ENS 5303220 October 2017 17:00:00Agreement StateAgreement State Report - Gamma Knife MalfunctionOn 10/20/17, at 1000 PDT, a Gamma knife treatment was in process when the machine malfunctioned and recorded an error. The backup battery on the UPS (Uninterruptible Power Source) was low and resulted in the machine pausing and returning the source to the shielded position. The patient received approximately one third of the prescribed dose. The service provider was contacted and is scheduled to report to the hospital today (10/23/2017) to replace the backup battery. Elekta Leksell Gamma Knife Model Perfexion Serial number NM-001-NM201 A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.