Confirms Installation of post-accident Sampling Sys & Sys Operability as of 840601.Procedure for Obtaining post- Accident Samples of Primary Coolant Is Contained in Health Physics Procedure (HPP)-14, Analytical Instrumentation
ML20090E642
Person / Time
Site: Fort Saint Vrain 
Issue date: 07/02/1984
From: Warembourg D
PUBLIC SERVICE CO. OF COLORADO
To: Johnson E
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION IV)
References
TASK-2.B.3, TASK-TM P-84192, TAC-44443, NUDOCS 8407200006
Download: ML20090E642 (5)


Text

.

.s public Service Onnimny #4 Cnknda Q1 V'

16805 WCR 19 1/2, Platteville, Colorado 80651 0

9 July 2, 1984 Fort St. Vrain Unit #1 P-84192

$$k R l 2 $84

{h Mr. E.H. Johnsoa, Chief

)'

Reactor Project Branch 1 L

U.S. Nuclear Regulatory Commission

\\

611 Ryan Plaza Drive, Suite 1000

. L_

Arlington, Texas 76011 j

SUBJECT:

Post Accident Sampling System Implementation

Reference:

1) P-82423 DW Warembourg to RA Clark, dated September 27, 1982
2) G-83349, GL Madsen to OR Lee, dated September 22, 1983
3) P-83352, FJ Borst to GL Madsen, dated October 28, 1983

Dear Mr. Johnson:

This is to confirm that the Fort St. Vrain Post Accident Sampling System is currently installed and operable, effective June 1,

1984.

As you are probably aware, our post accident sampling system utilized the identical sample line as our normal primary coolant sampling system; thus we are able to veriff on a continuing basis that the sampling system is operable.

Per REFERENCE 3,

however, it was necessary to make one modification to the collection system to enable us to obtain the small (approximately 0.04 SCC) sample required.

This modification was made via Temporary Configuration Report number 83-1103 and-involved installing a more sensitive pressure gauge in the system. Design Change. Action Request 467, dated 5/15/84 was also issued for this change.

The procedure for obtaining post accident samples of primary coolant is contained in Health Physics Prc:edure (HPP)-14, Analytical Instrumentation.

Mr.

Ron Baer of your office is on controlled distribution for this procedure.

As e courtesy, a copy of the applicable portion of HPP-14 is attached.

f$000?ol&ff,'

b

/

O m

yq t

~.

4 2--

If you have any further questions on this matter, please contact Mr. Ted Borst of my staff at (303) 571-7436.

Very truly yours, W

Oon Warembourg Manager, Nuclear Production Fort St. Vrain Nuclear Generating Station OW/dal i

Attachment i

?

i 4

D 1

a 3

+

.D

-e e

e

-_w..

wg.

p

.e O,i g 9,

y y

4

g PUBLIC SERVICE COMPANY OF COLORADO 3

FORT ST. VRAIN NUCLEAR GENERATING STATION page 21 of 24 g

t 5.12 PRIMARY COOLANT SAMPLING DURING ACCIDENT CONDITIONS 5.12.1 Initial Requirements a)

Determine the expected radiological conditions in the access to and in the Analytical Instrumentation Room.

b)

Determine the expected activity of the Primary Coolant, as required.

c)

Stop all the sample flows to I-9325 from the Control Room, d)

Don the proper protective clothing based on the determinations in STEPS a) and b).

e)

Monitor radiation levels when entering the A.I. Room.

f)

Set up a portable shielded container, if needed, as not to exceed 3 Rem (Whole Body) or 18.75 Rem (Extremities) to any individual involved.

(EMERGENCY LIMIT:

25 REM, WHOLE g

BODY) g)

Proceed to the A.I.

Room and establish the valve lineup per Attachment HPP-14B.6.

h)

Set up A.I.

panel to collect sample as per following procedure.

1)

Notify Radiochemistry Lab personnel of pending sample.

5.12.2 2 ml Sample Collection a)

Place 2 mi vial into lucite dummy and insert dummy into facility such that the needle will pierce the center of the vial cap.

Lucite dummy must be inserted one way only, and dummy is marked as to top.

b)

CLOSE facility door and tighten latch firmly.

c)

Ensure valves HS-93433, HS-93434, HS-93413, HS-93432, HS-93414, and HS-93416 are CLOSED.

d)

OPEN valve HS-93415.

If micro switches ZS-93416 and ZS-93415 are actuated, the valve will open.

FORM 372 22 3643

,--,...--.y.-

(~14 3

PUBLIC SERVICE COMPANY OF COLCRADO g

FORT ST. VRAIN NUCLEAR GENERATING STATION Page 22 of 24 D

e)

OPEN valves HS-93431 and HS-93435, then evacuate system using P-9301.

Evacuate to 20-25" Hg.

f)

SHUT off P-9301.

g)

CLOSE HS-93435 and HS-93431.

h)

Bypass the RT 7312 control actions, if required, to allow'the opening of the Primary Coolant sample line to I-9325, then OPEN the appropriate sample isolation valve (HS-9316 or HS-93256).

1)

OPEN HS-93433 and HS-93434. Adjust pressure to a maximum of 2.5 psig as indicated on PI-93405 using PCV-93412.

j)

Purge sample line >100 cc/ min for a minimum of one minute, as indicated on FI-93607.

k)

Record initial vacuum indication on PI 93406.

1)

Open sample metering valve very slowly until approximately a 0.4 inch decrease in vacuum is

(

observed on PI-93406.

m)

Close the metering valve.

n)

Record final vacuum indication on PI 93406.

o)

CLOSE HS-93415 and HS-93435, p)

Remove the sample bottle from the collection facility and place in the shielded container, if necessary.

q)

SHUT the sample isolation valve from the Control Room, r)

Transport the sample by the most direct route to the on-site analysis area, and place directly into the shielded storage area, if required, until sample is to be analyzed.

PORM 372 22 3643

._..y..,.

Attach' HPP-24K PUBLIC SERVICE COMPANY OF COLORADO Issue 18 1

j FORT ST. VRAIN NUCLEAR GENERATING STATION Page 2 of 2 j

VOLUME COLLECTED IN VARIOUS SERUM VIALS BASED ON EVACUATION AND nACYJILL MI'JiOD DELTA V(3)

Vt125)

DELTA Vt31 V(125)

DELTA V(3)

V(125)

IN.CH.

see sec IN.CH.

see see IN.CH.

see see 12.00_,.__ 1.09 48.65 8.00 0.72___ 32.43 4.00 0.36.___ 16.22 11.90 1.08_._,,48.24-7.90 0.72_,,,

32.03 3.90 0.35 __. 15.81 11.80,__ 1.07_,,_,47.84 7.80

0. 71,,,,,, 31. 6 2 3.80 0.34_,15.40 11.70_,,_ 1.06 47.43 7.70 0.70.,,_. 31.22 3.70
0. 3 4__.,, 15. 0 0 11.60__,,, 1.05 47.03 7.60 0.69 _ 30.81 3.60 0.33,__ 14.59 11.50,___ 1.04_

46.62 7.50 0.68_,_, 30.40 3.50

0. 3 2_,_,, 14.19 11.40___ 1.03_ _ 46.22 7.40 0.67 _ 30.00 3.40 0.31_,_ 13.78 11.30___ 1.02___ 45.81 7.30 0.66_,,,_ 29.59 3,30
0. 3 0_,,,,_

13. 3 8 11.20_,.,_ 1.01 _ 45.40 7.20 0.65_

29.19 3.20

0. 29_,,,,, 12. 9 7 11.10 __ 1.01 _ 45.00 7.10 0.64___ 28.78 3.10 0.28 __ 12.57 11.00_,

1.00_ _ 44.59 7.00 0.63.___ 28.38 3.00 0.27.,,,_ 12.16 10.90___ 0.99,_ 44.19 6.90 0.62.__ 27.97 2.90

0. 26_,,__ 11. 76 10.80._._ 0.98

_,,43.78 6.80

0. 62._,,,,, 2 7. 5 7 2.80 0.25_,

11.35 10.70___ 0.97 _ 4~.38 6.70 0.61__,,27.16 2.70 0.24, 10.95 10.60,_,0.96._,_ 42.97 6.60 0. 6 0.,,,,_,

2 6. 7 6 2.60 0.24_,_,10.54 10.50_ _,0.95

,_ 42.57 6.50 0.59 _ 26.35 2.50

0. 2 3.,,_, 10.13 10.40___ 0.94.,_,,42.16 6.40 0.58_ _ 25.95 2.40 0.22___

9.73

10. 3 0,,, _, 0. 9 3.,_,_, 41. 7 6 6.30 0.57___ 25.54 2.30
0. 21,

, 9.32

10. 2 0_,,,_ 0. 9 2._,,, 41. 3 5 6.20 0.56_ _ 25.13 2.20
0. 2 0_ _, 8.92 10.10 _,,,0.91

__ 40.94 6.10 0.55 __ 24.73 2.10 0.19__,, 8.51

10. 0 0,,,,._ 0. 91,,,, 4 0. 5 4 6.00 0.54 24.32 2.00 0.18 _,, 8.11 9.90,__ 0.90,_,_ 40.13 5.90 0.53._,

23.92 1.90 0.17___

7.70 9.80.,,,_ 0.89 ___ 39.73 5.80

0. 5 3_,,,_ 2 3. 51 1.80 0.16__,,

7.30 9.70,_, 0.88 _ 39,32 5.70 0.52_ _ 23.11 1.70 0.15 _. 6.89 9. 60._,_ 0. 8 7_,,,,,,, 3 r. 9 2 5.60 0.51 _ 22.70 1.60 0.14_._,,

6.49 9.50,,,_,,0.86

_ 38.51 5.50 0.50,_ 22.30 1.50 0.14_,,,,_

6.08

9. 4 0,,,, _ 0. 8 5_,_, 3 8.11 5.40 0.49._,21.89 1.40 0.13___

5.68 9.30,,,_ 0.84___,37.70 5.30 0.43_,_ 21.49 1.30 0.12_,,,,,

5.27 9.20 _ 0.83,,,,_,,37.30 5.20 0.47,,,

21.08 1.20 0.11.,,_, 4.86 9.10__,_ 0.82___ 36.89 5.10 0.46_,,20.68 1.10 0.10,,_ 4.46 9.00 _,0.82 _ 36.49 5.00 0.45_,20.27 1.00 0. 0 9_,,,,_

4.05 8.90,,_ 0.81_,_ 36.08 4.90 0.44_,,,19.86 0.90 0.08_,_

3.65 8.80 _ 0.80,_ 35.67 4.80 0.43,__,19.46 0.80 0. 0 7,___

3.24 8.70_,,,_ 0.79.,__ 35.27 4.70 0.43.,_,19.05 0.70 0.06 _ 2.84 8.60___ 0.78 _ 34.86 4.60 0.42_.,,, 18.65 0.60

0. 05_._,

2.43 8.50._,_ 0.77_,,,_ 34.46 4.50 0.41_ _ 18.24 0.50 0.05,_ 2.03 I

8.40__ 0.76,_ 34.05 4.40 0.40 17.84 0.40 0.04 _ 1.62 8.30__

0.75.,,._,,33 65 4.30 0.39 17.43 0.30

0. 0 3,,,,, 1.22 8.20.,_, 0.74_,,_,33.24 4.20 0.38.,,_ 17.03 0.20 0.02 0.81 8.10___ 0.73 _ 32.84 4.10 0.37._,,_ 16.62 0.10 0.01_ _ 0.41 i

I

(

ASSUME ACITAL VOLUME OT 2.71 see FOR THE SMALL VIAL AND 121.3 see FOR THE LARGE VIAL.