ML19295B384
| ML19295B384 | |
| Person / Time | |
|---|---|
| Issue date: | 09/11/1980 |
| From: | James O'Reilly NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION II) |
| To: | Union Boiler Co. |
| References | |
| NUDOCS 8010080112 | |
| Download: ML19295B384 (1) | |
Text
v' pn anou
+
o UNITED STATES
~,'n NUCLEAR REGULATORY COMMISSION
,E REGION 11 0
101 MAR f ETTA ST, N.W., sUlTE 3100 o
ATLANTA, GEORGI A 30303 In Reply Refer To:
RII:JP0 UNION BOILER COMPANY HUNTINGTON TESTING DIVISION P,0, Box 1018 HUNTINGTON
_WV 25717 NRC LIC, 88 182-01 Gentlemen:
The enclosed Bulletin 80-22 is sent to you for action.
If there are any questions related to the requested actions, please contact this office.
Sincerely,
\\ _c o-James P. O'Reilly irector
Enclosures:
1.
IE Bulletin No. 80-22 w/ enc 1 2.
List of Recently issued IE Bulletins s
t-8 01 0 08 o l.I 1
SSINS No:
6820 Accession No.:
8006190050 IEB 80-22 UNITED STATES NUCLEAR REGULATORY COMMISSION OFFICE OF INSPECTION AND ENFORCEMENT WASHINGTON, D.C.
20555 D
b September 11, 1980 IE Bulletin No. 80-22: AUTOMATION INDUSTRIES, MODEL 200-520-008 SEALED-SOURCE CONNECTORS Description of Circumstances:
An accident recently occurred in the State of California (an Agreement State) that resulted in several serious overexposures (see enclosed Circular 79-16).
The accident involved the use of an Automation Industries (AI) Model 200-520-008 source pigtail to drive cable connection. The connector is identified by the manufacturer as its " quick disconnect" type of connector and is used with several models of cobalt-60 and iridium-192 source assemblies distributed by the manufacturer. The connection is a simple hook and eye connection (see Fig.
- 1) that will permit separation of the eye from the hook when they are oriented at right angles to each other without any further positive action. As a result of the design, a disconnection may occur any time the source assembly is cranked out of the exposure device without the guide tube being attached.
Since the Automation Industries Model 200-520-008 sealed-source assembly is authorized to contain up to 120 curies of iridium-192, a source disconnection creates a potential for a serious exposure.
A disconnect is very unlikely to occur if proper procedures are followed by users of Automation Industries sealed sources containing the " quick disconnect".
However, due to the large number of overexposures that occur in the radio-graphy industry due to the failure to follow proper procedures, the NRC has determined that continued use of the Automation Industries " quick disconnect" connector with its Model 200-520-008 sealed-source assembly or other assemblics can constitute a hazard to the public health, safety or interest.
As a result, the NRC has ordered Automation Industries to discontinue distribution of the AI Model 200-520-008 sealed-source assembly or any other sealed-source assembly that uses the " quick disconnect" type of connector.
Additionally, the NRC will not accept new applications for authorization to use the AI Model 200-520-008 or other sealed-source assemblies using the
" quick disconnect".
In effect, this will remove such iridium-192 source and pigtail assemblies from service as licensees exchange their diminished activity sources. Special arrangements should be made for removing from service, those connectors that are attached to longer lived sources such as cobalt-60.
The removal from service of the " quick disconnect" type of connector will require that those drive cables having the matching open book connector be modified.
Actions To Be Taken by Licensees To assure the safe operation of radiogriphy equipment in which the AI " quick disconnect" connector is used, all licensees authorized to use byproduct materials under 10 CFR Part 34 shall perform the following:
September 11, 1980 Page 2 of 2 (1) Determine the number of AI Model 200-520-008 and other assemblies in your inventory that have the " quick disconnect" connector and establish a disposal schedule. Your schedule for iridium-192 sources should be no longer than your usual exchange schedule for diminished activity sources.
Your exchange schedule for connector modification for cobalt-60 sources should be accomplished as soon as is conveniently possible. However, you should complete your disposal or modification by September 1, 1981.
(2)
In the interim, a warning mark or tag should be placed on each radio-graphic device in which AI " quick disconnect" type assemblies are being i
used.
(3) This matter should be reviewed with all radiographers. They should be cautioned of the disconnect possibility and be informed of the method used to identify equipment as marked in accordance with Item (2) above.
(4) Review and modify operating, maintenance, inspection and handling procedures as appropriate to preclude this type of accidental disconnect.
NOTE: You may want to contact the manufacturer for assistance in determining appropriate procedural changes or fixes to preclus'e a disconnect.
(5) Discuss with the radiographers the importance of making adequate surveys and the need for following procedures.
A report of your actions for each of the above numbered items sh ll be submitted to the director of the appropriate regional office within 30 days from the date of this Bulletin. A copy of the report shall also be sent to the Director, Division of Fuel Facility and Materials Safety Inspection, Office of Inspection and Enforcement, U.S. Nuclear Regulatory Commission, Washington, D.C.
20555.
Approved by GAO, B180225 (R0072); approval expired July 31, 1980.
(Application for renewal pending before GAO). Approval was given under a blanket clearance specifically for identified generic problems.
Enclosures:
1.
IE Circular No. 79-16 2.
Figure 1
Accesuuns No.
7903020542 S5145:6830 UNITED STATES NUCLEAR REGULATORY COMMISSION 3
m OFFICE OF INSPECTION AND ENFORCEMENT WASHINGTON, D.C.
20555 ua
- 2. 5 August 16, 1979 IE Circular No. 79-16 EXCESSIVE RADIATION EXPOSURES TO MEMBERS OF THE GENERAL PUBLIC AND A RAD 10GRAPHER Description of Circumstances:
During radiographic operations using 40 curies of iridium-192, the source became disconnected unbeknownst to the radiographer--he did not use his survey instrument. After the radiographer left the facility, an employee of the customer for which radiography was perfonned, saw the source and, not knowing what it was, picked it up and placed it in his hip pocket. He carried it about for approximately two hours, later giving it to his super-visor to examine. While making a determination that it was something which belonged to the radiographer, and while waiting for the radiographer to pick up the source, nine employees of the radiographer's customer were exposed. The source was also left with a secretary who was instructed to contact the radiographer.
The radiographer returned, examined and took the source assurring the customer's employees that there was no problem, stating that the source was a " detector".
On the evening of the event, the employee who had put the source in his pocket became nauseous and went to a hospital for treatment. At that time a blister was found on his buttock. The initial diagnosis and treatment was for an insect bite. Thirty one days after this initial treatment the indi-vidual was hospitalized for treatment of the injury to his buttock. At that time the individual asked the physicians if there could be any connection of the injury to the radiography that had been performed at his place of work one month previously. An investigation followed which disclosed the above information.
The individual who had carried the source in his pocket remains under medical care following surgery. The attending physician does not consider the exposure to be life threatening. Neither does amputation appear necessary.
The localized dose is estimated to be 1.5 millicn rem at skin surface, 60,000 rem at 1 cm depth and 7,000 rem at 3 cm depth.
Estimated whole body doses to other individuals ranged from 1 to 60 rem.
Hand doses ranged to 5,000 rem. The radiographer received estimated doses of 14 rem to the whole body and 50 rem to the hands.
These are serious radiation overexposures. However, another important aspect of the case, second only to the physiological effects of the exposures is the radiographer's apparent disregard for the health and safety of the exposed individuals and for his own personal safety. The radiographer's failure to inform the involved individuals and to report the event to responsible management within his own and the customer's company is a serious disregard for safety and denied the exposed individuals early medical attention.
August 16, 1979 1 Circular No. /9-16 rage 2 of 3 e
diation The priinry cause of this incident was the failure to perfonn a ra in the radiography survey, a comon underlying cause in radiationi overexposures h
The number of radiation overexposures experienced in the radiogra h
ingle industry over the past several years has been higher than for any ot er To inform radiography licensees of HRC's concern for industry.
tatives met with a
these recurring overexposure incidents, NRC staff represen group of NRC licensees.
i d December licensees in a series of five regional meetings during the per oThe ma 1977 throught Nrch 1978.
NRC's concern for the high incidence of overexposures, anc ffort to achieve the common goal of improved radiation safety.those lic Meeting on Radiation Safety for Industrial Radiographers".
i which sent matted to each NRC radiography licensee and to other compan es representatives to the meetings.
tings The remarks presented by the staff and subjects discussed at those me into radiography included, among others, ways and means of incorporating safotyidents, wit operations, and case histories of overexposure incIn a discussion of the causes o the causes and possible preventions.
exposures, a presentation of statistics at the meetings diographic d
tal and exposure was by f ar the most prevalent cause.
by regulation in 10 CFR 34.43(b), they are also the most basic, fu inherent in a typical l
common-sense thing to do when dealing with radiation leve sFailure to h
treme importance ing intensive enough to permanently instill in radiographers t e ex radiography operation.
l Some of the case of surveys for protection of both themselves and other peop e.
injury to hands histories discussed in NUREG-0495 resulted in pai s
Notice to Radiography Licensees:
10 CFR Part 34 are NRC licensees authorized to use byproduct material under requested to take the following actions:
histories Review the event described in this Circular (and t l date; 1.
discuss and emphasize:
i tection the extreme importance of radiation surveys in assur ng pro of themselves and of other people, and a,
l events or circumstances the importance of reporting promptly any unusua b.
to responsible management.
is placed Review your training to assure that appropriate e ii courses.
2.
a 6
IE Circular No. 79-16 August 16, 1979 Page 3 of 3 Review your internal audit program to assure that appropriate emphasis 3.
is placed on these sane subjects, particularly the requirement for auditors to observe radiographic operations to assure the proper coadu;t of radiation surveys.
i No written response to this Circular is required.
If you need additional information regarding this subject, please contact the Director of'the appropriate NRC Regional Office.
o O
9
.r
/
A k\\\\\\
m) 2
- * ' " "' ' ' ^" "=3 27 TOP VI E W co m oi m ie f
s__-
l c ;_
~
\\\\\\\\
'N SIDE VIEW
~
&' l
%b
$ l.__]
-q 1
DI SEN G A GE M ENT T
POSSIBLE j
u F1CrL1R E L
I IEB 80-22 September 11, 1980 RECENTLY ISSUED IE BULLETINS Bulletin No.
Subject Date Issued Issued To 80-22 Automation Industries, Model 9/11/80 All radiography 200-520-008 Scaled-Source licensees Connectors 79-26 Boron loss from BWR 8/29/80 All BWR power Revision I control blades facilities with an OL 80-20 Failures of Westinghouse 7/31/80 To each nuclear Type W-2 Spring Return power facility in to Neutral Control Switches your region having an OL or a CP 80-19 Failures of Mercury-7/31/80 All nuclear power Wetted Matrix Relays in facilities having Reactor Protective Systems either an OL or a CP of Operating Nuclear Power Plants. Designed by Combus-tion Engineering 80-18 Maintenance of Adequate 7/24/80 All PWR power reactor Minimum Flow Thru Centrifugal facilities holding OLs Charging Pumps Following and to those PWRs Secondary Side High Energy nearing licensing Line Rupture Supplement 2 Failures Revealed by 7/22/80 All BWR power reactor to 80-17 Testing Subsequent to facilities holding OLs Failure of Control Rods to Insert During a Scram at a BWR Supplement 1 Failure of Control Rods 7/18/80 All BWR power reactor to 80-17 to Insert During a Scram facilities holding OLs at a BWR 80-17 Failure of Control Rods 7/3/80 All BWR power reactor to Insert During a Scram facilities holding OLs at a BWR 80-16 Potential Misapplication of 6/27/80 All Power Reactor Rosemount Inc., Models 1151 Facilities with an and 1152 Pressure Transmitters OL or a CP with Either "A" or "D" Output Codes 80-15 Possible Loss Of Hotline 6/18/80 All nuclear facilities With Loss Of Off-Site Power holding OLs