Amended Registry of Radioactive Sealed Sources & Devices SE of Continuous Level Indicator for Model 13320. Certificate:NR-321-D-105-S
ML20198G047
Person / Time
Issue date: 08/06/1997
From: Steven Baggett, Michele Burgess
NRC
To:
Shared Package
ML20198G014 List:
References
SSD, NUDOCS 9708130269
Download: ML20198G047 (6)


Text

-

~

s 1

REGISTRY OP: RACIOACTIVE. SEALED SOURCP.S #?D DEVICES 2

SAFETY EVALUATION'OF DEVIT

1

-(AMENDED IN:ITS ENTIRETY)~

L N_(L_1, NR-3 21 -D-105 - S DATyg August;6, 1997 PAGE 1 OF-4

{

e

' DEVICE TYPE:

Continuous level Indicator i

MODEL: 13320 f

MANUFACTURERlDISTRIBUTOR:

-General-Nucleonics, Inc.

l 2807 Metropolitan Place Pomona, CA 91767 1

i SEALED SOURCE MODEL DESIGNATION:

General Nucleonics Model i

No.

13326 i

f i-ISOTOPE:

MAXIMUM ACTIVITY:

Krypton-85 10.8 curies (400 GBq)

(Total for the seven sources used in the device.

No source shall

}

exceed 2.5 curies (92.5 GBq))

?-

LEAK TEST-FREOUENCY:

Not Required PRINCIPAL USE: (E) Beta Gauge CUSTOM DEVICE:

X YES NO CUSTOM USER:

BHP Petroleum Americas Refinery, Inc.

(formerly Hawaiian Independent Refinery)91-325 Komohana Street Kapolei~, HI 96707 9708130269 970008 PDR RC SSD PDR

.j.,

REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF DEVICE (AMENDED IN ITS ENTIRETY).

NO.: NR-321-D-105-S DATE:

August 6, 1997 PAGE 2 OF 4 DEVICE TYPEr-Continuous level Indicator DESCRIPTIONt This system utilizes a continuous source holder to gauge-the height of material over a six foot (1.8 m) span in the BHP Petroleum Americas Refinery, Inc., T-175 vacuum tower. The Model 13320 source enclosure is a 7-1/2 foot (2.3 m) tall, 6 inch (152.4 mm) diameter stainless steel cylinder (with one side flat) which contains a 4 inch (101.6 mm) O.D.

inner cylinder welded to

~the same flattened side (see attached sketch).

The volume between these two cylinders is lead filled. A third cylinder fits inside the 4 inch 0.D.

tube and rotates on a center shaft.

This cylinder has a 1 inch _(25.4 mm) deep grove to hold the source tubes.

Rotating the innermost cylinder 180* carries the source tubes from "ON" (source shine -into-vessel) to " SAFE" position in which the-source is shielded in all directions.. The source can be locked in the " SAFE" position whenever. someone-must work inside the vessel or if the source ~ enclosure has to be moved or relocated.

The sources are 3/4 inch' (19.1 mm) _ diameter by 1 foot (304.8 mm) long tubes filled with gaseous Krypton-85 and sealeduwith-welds at both ends.

There are 7, tubes in all with different amounts of Kr-85.

The strongest-at the lowest'end of-the source enclosure, contains 2.5 times more Kr-85-than the top one for instance, so-the-response to_ changes-in height of reduced-crude is linear over the 6 foot range.'

-LABELING:

o The device is affixed with two labels that contain the information required by 10 CFR<20.1901 and 10 CFR'20.1904.

DIAGRAM:

-See attachments 1 and 2.

CONDITIONS OF NORMAL USE:

.The' device is used to monitor the level of vacuum tower number T-175.

The device is mounted to the tower at approximately 25

- feet ' (7. 6 m) above the ground and the only access is by ladder.

LThe_ device is subjected to environmental conditions intrinsic to Hawaii.

l a

-REGISTRY OF. RADIOACTIVE SEALED? SOURCES END DEVICES SAFETY EVALUATION OFcDEVICE (AMENDED:IN;ITS. ENTIRETY)

NO.: NR-321-D-105-S.

DATE:

August 6,,1997 PAGE 3 OF 4 7-j-DEVICE TYPE:

Continuous level: Indicator i-f'.

PROTOTYPE TESTING:

No testing was done'on this custom unit.

The device is very similar with the exception of activity and shielding dimensions to a gauge design-approved by the State _of California in 1977.

We have no reason to believe,that this custom unit would behave any differently _when subjected to the specified conditions of

[

use, than the currently approved unit.

i e

EXTERNAL RADIATION LEVELS:

)

The radiation dose rate near the source enclosure will be less than-. 2. mR/hr (20 psv/hr) on the surface-and less than 0.5 mR/hr (5 psv/hr) at 2 ft. (609.6 um), so personnel working-near the 4

system will not need to wear film badges or other monitoring devices.-

Only inside the vessel, with the source in the "ON" position,-will the radiation levels exceed the 2 mR/hr that would.

require wearing. film badges.

The_ manufacture also reported a dose : rate. of 5 mR/hr (50 psv/hr) at 30 ft. (9.1 m), in the beam path inside the vessel.

OUALITY ASSURANCE AND CONTROL:

This device is manufactured by General Nucleonics of California.

They have been in the gauging-business for several years.

The

= State of California has.found their Quality Program acceptable for licensing purposes.

LIMITATIONS AND/OR OTHER CONSIDERATIONS OF USE:

This' device shall only bte distributed to BHP Patroleum

-Americas Refinery Inc.

Handling, storage, use, transfer, and disposal: To be determined by the licensing authority.

The device shall be installed and initially tested for proper operation of the source exposure mechanism, safety

+.

warning components, labels, external radiation levels (source exposed,~ source shielded) and leak tested by General

-Nucleonics, Inc._ or other persons specifically licensed'by the NRC orfan Agreement State.

._.m

.____.___m._.

s f

REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF DEVICE (AMENDED.IN ITS' ENTIRETY)

HQ21-NR-321-D-105-S DATES-~ August 6, 1997 PAGE 4 OF 4-4 4

DEVICE TYEEl-Continuous level Indicator LIMITATIONS AND/OR OTHER CONSIDERATIONS OF USE: (Cont.)

This registration-sheet and the information contained within the references shall not be changed or transferred without the written consent of the NRC.

SAFETY ANALYSIS SUl@%RY:

Based on our review of the information cited below, we continue to conclude that the Model 13320 continuous level indicator design'is acceptable for custom-licensing purposes.

Furthermore, based on operational-history of a similar device registered by the State of California, we continue to conclude that this device would be expected to maintain its integrity'for normal conditions 1

Jof use which might occur during uses specified in this certificate.

f REFEREECES1 i

-The following supporting documents for the Model 13320.are hereby incorporated bycreference and are made a part of this registry document.

Hawaiian Independent Refinery, Inc.

letters dated July 26, i

-1985-and September 30, 1985 with enclosure thereto.

l General Nucleonics certificate of registration-Number

CA-321-S-102-S dated April 20,-1972.

j BHP Petroleum Americas Refinery, Inc.,' letter dated July 22, 1997.-

i 4

i ISSUING AGENCY:

U.S.'LNuclear Regulatory Commission Date:

Auoust 6, 1997 Reviewer:

'3b4M ow%

Michele L.

Burgess Date:

Aucust 6.

1997

- Concurrence

./

Xd6

/ Steven L.

'Baggett i

4

.i..

-REGISTRY OF RADIOACTIVE-'-SEALED SOURCES 3AND DEVICES SAFETY EVALUATION OF DEVICE (AMENDED IN-ITS: ENTIRETY)

NO.t NR-321-D-105-S DATE:

August'6, 1997 ATTACHMENT l' t lb" RC e' o u Rc c SrecN Ca'T H Os _-

I hJ CUfR'CS(M^t.)

I I.O 2

I.I 3

/.2 4

1.4 o'

l'. G G

2.a 7

2. 5 Y

%YhLL cour.

WrL O N V

g,g p pg CsLD Wr6D y

N..' '

- h--

j r

e h

l l2.00

~

.)i i

j

}

1 l

i i

i_

i J

t

.7i

+

o.o.

REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF DEVICE (AMENDED IN ITS ENTIRETY)

NO.: NR-321-D-105-S DATE1 August 6, 1997 ATTACHMENT 2 46.,3V D ),','f s 8A e r p.<

/-a v o,c.g"w, s

souece e.rar, o u

.,, s s

s.

fxx drAo Fou.

s

\\'

I.

.;/ s r -GQ

'/

,:.4_ ~ : = _.: 'f.,5

's N;/

's,

j th

7 l

k'

^ '

c. Ar...

v.,J!

7, Sovrtcc Hou s sua Ass c.nc s s S e c*r u o u - S o u s e s oveM

/

Soveer Inptx //avocr f

I

!l l I

s e tc 5-. s. Leen

^

, y l

ilI 4

ii l

l_ 7(sor s enos) 2 mao cAe

!j i

f i

lll l

i

+o %

l 3,.

I i

~l I'{ h l _,

[2"oo.

J j

[

~

i I

f ll l

ll

~

L a

1 l

Ij i

i l,

,l r

i l.

I l

'l ll l l.

lij M o p s,_

/332o 6/

i sovec.c cacs a s u n c l

.i i

i

'I

+_. a

_..w.

i g