.s i.
l Appendix A NOTICE OF VIOLATION Shores Nuclear and Ultrasound Diagnostics License No. 21-17223-01 As a result of the inspection conducted on May 12, 1981, and in accordance with ter Interim Enforcement Policy, 45 FR 66754 (October 7,1980), the
]
following violation was identified:
10 CFR 35.14(e)(1)(i) states any licensee who possesses sealed sources as calibration or reference standards, which are greater than 100 microcuries, shall not use the sources in the absence of a certificate from the transferor indicating that a leak test has been made within six months prior to the transfer.
Contrary to the above, a 250 microcurie cesium-137 source and a 250 microcurie barium-133 source were placed into use on February 2,1981, without the leak test certit?. cates.
{
This is a Severity Level V violation (Supplement VII).
Pursuant to the provisions of 10 CFR 2.201, you are required to submit to this office within twenty-five days of the date of this Notice a written statement or explanation in reply, including for each item of noncompliance:
(1) corrective action taker and the results achieved; (2) corrective action to be taken to avoid further noncompliance; and (3) the date when full com-pliance will be achieved. Under the authority of Section 182 of the Atomic Energy Act of 1954, as amended, this response shall be submitted under oath or af firmation.
s/asler D \\. fQ Date D. J. frpniawski, Chief Materi m Radiation Protection Section 2 l
l 4
8107130389 810601 NMS LIC30 21-17223-01 PDR p.v-pre--.
,-.--4<.4m---,%w-m-.-----
---e-m