The following information was obtained froThe following information was obtained from the State of Pennsylvania via facsimile:</br>It was determined that the remote shutter actuator on a device is inoperable due to foreign material in the shutter mechanism. The device is identified as Texas Nuclear, Model 5197 (Serial No. 81637) with approximately 100 mCi of Cs-137.</br>The cause of the event was equipment failure.</br>The shutter is locked in the closed position and shall remain so until repairs are made. Licensee will notify the Department (Pennsylvania Dept. of Environmental Protection) when licensed repairs are made.</br>PA Event Report ID: PA110009</br>* * * UPDATE FROM JOE MELNIC TO DONALD NORWOOD AT 1431 EDT ON 5/9/2011 VIA FACSIMILE * * *</br>Upon further investigation it was determined that the remote actuator on the device is inoperable due to a crack in the mechanism that holds the actuator to the gauge. This prevents the shutter from operating as designed by the manufacturer. The device is identified as Texas Nuclear, Model 5197 (Serial No. B1637) with approximately 100 mCi of Cs-137.</br>Notified R1DO (DeFrancisco) and FSME (Diaz-Toro).d R1DO (DeFrancisco) and FSME (Diaz-Toro).  +
14:30:00, 3 May 2011  +
46,818  +
09:58:00, 4 May 2011  +
14:30:00, 3 May 2011  +
The following information was obtained froThe following information was obtained from the State of Pennsylvania via facsimile:</br>It was determined that the remote shutter actuator on a device is inoperable due to foreign material in the shutter mechanism. The device is identified as Texas Nuclear, Model 5197 (Serial No. 81637) with approximately 100 mCi of Cs-137.</br>The cause of the event was equipment failure.</br>The shutter is locked in the closed position and shall remain so until repairs are made. Licensee will notify the Department (Pennsylvania Dept. of Environmental Protection) when licensed repairs are made.</br>PA Event Report ID: PA110009</br>* * * UPDATE FROM JOE MELNIC TO DONALD NORWOOD AT 1431 EDT ON 5/9/2011 VIA FACSIMILE * * *</br>Upon further investigation it was determined that the remote actuator on the device is inoperable due to a crack in the mechanism that holds the actuator to the gauge. This prevents the shutter from operating as designed by the manufacturer. The device is identified as Texas Nuclear, Model 5197 (Serial No. B1637) with approximately 100 mCi of Cs-137.</br>Notified R1DO (DeFrancisco) and FSME (Diaz-Toro).d R1DO (DeFrancisco) and FSME (Diaz-Toro).  +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 9 May 2011  +
PA-G0283  +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
23:20:27, 24 November 2018  +
09:58:00, 4 May 2011  +
0.811 d (19.47 hours, 0.116 weeks, 0.0267 months)  +
14:30:00, 3 May 2011  +
Agreement State Report - Shutter Mechanism Failure  +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.