1 TABLE 4.21.A.2 RADIOACTIVE GASEOUS. EFFLUENT MONITORING INSTRUMENTATION SURVEILLANCE REQUIREMEhTS CHANNEL CHANNEL SOURCE CHANNEL FUNCTIONAL INSTRUMENT CHECK #

CHECK #

CALIBRATION #

TEST #

'l 1.

Steam Jet Air Ejector a.

Noble Gas Activity Monitor D

M R(3)

Q(2) R(1) b.

Effluent System Flow Rate Measuring Device D

NA R

Q 2.

Augmented Offgas Treatment System Explosive Gas Monitoring System a.

Hydrogen Monitor (2% Monitor)

D NA Q(4)

M 3.

Reactor Building Ventilation Monitoring System i

a.

Noble Gas Activity Monitor (KAMAN)

D M'

R(3)

Q(5)

Z

~.

b.

Iodine Sampler Cartridge W

NA NA NA c.

Particulate Sampler Filter W

NA NA NA d.

Effluent System Flow Rate Measuring Device D

NA R

Q e.

Sampler Flow Rate. Measuring Device D

NA R

Q f.

Isolation Monitor (GE)

D Q

R(3)

R(1) 4.

(****)

a.

Noble Gas Activity Monitor (KAMAN)

D M

R(3)

Q(5) b.

Iodine Sampler W

NA NA NA c.

Particulate Sampler W

NA NA NA d.

Effluent System Flow Rate Measuring Device D

NA R

Q e.

Sampler Flow Rate. Monitor D

NA R

Q 8701050559 861229 PDR ADOCK 05000298 P

PDR

~

l TABLE 4.21.A.1 RADIOACTIVE LIQUID EFFLUENT MONITORING INSTRUMENTATION SURVEILLANCE REQUIREMEh'IS CHANNEL CHANNEL SOURCE CHANNEL FUNCTIONAL INSTRUMENT CHECK #

CHECK #

CALIBRATION #

TEST #

l 1.

Gross Beta or Gamma Radioactivity Monitors Providing Alarm and Automatic Isolation a.

Liquid Radwaste Effluents Line D

P R(3)

Q(1) 2.

Gross Beta or Gamma Radioactivity Monitors Providing Alarm but not Providing Auto-matic Isolation a.

Service Water System Effluent Line D

M' R(3)

Q(2) h

~~

E i

4 3.

Flow Rate Measurement Devices 1

a.

Liquid Radwaste Effluent Line D(4)

NA R

SA t

3 s

w