Notice of Violation from Insp on 811119.Noncompliance Noted: No Linearity Checks Performed on New Dose Calibrator, Victoreen Model 444 Not Calibr as Required & Sealed Source Not Leaked Tested in More than 6-months
ML20038C204
Person / Time
Issue date: 12/04/1981
From: Sreniawski D
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
Shared Package
ML20038C202 List:
References
NUDOCS 8112100261
Download: ML20038C204 (2)


Text

'in

{s.

~

,h Append x NOTICE OF VIOLATION i

Wyan,$t,teGeneralHospital License No. 21-12930-01 i

L

\\

.s

,,As a' result of'the inspection conducted on November 19, 1981, and in sTaccorda. ice with the Interim Enforcement Policy, 45 FR 66754 (October 7,

c. ;

1 sg 1980), th.e following violations were identified:

\\

b License Coridition No. 19 requires that licensed material be possessed s.,

g T

and used in accordance with statements, representations, and procedures s

. ' N, contained in the application dated August 7, 1978.

ms 4

'x Item 10 of this, application states that the procedures described in Appendix D Section 2 of the NRC Medical Licensing Guide, NUREG-0338, Revision 1, will be used for calibration of the dose calibrator.

c (

Appendix D requires that quartecly linearity checks of the dose calibrator be performed and recorded.

Contrary to this requirement, no linearity checks have been performed on your new dose calibrator, put into service in early 1981.

This is a Severity Level V violation (Supplement VII).

2.

License Condition No. 19 requires that licensed material be possessed and used in accordance with statements, representations, and procedures contained ~1n the application dated August 7, 1978.

Item 10lof this application states that survey instruments will be calibrated at least annually and following repair.

Contrary to this requirement, your Victoreen Model 444, Serial No.

f389, was last calibrated on May 23, 1979, a period of excess of one

, gear.

6 T1.is is a Severity Level V violation (Supplement VII).

3.

10 CFR 35.14(e) requires that sealed calibration or reference sources possessed pursuant to 10 CFR 35.14(d) be tested for leakage and/or contamination at intervals not to exceed six months.

Contrary to this requirement, as of the day of the inspection, you have not leak tested your nominal 214 microcurie cesium-137 sealed reference source since April 1980, a interval of more than six months.

This is a Severity Level V violation (Supplement VII).

8112100261 811207^

NMS LIC30 21-12930-01 PDR'

- Appendix 2-Pursuant to the provisions of 10 CFR 2.201, you are required to submit to this office within thirty days of the date of this Notice a written state-ment or explanation in reply,. including for each item of noncompliance:

(1) corrective action taken and the results achieved; (2) corrective action to be taken to avoid further noncompliance; and (3) the date when full com-pliance will be achieved. Under the authority of Section 182 of the Atomic-Energy Act of 1954, as amended, this response shall be submitted under oath or affirmation. Consideration may be given to extending your response time for good cause shown.

(Ry1/SI b

Dated D. J.

r4niawski, Chief Materi s Radiation Protection Section 2 i