Notice of Violation from Insp on 810605.Noncompliance Noted: Daily Constancy Checks of Dose Calibrator on Cs-137 Setting & Wipe Tests of Final Source Containers Not Performed
ML20005A803
Person / Time
Issue date: 06/22/1981
From: Wiedeman D
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
Shared Package
ML20005A802 List:
References
NUDOCS 8107010281
Download: ML20005A803 (2)


Text

s a

~

4

~

Appendix A NOTICE OF VIOLATION Lincoln County Memorial Hospital License No. 24-18689-01 As a result of the inspection conducted on June 5, 1981, and in accordance with the Interim Enforcement Policy, 45 FR 66754 (October 7, 1980), the following violations were identified:

1.

License Condition No. 16 requires that licensed material be possessed and used in accordance with statements, representations, and procedures contained in application dated March 28, 1979.

License application dated Barch 28, 1979, states that the procedures described in Appendix D, Section 2, of the NRC Draft Medical Licensing Guide dated November 1977, shall be followed for calibration of the dose calibrator. Appendix D, Section 2 states that daily constancy checks of the dose calibrator shall be performed on the cesium-137 setting.

Results of the quarterly linearity checks of the dose calibrator should be within i 5% of the predicted values.

Errors greater than i 5% require repair or adjustment of the dose calibrator.

A geocetrical variation check of the dose calibrator shall be performed at installation.

Contrary to these requirements, it was learned through statements of licensee representatives and a revi2w of records that daily constancy checks of the dose calibrator on the cesium-137 setting was not performed since the date of license issuance.

The resultc of quarterly linearity checks of che dose calibrator varied more than i 5% of the predicted values and repair or adjustment of the dose calibrator was not performed. A geometrical variation of the dose calibrator has not been performed since the date of license issuance.

This is a Severity Level V violation (Supplement VII).

2.

Licance Condition No. 16 requires that licensed material be possessed and used in accordance with statements, representations, and procedures contained in application dated March 28, 1979.

License application dated March 28,'1979, states that the procedures described in Appendix F of the NRC Draft Medical Licensing Guide dated November 1977, will be followed for opening packages containing radioactive materials. Appendix F states that a wipe test of the external surface of the final source container shall be performed andtheresu{_tsrecorded.

8107010281 810626 NMS LIC30 24-18689-01 PDR

4.

' h..

Appendix A

.- 2:-

Contrary to this requirement, it was learned through statements of licensee representatives that a wipe test of the external surface.

of the final source containers was not performed since the date of license issuance.

This-is a Severity Level V violation (Supplement VII).

. Pursuant to the provisicns of 10 CFR 2.201, you are required to submit to this office within thirty days of the date of this Notice a written state-ment or explanation in reply, including for each item of noncompliance:

(1) corrective action taken and the results achieved; (2) corrective action to be taken to avoid further noncompliance; and (3) the date when full com-pliance will be achieved. Under the authority of Section 182 of the Atomic Energy Act of 1954, as amended, this response shall be submitted under oath or affirmation. Consideration may be given to extending your response time for good cause shown.

0N b

Dated /

/

D. G.' Wiedeman, Acting Chief Materials Radiation Protection Section 1 8

%