SEABROOK STATION Engineering Office i

Pub 5c Service of New Hampshke Mew Hampshire Yankee Division April 10, 1986 SBN-1000 T.F. Q2.2.2 United States Nuclear Regulatory Cor:aission Region I 631 Park Avenue King of Prussia, PA 19406 Attention:

Mr. Richard W. Starostecki, Director Division of Project and Resident brograms

References:

(a) Construction Permitc CPPR-135 and CPPR-136, Docket Nos. 50-443 and 50-444 (b) Telecon of March 7, 1986, J.M. O'Connor, W.J. Daley, A.B. Dubison (YNSD) to J.P. Rogers (Region I)

Subject:

Interin 10CFR50.55(e) Report:

Pressure Transmitters (CDR 86-00-02)

Dear Sir:

On March 7, 1986 we reported a 10CFR50.55(e) item to Region I (Reference (b)).

The following information is provided under the reporting requirements of 10CFR50.55(e).

Description of Deficiency During testing the Veritrak 76PH2 Pressure Transitters, calibration shifts due to ambient temperatures changes in excess of the 1% per 100 F specified maximum were recorded.

Analysis of Safety Implication These transmitters provide input to the reactor trip and safety injection actuation logic. The trip cet points for the above are based on the maximum assu.ned error plus a margin. The excessive ambient temperature calibration shifts yauld create a condition that would violate allowable technical specification limits.

Status of Investigation The transmitters in question have been returned to Westinghouse so that they may, (1) analyze the cause and (2) repair the transmitters.

8604250256 860410 PLR ACOCK 05000443 S

PDR P.O. Box 300 + Soobrook,NH03874 Tetephone(603)474-9521 l

A njO 3

u o

United States Nuclear Regulatory Commission Attention:

Mr. Richard W. Starostecki Upon completion of the Westinghouse analysis and repair we will file an additional report.

Very truly yo rs, W

y s)

John DeVincentis, Director Engineering and Licensing cc: Atomic Safety and Licensing Board Service List s

Diane Curran; Esquire Calvin A. Canney Harmon & Weiss City Manager 2001 S. Street, N.W.

City Hall Suite 430 126 Daniel Street Washington, D.C.

20009 Portsmouth, NH 03801 Sherwin 2. Turk, Esq.

Stephen E. Merrill, Esquire Office of the Executive Legal Director Attorney General U.S. Nuclear Regulatory Commission George Dana Bisbee, Esquire Tenth Floor Assistant Attorney General Washington, DC 20555 Office of the Attorney General 25 Capitol Street Robert A. Backus, Esquire Concord, Nn 03301-6397 116 Lowell Street P.O. Box 516 Mr. J. P. Nadeau Manchester, NH 03105 Selectmen's Office 10 Central Road Philip Ahrens, Esquire Rye, NH 03870 Assistant Attorney General Department of The Attorney General Mr. Angie Machiros Statehouse Station #6 Chairman of the Board of Selectmen Augusta, ME 04333 Town of Newbury Newbury, MA 01950 Mrs. Sandra Gavutis Chairman, Board of Selectmen Mr. William S. Lord RFD 1 - Box 1154 Board of Selectmen Kennsington, NH 03827 Town Hall - Friend Street Amesbury, MA 01913 Carol S. Sneider, Esquire Assistant Attorney General Senator Gordon J. Humphrey Department of the Attorney General 1 Pillsbury Street One Ashburton Place, 19th Floor Concord, NH 03301 Boston, MA 02108 (ATTN:

Herb Boynton)

Senator Gordon J. Humphrey H. Joseph Flynn, Esquire U.S. Senate Office'of General Counsel Washington, DC 20510 Federal Emergency Management Agency (ATTN: Tom Burack) 500 C Street, SW i

Washington, DC 20472 Richard A. Hampe Esq.

Hampe and McNicholas Paul McEachern Esquire 35 Pleasant Street Matthew T. Brock, Esquire Concord, NH 03301 Shaines & McEachern 25 Maplewood Avenue Donald E. Chick P.O. Box 360 Town Manager Portsmouth, NH 03801 Town of Exeter 10 Front Street Gary W. Holmes, Esq.

Exeter, NH 03833 Holmes & Ells 47 Winnacunnet Road Breritwood Board of Selectmen Hampton, NH 03841 RFD Dalton Road Brentwood, NH 03833 Mr. Ed Thomas FEMA Region I Peter J. Mathews, Mayor 442 John W. McCormack PO & Courthouse City Hall Boston, MA 02109 Newburyperi, MA 31950 Stanley W. Knowles, Chairman Board of Selectmen P.O. Box 710 Nort{h fn, itsib M 03862