ML20029A683

From kanterella
Jump to navigation Jump to search
Part 21 Rept Re Dimensional Defect in Front End Plate in Amersham Model 920 Camera.Guide Tube May Become Disconnected from Camera While Source in Exposed Position.Initially Reported on 910107.All Cameras Will Be Inspected
ML20029A683
Person / Time
Site: San Onofre, Ginna  Southern California Edison icon.png
Issue date: 01/17/1991
From: Mcdaniel W
AMERSHAM CORP.
To: Jason White
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION I)
References
REF-PT21-91 NUDOCS 9103010216
Download: ML20029A683 (8)


Text

.

f JAN 28 '9.1 13J 07 NRC RI DRSS P02 Amersham Corporation 60 Nortr) Avenue Burlington, Massachu6ette 01603 Telephone (617) 212 2000 January 17, 1991 Mr. Juan 1 nite, Chief Nucinar l'aterials Safety Section C Divisien of Midiation Safety and Safeguartin U.S. H ut:1 it Angulatory Commission Region 1 475 A11endt'a 30s.d King of Prun.*.it, PA 194 0ti ame r Mr. W(v to t

s let,tcr is the formal written notificat4?n of the Part 21 defect reported to you by telephone on January h 1991.

The defect involves a component used in the mam rr':ture of the Amorsham Model 920 camera. The component involved

.s the front end pintu

  • hat incorporates the male " quick disconnw t" fitting (Aneromm 9
    • h Number 90001-4 ). This fitting is for unnecting tM gu.4 ja tuta. and related accessories.

A dimensienal 0. Rect -in the male fitting on a sLall nurthu of 920 cameras may pri.3 unt a proper secura connection of the su 4.de Wbe.

Thow is a chan 101 on the male fitting th:t, engages the beatings on the quick dlaconnect to allow a secu'/e connection. This chan-nel is not properly located on some of the defective parts. Due to this it is possible thrat the guide tube could become discon-nected from the camera while the source is lit the exposed posi-tion. We feel that this 14 unlikely, but is still possible.

We d isW'</. red this proh'lan as a _ result of investigating a 920 d Wloe 4 &'h c00 sed a euutom3r to have problems connecting their guide tub 4L. This was not, related to any radiological incident.

This part histei ically ad been manufactured from an item brought in "of f-the-shel " and then modified. Our drawings only included d: stalled specif.W ons for the portions we modified.

The way this.part is manu 44)ared was changed in December 1987. At that (K.

time we arranged for a local machine shop to manufacture the part f ow, raw stock. The ora'ing we sup catiens for the channel location. plied did not include specifi-We have since tightened our controls-in Engineering so that this type of problem is not likely to re-occur.

t

p s uhSp AmerShd.h 5

JAN 28 '9,1 13108 HRC R1 DRSS P03

(

4 Page 2 of 3 The first shipment of parts manut.:mtured using this method were roccivee. oec.mber 29, 1982. We havo manuf actured sixt.en (1 3 920 cameras since that time. The serial numbers possibly affected are 51 through 66. Thirtoon of these have been obipped to domontic customora, two to Amersham International in the U.K.,

and one remains in stock (Serial / 66). All domestic customorn have been identified.

In addition wo are checking our service records to datormine if any cameras have boon servicod to include replacement of this part since January 1988.

Corrective actions are as follows:

1.

Contact-all customers and distributorn who havo re-coivod che cameras in question. Tolophone contact was initie'4ed by Amorcham on January 16, 1991. One customer was contacted on January 15, 1991 by NRC Region 4.

2.

Follow t.p on thra telephone contact in writing.

/

1.

Fvartne all camerne that may pessmen the daf6e. Liv.

comptnent. Thin may be accomplished in one of two waynt We have designau and manufactured a Co-/No-Go gauge that will be dont to our Hounton and Baton Rougo service centers, au well as all of our industrias products Fiold Salen Representatives. Thia gauge can be used by Amersham personnel to check cameras in the field or at the service contors.

The cameras can also be returned to the Burlington facility to be inspected.

4.

All cameran that possean the defective component will be nood to be returned to the Durlington facility for replacement of the fitting.

We will koop records of the telephone contacto, units checked, and units requiring replacement of a defectivo component.

x l

JAN 2891 13:08 NRC RI DRSS PO4

,[

t

  • g -

Page 3 of 3 Attached are the followingt Telephone script used to contact customers.

Follow-up letter.

Listing of serial numbers and locations these units were shipped to.

Front Plug Gauge Diagram If you have any questions regarding this notification please give me a call'.

Sincerely, William M. McDaniel Facility Manager

- Attachments - 4 s

Send By FAX and Mail.

Copy To:

C.Roughan; Q.A. Manager /R.S.O.

L.Podolak; Radiation Safety Specialist J.Buchenberger; Radiography Products Marketing-Manager

\\.

I s

__m____._.m_

JAN 28 ?91 13,t09 NRC RI DRSS P05 Attachment _L _

TELEPHONE SCRIPT i

customer:

Date called We are calling you to inform you of a possible component defect on a small number of the Model 920 exposure devices. Duc to the nature of the problem, we have classified this as a Part 21 defect and the NRC has been informed.

The defect may prevent a secure connection when the guide tube or related accessory is connected to the front of the Model 920. In most cases the connection is secure. In nosu cases the connection is secure. However, it is possible, but ver/ 'anlikely, that the connection may not be properly secured and the guide tube-could-become disconnected from the 920 while the soutte is within the guide tube, only a small' group of 920 cameras may be affected by this. The serial numbers of the devices in your possession that need to be inspected are:

Serial. Number (a):

We are also checking our files to see if any cameras you have sent in for repair need to be checked.

As a corrective action we need to arrange for one of the follow-ing:

- We can arrange for the return of the device (a) in your possession that need to be checked to cur Burling-ton Ma. service conter. We will inspect the device for this defect, and if the defect is found, the component will be replaced.

or

- We can have your regional Sales Representative visit

-your facility and they can.do a quick and simplo check to determine if your unit has the defective conponent.

If the component is found to be defective, it will have to be returned to our Burlington, Ma. service center for repair.

We will make any transport arrangemente for you. Your camera Will be returned promptly. If you have any special requirements we will try-to accommodate them.

Please accept our apologies for any inconvenience this may have

-caused for you. Safety le our primary concern and we want to resolve this situation as quickly as possible.

'JAN 28 '9,1 13,810 NRC RI DRSS P06 January 17, 1991

Dear (CUSTOMER),

This letter is a follow up to,the notification we made by tele-phone on January 16 and 17, 1991 regarding the Part 21 Notifica-tion for the male fitting on the front of a small number the Amersham Model 920 exposure devices. The NRC has been appropri-ately intornud.

The defect may prevent a nacure connection when the guide tube or related accessory is connectea to tne Iront or une Mocal vzu. In most cases the connection is secure, llowever, it is possible, but very unlikely, that the connection may not be properly secured and the guide tube could become disconnected from the 920 while the source is within the guide tube.

Only a small group of 920 cameras may be affected by thin. The serial numbers of the devicea in your possosalon that need to be inspected are:

Seria1 Number As a corrective action we need to arrange for one of the follow-ing:

- We can arrange for the return of the device (s) in your possession that need to be checked to our Burling-ton Ma. service center. We will inspect the device for this defect, and if the defect is found, the component will be replaced, or

- We can have your regional Sales Rspresentative visit your facility and they can do a quick and einple check to determine if your unit has the defective component.

If the component is found to be defective, it will nave to be returned to our Burlington, Ma. service center

.for repair.

If your camera is going to be returned to Burlington, MA for in-spection and/or repair, our custome.r Service department will make the arratagements for you. We guarantee to return your camera promptly. If you have any special requirements, please inform the customer service representative when they contact you.

i JAN 28 '91 13:11 NRC-RI DRSS P07

... i.

. '(cont'd)

Please accept our' apologies for any inconvenience this may have caused for you. Safety is our primary concern and we want to resolve this-situation as quickly as possible.

If you have' any. questions please feel free to contact our customer Service Supervisor, Jeanne Price, at 1-800-225-1383, extension *222, or myself.

William M. McDaniel Facility Manager

I

-J A N ' 2 8 ' 9,1 ' 1 3,8 1 1 NRC RI DRSS P08 il i

-+

f--

5 Serial #

Customer 51',

52, 53 Western Stress Texas, USA 54 Q.N.D.E.

Canada 55 Western Stress Texas, USA 56, 57 J.A. Jones Colorado, USA 58 E.B.A.S.C.O.

^

New Jersey, USA

[

_1'u l \\1 59 Rochester Gas and Electric New York, USA 60 Q.N.D.E.

Canada 61 Rochester Gas and Electric New York, USA J O h' 62 Southern California Edison

~

California, USA 63, 64 Amersham International United Kingdom 65 Rockwell/E.G. & G.

Colorado, USA i-I' l

i L

i

I, JAN 28 '9,1 13,812 NRC RI DRSS P09 REV ENGINEER DATE DESCRIPTION At t achroe nt 4 g MARK " MIN" TOP SURFACE

-*-- 155 1.002 W 716 i.002 THRU C' SINK 45' X W.81 y,06

~

DO NOT CHAM. EDGES THIS SIDE L

4 74

'~

.210 t.002

~

MARK " MAX" TOP SURFACE MIN GAGE THIS EDGE MUST DE WITHIN MIN & h%X OF GAGE

_f 900 SERIES FRONT CONN.

~

v-N A,

MAX USED ON:

RELEASED FOR PRODUCTION ON BY MATERIALS:

AMERSHAM ' CORPORAT10N Amersham 1" HEX STEEL BURLINGTON, MA 01803 FINISH DWO. TITLE BLACK OXIDE A"

/ "

MM=

FRONT PLUG GAGE Pfttp#tED C.PARCONS J J__,x go,j l

O CHEUt G. PARSONS )f

.XX

  • 0.01 cHEcean 0.0VNCAN$0N APPR

.XXX 1 0.005 cussincAnoH SITE DWG. NO.

REV ff/fo ANGLES

  • 1-A T900-11 di

' vo h FRACT i1/64 SCALE 1:1 l SHEET

  • _

1_

OF 1

/