (Cold Reduction Mill).</br>CAUSE OF THE EVENT: A faulty actuator cylinder.</br>ACTIONS: The shutter mechanism was replaced, tested, and confirmed as operating properly. The actuator cylinder will be placed on a preventive maintenance schedule following this event.</br>PA Report No.: PA110022lowing this event. PA Report No.: PA110022  +
04:00:00, 2 August 2011  +
47,226  +
14:41:00, 31 August 2011  +
04:00:00, 2 August 2011  +
The following information was obtained froThe following information was obtained from the Commonwealth of Pennsylvania via facsimile:</br>EVENT DESCRIPTION: Failure of the shutter mechanism on one of its generally licensed thickness gauges was identified immediately to operators through a computer based warning message and visually by a red status light adjacent to the gauge location. Adjustment of the air supply allowed the shutter to close but still not operate properly. Routine maintenance activities were not successful in allowing the shutter to return to operational status and Applied Health Physics were notified to provide radiological support. No radiation exposure to personnel ensued during this event. The device is identified as: Manufacturer (IRMS), Model (TG-2), Serial# (00M0397-I5), Isotope (Americium-241), Activity (3 Ci), Location (Cold Reduction Mill).</br>CAUSE OF THE EVENT: A faulty actuator cylinder.</br>ACTIONS: The shutter mechanism was replaced, tested, and confirmed as operating properly. The actuator cylinder will be placed on a preventive maintenance schedule following this event.</br>PA Report No.: PA110022lowing this event. PA Report No.: PA110022  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 31 August 2011  +
PA-G0309  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
23:19:04, 24 November 2018  +
14:41:00, 31 August 2011  +
29.445 d (706.68 hours, 4.206 weeks, 0.968 months)  +
04:00:00, 2 August 2011  +
Agreement State Report - Shutter Mechanism Failure  +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.