s ST-RM-3-1 Fort Calhoun Station Unit No. 1 SURVEILLANCE TEST ST-RM-3 Emergency Plan Radiation Instruments and Equiement A.

PURPOSE 1.

To verify the operability and the current calibration status of the Emergency Plan Radiation Instruments.

2.

To verify the satisfactory condition and operability of respiratory protection equipment.

3 To inventory emergency plan supplies and equipment.

/

B.

REFERENCES 1.

Fort Calhoun Station Unit No. 1 Standing Order No. T-13 " Quality Control Program for Chemistry and Radiation Protection Equipment".

2.

Technical Specification Table 3-3, Item 4.

3 NUREG 00hl C.

SPECIAL INSTRUMENTATION / EQUIPMENT 1.

8 pC Cesium 137 Button Source.

D.

PREREQUISITES l

l None l

E.

PRECAUTIONS 1.

Sources must be handled in accordance with the Radaition

(

Protection Manual.

t l

ISSllED NOV 181982 R16 11-18-82 8304040343 830330 PDR ADOCK 05000285 G

PDR l

t ST-RM-3-2 F.

TEST PROCEDURE 1.

Emergency Plan Radiation Instruments Operational Checks Frequency:

Monthly Responsibility:

Chemical and Radiation Protection a.

Initial Conditions None b.

Test Procedure (1)

Operationally test, check and record items listed in the Test Record.

c.

Remarks:

Emergency Operations Facility

/

Check Completed By

/

Date l

1 Fort Calhoun Control Room

/

Check Completed By

/

Date Alternate Emergency Control

/

Facility Check Completed By

/

Date 1

1 Results Reviewed By

/

Plant Health Physicist

/

Date l

l ISSllED NOV 181982 R16 11-18-82 t

f<?$

gg ST-RM-3-3

<- m

  • im TEST RECORD g ;' %

j Section F.1 Emergency Operations Facility f

Operational Function Check /

Serial Battery Source Test Calibration Remarks Instrument Number Check Used Acceptable Due Date-(1)

Eberline 120 (Off-site Monitor Kit)

(1)

Eberline 120 (Off-site Monitor Kit )

(1)

Eberline 120 (Counting Kit)

(1)

Eberline 120 (Room 6 Cabinet)

Eberline 520/530 (Rescue Squad Kit)

(2)

RM-lh/15/19 (Gate Monitor Kit)

(2)

RM-lh/15/19 (Gate Monitor Kit)

RM-lh/15/19 (Decon)

High Range Ion Chamber or GM (Room 6 Cabinet)

Dosimeter Charger g

(Counter / Dosimeter Kit)

N/A N/A.

N/A s

e Tape Recorder 4

(Room 11 Cabinet)

[

(1) Must have conversion factors taped to instrument housing (cpm to micro-curies /ml or dpm).

(2) Conversion epm to dpm must be attached.

se cuace M

E

~

ST-RM-3-4 S CA Qg TEST RECORD Section F.1 Emergency Operations Facility Operational Serial Battery Source Test Remarks Instrument Number Check Used Acceptable Calculator - TI-59 (Room 11 Cabinet)

N/A N/A Personal Air Sampler (Rescue Squad Kit)

N/A N/A l

Pencil Dosimeter 0-500 MRD4 Quantity (Dosimeter Kit)/Rese Squad 12/6 N/A N/A N/A i{

Pencil Dosimeter 0-50 RD4 Quantity (Dosimeter Kit) 12 N/A N/A N/A 1

TLD Badges Quantity (Dosimeter Kit) 12 N/A N/A N/A Date Replacement Due - - - - -

Thyro-Block Tablets Quantity 3

(Room 11 Cabinet) 2 boxes N/A N/A N/A Date Replacement Due - - - - -

Radeco E-809V Air Sampler (Emergency Vehicle)

N/A N/A Radeco E-809V Air Sampler N/A N/A l

(Emergency Vehicle) l Radeco E-809V Air Sampler N/A N/A (Room 6)

Radeco E-809V Air Sampler N/A N/A l

(Room 6)

SAM-2 i

N/A (Room 6)

SAM-2 (Room 6)

Kib 11-18-82

ST-RM-3-5

-TEST RECORD Section F.1 Emergency Operations Facility Respiratory Equipment RESPIRATOR SERIAL TEST TYPE NUMBER INSTRUCTIONS RESULTS REMARKS Pressure Demand SCBA EOF-1 Check that air cylinders are fully charged (>2000 psi),

Pressure facepiece assemblies are Demand SCBA EOF-2 totally functional & properly stored, harnesses are in good Pressure condition & regulators &

Demand SCBA EOF-3 varning devices are function-ing properly.

Initial Pressure inspection sticker is found Demand SCBA EOF-h satisfactory.

Pressure Demand SCBA EOF-5 Pressure Demand SCBA EOF-6 Full Face Mask N/A Check tightness of connec-tions & conditions of face-Full Face piece, head bands, valves, Mask N/A connecting tube & canister.

Check rubber or elastomer Full Face parts to ensure that they Mask N/A are pliable, flexible and not deteriorating or taking Full Face a set during storage.

l Mask N/A Initial inspection sticker, l

rebag & seal if found sat-l Full Face isfactory.

l Mask N/A l

Full Face Mask N/A I

Full Face Mask N/A i

l l

Full Face Mask N/A Full Face

(

Mask N/A l

R16 11-18-82 NOV 181982 i

.-._,= = - -

ST-RM-3-6 TEST RECORD Section F.1 Emergency Operations Facility Fespiratory Equipment RESPIRATOR SERIAL TEST TYPE

-NUMBER INSTRUCTIONS RESULTS REMARKS Check tightness of connec-

^

tions & conditions of face--

piece, head bands, valves, Full Face connecting tube & canister.

Mask N/A Check rubber or elastomer parts to ensure that they FC Fue are Pliable, flemle and Mask N/A not deteriorating or taking a set during storage.

  • FC Fe n

a nspecM on sM ch r, Mask N/A rebag & seal if found sat-

  1. 7'
  • Full Face Mask N/A
  • Full Face Mask N/A
  • Full Face Mask N/A
  • Full Face Mask N/A
  • Full Face Mask

.N/A

(

These Full Face Masks are located in the Emergency Vehicles, 3 in each vehicle.

l I

ISSilED NOV 181982 R16 11-18-82 i

"l

~

ert?Jass g

N ST-RM-3-7 f? A

~

{

(<mm8

,.j TEST RECORD Section F.1 g (!{$

Emergency Operations Facility l

Inventory of Additional Supplies and Equipment ITEM QUANTITY INVENTORY REMARKS Off-Site Monitor 2

Each Containing:

(1) E-120 survey instrument, HP-190 Team Kit s probe and SH-3 sample holder.

(1) Sector map and book of sample locations.

(5) Small plastic bags.

(10) Air sample filters-(Gelman or equivalent.).

(4) Pair surgeon gloves.-(2) Spare

'D' Cell Batteries.

(10) Silver zeolite cartridges.

(1) Calculator.

(30) Smear papers.

(1) Clipboard with pencils or pens and paper.

(1) Flash-light.

Rescue Squad 1

(1) Survey instrument.

(1) Lapel air sampler.

Kit (6) Pencil Dosimeters -(0-500 MR).

(1) Dosimeter Charger.

(1) Note pad and pen or pencil.

Emergency Kit 1

(12) Sets of protective clothing.

(12) Full face masks.

(Roll Around)

(3) Flashlights.

(10) Spare 'D' cell batteries.(5) Poly bags.

(6) Rolls of masking tape.

(6) Pencils or pens.

(6) Clipboards with paper.

(100) Smear papers.

(100 ft.) Radiation barrior rope.

(5) Radiation area caution signs.

(5) Grease pencils.

(100) Plastic shoe covers.

(2) Bundles surgeons gloves.

Personnel Decon (1) & (2)

Split between two lockers:

Lockers (24) Towels.

(6) Bara of soap. (6) Hand brushes.

(6) Step off pads.

(12) Disposable coveralls.

(2) Bundles surgeon gloves.

(2) Large plastic bags-(for dirty clothes).

(50) Plastic shoe covers.

(1) Clipboard with paper, pen or pencil.

(4) 5 pound bottles sodium citrate.

R16 11-18-82

t ? css 5 a'f2

< S. WD ST-RM-3-8 e

cp

.._.m:::s TEST RECORD

    1. M Section F.1
IYd

'] ]

Emergency Operations Facilities Inventory of Additional Supplies and Equipment (Continued)

ITEM QUANTITY INVENTORY REMARKS Gate Monitor Kit 2

Each Containing:

(1) Rm-14/35/19 frisker with pan-

"Frisker may be found out cake probe.

(4) Pair surgeons gloves.

(1) Clip-of kit if battery is being board with paper, pencil or pen.

charged."

Counter / Dosimetry 1

(1) E-120 survey instrument with HP-190 probe and Kit SH-3 sample holder.

(10) Small plastic bags.

(2) Spare

'D' cell batteries.

(12) TLD badges.

(12) Pencil dosimeters-(0-500 MR). (12) Pencil dosimeters-(0-50 R).

(1) Dosimeter charger.

(1) Clipboard with paper, pencil or pen.

(4) Pair surgeons gloves.

(1) Calculator.

6 Room 4 Public Infor-1 (1) RERP-(Copy #135).

(1) EPIP-(Copy #135).

mation Book Shelf (1) Set Updated Safety Analysis Report-(Copy #39).

t 1

(1) Systems Description Volumes I, II & III-(Copy l

  1. 77).

(1) State of Nebraska Emergency Plan-(Copy j

  1. 6).

(1) State of Iowa Emergency Plan-(Copy #6).

R (1) Harrison County Iowa Emergency Plan-(Copy #6).

t (1) Pottawattamie County Iowa Emergency Plan-(Copy

  1. 6).

(1) P&ID-(Copy #212).

J EOF-Room 6 Cabinet 1

(2) Gate Monitor Kits. (1) Counter / Dosimetry Kit.

(1) Rescue Squad Kit.

(2) Off-Site Monitor Team Kit.

(1) High Range GM or Ion Chamber Survey Instrument.

(10) Silver zeolite cartridges.

(1) E-120 survey instrument.

(1) RM-14/15/19 (For Decon Team). (2) RADECO H-809V Air Sampler.

(2) SAM-2 Iodine Sample Counters for Off-Site use.

4 R16 11-18-82

f

~

)

summ===

&5 ST-RM-3-9' TEST RECORD g

Section F.1 g

Emergency Operations Facilities Inventory of Additional Supplies and Equipment (Continued)

ITEM QUANTITY INVENTORY REMARKS EOF Room 11 Cabinet 1

Each Containing:

(1) EDO/ Recovery Manager Record'

-Book.

(1) Cassette tape recorder with 3 tapes.

(1) TI-59 Calculator.

(1) Complete set operating Manuals-(Copy #50).

(1) Set EPIP and RERP-(Copy

  1. 12).

(1) Set EPIP and RERP-(Copy #105).

(1) Set Updated Safety Analysis Report-(Copy #38).

(200) Bottles Thyro-BlocA.

(1) P&ID-(Copy #189).

(1) Operating License and Tech Specs-(Copy #83).

(1) Communications Log.

EOF Room 11 1

(6) Pressure Demand SCBA's.

(4) Spare SCBA air (General) bottles.

(1) Lunch box radio.

(2) Plastic trash containers for Decon Area.

(1) Large chemist bottle for Decon Area - Sodium Citrate mixing.

(1) Tag Board - insure all tags are present -

insure all keys are present. (A) Key to Radio Room located behind Radio Operator Tag.

(B) Keys to Decon locker located behind Personnel Decon Team Tag.

(C) Keys to Monitor Assembly Area and Room 6 Cabinet located near bottom of the board.

(D) Keys for Emergency Vehicles located near bottom of tag board. (1) Latest revision of Emergency Call List posted on bulletin board adjacent to tag board.

R16 11-18-82

~ __.. -.-

.usamme 2:

ST-RM-3-10 k-C TEST RECORD

" gmqyng Section F.1 gg'Eg)

Emergency Operations Facility Inventory of Additional Supplies'and Equipment (Continued) l' ITEM QUANTITY INVENTORY REMARKS Emergency Vehicles 2

Each Containing:

(3) Full face respirators.

for Off-Site Monitor (12) Disposable coveralls.

(12) Pair plastic Teams shoe covers.

(12)-Disposable surgeon caps.

(12) Disposable hoods. '(1) Roll masking tape.

(3) Bottles Thyro-Block.

(3) Pair rubber shee covers.

(3) Plastic bags-large.

(3) Plastic b

(1) Plastic trash container.

I(ags-small.

1) RADECO H-809V air sampler.

(1) Bundle surgeon gloves.

Check vechicle will start and fuel level

>3/4 tank. Check SAM-2 Pig undamaged and secure.

I a

R16 11-18-82 4

d i

2

j f(:.fs h

n o'

ST-RM-3-ll ra%

TEST RECORD Section F.1 Fort Calhoun Control Room OPERATIONAL FUNCTION CHECK /

SERIAL BATTERY SOURCE TEST

. CALIBRATION INSTRUMENT NUMBER CHECK USED ACCEPTABLE DUE DATE REMARKS Eberline 120 (1)

(Emergency Locker) f Portable Air Sampler (Emergency Locker)

N/A N/A Teletector (Emergency Locker)

Dosimeter Charger (Emergency Locker)

N/A N/A N/A-Pencil Dosimeter Quantity (0-500 MREM) 12 N/A N/A N/A (Emergency Locker)

Pencil Dosimeter Quantity (0-50 REM) 12 N/A N/A N/A (Emergency Locker)

TLD Badges Quantity Date Replacement Due (Emergency Locker) 12 N/A N/A N/A Thyro-Block Tablets Quantity Date Replacement Due (Emergency Locker) 12 N/A N/A N/A (1) Must have conversion factors taped to instrument housing (cpm.to micro-curies /ml or dpm)

R16 11-18-82

l n==.:*. r.

5DU

< hfh ST-RM-3-12 (R

TEST RECORD

{

!b Section F.1

\\

EYN3 i

Fort Calhoun Control Room f

Inventory of Additional Supplies and Equipment J

ITEM QUANTITY INVENTORY REMARKS

\\'

Sample Counting 1

Containing:

(1) E-120 Survey Instrument, HP-190 Probe Kit and SN-3 Sample Holder.

(1) Clipboard with pen or pencil I

and paper.

(5) Small plastic bags.

(14) Pair surgeons gloves.

(2) Spare

'D' cell batteries.

(1) Calculator Emergency Locker 1

(12) TLD badges.

(12) Pocket dosimeters 0-500 MREM.

(12) Pocket dosimeters 0-50 R.

(12) Sets of protective clothing.

(6) Full face masks.

(1) Teletector.

(1) Portable Air sampler.

(12) Air sample filters.

(1) Dosimeter charger.

(5) Flashlights.

(10) Flashlight batteries.

(5) Polyethylene bags.

(6) Rolls of masking tape.

(6) Pencils or pens.

(3) Clipboards - paper and survey sheets.

(100) nucon smears.

(100) feet of radiation caution rope.

(5) Radiation area signs.

(5) Grease pencils. Plant and area niaps.(1) Sample Ctg. Kit Emergency Plan and 1

Includes latest revision and is up to date.

I Implementing RERP Copy #107; EPIP Copy #107; P&ID l-Procedures i

R16 11-18-82

ST-RM-3-13 TEST RECORD Section F.1 Fort Calhoun Control Room Respiratory Equipment RESPIRATOR SERIAL TEST TYPE NUMBER INSTRUCTIONS RESULTS REMARKS Pressure Control Check that air cylinders are Demand SCBA Room 1 fally charged (>2000 psi),

facepiece assemblies are Pressure Control totally fun tional & properly Demand SCBA Room 2 stored, harnesses are in good condition & regulators Pressure Control

& warning devices are func-Demand SCBA Room 3 tioning properly.

Initial inspection sticker if found Pressure Control satisfactory.

Demand SCBA Room 4 Pressure Control Demand SCBA Room 5 Pressure Control Demand SCBA Room 6 Full Face Check tightness of connec-N/A Mask tions & condition of face-l piece, headbands, valves, Full Face nnecting tube & canister.

N/A Mask Check rubber or elastoner l

parts to ensure that they Full Face are pliable, flexible &

Mask N/A not deteriorating or taking i

a set during storage.

Ini-Full Face tial inspection sticker, re-Mask N/A bag & seal if found satis-factory.

Full Face Mask N/A Full Face Mask N/A

{fb

" 15' y a i h f,DV 131962 TEST RECORD tra S AUXILIARY BUILDING I

Emergency 1

Check that carrier is in good Personnel condition & with the following Carrier equipment inside:

stretcher, 5 gallon bottle, 6 pairs rubber gloves, 6 pairs shoe covers, HEPA filter & check operation of installed blower.

R16 11-18-82

.c :.z.a

@ g?sT) s,pgr.

ST-RM-3-1L k

TEST RECORD

% gj Section F.1 p

Fort Calhoun Technical Support Center OPERATIONAL FUNCTION CHECK /

SERIAL BATTERY SOURCE TEST CALIBRATION INSTRUMENT NUMBER CHECK USED ACCEPTABLE DUE DATE REMARKS SAM-2 N/A Air Sampler RADECO H-809V N/A h/A l

Eberline 120 VAMP a

i R16 11-18-82

a arr.J h'

hb ST-RM-3-15 o>

.. u I

TEST RECORD

$(d Section F.1 Fort Calhoun Technical Support Center Inventory of Additional Supplies and Equipnent ITEM QUANTITY INVENTORY REMARKS Sample Counting 1

(1) E-120 instrument, HP-190 Probe and SH-3 Kit sample holder.

(5) Small plastic bags. (h) Plastic gloves.

(2) Spare

'D' cell batteries. Pencils, paper and clipboard.

Technical Support 1

(1) Case disposable coveralls.

(1) Box disposable Center Closet surgeons caps.

(1) Box surgeons gloves. (1) Bundle cotton gloves.

(5) Bundles cotton insert gloves.

(5) Packages of plastic shoe covers.

(6) Rolls 3

masking tape.

(25) Disposable hoods. (10) Silver i

zeolite cartridges.

(1) Box Gelman filters.

(1) TI-59 Calculator.

(10) Pressure Demand SCBA's.

(10) Full face respirator. (1) Sanple Counting Kit.

R16 11-18-82

O ST-RM-3-16 TEST RECORD Section F.1 Fort Calhoun Technical Support Center Respiratory Equipment RESPIRATOR SERIAL TEST TYPE NUMBER INVENTORY RECORD REMARKS Pressure Check that air cylinders are Demand SCBA TSC-1 fully charged (>2000 psi),

facepiece assemblies are Pressure totally functional & properly Demand SCBA TSC-2 stored, harnesses are in good condition & regulators Pressure

& varning devices are func-Demand SCBA TSC-3 tioning properly. Initial inspection sticker if found Pressure satisfactory.

Demand SCBA TSC-4 Pressure Demand SCBA TSC-5 Pressure Demand SCBA TSC-6 Pressure Demand SCBA TSC-7 Pressure Demand SCBA TSC-8 Pressure Demand SCBA TSC-9 Pressure Demand SCBA TSC-10 Full Face Check tightness of connections Mask N/A

& condition of facepiece, headbands, valves, connecting Full Face tube & canister. Check rubber Mask N/A or elastomer parts to ensure that they are pliable, flex-Full Face ible & not deteriorating or Mask N/A taking a set during storage.

Initial inspection sticker, Full Face rebag & seal if found sat-Mask N/A isfactory.

Full Face Mask N/A NOV 181302 K16 11-18-82

3 ST-RM-3 17 TEST RECORD Section F.1 Fort Calhoun Technical Support Center Respiratory Equipment RESPIRATOR SERIAL TEST TYPE NUMBER INVENTORY RECORD REMARKS Full Face Check tightness of connections Mask N/A

& conditions of facepiece, headbands, valves, connecting Full Face tube & canister. Check Mask N/A rubber or elastomer parts to encure that they are pliable, Full Face flexible & not deteriorating Mask N/A or taking a set during stor-age.

Initial inspection Full Face sticker, rebag & seal if Mask N/A found satisfactory.

Full Face Mask N/A 9d(ud[Ef 2 ffI E

CU NOV 18 B62 R16 11-18-82

au a'.a

'h M ST-RM-3-18

-C on TEST RECORD

~

Section F.1 B m*

North Omaha Station OPERATIONAL FUNCTI0d CHECK /

SERIAL BATTERY SOURCE TEST CALIBRATION INSTRUMENT NUMBER CHECK USED ACCEPTABLE DUE DATE REMARKS (1)

I Eberline 120 (Sample Counter Kit)

Porcable Air Sampler (Emergency Kit)

N/A N/A High Range Ion Chamber or GM (Emergency Kit)

Dosimeter Charger (Emergency Kit)

N/A N/A N/A Pencil Dosimeter Quantity (0-500 MREM) 12 N/A N/A N/A (Emergency Kit)

Pencil Dosimeter Quantity

(

y, (0-50 REM) 12 N/A N/ *.

N/A tj (Emergency Kit)

!~

TLD Badges Quantity (Emergency Kit) 12 N/A N/A N/A Date Replacement Due (1) Must have conversion factors taped to instrument housing (cpm to micro-curies /ml or dpm) i i

i 1

R16 11-18-82

O O

ST-RM-3-19 TEST RECORD Section F.1 North Omah'a Station Respiratory Equipment RESPIRATOR SERIAL TEST TYPE NU'4BER INSTRUMENT RESULTS REMARKS Full Face Check tightness of connec-Mask N/A tions & conditions of face-piece, headbands, valves, Full Face connecting tube & canister.

Mask N/A Check rubber or elastomer parts to ensure that they Full Face are pliable, flexi' ale &

Mask N/A not deteriorating or taking a set during storage.

Full Face Initial inspection sticker, Mask N/A rebag & seal if found satisfactory.

Full Face Mask N/A Full Face Mask N/A p@

  • 3 r>

NOV 181982 R16 11-18-82

T*9 b $[f5)

ST-RW 20

  • r

~

TEST RECORD Q 83 Section F.1 b

North Omaha Station Inventory of Additional Supplies and Equipment ITEM QUANTITY INVENTORY RD! ARKS Sample Counting 1

Containing:

(1) E-120 Survey instrument with HP-190 Kit Probe and SH-3 sample holder.

(5) Small plastic bags.

(14 ) Pair surgeon gloves.

(2) Spare

'D' cell batteries.

(1) Calculator.

Emergency Locker 1

(12) TLD badges.

(12 Pocket dosimeters 0-500 MREM.

(12) Pocket dosimeters 0-50 R.

'(12) Sets of pro-tective clothing.

(6) Full face masks.

(1) High range Ion chamber or hig range GM.

(1) Air sampler.

(12) Air sample filters.

(1) Dosimeter charger.

(5) Flashlights.

(10) Flashlight batteries.

(5) Polyethylene bags.

(6) Rolls of masking tape.

(6) Pencils or pens.

(3) Clipboards; paper and survey sheets.

(100) Nucon smears.

(100) feet of radiation caution rope.

(5) Radiation area signs.

(5) Grease pencils.

(2) 1 Pint water sampling bottles. Plant and f

area maps, area overlay and map.(1) Sample Counting Kit.

Energency Plan 1

Includes latest revision and is up to date.

and Implementing RERP Copy numbers 13 and 106; EPIP Copy numbers 13 and Procedure 106.

R16 11-18-82

H-15 Table H-1

" Fort Calhoun Station Emergency Kits" C0NTENTS Location Name of Kit Radiological Monitoring Equipment Emergency Supplies Fort Calhoun Sample Counting Kit E-120 Survey Instrument with Clipboard with pen or pencil Station-Control HP-190 Probe with SH-3 and paper (1)

Room Sample Holder (1)

Small Plastic Bags (5)

Surgeon Gloves (4) pair "0" cell batteries (2)

Calculator (1)

Fort Calhoun Emergency Locker TLD Badges (12)

Protective Clothing Sets (12)

Station-Control (1)

Room Pencil Dosimeters, 0-50 R Full Face Masks (6)

(12)

Flashlights (5)

Pencil Dosimeters, 0-500 MREM (12)

Flashlight Batteries (10)

Teletector (1)

Polyethylene Bags (5)

Portable Air Sampler (1)

Rolls of Masking Tape (6)

Air Sample Filters (12)

Plant and Area Maps, Area Overlay and Map Potassium Iodine tablets

( 200 bottles, 14 tablets per bottle)

ISSilED NOV 181982 R1 11-18-82

H-16 Table H-1

" Fort Calhoun Station Emergency Kits" (continued)

C0NTENTS Location Name of Kit Radiological Monitoring Equipment Emergency Supplies Fort Calhoun Emergency Locker Smears Paper (100)

Clipboards w/ paper Station - Control (Continued) and Survey Sheets (3)

Room Dosimeter Charger (1)

Radiation Caution Rope (100 ft)

Radiation Area Signs (5)

Sample Counting Kit (1)

Grease Pencils (5)

Pencils / Pens (6)

RERP and EPIP Copy No. 107 Fort Calhoun N/A Pressure Demand Station Control SCBA's (6)

Room - Located on Floor Behind Control Panel Emergency Offsite Monitor (Each kit contain those items listed)

Clipboard w/ Pencils, Paper Operations Team Kits (2) i Facility - Fort E-120 Survey Instrument Small Plastic Bags (5)

Calhoun Station with HP-190 Probe and SH-3 Sample Holder (1)

Plastic Surgeons Gloves (4 pair)

"D" Cell Batteries (2)

Air Sample Filters Calculator (1) i (Gelman or equivalent) (10)

Smears Paper (30)

Sector Map and Book ISSUED of sample locations Silver Zeolite Cartridges (10)

NOV 181982 R1 11-18-82

H-17 Table H-1

" Fort Calhoun Station Emergency Kits" (Continued)

C0NTENTS Location Name of Kit Rad M ogical Monitoring Equipment Emergency Supplies Fort Calhoun Emergency Vehicles (2)

(each vehicle contains)

Full face respirators (3)

Station SAM II Monitor Detector Shield (1)

Disposable Coveralls (12)

RADEC0 H-809V Air sampler (1)

Plastic shoe covers (12)

Disposable Surgeon caps (12)

Disposable hoods (12)

Roll masking tape (1)

Bottle Thyro-Bloc (3)

Rubber Shoe Covers (3 pair)

Plastic Bags, Large (3)

Plastic Bags, Small (3)

Plastic trash containers (1)

ISSilED NOV 181982 R1 11-18-82

H-18 Table H-1 " Fort Calhoun Station Emergency Kits" (continued)

I C0NTENTS Location Name of Kit Radiological Monitoring Equipment Emergency Supplies Emergency Operations Rescue Squad Survey Instrument (1)

Note pad and Pencil (1)

Facility - Fort Monitor Kit (1)

Calhoun Station Lapel Air Sampler (1)

Pencil Dosimeters, 0-500 MREM (6)

SosimeterCharger(1)

Emergency Operations Room 6, Cabinet Gate Monitor Kits (2)

Facility - Fort Calhoun Station Counter Dosimetry Kit (1)

Rescue Squad Kit (1)

Offsite Monitor Kits (2) t E-120 Survey Instrument (1)

High Range GM or Ion Chamber Silver Zeolite Cartridges (10)

RM-14/15/19 Frisker (1) (For decon team)

SAM-II Iodine Sample Counters (2)

RADEC0 H-809V Air Sampler (2)

ISSUED NOV 181982 R1 11-18-82

H-19 Table H-1 " Fort Calhoun Station Emergency Kits"1 (continued)

~

C0NTENTS Location Name of Kit Radiological Monitoring Equipment Emergency Supplies I

Emergency Room 11 - Book E00/ Recovery Mgr Record Book (1)

Operations Cabinet (1)

Facility -

Cassette Tape Recorder w/

Fort Calhoun three tapes (1)

Station TI-59 Calculator (1)

Complete Set Operating Manuals (Copy No. 50) (1)

RERP and EPIP Copies 12 and 105 Updated Safety Analysis Report (copy No. 38) (1 set)

Thyro-Block tablets (200 btis.)

P[ID(copyNo.189)

Operating License and Tech Spec (copy No. 83) (1 set)

Communications Log (1)

Emergency Room 11 Pressure demand SCBA's (6)

Operations (General Area)

Facility -

Spare SCBA Bottles (4)

Fort Calhoun Station Portable Radio (1)

Plastic trash containers (for decon room) (2)

Chemist bottle (sodium cetrate ED mixing) (for decon room)

, t's NOV 181982 RI 11-18-82

H-20 Table H-1 " Fort Calhoun Station Emergency Kits" (continued)

C0NTENTS Location Name of Kit Radiological Monitoring Equipment Emergency Supplies i

Emergency Room 11 Tag board (1)

Operations (General Area -

Facility -

Continued)

Posted Emergency Call Fort Calhoun List (1)

Station Keys for Emergency i

Team Use Emergency Room 4 RERP and EPIP copies No. 135 Operations (Information (1 set)

Facility -

Specialist)

Fort Calhoun Updated Safety Analysis Report Station (Copy No. 39) (1 set) 4 Safety Systems Description Vol I, II and III (copy 77)

(1 set)

State of Nebraska Emergency Plan (Copy No. 6)

State of Iowa Emergency Plan (Copy No. 6)

Pottawattamie County Emergency Plan (Copy No. 6) f 3

Harrison County Emergency Plan (Copy No. 6)

P and ID (copy 212) (1)

Washington County Emergency Plan (Copy No. 6)

ISSUED NOV 181982 m n_in_n2

H-21 Table H-1 " Fort Calhoun Station Emergency Kits" (continued)

C0NTENTS Location Name of Kit Radiological Monitoring Equipment Emergency Supplies Emergency Personnel Decon Operations Lockers (2)

Towels (24)

Facility - Fort (Items split between Calhoun Station the two lockers)

Soap (6 bars) 4 Hand Brushes (6)

Step Off Pads (6)

Disposable Coveralls (12)

Surgeon Gloves (2 bundles)

Large Plastic Bags (2)

Clipboard, Pad and Pencils Sodium Citrate, 5 lb bottle (4)

Plastic Shoe Covers (50)

Emergency Gate Monitor (Each kit contains items listed)

Clipboard w/ pen or pencil and Operations Kits (2) paper Facility - Fort RM-14/15/19 Frisker with pancake probe Calhoun Station Surgeon Gloves (4 pair)

ISSUED NOV 181982 91 11 1A 99

H-22 Table H-1 " Fort Calhoun Station Emergency Kits" (continued)

C0NTENTS Location Name of Kit Radiological Monitoring Equipment Emergency Supplies Emergency Counter / Dosimetry E-120 Survey Instrument with Clipboard w/ paper, Operations Kit (1)

HP-190 Probe and SH-3 pen / pencils Facility - Fort Sample Holder (1)

Calhoun Station Small Plastic Bags (10)

TLD Badges (12)

"D" Cell Batteries (2)

Pencil Dosimeters (0-500 MREM) (12)

Surgeon Gloves (4 pair)

Pencil Dosimeters (0-50R) (12)

Calculator (1)

Dosimeter Charger (1)

Alternate Sample Counting Kit E-120 Survey Instrument with Clipboards, w/ paper, pens / pencils Emergency (1)

HP-190 Probe and SH-3 sample Holder Operations Small Plastic Bags (5)

Facility-North Omaha Station Surgeon Gloves (4 pair)

"D" Cell Batteries (2)

Calculator (1)

ISSllED NOV 181982 D1 11.1R.99

H-23 Table H-1 " Fort Calhoun Station Emergency Kits" (continued)

CONTENTS Location Name of Kit Radiological Monitoring Equipment Emergency Supplies l

Alternate Emergency Kit TLD Badges (12)

Protective Clothing (12 sets)

Emergency (Roll around)

Operations Pocket Desimeter, 0-500 MREM (12)

Full Face Masks (6)

Facility - North Omaha Station Pocket Dosimeters, 0-50 R (12)

Flashlights (5)

High Range Ion Chamber or GM Survey Spare "D" Cell Batteries (10) i Instrument Polyethylene Bags (5)

Portable Air Sampler (1)

Masking Tape (6 rolls)

Air Sampler Filters (12) i Plant and Area Maps Dosimeter Charger (1)

Area Overlay and Map Smear Papers (100)

Pens / Pencils (6)

Clipboards w/ paper and Survey Sheets (3)

Radiation Barrier Rope (100 feet)

Radiation Area Caution Signs (5)

Grease Pencils (5)

One Pint Water Sample Bottle (1) 3 RERP and EPIP,(Copy No. 13)

RERP and EPIP (Copy No. 106)

IRRilFD NOV 181982

H-24 Table H-1 " Fort Calhoun Station Emergency Kits" (continued)

CONTENTS Location Name of Kit Radiological Monitoring Equipment Emergency Supplies Technical Emergency Locker /

SAM-II Iodine Counter (1)

Coveralls (disposable) (1 case)

Support Center Closet Fort Calhoun RADEC0 H-809V Air Sampler (1)

Disposable Surgeon Cap (1 box)

Station Silver Zeolite Cartridges (10)

Surgeon Gloves (1 box)

Gelman Filters (or equivalent)

Cotton Gloves (1 bundle)

(1 box)

Cotton Insert Gloves (5 bundles)

Plastic Shoe Covers (5 packs)

Rolls Masking Tape (6)

Disposable Hoods (25)

TI-59 Calculator (1)

Pressure Demand SCBA's (10)

Full Face Respirators (10)

Technical Sample Counting E-120 Survey Instrument with Small Plastic Bags (5)

Support Center Kit (1)

HP-190 Probe and SH-3 Sample Holder Fort Calhoun Surgeon Gloves (4 pair)

Station "D" Cell Batteries (2)

Clipboard, pens and pencils Calculator I

ISSllED NOV 181982 R1 11-18-82

H-25 Table H-1 " Fort Calhoun Station Emergency Kits" (continued)

C0NTENTS Location Name of Kit Radiological Monitoring Equipment Emergency Supplies Technical Support Instrument Room Particulate Iodine and Noble Gas Center - Fort (Room 10)

Monitor (PING) (1)

Calhoun Station Detector Shield for SAM-II Counting System (1)

Vamp or equivalent area radiation monitor Auxiliary Emergency Personnel Stretcher (1)

Building - Fort Carrier Calhoun Station 5 Gallon Bottle (1)

Rubber Gloves (5 pair)

Shoe Covers (6 pair)

HEPA Filter ISSUED NOV 181982 R1 11-18-82 J