UNITED STATES NUCLEAR REGULATORY COMMISSION WASHINGTON, D.C. 20555-0001 Eva Nair, Program Manager Radiological Health Program Air and Radiation Administration Maryland Department of the Environment 1800 Washington Boulevard Baltimore, Maryland 21230
Dear Ms. Nair:
We have reviewed the proposed and revised final revisions to the Maryland regulations COMAR 26.12.01.01 Supplement 32, received by our office on March 29, 2022. These regulations were reviewed by comparison to the equivalent U.S. Nuclear Regulatory Commission (NRC) rules and Regulation Amendment Tracking System Identification Numbers (RATS IDs) 2018-1, 2018-3, 2019-1, 2020-1, 2020-2, and 2020-3 as identified in the enclosed State Regulation Status (SRS) Data Sheet. We discussed our review of the regulations with you on May 10, 2022.
As a result of our review, we have six comments that have been identified in the enclosure.
Please note that we have limited our review to regulations required for compatibility and/or health and safety. Under our current procedure, a finding that the Maryland regulations meet the compatibility and health and safety categories of the equivalent NRC regulation may only be made based on a review of the final Maryland regulations. However, we have determined that if your proposed regulations were adopted, incorporating our comments and without other significant change, they would meet the compatibility and health and safety categories established in the Office of Nuclear Material Safety and Safeguards (NMSS) Procedure SA-200, Compatibility Categories and Health and Safety Identification for NRC Regulations and Other Program Elements.
We request that when the proposed regulations are adopted and published as final regulations, a copy of the as published regulations be provided to us for review. As requested in NMSS Procedure SA-201, Review of State Regulatory Requirements, please highlight the final changes and provide a copy to Division of Material Safety, Security, State, Tribal Programs, NMSS.
The SRS Data Sheet summarizes our knowledge of the status of other Maryland regulations, as indicated. Please let us know if you note any inaccuracies, or have any comments on the information contained in the SRS Data Sheet. This letter, including the SRS Data Sheet, is posted on the NMSS State Communication Portal: https://scp.nrc.gov/rulemaking.html.
May 18, 2022
E. Nair If you have any questions regarding the comments, the compatibility and health and safety categories, or any of the NRC regulations used in the review, please contact Michelle Beardsley, State Regulation Review Coordinator, at 301-415-0275 (Michelle.Beardley@nrc.gov) or Farrah Gaskins at 610-337-5143 (Farrah.Gaskins@nrc.gov).
Sincerely, Brian C. Anderson, Chief State Agreement and Liaison Programs Branch Division of Materials Safety, Security, State and Tribal Programs Office of Nuclear Material Safety and Safeguards
Enclosures:
- 1. Compatibility Comments
ML22097A063 pkg and ML22136A308 letter OFFICE RI NMSS/MSST OGC NMSS/MSST NAME FGaskins MBeardsley N/A BAnderson DATE 05/18/22 05/18/22 05/18/22
COMPATIBILITY COMMENTS ON MARYLAND PROPOSED REGULATIONS STATE SECTION NRC SECTION RATS ID CATEGORY SUBJECT and COMMENTS 1
G.690(a)(1)
§ 35.690 2020-3 B
Training for use of remote afterloader units, teletherapy units, and gamma stereotactic radiosurgery units Maryland regulations state, Council on Postgraduate This should state, Council on Postdoctoral.
Maryland needs to make the change indicated above to meet the Compatibility Category B designation assigned to 10 CFR 35.690.
2 D.1201
(ii) 2020-3 C
Reports of theft or loss of licensed material Maryland omits an equivalent provision to § 20.2201(a)(2)(ii).
Maryland needs to include the equivalent of 10 CFR 20.2201(a)(2)
(ii) in order to meet the Compatibility Category C designation assigned to 10 CFR 20.2201(a)(2)(ii).
3 D.1202
§ 20.2202(d)(2) 2020-3 C
Notification of incidents Maryland omits an equivalent provision to § 20.2202(d)(2).
Maryland needs to include the equivalent of 10 CFR 20.2202(d)(2) in order to meet the Compatibility Category C designation assigned to 10 CFR 20.2202(d)(2).
4 D.1211
§ 30.50(c)(1 2020-3 C
Reporting Requirements Maryland omits an equivalent provision to § 30.50(c)(1).
Maryland needs to include the equivalent of 10 CFR 30.50(c)(1) in order to meet the Compatibility Category C designation assigned to 10 CFR 30.50(c)(1).
STATE SECTION NRC SECTION RATS ID CATEGORY SUBJECT and COMMENTS 5
§ 40.60(c)(1) 2020-3 C
Reporting requirements Maryland stated there is no equivalent COMAR to meet this regulation. This regulation addresses reporting requirements for source material.
Maryland needs to submit equivalent provisions to 10 CFR 40.60(c)(1) or confirm that reporting of these materials are included in D.1211, to meet the Compatibility Category C designation assigned to 10 CFR 40.60(c)(1) 6
§ 70.50(c)(1) 2020-3 C
General provisions and scope Maryland stated there is no applicable COMAR to meet this regulation. This regulation addresses reporting requirements for special nuclear material (which includes depleted uranium as shielding).
Maryland needs to submit equivalent provisions to 10 CFR 70.50(c)(1) or confirm that reporting of these materials are included in D.1211, to meet the Compatibility Category C designation assigned to 10 CFR 70.50(c)(1).
STATE REGULATION STATUS State: Maryland Tracking Ticket Number: 22-25 Date: May 18, 2022
[ # amendment(s) reviewed identified by a
- at the beginning of the equivalent NRC requirement.]
RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1991-1 Safety Requirements for Radiographic Equipment Part 34 55 FR 843 (Superceded by 1997-5) 01/10/1994 Final No Comments 09/09/1995 Maryland has adopted Final Regulations equivalent to RATS ID:
1997-5.
1991-2 ASNT Certification of Radiographers Part 34 56 FR 11504 (Superceded by 1997-5) none Final No Comments 09/09/1995 Maryland has adopted Final Regulations equivalent to RATS ID:
1997-5.
1991-3 Standards for Protection Against Radiation Part 20 56 FR 23360; 56 FR 61352; 57 FR 38588; 57 FR 57877; 58 FR 67657; 59 FR 41641; 60 FR 20183 01/01/1994 Final No Comments 02/28/1997 1991-4 Notification of Incidents Parts 20, 30, 31, 34, 39, 40, 70 56 FR 64980 10/15/1994 Final No Comments 04/15/1998 1992-1 Quality Management Program and Misadministrations Part 35 56 FR 34104 (Superceded by 2002-2) 01/27/1995 Final No Comments 09/09/1995 Maryland has not yet adopted Final Regulations equivalent to RATS ID:
2002-2.
1992-2 Eliminating the Recordkeeping Requirements for Departures from Manufacturer's Instructions Parts 30, 35 57 FR 45566 none Not Required Not Required These regulation changes are not required to be adopted for purposes of Compatibility.
2 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1993-1 Decommissioning Recordkeeping and License Termination: Documentation Additions [Restricted areas and spill sites]
Parts 30, 40 58 FR 39628 10/25/1996 Final No Comments 10/19/1995 1993-2 Licensing and Radiation Safety Requirements for Irradiators Part 36 58 FR 7715 07/01/1996 Final No Comments 03/20/1997 1993-3 Definition of Land Disposal and Waste Site QA Program Part 61 58 FR 33886 07/22/1996 Not Applicable1 Not Applicable Maryland does not have any licensees subject to these regulations. (See SECY-95-112) 1994-1 Self-Guarantee as an Additional Financial Mechanism Parts 30, 40, 70 58 FR 68726; 59 FR 1618 none Not Required Not Required These regulation changes are not required to be adopted for purposes of Compatibility.
1994-2 Uranium Mill Tailings Regulations: Conforming NRC Requirements to EPA Standards Part 40 59 FR 28220 07/01/1997 Not Applicable Not Applicable Maryland does not have authority to regulate this material under its Agreement.
1994-3 Timeliness in Decommissioning Material Facilities Parts 30, 40, 70 59 FR 36026 08/15/1997 Final No Comments 08/08/2000 1995-1 Preparation, Transfer for Commercial Distribution, and Use of Byproduct Material for Medical Use Parts 30, 32, 35 59 FR 61767; 59 FR 65243; 60 FR 322 01/01/1998 Final No Comments 08/08/2000
3 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1995-2 Frequency of Medical Examinations for Use of Respiratory Protection Equipment Part 20 60 FR 7900 03/13/1998 Final No Comments 11/03/1997 1995-3 Low-Level Waste Shipment Manifest Information and Reporting Parts 20, 61 60 FR 15649; 60 FR 25983 03/01/1998 Final No Comments 12/28/1997 1995-4 Performance Requirements for Radiography Equipment Part 34 60 FR 28323 (Superceded by 1997-5) 06/30/1998 Final No Comments 08/08/2000 Maryland has adopted Final Regulations equivalent to RATS ID:
1997-5.
1995-5 Radiation Protection Requirements: Amended Definitions and Criteria Parts 19, 20 60 FR 36038 08/14/1998 Final No Comments 08/08/2000 1995-6 Clarification of Decommissioning Funding Requirements Parts 30, 40, 70 60 FR 38235 11/24/1998 Final No Comments 08/08/2000 1995-7 Medical Administration of Radiation and Radioactive Materials Parts 20, 35 60 FR 48623 (Superceded by 2002-2 and 2005-2) 10/20/1998 Final No Comments 08/08/2000 Maryland has not yet adopted Final Regulations equivalent to RATS IDs: 2002-2 and 2005-2.
1996-1 Compatibility with the International Atomic Energy Agency Part 71 60 FR 50248; 61 FR 28724 (Superceded by 2004-1) 04/01/1999 Final No Comments 08/08/2000 Maryland has not yet adopted Final Regulations Equivalent to RATS ID:
2004-1.
4 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1996-2 One Time Extension of Certain Byproduct, Source and Special Nuclear Materials Licenses Parts 30, 40, 70 61 FR 1109 02/15/1999 Not Required Not Required These regulation changes are not required to be adopted for purposes of Compatibility.
1996-3 Termination or Transfer of Licensed Activities:
Record keeping Requirements Parts 20, 30, 40, 61, 70 61 FR 24669 06/17/1999 Final No Comments 08/08/2000 1997-1 Resolution of Dual Regulation of Airborne Effluents of Radioactive Materials; Clean Air Act Part 20 61 FR 65120 01/9/2000 Final ML051380408 No Comments 06/27/2005 ML051800055 1997-2 Recognition of Agreement State Licenses in Areas Under Exclusive Federal Jurisdiction Within an Agreement State Part 150 62 FR 1662 02/27/2000 Final ML051380408 No Comments 06/27/2005 ML051800055 1997-3 Criteria for the Release of Individuals Administered Radioactive Material Parts 20, 35 62 FR 4120 05/29/2000 Final ML021120426 No Comments 05/10/2002 ML021340513 1997-4 Fissile Material Shipments and Exemptions Part 71 62 FR 5907 (Superceded by 2004-1) 02/10/2000 Not Required Not Required These regulation changes are not required to be adopted for purposes of Compatibility. (See STP-97-078) 1997-5 Licenses for Industrial Radiography and Radiation Safety Requirements for Industrial Radiography Operations Parts 30, 34, 71, 150 62 FR 28947 06/27/2000 Final ML021120426 No Comments 05/10/2002 ML021340513
5 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1997-6 Radiological Criteria for License Termination Parts 20, 30, 40, 70 62 FR 39057 08/20/2000 Final ML021120426 No Comments 05/10/2002 ML021340513 1997-7 Exempt Distribution of a Radioactive Drug Containing One Micro curie of Carbon-14 Urea Part 30 62 FR 63634 01/02/2001 Final ML051380408 No Comments 06/27/2005 ML051800055 1998-1 Deliberate Misconduct by Unlicensed Persons Parts 30, 40, 61, 70, 71, 150 63 FR 1890; 63 FR 13773 02/12/2001 Final ML051380408 No Comments 06/27/2005 ML051800055 1998-2 Self-Guarantee of Decommissioning Funding by Nonprofit and Non-Bond-Issuing Licensees Parts 30, 40, 70 63 FR 29535 07/01/2001 Not Required Not Required These regulation changes are not required to be adopted for purposes of Compatibility.
1998-3 License Term for Medical Use Licenses Part 35 63 FR 31604 (Superceded by 2002-2) 07/10/2001 Not Required Not Required These regulation changes are not required to be adopted for purposes of Compatibility. (See STP-98-074) 1998-4 Licenses for Industrial Radiography and Radiation Safety Requirements for Industrial Radiographic Operations Part 34 63 FR 37059 07/09/2001 Final ML051100241 No Comments 05/13/2005 ML051360003 1998-5 Minor Corrections, Clarifying Changes, and a Minor Policy Change Parts 20, 32, 35, 36, 39 63 FR 39477; 63 FR 45393 10/26/2001 Final ML051100241 No Comments 05/13/2005 ML051360003
6 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 1998-6 Transfer for Disposal and Manifests: Minor Technical Conforming Amendment Part 20 63 FR 50127 11/20/2001 Final ML051100241 No Comments 05/13/2005 ML051360003 1999-1 Radiological Criteria for License Termination of Uranium Recovery Facilities Part 40 64 FR 17506 06/11/2002 Not Applicable Not applicable Maryland does not have authority to regulate this material under its Agreement.
1999-2 Requirements for Those Who Possess Certain Industrial Devices Containing Byproduct Material to Provide Requested Information Part 31 64 FR 42269 10/04/2002 Not Required Not Required These regulation changes are not required to be adopted for purposes of Compatibility.
1999-3 Respiratory Protection and Controls to Restrict Internal Exposure Part 20 64 FR 54543; 64 FR 55524 02/02/2003 Final ML051100241 No Comments 05/13/2005 ML051360003 2000-1 Energy Compensation Sources for Well Logging and Other Regulatory Clarifications Part 39 65 FR 20337 05/17/2003 Final ML051100241 No Comments 05/13/2005 ML051360003 2000-2 New Dosimetry Technology Parts 34, 36, 39 65 FR 63750 01/08/2004 Final ML071710580 No Comments 09/04/2007 ML072470495 2001-1 Requirements for Certain Generally Licensed Industrial Devices Containing Byproduct Material Parts 30, 31, 32 65 FR 79162 02/16/2004 Final ML071980288 No Comments 08/03/2007 ML072210120
7 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2002-1 Revision of the Skin Dose Limit Part 20 67 FR 16298 04/05/2005 Final ML071710580 No Comments 09/04/2007 ML072470495 2002-2 Medical Use of Byproduct Material Parts 20, 32, 35 67 FR 20249 10/24/2005 Final ML083290007 No Comments 12/23/2008 ML083460488 2003-1 Financial Assurance for Materials Licensees Parts 30, 40, 70 68 FR 57327 12/03/2006 Final ML071710580 No Comments 09/04/2007 ML072470495 2004-1 Compatibility With IAEA Transportation Safety Standards and Other Transportation Safety Amendments Part 71 69 FR 3697 10/01/2007 Final ML093500525 No Comments 01/27/2010 ML100140061 2005-1 Security Requirements for Portable Gauges Containing Byproduct Material Part 30 70 FR 2001 07/11/2008 Final ML083290007 No Comments 12/23/2008 ML083460488 2005-2 Medical Use of Byproduct Material - Recognition of Specialty Boards Part 35 70 FR 16336; 71 FR 1926 04/29/2008 Final ML083290007 No Comments 12/23/2008 ML083460488 2005-3 Increased Controls for Risk-Significant Radioactive Sources (NRC Order EA-05-090) 70 FR 72128 12/01/2005 License Condition ML053040172 No Comments 11/01/2005 ML053050373 2006-1 Minor Amendments Parts 20, 30, 32, 35, 40 and 70 71 FR 15005 03/27/2009 Final ML093500525 No Comments 01/27/2010 ML100140061
8 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2006-2 National Source Tracking System - Serialization Requirements Part 32 with reference to Part 20 Appendix E 71 FR 65685 02/06/2007 Final ML093500525 No Comments 01/27/2010 ML100140061 2006-3 National Source Tracking System Part 20 71 FR 65685, 72 FR 59162 01/31/2009 Final ML112260004 License Condition ML083390004 No Comment 09/22/2011 ML112411249 No Comment 12/11/2008 ML083390024 2007-1 Medical Use of Byproduct Material - Minor Corrections and Clarifications Parts 32 and 35 72 FR 45147, 54207 10/29/2010 Final ML093500525 No Comments 01/27/2010 ML100140061 2007-2 Exemptions From Licensing, General Licenses, and Distribution of Byproduct Material: Licensing and Reporting Requirements Parts 30, 31, 32, 150 72 FR 58473 12/17/2010 Final ML112260004 No Comment 09/22/2011 ML112411249 2007-3 Requirements for Expanded Definition of Byproduct Material Parts 20, 30, 31, 32, 33, 35, 61, 150 72 FR 55864 11/30/2010 Final ML113310005 No Comments 01/10/2012 ML113530474 2007-4 Order Imposing Fingerprinting Requirements and Criminal History Records Check Requirements for Unescorted Access to Certain Radioactive Material NRC Order EA-07-305 72 FR 70901 06/05/2008 License Condition ML081160371 No Comments 05/02/2008 ML081230186
9 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2008-1 Occupational Dose Records, Labeling Containers, and Total Effective Dose Equivalent Parts 19, 20 72 FR 68043 02/15/2011 Final ML113310005 No Comments 01/10/2012 ML113530474 2009-1 Medical Use of Byproduct Material - Authorized User Clarification Part 35 74 FR 33901 09/28/2012 Final ML13061A002 No Comments 03/28/2013 ML13070A113 2011-1 Decommissioning Planning Parts 20, 30, 40, 70 76 FR 35512 12/17/2015 Proposed ML12347A347 Final ML15077A029 Revised Final ML21209A039 No Comments 01/16/2013 ML13002A081 Comments 05/07/2015 ML15077A026 No Comments 08/08/2021 ML21209A028 2011-2 Licenses, Certifications, and Approvals for Materials Licensees Parts 30, 36, 39, 40, 70, and 150 76 FR 56951 11/14/2014 Proposed ML12347A347 Final ML15077A029 No Comments 01/16/2013 ML13002A081 No Comments 05/07/2015 ML15077A026 2012-1 Change of Compatibility of 10 CFR 31.5 and 31.6 (See RATS ID: 2001-1 for Rule text) 77 FR 3640 01/25/2015 R2 ML13347A977 Revised Final ML15077A977 Comments 01/17/2014 ML13347A961 No Comments 05/07/2015 ML15077A026
10 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2012-2 Advance Notification to Native American Tribes of Transportation of Certain Types of Nuclear Waste Part 71 77 FR 34194 08/10/2015 Proposed ML15119A462 Final ML15321A146 No Comments 05/28/2015 ML15119A456 No Comments 12/21/2015 ML15321A140 2012-3 Technical Corrections Part 30, 34, 40 and 71 77 FR 39899 08/06/2015 Proposed ML13347A977 Final ML15077A029 No Comments 01/17/2014 ML13347A961 No Comments 05/07/2015 ML15077A026 2012-4 Requirements for Distribution of Byproduct Material Parts 30, 31, 32, 40 and 70 77 FR 43666 10/23/2015 Proposed ML14176A138 Final ML15281A377 Comments 08/19/2014 ML14171A465 No Comments 11/25/2015 ML15281A372 2013-1 Physical Protection of Byproduct Material, 10 CFR Parts 20, 30, 32, 33, 34, 35, 36, 37, 39, and 71 78 FR 16922 03/19/2016 Proposed ML14356A521 Final ML15281A377 Comments 02/09/2015 ML14356A459 No Comments 11/25/2015 ML15281A372
11 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2013-2 Distribution of Source Material to Exempt Persons and to General Licensees and Revision of General License and Exemptions, 10 CFR Parts 30, 40, and 70 78 FR 32310 08/27/2016 Proposed ML15119A462 Final ML15321A146 No Comments 05/28/2015 ML15119A456 No Comments 12/21/2015 ML15321A140 2015-1 Domestic Licensing of Special Nuclear Material -
Written Reports and Clarifying Amendments Part 70 79 FR 57721, 80 FR 143 01/26/2018 Proposed ML16109A104 Final ML16334A253 Revised Final ML18142B495 No Comments 05/16/2016 ML16109A072 No Comments 01/24/2017 ML16334A248 No Comments 06/06/2018 ML18142B494 2015-2 Safeguards Information - Modified Handling Categorization, Change for Materials Facilities Parts 30, 37, 73, and 150 79 FR 58664, 80 FR 3865 01/28/2018 Proposed ML16109A104 Final ML16334A253 Revised Final ML18142B495 No Comments 05/16/2016 ML16109A072 No Comments 01/24/2017 ML16334A248 No Comments 06/06/2018 ML18142B494
12 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2015-3 Revisions to Transportation Safety Requirements and Harmonization with International Atomic Energy Agency Transportation Requirements Part 71 80 FR 33987 07/13/2018
- extended to 08/15/2020 See STC 17-060 Proposed ML16109A104 Final ML16334A253 Revised Final ML18142B495 No Comments 05/16/2016 ML16109A072 Comments 01/24/2017 ML16334A248 No Comments 06/06/2018 ML18142B494 2015-4 Miscellaneous Corrections, Parts 37 and 40 80 FR 45841 09/02/2018 Proposed ML16109A104 Final ML16334A253 Revised Final ML18142B495 No Comments 05/16/2016 ML16109A072 No Comments 01/24/2017 ML16334A248 No Comments 06/06/2018 ML18142B494 2015-5 Miscellaneous Corrections, 10 CFR Parts 19, 20, 30, 32, 37, 40, 61, 70, 71, and 150 80 FR 74974 12/31/2018 Proposed ML16109A104 Final ML16334A253 Revised Final ML18142B495 No Comments 05/16/2016 ML16109A072 No Comments 01/24/2017 ML16334A248 No Comments 06/06/2018 ML18142B494
13 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes
- 2018-1 Medical Use of Byproduct Material - Medical Event Definitions, Training and Experience, and Clarifying Amendments, 10 CFR Parts 30, 32 and 35 01/14/2022 Proposed ML21166A324 Final ML22010A074 Revised Final ML22060A173 Revised Final ML22097A073 Comments 07/27/2021 ML21166A314 Comment 02/17/2022 ML22010080 Comment 03/16/2022 ML22060A170 No Comments 05/18/2022 ML22097A063 2018-2 Miscellaneous Corrections -
Organizational Changes 10 CFR Parts 37, 40. 70 and 71 12/21/2021 Proposed ML21166A324 Final ML22010A074 Revised Final ML22060A173 Comments 07/27/2021 ML21166A314 Comment 02/17/2022 ML22010A080 No Comments 03/16/2022 ML22060A170 2018-3 Miscellaneous Corrections Parts 1, 2, 34, 37, 50, 71, 73, and 140 7/30/2022 2019-1 Miscellaneous Corrections Parts 2, 21, 37, 50, 52, 73, and 110 12/18/2022
14 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes 2019-2 Organizational Changes and Conforming Amendments Parts 1, 2, 37, 40, 50, 51, 52, 55, 71, 72, 73, 74, 100, 140, and 150 12/30/2022 2020-1 Individual Monitoring Devices 10 CFR Parts 34, 36, and 39 06/16/2023 2020-2 Social Security Number Fraud Prevention 10 CFR Parts 9 and 35 08/17/2023 2020-3 Miscellaneous Corrections 10 CFR Parts 1, 2, 19, 20, 21, 30, 34, 35, 40, 50, 51, 52, 60, 61, 62, 63, 70, 71, 72, 73, 74, 75, 76, 110, and 140 11/16/2023 2021-1 Miscellaneous Corrections 10 CFR Parts 2, 11, 20, 25, 32, 35, 37, 50, 52,55, 70, 72, 73, 95, and 110 09/08/2024 2021-2 Miscellaneous Corrections 10 CFR Parts 9, 37, 40, 50, 51, 52, 55, 71, 73, and 110 12/30/2024 N/A 10 CFR 20.1301, 32.72, 35.80 and 35.3045 N/A Final ML113310005 No Comments 01/11/2012 ML113530474 N/A 10 CFR 31.5, 35.60, 35.63 35.75, and 35.80 N/A Final ML13061A002 No Comments 03/28/2013 ML13070A113 N/A 10 CFR 35.61 N/A Proposed ML13347A977 No Comments 01/17/2014 ML13347A961
15 RATS ID NRC Chronology Identification Date Due for State Adoption Incoming Letter Outgoing Package Notes N/A Legally Binding Requirements, i.e., license conditions N/A License conditions ML21218A103 No Comments 09/14/2021 ML21218A094 1 IMPEP Team: verify that Maryland does not have any licensees subject to these regulations during each review.
2 R stands for revision to final rule