ML23040A023
| ML23040A023 | |
| Person / Time | |
|---|---|
| Issue date: | 02/10/2023 |
| From: | Bethany Cecere NRC/NMSS/DMSST/ASPB |
| To: | Blackwell R State of LA, Dept of Environmental Quality |
| References | |
| Download: ML23040A023 (19) | |
Text
UNITED STATES NUCLEAR REGULATORY COMMISSION WASHINGTON, D.C. 20555-0001 Richard S. Blackwell Senior Environmental Scientist LA DEQ/OEC/ERSD/Radiation Section Post Office Box 4312 Baton Rouge, Louisiana 70821-4312
Dear Mr. Blackwell:
We have reviewed the final revisions to the Louisiana Department of Environmental Quality rules in the Louisiana Administrative Code, Title 33, Part XV, Radiation Protection, received by our office on January 20, 2023. These regulations were reviewed by comparison to the equivalent U.S. Nuclear Regulatory Commission (NRC) rules and Regulation Amendment Tracking System Identification Numbers (RATS IDs) 2018-1, 2018-3, 2019-1, 2019-2, 2020-1, 2020-3 and 10 CFR Part 34, as identified in the enclosed State Regulation Status (SRS) Data Sheet. We discussed our review of the regulations with you on February 3, 2023.
As a result of our review, we have no comments. Please note that we have limited our review to regulations required for compatibility and/or health and safety. We have determined that the regulations, as adopted, meet the compatibility and health and safety categories established in the Office of Nuclear Material Safety and Safeguards (NMSS) Procedure SA-200, Compatibility Categories and Health and Safety Identification for NRC Regulations and Other Program Elements.
The SRS Data Sheet summarizes our knowledge of the status of other Louisiana regulations, as indicated. Please let us know if you note any inaccuracies, or have any comments on the information contained in the SRS Data Sheet. This letter, including the SRS Data Sheet, is posted on the NMSS State Communication Portal: https://scp.nrc.gov/rulemaking.html.
February 10, 2023
R. Blackwell If you have any questions regarding the review results, the compatibility and health and safety categories or any of the NRC regulations used in the review, please contact Michelle Beardsley, State Regulation Review Coordinator, at (301) 415-0275 (Michelle.Beardsley@nrc.gov) or Joe OHara at (301) 415-6854 (Joe.OHara@nrc.gov).
Sincerely, Bethany K. Cecere, Acting Chief State Agreement and Liaison Programs Branch Division of Materials Safety, Security, State and Tribal Programs Office of Nuclear Material Safety and Safeguards
Enclosure:
Louisiana SRS Data Sheet Signed by Cecere, Bethany on 02/10/23
ML pkg ML23030B064 and letter ML23040A023 OFFICE NMSS/MSST NMSS/MSST OGC NMSS/MSST NAME JOHara MBeardsley N/A BCecere DATE 01/31/2023 02/09/2023 N/A 02/10/2023
STATE REGULATION STATUS State: Louisiana Tracking Ticket Number: 23-15 Date: February 10, 2023
[ # amendment(s) reviewed identified by a
- at the beginning of the equivalent NRC requirement.]
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1991-1 Safety Requirements for Radiographic Equipment Part 34 55 FR 843 (Superceded by 1997-5) 01/10/1994 Not Required Not Required Louisiana has adopted Final Regulations equivalent to RATS ID:
1997-5.
1991-2 ASNT Certification of Radiographers Part 34 56 FR 11504 (Superceded by 1997-5) none Not Required Not Required Louisiana has adopted Final Regulations equivalent to RATS ID:
1997-5.
1991-3 Standards for Protection Against Radiation Part 20 56 FR 23360; 56 FR 61352; 57 FR 38588; 57 FR 57877; 58 FR 67657; 59 FR 41641; 60 FR 20183; 01/01/1994 Final ML14266A408 No Comments 11/19/2014 ML14266A336 1991-4 Notification of Incidents Parts 20, 30, 31, 34, 39, 40 and 70 56 FR 64980; 10/15/1994 Final ML12361A157 No Comments 03/29/2013 ML13044A340 1992-1 Quality Management Program and Misadministrations Part 35 56 FR 34104 (Superceded by 2002-2) 01/27/1995 Not Required Not Required Louisiana has adopted Final Regulations equivalent to RATS ID:
2002-2.
1992-2 Eliminating the Recordkeeping Requirements for Departures from Manufacturer's Instructions Parts 30 and 35 57 FR 45566 none Not Required Not Required These regulations are not required to be adopted for purposes of Compatibility.
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1993-1 Decommissioning Recordkeeping and License Termination: Documentation Additions [Restricted areas and spill sites]
Parts 30 and 40 58 FR 39628 10/25/1996 Final ML12361A157 No Comments 03/29/2013 ML13044A340 1993-2 Licensing and Radiation Safety Requirements for Irradiators Part 36 58 FR 7715 07/01/1996 Not Applicablei Not Applicable Louisiana does not have any licensees subject to these regulations. (See SECY-95-112) 1993-3 Definition of Land Disposal and Waste Site QA Program Part 61 58 FR 33886 07/22/1996 Not Applicable1 Not Applicable Louisiana does not have any licensees subject to these regulations. (See SECY-95-112) 1994-1 Self-Guarantee as an Additional Financial Mechanism Parts 30, 40 and 70 58 FR 68726; 59 FR 1618 none Not Required Not Required These regulations are not required to be adopted for purposes of Compatibility.
1994-2 Uranium Mill Tailings Regulations: Conforming NRC Requirements to EPA Standards Part 40 59 FR 28220 07/01/1997 Not Applicable Not Applicable Louisiana does not have authority to regulate this material under its Agreement.
1994-3 Timeliness in Decommissioning Material Facilities Parts 30, 40 and 70 59 FR 36026 08/15/1997 Final No Comments 01/06/1999 1995-1 Preparation, Transfer for Commercial Distribution, and Use of Byproduct Material for Medical Use Parts 30, 32 and 35 59 FR 61767; 59 FR 65243; 60 FR 322 01/01/1998 Final No Comments 01/06/1999
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1995-2 Frequency of Medical Examinations for Use of Respiratory Protection Equipment Part 20 60 FR 7900 03/13/1998 Final No Comments 01/06/1999 1995-3 Low-Level Waste Shipment Manifest Information and Reporting Parts 20 and 61 60 FR 15649; 60 FR 25983 03/01/1998 Final No Comments 01/06/1999 1995-4 Performance Requirements for Radiography Equipment Part 34 60 FR 28323 (Superceded by 1997-5) 06/30/1998 Final No Comments 01/06/1999 Louisiana has adopted Final Regulations equivalent to RATS ID:
1997-5.
1995-5 Radiation Protection Requirements: Amended Definitions and Criteria Parts 19 and 20 60 FR 36038 08/14/1998 Final No Comments 01/06/1999 1995-6 Clarification of Decommissioning Funding Requirements Parts 30, 40 and 70 60 FR 38235 11/24/1998 Final No Comments 01/06/1999 1995-7 Medical Administration of Radiation and Radioactive Materials Parts 20 and 35 60 FR 48623 (Superceded by 2002-2 and 2005-2) 10/20/1998 Final No Comments 01/06/1999 Louisiana has not yet adopted Final Regulations equivalent to RATS ID:
2005-2.
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1996-1 Compatibility with the International Atomic Energy Agency Part 71 60 FR 50248; 61 FR 28724 (Superceded by 2004-1) 04/01/1999 Final ML020280299 No Comments 02/15/2002 ML020460125 Louisiana has not yet adopted Final Regulations equivalent to RATS ID:
2004-1.
1996-2 One Time Extension of Certain Byproduct, Source and Special Nuclear Materials Licenses Parts 30, 40 and 70 61 FR 1109 02/15/1999 Not Required Not Required These regulations are not required to be adopted for purposes of Compatibility.
1996-3 Termination or Transfer of Licensed Activities:
Record keeping Requirements Parts 20, 30, 40, 61 and 70 61 FR 24669 06/17/1999 Final ML12361A157 No Comments 03/29/2013 ML13044A340 1997-1 Resolution of Dual Regulation of Airborne Effluents of Radioactive Materials; Clean Air Act Part 20 61 FR 65120 01/9/2000 Final ML011510385 No Comments 11/15/2001 ML013100191 1997-2 Recognition of Agreement State Licenses in Areas Under Exclusive Federal Jurisdiction Within an Agreement State Part 150 62 FR 1662 02/27/2000 Final ML011510385 No Comments 11/15/2001 ML013100191 1997-3 Criteria for the Release of Individuals Administered Radioactive Material Parts 20 and 35 62 FR 4120 05/29/2000 Final No Comments 01/06/1999
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1997-4 Fissile Material Shipments and Exemptions Part 71 62 FR 5907 (Superceded by 2004-1) 02/10/2000 Not Required Not Required These regulations are not required to be adopted for purposes of Compatibility. (See STP-97-078) 1997-5 Licenses for Industrial Radiography and Radiation Safety Requirements for Industrial Radiography Operations Parts 30, 34, 71 and 150 62 FR 28947 06/27/2000 Final ML14266A408 Revised Final ML14364A230 Revised Final ML15212A809 Comments 11/19/2014 ML14266A336 No Comments 01/29/2015 ML14364A226 No Comments 08/31/2015 ML15212A806 1997-6 Radiological Criteria for License Termination Parts 20, 30, 40 and 70 62 FR 39057 08/20/2000 Final ML011510385 No Comments 11/15/2001 ML013100191 1997-7 Exempt Distribution of a Radioactive Drug Containing One Micro curie of Carbon-14 Urea Part 30 62 FR 63634 01/02/2001 Final No Comments 01/06/1999 1998-1 Deliberate Misconduct by Unlicensed Persons Parts 30, 40, 61, 70, 71 and 150 63 FR 1890; 63 FR 13773 02/12/2001 Final ML011510385 No Comments 11/15/2001 ML013100191
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1998-2 Self-Guarantee of Decommissioning Funding by Nonprofit and Non-Bond-Issuing Licensees Parts 30, 40 and 70 63 FR 29535 07/01/2001 Not Required Not Required These regulations are not required to be adopted for purposes of Compatibility.
1998-3 License Term for Medical Use Licenses Part 35 63 FR 31604 (Superceded by 2002-2) 07/10/2001 Not Required Not Required These regulations are not required to be adopted for purposes of Compatibility. (See STP-98-074)
Louisiana has adopted Final Regulations equivalent to RATS ID:
2002-2.
1998-4 Licenses for Industrial Radiography and Radiation Safety Requirements for Industrial Radiographic Operations Part 34 63 FR 37059 07/09/2001 Final ML062480284 No Comments 10/23/2006 ML062960278 1998-5 Minor Corrections, Clarifying Changes, and a Minor Policy Change Parts 20, 32, 35, 36 and 39 63 FR 39477; 63 FR 45393 10/26/2001 Final ML14266A408 No Comments 11/19/2014 ML14266A336 1998-6 Transfer for Disposal and Manifests: Minor Technical Conforming Amendment Part 20 63 FR 50127 11/20/2001 Final ML062480284 No Comments 10/23/2006 ML062960278 1999-1 Radiological Criteria for License Termination of Uranium Recovery Facilities Part 40 64 FR 17506 06/11/2002 Not Applicable Not Applicable Louisiana does not have the authority to regulate this material under its Agreement.
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1999-2 Requirements for Those Who Possess Certain Industrial Devices Containing Byproduct Material to Provide Requested Information Part 31 64 FR 42269 10/04/2002 Not Required Not Required These regulations are not required to be adopted for purposes of Compatibility.
1999-3 Respiratory Protection and Controls to Restrict Internal Exposure Part 20 64 FR 54543; 64 FR 55524 02/02/2003 Final ML061720127 No Comments 07/27/2006 ML062120471 2000-1 Energy Compensation Sources for Well Logging and Other Regulatory Clarifications Part 39 65 FR 20337 05/17/2003 Final ML080800500 No Comments 04/14/2008 ML081050467 2000-2 New Dosimetry Technology Parts 34, 36 and 39 65 FR 63750 01/08/2004 Final ML041880305 No Comments 08/26/2004 ML042390322 2001-1 Requirements for Certain Generally Licensed Industrial Devices Containing Byproduct Material Parts 30, 31 and 32 65 FR 79162 02/16/2004 Final ML14083A568 No Comments 06/12/2014 ML14083A554 2002-1 Revision of the Skin Dose Limit Part 20 67 FR 16298 04/05/2005 Final ML062480284 No Comments 11/27/2006 ML063180075 2002-2 Medical Use of Byproduct Material Parts 20, 32 and 35 67 FR 20249 10/24/2005 Final ML061720127 No Comments 07/27/2006 ML062120471
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2003-1 Financial Assurance for Materials Licensees Parts 30, 40 and 70 68 FR 57327 12/03/2006 Final ML062480284 No Comments 10/23/2006 ML062960278 2004-1 Compatibility With IAEA Transportation Safety Standards and Other Transportation Safety Amendments Part 71 69 FR 3697 10/01/2007 Final ML090480013 No Comments 04/01/2009 ML090710728 2005-1 Security Requirements for Portable Gauges Containing Byproduct Material Part 30 70 FR 2001 07/11/2008 Final ML081080146 No Comments 05/29/2008 ML081500119 2005-2 Medical Use of Byproduct Material - Recognition of Specialty Boards Part 35 70 FR 16336; 71 FR 1926 04/29/2008 Final ML090480013 No Comments 04/01/2009 ML090710728 2005-3 Increased Controls for Risk-Significant Radioactive Sources (NRC Order EA-05-090) 70 FR 72128 12/01/2005 License Condition ML052900105 No Comments 10/20/2005 ML052930327 2006-1 Minor Amendments Parts 20, 30, 32, 35, 40 and 70 71 FR 15005 03/27/2009 Final ML14266A408 No Comments 11/19/2014 ML14266A336 2006-2 National Source Tracking System - Serialization Requirements Part 32 with reference to Part 20 Appendix E 71 FR 65685 02/06/2007 Final ML080800500 No Comments 04/14/2008 ML081050467
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2006-3 National Source Tracking System Part 20 71 FR 65685, 72 FR 59162 01/31/2009 Final ML080800500 No Comments 04/14/2008 ML081050467 2007-1 Medical Use of Byproduct Material - Minor Corrections and Clarifications Parts 32 and 35 72 FR 45147, 54207 10/29/2010 Final ML14083A603 No Comments 07/29/2014 ML14083A595 2007-2 Exemptions From Licensing, General Licenses, and Distribution of Byproduct Material: Licensing and Reporting Requirements Parts 30, 31, 32 and 150 72 FR 58473 12/17/2010 Final ML14266A408 No Comments 11/19/2014 ML14266A336 2007-3 Requirements for Expanded Definition of Byproduct Material Parts 20, 30, 31, 32, 33, 35, 61 and 150 72 FR 55864 11/30/2010 Final ML14083A603 Proposed Revision to Final ML14364A230 Revised Final ML15212A809 Comments 07/29/2014 ML14083A595 Comments 01/29/2015 ML14364A226 No Comments 08/31/2015 ML15212A806 2007-4 Order Imposing Fingerprinting Requirements and Criminal History Records Check Requirements for Unescorted Access to Certain Radioactive Material NRC Order EA-07-305 72 FR 70901 06/05/2008 License Condition ML080800499 No Comments 03/27/2008 Ml080870350
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2008-1 Occupational Dose Records, Labeling Containers, and Total Effective Dose Equivalent Parts 19 and 20 72 FR 68043 02/15/2011 Final ML14266A408 No Comments 11/19/2014 ML14266A336 2009-1 Medical Use of Byproduct Material - Authorized User Clarification Part 35 74 FR 33901 09/28/2012 Final ML14266A408 No Comments 11/19/2014 ML14266A336 2011-1 Decommissioning Planning Parts 20, 30, 40 and 70 76 FR 35512 12/17/2015 Proposed ML15054A380 Final ML15300A512 No Comments 03/12/2015 ML15054A372 No Comments 11/18/2015 ML15300A511 2011-2 Licenses, Certifications, and Approvals for Materials Licensees Parts 30, 36, 39, 40, 70 and 150 76 FR 56951 11/14/2014 Final ML15021A524 No Comments 02/19/2015 ML15021A520 2012-1 Change of Compatibility Parts 31.5 and 31.6 (See RATS ID: 2001-1 for Rule text) 77 FR 3640 01/25/2015 Final ML14083A568 No Comments 06/12/2014 ML14083A554 2012-2 Advance Notification to Native American Tribes of Transportation of Certain Types of Nuclear Waste Part 71 77 FR 34194 08/10/2015 Final ML15021A524 No Comments 02/19/2015 ML15021A520
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2012-3 Technical Corrections Parts 30, 34, 40 and 71 77 FR 39899 08/06/2015 Final ML15021A524 No Comments 02/19/2015 ML15021A520 2012-4 Requirements for Distribution of Byproduct Material Parts 30, 31, 32, 40 and 70 77 FR 43666 10/23/2015 Proposed ML14364A230 Final ML15212A809 Comments 01/29/2015 ML14364A226 No Comments 08/31/2015 ML15212A806 2013-1 Physical Protection of Byproduct Material, Parts 20, 30, 32, 33, 34, 35, 36, 37, 39 and 71 78 FR 16922 03/19/2016 Proposed ML15126A361 Final ML16012A155 Revised Final ML18183A438 No Comments 06/11/2015 ML15124A889 Comment 02/05/2016 ML16012A148 No Comments 08/03/2018 ML18183A434 2013-2 Distribution of Source Material to Exempt Persons and to General Licensees and Revision of General License and Exemptions Parts 30, 40 and 70 78 FR 32310 08/27/2016 Proposed ML15126A361 Final ML16012A155 No Comments 06/11/2015 ML15124A889 No Comments 02/05/2016 ML16012A148
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2015-1 Domestic Licensing of Special Nuclear Material -
Written Reports and Clarifying Amendments Part 70 79 FR 57721, 80 FR 143 01/26/2018 Proposed ML15126A361 Final ML16012A155 No Comments 06/11/2015 ML15124A889 No Comments 02/05/16 ML16012A148 2015-2 Safeguards Information - Modified Handling Categorization, Change for Materials Facilities Parts 30, 37, 73 and 150 79 FR 58664, 80 FR 3865 01/28/2018 Proposed ML15126A361 Final ML16012A155 Revised Final ML18183A438 No Comments 06/11/2015 ML15124A889 Comment 02/05/16 ML16012A148 No Comments 08/03/2018 ML18183A434 2015-3 Revisions to Transportation Safety Requirements and Harmonization with International Atomic Energy Agency Transportation Requirements Part 71 80 FR 33987 07/13/2018
- extended to 08/15/2020 See STC 17-060 Proposed ML19038A132 Final ML19276E404 Revised Final ML21245A058 Comments 03/22/2019 ML19037A026 Comments 10/23/2019 ML19276E370 No Comments 10/15/2021 ML21245A054
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2015-4 Miscellaneous Corrections Parts 37 and 40 80 FR 45841 09/02/2018 Proposed ML18183A438 Final ML18360A588 No Comments 08/03/2018 ML18183A434 No Comments 01/30/2019 ML18355A459 2015-5 Miscellaneous Corrections Parts 19, 20, 30, 32, 37, 40, 61, 70, 71 and 150 80 FR 74974 12/31/2018 Proposed ML18183A438 Final ML18360A588 No Comments 08/03/2018 ML18183A434 No Comments 01/30/2019 ML18355A459
- 2018-1 Medical Use of Byproduct Material - Medical Event Definitions, Training and Experience, and Clarifying Amendments, Parts 30, 32 and 35 01/14/2022 Proposed ML21245A058 Final ML22004A159 Revised Final ML23030B826 Comments 10/15/2021 ML21245A054 Comments 01/25/2022 ML22004A156 No Comments 02/10/2023 ML23030B064
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2018-2 Miscellaneous Corrections -
Organizational Changes Parts 37, 40. 70 and 71 12/21/2021 Final ML18339A019 No Comments 01/03/2019 ML18339A018
- 2018-3 Miscelleaneous Corrections Parts 1, 2, 34, 37, 50, 71, 73, and 140 07/30/2022 Proposed ML22181A057 Final ML23030B826 No Comments 07/12/2022 ML22180A301 No Comments 02/10/2023 ML23030B064
- 2019-1 Miscellaneous Corrections Parts 2, 21, 37, 50, 52, 73, and 110 12/18/2022 Proposed ML22181A057 Final ML23030B826 No Comments 07/12/2022 ML22180A301 No Comments 02/10/2023 ML23030B064
- 2019-2 Organizational Changes and Conforming Amendments Parts 1, 2, 37, 40, 50, 51, 52, 55, 71, 72, 73, 74, 100, 140, and 150 12/30/2022 Proposed ML22181A057 Final ML23030B826 No Comments 07/12/2022 ML22180A301 No Comments 02/10/2023 ML23030B064
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes
- 2020-1 Individual Monitoring Devices 10 CFR Parts 34, 36, and 39 06/16/2023 Proposed ML22181A057 Final ML23030B826 No Comments 07/12/2022 ML22180A301 No Comments 02/10/2023 ML23030B064 2020-2 Social Security Number Fraud Prevention 10 CFR Parts 9 and 35 08/17/2023 Proposed ML21245A058 Final ML22004A159 No Comments 10/15/2021 ML21245A054 No Comments 01/25/2022 ML22004156
- 2020-3 Miscellaneous Corrections 10 CFR Parts 1, 2, 19, 20, 21, 30, 34, 35, 40, 50, 51, 52, 60, 61, 62, 63, 70, 71, 72, 73, 74, 75, 76, 110, and 140 11/16/2023 Proposed ML22181A057 Final ML23030B826 No Comments 07/12/2022 ML22180A301 No Comments 02/10/2023 ML23030B064 2021-1 Miscellaneous Corrections 10 CFR Parts 2, 11, 20, 25, 32, 35, 37, 50, 52,55, 70, 72, 73, 95, and 110 09/08/2024 2021-2 Miscellaneous Corrections 10 CFR Parts 9, 37, 40, 50, 51, 52, 55, 71, 73, and 110 12/30/2024
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2022-1 Miscellaneous Corrections 10 CFR 1, 2, 20, 30, 40, 50, 55, 70, 73, and 170 none Provisions are not required for compatibility.
2022-2 Miscellaneous Corrections 10 CFR Parts 20, 35, 50, 51, 52, 72, 73, 110, and 150 none Provisions are not required for compatibility.
NA Louisiana legally binding requirements,( i.e., license conditions to 10 CFR Parts 34 and 35)
NA License Conditions ML21026A258 No Comments 02/02/2021 ML21026A220 NA 10 CFR Parts 35.24(g), 35.404, 35.406, 35.410, 35.415, 35.2404, 35.2406, 35.2310, and 35.3045 NA Proposed ML21245A058 Final ML22004A159 No Comments 10/15/2021 ML21245A054 No Comments 01/25/2022 ML22004A156
- NA 10 CFR Part 34 NA Proposed ML18324A731 Final ML20017A198 Revised Final ML23030B826 Comments 01/18/2019 ML18324A728 Comments 03/02/2020 ML20017A189 No Comments 02/10/2023 ML23030B064 i IMPEP Team: verify that Louisiana does not have any licensees subject to these regulations during each review.