The following was received via fax: The WiThe following was received via fax:</br>The Wisconsin Department of Health Services (DHS) received a phone call from the Radiation Safety Officer (RSO) on August 8, 2011 that a patient received a dose of 1.05 GBq of Y-90 Theraspheres on July 7, 2011 to the wrong side of the liver as documented on the written directive. The patient was scheduled for treatment of 1.04 GBq Y-90 Theraspheres for multinodular hepatocellular cancer (HCC) to the left lobe of the liver. The dosimetry for Y-90 Theraspheres was based on volume (mass) of the lobe bearing the tumors (the patient has tumors on both lobes, right and left). A treatment plan was created for the left lobe (using the volume of the left lobe) but during the procedure the right lobe was treated with the prescribed dose for the left lobe.</br>DHS will be conducting an investigation on August 12, 2011 and the licensee will be submitting a 15 day written report concerning the medical event.</br>WI Event Number: 110012</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
05:00:00, 7 August 2011  +
47,135  +
12:26:00, 9 August 2011  +
05:00:00, 7 August 2011  +
The following was received via fax: The WiThe following was received via fax:</br>The Wisconsin Department of Health Services (DHS) received a phone call from the Radiation Safety Officer (RSO) on August 8, 2011 that a patient received a dose of 1.05 GBq of Y-90 Theraspheres on July 7, 2011 to the wrong side of the liver as documented on the written directive. The patient was scheduled for treatment of 1.04 GBq Y-90 Theraspheres for multinodular hepatocellular cancer (HCC) to the left lobe of the liver. The dosimetry for Y-90 Theraspheres was based on volume (mass) of the lobe bearing the tumors (the patient has tumors on both lobes, right and left). A treatment plan was created for the left lobe (using the volume of the left lobe) but during the procedure the right lobe was treated with the prescribed dose for the left lobe.</br>DHS will be conducting an investigation on August 12, 2011 and the licensee will be submitting a 15 day written report concerning the medical event.</br>WI Event Number: 110012</br>A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient.  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 9 August 2011  +
025-1323-01  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
02:06:20, 2 March 2018  +
12:26:00, 9 August 2011  +
2.31 d (55.43 hours, 0.33 weeks, 0.0759 months)  +
05:00:00, 7 August 2011  +
Agreement State Report - Therasphere Treatment to Wrong Lobe of Liver  +