J REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF SEALED SOURCE 3
4 4
NO.: NR-420-S-806-S DATE:
September 8, 1995 PAGE 1 OF 4 i
SEALED SOURCE TYPE:
Sealed Source MODEL:
RXA, RXB, RXC, RXL, RXM, RXN i
i MANUFACTURER / DISTRIBUTOR:
LFE Industrial Systems Corporation 55 Green Street Clinton, MA 01510 i
ISOTOPE:
MAXIMUM ACTIVITY:
l Cadmium-109 10 millicuries (370 MBq) s Iron-55 15 millicuries (555 MBq)
Americium-241 30 millicuries (1110 MBq)
- see description for activity for each source LEAK TEST FREOUENCY:
6 Months i
l PRINCIPAL USE:
(U) X-Ray Fluorescence l
CUSTOM SOURCE:
YES X
NO 9702240240 950908 PDR RC SSD PDR
REGISTRY.OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF SEALED SOURCE l
NO.: NR-420-S-806-S DATE:
September 8, 1995 PAGE 2 OF 4 SEALED SOURCE TYPE:
Sealed Source i
DESCRIPTION:
The cadmium and iron sources are electroplated on copper foil and annealed in. a hydrogen at'nosphere.
The americium is deposited as the nitrate onto an A1,0
>r beryllium window to a monel body 3
using eccobord 276 epoxidt binder.
The radioactive source is placed against the window in the capsule and a rear plug is heliarc welded to the body.
Dimensions Capsule Length
_ Diameter Wall Thickness inches mm inches mm inches mm RXA, RXL 0.2 5.1 0.32 8.1 0.06 1.5 RXB, RXM 0.2 5.1 0.43 10.9 0.1 2.5 RXC, RXN 0.2 5.1 0.6 15.2 0.1 2.5 Note:
RXA, RXB and RXC capsules have 0.004" (0.1 mm) Be windows.
RXL, RXM and RXN capsules have 0.040" (1.0 mm) Al windows.
Isotooe Maximum Activity Capsule Desionations mci /MBc Cd-109 10/370 RXA, RXB, RXC Fe-55 15/555 RXA, RXB, RXC Am-241 10/370 RXA, RXB, RXC Am-241 30/1110 RXL, RXM, RXN For their own administrative purposes, the manufacturer will identify sources using a designation which identifies the source capsule type, isotope and activity.
For example, an RXA-Am-10 source would contain 10 millicuries of americium-241.
LABELING:
Each source is engraved with its serial number, isotope, and activity.
j
REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES 1
SAFETY EVALUATION OF SEALED SOURCE NO.: NR-420-S-806-S DATE:
September 8, 1995 PAGE 3 OF 4 i
SEALED SOURCE TYPE:
Sealed Source 5
CONDITIONS OF NORMAL USE:
j The sealed sources are designed for use in x-ray fluorescence analyzers and other laboratory equipment.
e PROTOTYPE TESTING:
A prototype of the source designs were tested to the specifications of ANSI N5.10 1968 and achieved a classification of C32231.
l LIMITATIONS AND/OR OTHER CONSIDERATIONS OF USE:
4
- The sealed source may be used by specific licensees of NRC or Agreement States.
?
- Handling, storage, use, transfer, and disposal:
To be determined by the licensing authority.
- Each source shall not be subjected to conditions which exceed its ANSI classification.
- The sealed source shall be leak tested at intervals not to exceed 6 months using techniques capable of detecting 4
0.005 microcurie (185 Bq) of removable contamination, a
- This registration sheet and the information contained within the references shall not be changed without the written consent of the NRC.
SAFETY ANALYSIS
SUMMARY
LFE distributed this device prior to 1969.
However, in 1969'the product line was sold and all the information that LFE has concerning the device is contained in this certificate.
LFE will not accept return of the sources.
_. _. -. - _ _. =
l REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF SEALED SOURCE i
5 NO.: NR-420-S-806-S DATE:
September 8, 1995 PAGE 4 OF 4 j
SEALED SOURCE TYPE:
Sealed Source i
REFERENCES:
The following supporting documents for the Models RXA, RXB, RXC, RXL, RXM and RXN sealed source are hereby incorporated by reference and are made a part of this registry document.
- LFE's letters dated July 12, 1995, September 16, 1991, and August 6, 1991, with enclosures thereto.
ISSUING AGENCY:
U.S. Nuclear Regulatory Commission Date:
Sectember 8, 1995 Reviewer:
/~
John W. Lubinski Date:
Seotember 8, 1995 Concurrence:
/ Eft'even L.'Baggett
)
i INFORMAT.ON REC,3D i
TYPE:
Note to LFE files for registration certificates made inactive September 1995.
ORGANIZATION:
LFE DATE:
09/21/95 i
i
SUMMARY
i Existing AEC/NRC files (registration files, license files, NUDOCS, and archives) concerning these products i
were examined for information that supports registration of the products.
Copies of all applicable information located is included in this file.
i The manufacturer, LFE, was also asked for additional information concerning the products.
However, man the products have not been distributed for years, y of the product lines were sold in the late 1960's, or and LFE provided all additional information they have concerning the products.
NRC has transferred the registration certificates to inactive, as requested by LFE, and continues to conclude the products are acceptable for licensing purposes based on AEC's original evaluation of the products and the operational history of the products.
i PERSONS DOCUMENTING THE SIGNATURE?
INFORMATION:
DATE:
/
j)/9t?
^
John W. Lubinski
/
7
/
09/21/95 Steven L. Baggett
[
09/21/95
=
1 i
l I
l
William R. Prendergast LFE Industrial Systems Corporation 55 Green Street Clinton, MA 01510
Dear Mr. Prendergast:
This letter is in reference to your letter dated August 6,1991, which requested that we transfer the registration certificates listed in column 1 of Enclosure I to inactive status.
The certificates have been transferred to inactive status as you have requested.
Please note that. transfer of the registration certificates to inactive status means the sources or devices included on the certificates may no longer be commercially distributed.
The registration numbers for the certificates have been changed.
Enclosure I specifies the old registration number, the new registration number, and the model numbers.
Please note that some old certificates have been combined into one new inactive certificate.
In addition, two registration certificates have been deleted from the system since they were duplicates of other certificates.
Please review the registration certificates (copies enclosed in their entirety and notify us immediately if there are any errors or) omissions.
If you have any questions, please call me at (301) 415-7868 or Mr. Steven Baggett at (301)_415-7273.
Sincerely,
/3/
John W. Lubinski, Mechanical Engineer Sealed Source Safety Section Source Containment and Devices Branch Division of Industrial and Medical Nuclear Safety Office of Nuclear Material Safety And Safeguards
Enclosures:
As stated cc: S. Kimberley, LFDCB Distribution:
SCDB r/f SSD-91-86 NE01 SSD Files listed in column 1 of enclosure 1 of this letter DOCUMENT NAME: C:\\ FILES \\SSDS\\LFE\\LFE. CMP To,eo:fve a copy of this document, Indicate in the boa: 'c" = Copy without attachment / enclosure
- E* = Copy with attachment / enclosure
'N' = No copy 0FFICE SCDB l.L l
l l
l NAME JLubiriski/jl DATE 09/ A /95 0FFICIAL RECORD COPY hW
a ENCLOSURE 1 Old Reg. Number New Reg. Number Model Number (s)
NR-420-D-101-0 NR-420-D-801-S Keleket-Barnes Rotary NR-420-D-102-U NR-420-D-802-S Keleket-Barnes Shperay NR-420-D-103-U NR-420-D-803-S Keleket-Barnes Ceiling Mounted Unit NR-420-S-104-S NR-420-S-804-S S-10 NR-420-D-106-U NR-420-D-805-S Keleket-Barnes Flexaray NR-420-S-107-U NR-420-S-806-S RXA, RXB, RXC, RXL, RXM, RXN NR-420-D-108-U NR-420-D-807-B SCL-ID NR-420-D-109-B NR-420-D-808-B BGL-7C certificates were BGL-7 NR-420-D-110-G combined BGL-lC NR-420-D-Ill-B NR-420-D-112-8 NR-420-D-809-B BGL-1 NR-420-D-113-B NR-420-D-810-B S0-70, S0-7E certificates were 50-7A, S0-78, 50-7C NR-420-D-Il4-8 combined NR-420-D-Il5-B NR-420-D-811-B S0-1A, S0-18, S0-1C NR-420-D-116-B NR-420-D-812-B SN-7D, SN-7E certificates were SN-7A, SN-78, SN-7C NR-420-D-117-B combined NR-420-D-118-B NR-420-D-813-B SN-1A, SN-1B, SN-lC, SN-1D, SH-1E, SH-lF NR-420-D-119-B NR-420-D-814-B SC-70, SC-7E NR-420-D-120-B NR-420-D-815-B SC-7A, SC-78, SC-7C NR-420-D-121-B NR-420-D-816-B SC-1A, SC-1B, 5C-10, SC,1D, SC-1E, SC-lF NR-420-D-122-8 NR-420-D-817-B SN-P7B NR-420-D-124-B NR-420-D-818-B SCL-1A and SCL-1B NR-420-D-125-G NR-420-D-819-B SN-P7A4 NR-420-D-126-U NR-420-D-820-B HSB-76 and HSB-77 NR-420-D-127-G NR-420-D-821-G HSB-76A and HSB-77A 1
i l
ENCLOSURE 1 Old Reg. Number New Reg. Number Model Number (s)
\\
NR-420-D-128-G NR-420-D-822-G BGL-7A NR-420-0-132-G
_NR-420-D-823-G SNF-18 NR-420-D-133-G NR-420-D-824-G SSL-3 l
NR-420-S-137-U NR-420-S-826-S R-30, R-31 l
NR-420-S-138-U NR-420-S-827-S CD-1 NR-420-0-139-U NR-420-D-828-S SC-11 NR-420-D-140-U NR-420-D-829-B DH-75 l
NR-420-S-141-U NR-420-S-830-S S-75 NR-420-D-142-U NR-420-D-831-S AX-4 NR-420-D-143-U NR-420-D-832-S BG-107 NR-420-0-144-U NR-420-D-833-S BG-105 NR-420-D-145-U NR-420-D-834-S BG-7 NR-420-D-146-U NR-420-D-835-S BG-5 NR-420-0-147-U NR-420-D-836-S BG-2/5 NR-420-D-151-B deleted from system -
same as old registration NR-420-D-117-B NR-420-D-152-U NR-420-D-840-B SC-9C NR-420-S-154-U NR-420-S-842-S S-9 i
NR-420-S-155-U NR-420-S-843-S R-58, R-59 NR-420-S-156-U NR-420-S-844-S CS-15 NR-420-S-157-U NR-420-S-845-S CD-15 l
NR-420-S-158-U deleted from system -
same as old registration NR-420-S-156-U l
NR-420-D-159-U NR-420-D-846-S SC-11A NR-420-D-160-U NFet20-D-847-S
, Clathrate Gas Detector NR-420-S-163-U NR-420-S-849-S CD-11 NR-470-0-164-B NR-470-0-850-B SN-P1B J
2