The following information was obtained fro … The following information was obtained from the Commonwealth of Pennsylvania via email:</br>Dose administered to patient was only 44% of the prescribed amount of Y90.</br>Cause: the treatment catheter became occluded and prevented further delivery of the Y-90 microspheres to the patient. It is not known what caused the occlusion. A later radiation survey of the catheter indicated a relative increased amount of radioactivity within the last 2cm from the tip. The catheter will be sent to the manufacturer for further analysis.</br>Effect to the Patient: None</br>Notifications : The referring physician was verbally notified of the medical event by the Authorized User. The referring physician verbally notified the patient.</br>Corrective Action: Equipment returned to manufacturer for repair or disposal.</br>Dose given was 5.3 mCi of Y-90. Dose intended was 12.1 mCi.</br>PA Reference Number: PA080033</br>A Medical Event may indicate potential problems in a medical facilities use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient. +
05:00:00, 4 December 2008 +
44,800 +
17:04:00, 22 January 2009 +
05:00:00, 4 December 2008 +
The following information was obtained fro … The following information was obtained from the Commonwealth of Pennsylvania via email:</br>Dose administered to patient was only 44% of the prescribed amount of Y90.</br>Cause: the treatment catheter became occluded and prevented further delivery of the Y-90 microspheres to the patient. It is not known what caused the occlusion. A later radiation survey of the catheter indicated a relative increased amount of radioactivity within the last 2cm from the tip. The catheter will be sent to the manufacturer for further analysis.</br>Effect to the Patient: None</br>Notifications : The referring physician was verbally notified of the medical event by the Authorized User. The referring physician verbally notified the patient.</br>Corrective Action: Equipment returned to manufacturer for repair or disposal.</br>Dose given was 5.3 mCi of Y-90. Dose intended was 12.1 mCi.</br>PA Reference Number: PA080033</br>A Medical Event may indicate potential problems in a medical facilities use of radioactive materials. It does not necessarily result in harm to the patient.necessarily result in harm to the patient. +
Has query"Has query" is a predefined property that represents meta information (in form of a <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Subobject">subobject</a>) about individual queries and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
00:00:00, 23 January 2009 +
PA-0190 +
Modification date"Modification date" is a predefined property that corresponds to the date of the last modification of a subject and is provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a>.
23:27:26, 24 November 2018 +
17:04:00, 22 January 2009 +
49.503 d (1,188.07 hours, 7.072 weeks, 1.627 months) +
05:00:00, 4 December 2008 +
URL"URL" is a <a href="/Special:Types/URL" title="Special:Types/URL">type</a> and predefined property provided by <a rel="nofollow" class="external text" href="https://www.semantic-mediawiki.org/wiki/Help:Special_properties">Semantic MediaWiki</a> to represent URI/URL values.