ML20040F410

From kanterella
Jump to navigation Jump to search
Notice of Violation from Insp on 820127.Noncompliance Noted: Records of Leak Tests Performed Not Maintained & Required Calibr Has Not Been Performed
ML20040F410
Person / Time
Issue date: 02/02/1982
From: Sreniawski D
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
Shared Package
ML20040F404 List:
References
NUDOCS 8202090167
Download: ML20040F410 (1)


Text

_

?

, a Appendix NOTICE OF VIOLATION Rose-Hulman Institute of Technology License No. 13-17582-01 As a result of the inspection conducted on January 27, 1982, and in accordance with the Interim Enforcement Policy, 45 FR 66754 (October 7, 1980), the follow-ing violations were identified:

1.

License Condition 13.A. requires each sealed source containing licensed materialbetested,forleakageatintervalsnottoexceedsixmont)s.

License Condition 13.B. requires records of leak test results be kept in units of microcuries and mainta.ned for inspection by the Commission.

Contrary to the above, a licensee representative stated leak tests are performed, however, not at the required six month interval. Records of leak tests performed were not maintained on the day of inspection except j

the test performed on July 23, 1981', on your 100 millicurie americium-241 sealed source. This record was not maintained in microcurie units.

This is a Severity Level V violation (Supplement VII).

2.

License Condition No. 15 requires that licensed material be possessed and used in accordance with statements, representation, and procedures contained in application dated May 26, 1976, and letters dated January 26, 1973, and March 4, 1980.

Item No. 10 of the above referenced application supplementiary sheet 7

No. 1-states your survey meter will be checked for calibration every two months during periods in which'any sources are in use. This check j

will be done using a 5 millicurie cobalt-60 standard.

Contrary to'the above, a licensee representative stated the above required calibration has not been performed since the inception of this'1TcEnse.

This is a Sever'ity Level V' violation (supplement VII).

Pursuant to the provisions of 10 CFR 2.201, you are required to submit to this office with'in thirty days ~of the date of this Notice a written statement or explanation in reply, including for each item of' noncompliance:

(1) corrective action taken and the results achieved; (2) corrective action to be taken to i-avoid further noncompliance; and (3) the date when full compliance will be achieved. Under the authority of Section 182 of the Atomic Energy Act of 1954, 4

as amended, this response shall be submitted under oath or affirmation. Con-sideration may be given to. extending your response time for good cause shown.

2 a/s2

\\

UL Dated / [

Dr J.

3reniawski, Chief Mal (ri als Radiation Protection j

Section 2 i

8202090167 820203 NMS LIC30 13-17582-01 PDR

.~,

- -,