JAN 16195U APPENDIX A NOTICE OF VIOLATION Pharmatopes, Inc.
License No. 45-18378-01MD Based on the NRC inspection December 6, 1979, certain of your activities were apparently not conducted in full compliance with KRC requirements as indicated below. These items have been categorized as described in correspondence to you dated December 31, 1974.
A.
As required in part by Condition 20 of your license and your procedures dated December 18, 1979, linearity tests must be conducted on your dose calibrators every three months.
Contrary to the above, linearity tests were not conducted on your dose cali-brators between Apriit 1979 and November 1979.
This is an infraction.
B.
As required in part by 20.401(b), each licensee shall maintain records of disposal surveys that are made under 20.201(b).
Contrary tb the above, ret ids of radiation surveys perfnrmed on bags of syringes at the time of disposal have not been recorded.
This is a deficiency.
8008190 y