Matl/Byproduct License 48-20087-01 for Sentry Labs Authorizes Use of Ni-63 Foil in Perkin-Elmer Gas Chromatograph for Sample Analysis
ML20148T854
Person / Time
Issue date: 02/02/1981
From: Cooper J
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
References
NUDOCS 8102270861
Download: ML20148T854 (3)


Text

  • acis"f,

. ~[

U. S. NUCLEAR REGULATORY COMMISSION N ""

"R" l

,y MATERIALS LICENSE g3

6. -

F l

1>

j s-Pursuant to the Atomic Energy Act of 1954, as amended, the Energy Iteorganization Act of 1974 (Public Law 93-

[l '

438), and Title 10, Code of Federal Regulations, Chapter 1, Parts 30, 31, 32, 33, 31, 33, 36, 40 and 70, and in reliance on statements and representations heretofore made by the licensee, a license is hereby issued authorizing the licensee to receive, acquire, possess, and transfer byproduct, source, and special nuclear material designated below: to use such 6,

material for the purpose (s) and at the place (s) designated below: to deliver or transfer such material to persons authorized j!

to receive it in accordance with the regulations of the applicable Part(s); and to import such byproduct and source 5

material. This license shall ue deemed to contain the conditions specified in Section 183 of the Atomic Energy Act of 1954,

[

as amended, and is subject to all applicable rules, regulations and orders of the Nuclear llegulatory Commission now or i

hereafter in effect and to any conditions specified below.

.1 j

l.icensee 1

1, Sentry Laboratories 1

3. License number 43-20087-01

,q 1500 I! orth Point Drive 2*

Stevens Point, Ulsconsin 34481

4. Expiration date February 26, 1986 Docket or Reference No.

e

6. Byproduct, source, and/or
7. Chemical and/or physical
8. Maximum amount that licensee special nuclear material form may possess at any one time under this license A. Mickel-63 A. Foil source contalued A. Iso singic foil to in Perkin-Elraer exceed 15 Detector Cell uillicuries p g6 s

1:odel No. 0330-0119

  1. @h /s 3

s 6

1

9. Authoriced use

* UG ED y<s/Sg7 w a

e' A. To be used in Perkin-Eltaer sao chrenatograph for sample analysis.

(.

44 s,

%'dDcw 7 -r,7,

s' C0JDITIONS 4

d '}\\0 s/

i 10.

Licensed caterial shall be used only at the licensee's address stated in Item above.

t 11.

The licensee shall comply with the provisions of Title 10, Chapter 1, Code of 7

1 J

Federal Regulations, Part 19, "Motices, Instructions and P,cports to Workers; Inspections" and Part 20, " Standards for Protection Against Radiation."

1 12.

Licensed ciaterial shall be use ' Sy, or under the supervision of, P.obert C. Voborsky.

l e

p

?

m.w) c.w:

~

i THIS DOCUMENT CONTAINS POOR QUALITY PAGES 820eer 6@

o e

s.pulWE=M g

i x

'[.

~

roEu sr.C-3 MA U* S. NUCLEAR REGULATORY COMMISSION

-Page 9 of 1 Pages M

MATERIALS LICENSE Supplementary Sheet License Number 43-20087-01 Continuco f ro.a Page 1 Docket or Reference No.

~

CONDI'IJ.J ll l

13.

ir icu of using the conventional r tdiatiou caution. colm s (.aagenta or purple on ye i m bachground) as provided in Section 20.203(a)(1), Titic 10, Code'of Federal c ulacions, Part 20 t.he liccasee is hereby authorized to Jabei detector cells cell caths, cont.ninin; licenaec aaterial and nued in Lpc chroaatography devices, u.

cita ccaspicuously etched or stamped radiation caution nyabols without a calor j

regI1recent.

N

{

14.

A.

Each chrouato;raph detector cell containin; Mickel-63 shall be tested for 1eakaac and/or contanination at intervals not to exceed six (6) wonths.

In t

tite absence of a s.ertificate f rom a transferor indicatint; that a test has bcea cade within six (6) nonths prior to the transfer, a detector cell received froa another person chall not be put into uce until tested.

L 3.

The test shall be capable of detectin;; the presence of 0.005 nicrocurie of radioactive r~aterial on the test sauple.

The test cacple shall be taken froa the surfaces of the device in which the Hickel-63 is mounted or stored ou uhich one ui;;ht expect contacination to accumtlate.

nccords of leak test

~

recults shall be kept in units of microcuries and naintained for inspection by

[

the Coanicsion.

f f

C.

If the test reveals the presence of 0.005 alcrocurie or tore of reuovabic contaa-

[

luation, the liccacce chall i=cdiately uithdraw the detector cell f rom use and

)

chall cause it to be decontaainated and repaired or to be disposed of in accord-ance uith Conaission regulations. A report shall be filed within five (5) days of the test uitn tha U. 3. Cuclear Regulatory Cocniusion, nogion Ill, Office of Inspection and Enforcement, 799 Roosevelt Road, Glen Ellyn, Illinola 60137, des-l cribia; the eculpaent involve,, the test reaulto, and the corrective action taken.

1;.

Testa for leakane am.r contacination shall bc parforced by the licensee or

[

by other percons specitically authorized by the Cowuiauion or an Agrecuent

(

State to perfora such services.

I t

15.

Detector cells containing licensed catorial chall not be opened by the licensee.

i 16.

Maintenance, repair, cleaning, replaceuent and disposal of foils contained in detector celic chall be perforned only by tuo device ranufacturer or other persons specifically authorized by the Cocaission or an Agrecuent State to perform such ac rvicco.

I 17.

Ihaaust frou detector cells containin;; licensed catorial shall be vented through a laboratory fuae hood or other suitable acans designed, to reduce potential exposure to personnel to the lowest practicable level.

13. 1: cept as otherviae spacified in this liconac, the liccasce shall have availablo and folion the instructions contained in the aanufacturar's instruction aanual for the chromato3raphy device.

w.m-g' 9

.d

~...

- l I

].<

U. S. NUCLEAR REGULATORY COMMISSION Page 4 of 3 Pa8cs:-

FORM Nr.C 374A ts n2

+~

MATERIALS LICENSE Supplementary Sheet License Number 20087-01 Docket or Reference No.

r.

[

Continue.' ircu'Pa3a :

'ti N

C3.'iDI::.

' s.

19.

Incept au specifically provided otheraice vi tnis liccace, the licensec shall possess and tva liccused naterial described in Itcus 6., 7., a:4 3. of thia liccaso in accord-h ance.zith statcaents, reprecantations, and pr.>cedurce containea in application dated f

E ncary 3, liidl. The ::uclear Regulatory Cosission's regulatirans shall govern the licutsee's stater: cuts in applications or letterr, unless the statecents are note 1

restricH.ve than the redulaticac.