Official Transcript of Proceedings NUCLEAR REGULATORY COMMISSION
Title:
Draft Report of the ACMUI Training and Experience Requirements for All Modalities Subcommittee Docket Number: N/A Location: Teleconference Date: February 26, 2019 Work Order No.: NRC-0148 Pages 1-73 NEAL R. GROSS AND CO., INC.
Court Reporters and Transcribers 1323 Rhode Island Avenue, N.W.
Washington, D.C. 20005 (202) 234-4433
1 UNITED STATES OF AMERICA NUCLEAR REGULATORY COMMISSION
+ + + + +
ADVISORY COMMITTEE ON THE MEDICAL USES OF ISOTOPES
+ + + + +
TELECONFERENCE
+ + + + +
TUESDAY FEBRUARY 26, 2019
+ + + + +
The meeting convened by teleconference at 10:00 a.m., Christopher J. Palestro, M.D., ACMUI Chairman, presiding.
MEMBERS PRESENT:
CHRISTOPHER J. PALESTRO, M.D., ACMUI Chairman; Nuclear Medicine Physician DARLENE F. METTER, M.D., ACMUI Vice Chairman; Diagnostic Radiologist VASKEN DILSIZIAN, M.D., Nuclear Cardiologist RONALD D. ENNIS, M.D., Radiation Oncologist (Brachytherapy)
RICHARD L. GREEN, Nuclear Pharmacist MELISSA C. MARTIN, Nuclear Medicine Physicist MICHAEL D. OHARA, Ph.D., FDA Representative ZOUBIR OUHIB, Therapy Medical Physicist NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
2 A. ROBERT SCHLEIPMAN, Ph.D., Health Care Administrator MICHAEL SHEETZ, Radiation Safety Officer MEGAN L. SHOBER, Agreement State Representative LAURA M. WEIL, Patients Rights Advocate NON-VOTING MEMBER PRESENT:
HARVEY B. WOLKOV, M.D., Radiation Oncologist (GSR)
STAFF PRESENT:
ANDREA KOCK, Director, Division of Materials Safety, Security, States, and Tribal Programs (MSST)
CHRISTIAN EINBERG, ACMUI Designated Federal Officer (DFO), Chief, NMSS/MSST/MSEB SOPHIE HOLIDAY, ACMUI DFO KELLEE JAMERSON, ACMUI DFO; ACMUI Coordinator MARYANN AYOADE, NMSS/MSST/MSEB SAID DAIBES, Ph.D., NMSS/MSST/MSEB LISA DIMMICK, Medical Radiation Safety Team Leader, NMSS/MSST/MSEB SARA FORSTER, R-III/DNMS/MLB ROBERT GALLAGHAR, R-I/DNMS/MLAB EDWARD HARVEY, R-III/DNMS/MIB ESTHER HOUSEMAN, OGC/GCLR/RMR DONNA-BETH HOWE, Ph.D., NMSS/MSST/MSEB NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
3 IAN IRVIN, OGC/GCLR/RMR DONNA JANDA, R-I/DNMS/MLAB SARAH LOPAS, NMSS/MSST/MSEB KATHY MODES, NMSS/MSST/ASPB JANICE NGUYEN, R-I/DNMS/MLAB PATTY PELKE, R-III/DNMS/MLB ZAHID SULAIMAN, R-III/DNMS/MIB KATHERINE TAPP, Ph.D., NMSS/MSST/MSEB MEMBERS OF THE PUBLIC PRESENT:
MICHAEL BAXTER, American Pharmacists Association KENDALL BERRY, Fox Chase Cancer Center JANET BUKOVCAN, British Technology Group (BTG)
MARY BURKHART, Illinois Emergency Management Agency (IEMA)
WILLIAM CHEN, unaffiliated JOHN CHIPPO, Pennsylvania Department of Environmental Protection (PDEP)
THOMAS CONLEY, University of Kansas Medical Center WHITNEY COX, IEMA DAVID CROWLEY, North Carolina Department of Health and Human Services, Radiation Protection Section ARIEL DOUCET, Virtua Health BRIAN ERASMUS, BTG NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
4 LYNNE A. FAIROBENT, unaffiliated SHERRIE FLAHERTY, Minnesota Radioactive Materials Unit MIKE FULLER, Virginia Department of Health (VDH)
SANDY GABRIEL, unaffiliated WENDY GALBRAITH, University of Oklahoma Health Sciences Center BENNETT GREENSPAN, M.D., Medical College of Georgia, Augusta University MIGUEL de la GUARDIA, Cook Childrens Medical Center MICHAEL GUASTELLA, Council on Radionuclides and Radiopharmaceuticals, Inc. (CORAR)
STANLEY HAMPTON, Eli Lilly DAN HILL, Cardinal Health DANIEL JANUSESKI, Virtua Health TRACY JUE, California Department of Public Health SUE LANGHORST, Ph.D., unaffiliated RALPH LIETO, St. Joseph Mercy Health System CINDI LUCKETT-GILBERT, Shertech Pharmacy CAROL MARCUS, Ph.D, M.D., University of California at Los Angeles (UCLA)
RICHARD MARTIN, American Association of Physicists in Medicine (AAPM)
SAMUEL MEHR, M.D., Nebraska Cancer Specialists ASHLEY MISHOE, University of California, San NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
5 Francisco MARY MOORE, VDH JOSHUA MYERS, PDEP CHRISTOPHER OTT, PDEP BRANDON PATERSON, unaffiliated RICHARD PEROS, New Jersey Department of Environmental Protection (NJDEP)
MICHAEL PETERS, American College of Radiology (ACR)
CARMINE PLOTT, Novant Health ARIA RAZMARIA, M.D., UCLA Medical Center GLORIA ROMANELLI, ACR GEORGE SEGALL, M.D., American Board of Nuclear Medicine (ABNM)
BEN SEIBER, PDEP BETH SHELTON, Tennessee Department of Environment and Conservation JEFFRY SIEGEL, Ph.D., Nuclear Physics Enterprises DANIEL STROHMEYER, unaffiliated CINDY TOMLINSON, American Society of Radiation Oncology (ASTRO)
MICHAEL UJHELYI, BTG PAUL WALLNER, M.D., 21st Century Oncology, Inc.
MATTHEW WILLIAMSON, Memorial Sloan Kettering Cancer Center MELONIE WISSING, VDH NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
6 JOHN WITKOWSKI, United Pharmacy Partners (UPPI) 1 NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
7 1 P-R-O-C-E-E-D-I-N-G-S 2 (10:15 a.m.)
3 MR. EINBERG: Thank you. This is Chris 4 Einberg. I am the Branch Chief of the Medical Safety 5 and Events Assessment Branch, and I wanted to 6 apologize for the technical difficulties that we have 7 been having right now getting the webinar running.
8 As Sophie mentioned, the slides will be 9 available on the ACMUI public website. So were going 10 to try to get the webinar running, but if you cannot 11 access the -- or if we cant get the webinar running, 12 then please access the slides from the -- from the 13 public SharePoint -- or from the public website.
14 So I am going to start with the opening 15 remarks here. As the Designated Federal Officer for 16 this meeting, I am pleased to welcome you to the 17 public meeting of the Advisory Committee on the 18 Medical Uses of Isotopes.
19 Once again, my name is Chris Einberg. Im 20 the Branch Chief of the Medical Safety and Events 21 Assessment Branch, and I have been designated as the 22 Federal Officer for this Advisory Committee in 23 accordance with 10 CFR Part 7.11.
24 Present today as the Designated Officer 25 is Sophie Holiday. And, likewise, as a Designated NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
8 1 Federal Officer, I would like to introduce Kellee 2 Jamerson, who was assigned to the role of the ACMUI 3 Coordinator in December.
4 This is an announced meeting of the 5 committee. It has been held in accordance with the 6 rules and the regulations of the Federal Advisory 7 Committee Act and the Nuclear Regulatory Commission.
8 This meeting is being transcribed by the NRC and may 9 also be transcribed or recorded by others.
10 The meeting was announced on -- or in the 11 February 4, 2019, edition of the Federal Register, 12 Volume 84, page 1521.
13 The function of the committee is to 14 advise the staff on issues and questions that arise 15 on the medical use of byproduct material. The 16 Committee provides counsel to the staff, but does not 17 determine or direct the actual decisions of the staff 18 or the Commission. The NRC solicits the views of the 19 Committee and values their opinions.
20 I request that whenever possible we try 21 to reach a consensus on the various issues that we 22 will discuss today, but I also recognize that there 23 may be minority or dissenting opinions. If you have 24 such opinions, please allow them to be read into the 25 record.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
9 1 At this point, I would like to perform a 2 roll call of the ACMUI members participating today.
3 First is Dr. Christopher Palestro, Chairman, Nuclear 4 Medicine Physician.
5 CHAIR PALESTRO: Present.
6 MR. EINBERG: Dr. Darlene Metter, Vice 7 Chairman, Diagnostic Radiologist.
8 VICE CHAIR METTER: Present.
9 MR. EINBERG: Dr. Vasken Dilsizian, 10 Nuclear Cardiologist.
11 MEMBER DILSIZIAN: Present.
12 MR. EINBERG: Dr. Ronald Ennis, Radiation 13 Oncologist.
14 MEMBER ENNIS: Here.
15 MR. EINBERG: Mr. Richard Green, Nuclear 16 Pharmacist.
17 MEMBER GREEN: Present.
18 MR. EINBERG: Ms. Melissa Martin, Nuclear 19 Medicine Physicist.
20 MEMBER MARTIN: Present.
21 MR. EINBERG: Dr. Michael OHara, FDA 22 Representative.
23 MEMBER O'HARA: Present.
24 MR. EINBERG: Mr. Zoubir Ouhib, Radiation 25 Therapy Physicist.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
10 1 MEMBER OUHIB: Present.
2 MR. EINBERG: Dr. A. Robert Schleipman, 3 Health Care Administrator.
4 MEMBER SCHLEIPMAN: Present.
5 MR. EINBERG: Mr. Michael Sheetz, 6 Radiation Safety Officer.
7 MEMBER SHEETZ: Present.
8 MR. EINBERG: Ms. Meghan Shober, State 9 Government Representative.
10 MEMBER SHOBER: Present.
11 MR. EINBERG: Ms. Laura Weil, Patients 12 Rights Advocate.
13 MEMBER WEIL: Present.
14 MR. EINBERG: Okay. We have a quorum.
15 On the phone we also have Dr. Wolkov.
16 Dr. Wolkov has been selected as the ACMUI Radiation 17 Oncologist. He is pending a security clearance, but 18 may participate in the meeting. However, he does not 19 have voting rights at this time.
20 And I now ask that NRC staff members who 21 are present to identify themselves. Ill start with 22 individuals in the room here. And, Sophie, if you 23 want to go first.
24 MS. HOLIDAY: Hi. Sophie Holiday, 25 Medical Radiation Safety Team.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
11 1 MS. JAMERSON: Kellee Jamerson, Medical 2 Radiation Safety Team.
3 MS. HOUSEMAN: Esther Houseman, OGC.
4 MS. LOPAS: Sarah Lopas, Medical 5 Radiation Safety Team.
7 DR. TAPP: Katie Tapp, Medical Radiation 8 Safety Team.
9 MR. EINBERG: Donna-Beth Howe is here as 10 well. And now Ill go to the NRC staff members on 11 the phone. Can you please identify yourselves?
12 MS. AYOADE: Maryann Ayoade, Medical 13 Radiation Safety Team.
14 MR. EINBERG: Okay. Thank you. Members 15 of the public who notified Ms. Jamerson that they 16 would be participating on the teleconference will be 17 captured in the transcripts. Those of you who did 18 not provide prior notification, please contact Ms.
19 Jamerson at kellee.jamerson@nrc.gov, and thats K-E-20 L-L-E-E dot J-A-M-E-R-S-O-N at nrc.gov, or at (301) 21 415-7408.
22 We have a bridge line available, and that 23 phone number is (888) 790-6447. The passcode to 24 access the bridge line is 3279476#. Once again, 25 3279476#.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
12 1 This meeting is also using the 2 GoToWebinar application to view the presentation 3 handouts real time. Hopefully, we are getting that 4 up.
5 Maryann, have you been able to get that 6 going?
7 MS. AYOADE: Yes. Its up and running.
8 MR. EINBERG: Okay. Thank you so much.
9 MS. AYOADE: If anyone on the line cant 10 see it, let me know.
11 MR. EINBERG: And you can access the 12 webinar by going to www.gotowebinar.com and searching 13 for meeting ID 657537587. Once again, thats 14 657537587.
15 The purpose of this meeting is to discuss 16 the draft report of the ACMUI Training and Experience 17 Requirements for All Modalities Subcommittee. In its 18 report, the subcommittee provides recommendations 19 with respect to the T&E requirements for all 20 modalities under 10 CFR Part 35, Medical Use of 21 Byproduct Material, with specific focus on Part 22 35.300 uses.
23 Individuals who would like to ask a 24 question or make a comment regarding a specific issue 25 the Committee has discussed should dial star one to NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
13 1 signal the operator that you wish to speak. Please 2 clearly state your first and last name for the record.
3 Comments and questions are usually addressed by the 4 Committee near the end of the presentation after the 5 Committee has fully discussed the topic. We will 6 notify the operator when we are ready for the public 7 comment period of the meeting.
8 I would also like to add that the handouts 9 and agenda for this meeting are available on the NRCs 10 public website.
11 At this time, I ask that everyone on the 12 call who is not speaking to place their phones on 13 mute. If you do not have the capability to mute your 14 phone, please press star six to utilize the conference 15 line mute and unmute functions.
16 I would also like to ask everyone to 17 exercise extreme care to ensure that background noise 18 is kept at a minimum as any stray background sounds 19 can be very disruptive on a conference call this 20 large.
21 At this point, I would like to turn the 22 meeting back to Dr. Palestro.
23 CHAIR PALESTRO: Thank you, Mr. Einberg.
24 This is Dr. Christopher Palestro, Chair of the ACMUI, 25 and I will now turn the meeting over to Dr. Darlene NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
14 1 Metter, who chairs the Subcommittee for Training and 2 Experience for All Modalities for presentation of the 3 subcommittees report. Dr. Metter?
4 VICE CHAIR METTER: Thank you, Dr.
5 Palestro, and thank you for letting us use this 6 conference call to present our Subcommittee report.
7 And before I start, I would like to thank the work by 8 Subcommittee members Dr. Ronald Ennis, Dr. Robert 9 Schleipman, Mr. Michael Sheetz, Ms. Megan Shober, and 10 Ms. Laura Weil.
11 While I present this presentation for our 12 document that we submitted, you will have slides that 13 will help point out the key points of this document.
14 The Subcommittee charge. In 2016, the 15 U.S. NRC ACMUIs Subcommittee on Training and 16 Experience Requirements for All Modalities was 17 charged to periodically review the training and 18 experience requirements for the medical use of 19 unsealed byproduct material under Title 10, Code of 20 Federal Regulations, Part 35, Subparts D to H, to 21 make recommendations for changes as needed.
22 Subcommittee subcharge. The Subcommittee 23 should reprioritize its work such that the review of 24 the T&E requirements for 10 CFR 35.300, Uses, is 25 conducted prior to the review of the T&E requirements NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
15 1 for 10 CFR 35.200. As part of the reprioritized work, 2 and in light of the NRCs tasking, the Subcommittee 3 should consider the development of a limited scope AU 4 pathway.
5 Now I will review some of the background.
6 In March 2016, the ACMUI held a public teleconference 7 meeting to discuss the report of the Subcommittee on 8 the Training and Experience Requirements authorized 9 users of alpha, beta, and gamma emitters under 10 CFR 10 35.390.
11 During this teleconference meeting, the 12 Committee unanimously endorsed the subcommittees 13 report and recommendations, which included 14 maintaining the existing 700-hour training and 15 experience requirements also termed the alternate 16 pathway.
17 Additionally, it was recommended that a 18 separate subcommittee be formed to conduct periodic 19 reviews of the training and experience requirements 20 for all modalities under 10 CFR Part 35.
21 The subcommittee developed a data-driven 22 standardized review template that would provide a 23 comparative format for future review and 24 reassessment. To optimize this review process, the 25 subcommittee intended to begin the review with NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
16 1 35.100, followed by 35.200, 35.300, and so on. The 2 subcommittee completed its review of 10 CFR 35.100 3 with no suggested revisions.
4 However, because of ongoing concerns 5 about patient access to unsealed byproduct material 6 for which a written directive is required, the 7 subcommittee was directed to review the T&E 8 requirements for 10 CFR 35.300 before reviewing 9 35.200.
10 In August 2017, the Commission voted on 11 the 10 CFR Part 35 rulemaking package and included 12 direction to the NRC staff to review 10 CFR 35 Subpart 13 E and evaluate the possibility of tailored training 14 and experience with different categories of 15 radiopharmaceuticals, delineate how these categories 16 would be created, recommend the appropriate training 17 and experience requirements, and whether these 18 requirements would be satisfied based on hours of 19 training or would require a formal assessment of 20 competency.
21 In January 2018, the U.S. Food and Drug 22 Administration approved a therapeutic 23 radiopharmaceutical, lutetium-177 dotatate, with the 24 potential for greater use than previously approved 25 therapeutic radiopharmaceuticals.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
17 1 In addition, there was a decrease in the 2 first-time candidates sitting for the American Board 3 of Nuclear Medicine or the ABNM certification exam.
4 These two observations raised concerns about a 5 potential authorized user or AU shortage in the 6 future. Thus, the ACMUI proposed the reconsideration 7 of an alternate AU pathway for 10 CFR 35.390.
8 Now, what Ill do is give the 9 Subcommittee review, comments, and recommendations, 10 and this was divided into three topics. The first 11 topic is a potential AU shortage. To address concerns 12 about a potential future shortage of AU, the 13 subcommittee reviewed the current pathways for AU 14 certification.
15 Traditionally, nuclear medicine, nuclear 16 radiology, diagnostic radiology, and radiation 17 oncology graduates of Accreditation Council for 18 Graduate Medical Education, or ACGME, approved 19 residencies, seek board certification; and, hence, 20 authorized user status, by the NRC-deemed Status 21 Boards with the American Board of Radiology, ABR; 22 Nuclear Medicine, ABNM; and Osteopathic Radiology.
23 In 2016, the ABR supported a redesigned 24 AU eligibility pathway consisting of 16 months of 25 nuclear radiology or nuclear medicine training NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
18 1 incorporated into the 48-month or four-year ACGME-2 accredited diagnostic radiology residency.
3 This revised program is a redesigned ABR 4 pathway and would satisfy the NRCs training and 5 experience requirements for 10 CFR 35.390 via the 6 alternate pathway.
7 Upon completion of the radiology 8 residency, the graduate trainee is then eligible to 9 sit for the board certification exams for the ABR in 10 nuclear radiology and/or the ABNM.
11 Now, to explore the concern for potential 12 AU shortage, the Subcommittee reviewed the 2018 and 13 2019 ACGME website, which provided the following 14 information on the current number of potential future 15 AUs in training. And the slide you have listed the 16 total number of residents in training and the 17 estimated graduates per year. The first is nuclear 18 medicine, which has 40 programs. The peculiar aspects 19 about that residency is that it has a one-, two-, and 20 three-year pathway, with the one-year pathway taken 21 by radiologists, which are the majority of the total 22 residents.
23 So the number 79 is the total number of 24 residents currently in training with an estimate of 25 approximately 50 graduates, 40 to 50 graduates per NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
19 1 year.
2 For nuclear radiology, total residents 3 11, with 11 graduates per year. The redesigned 4 pathway, which has -- I just went over -- currently 5 has 56 residents, and the graduates wont be coming 6 up for a few more years because this just started 7 about a year or so ago. So when it does mature, it 8 will be approximately with this current number of 14 9 per year, but likely it is likely to increase in 10 number.
11 Radiation oncology, 775 total residents 12 with 194 graduates per year. The diagnostic radiology 13 and osteopathic radiology residents are large in 14 number, with certain graduates per year. However, 15 the majority of them do not satisfy 35.390 16 qualifications.
17 So in the current pipeline for AUs in 18 training for 35.390, its over 900, which includes 19 trainees in nuclear medicine, nuclear radiology, 20 diagnostic radiology, and radiation oncology. As of 21 2018, the ABNM reported 3,591 practicing ABNM 22 diplomates. And along with the current practicing 23 authorized users, the nearly 270 annual 35.390 AU 24 graduates, and counting for the retiring AUs, the 25 Subcommittee concluded that there is no objective NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
20 1 data to support an authorized user shortage at the 2 present time.
3 And if you think about it, if you have 4 about 250, 270 graduates per year in the next four 5 years, you will add another 1,000 AUs for 35.390.
6 Topic 2, the limited scope AU pathway.
7 Although there is no evidence that there is a current 8 AU shortage, there is the likelihood that the number 9 of available therapeutic radiopharmaceuticals, and 10 the demand for these therapies will increase. The 11 subcommittee explored whether the NRC should consider 12 developing a limited scope AU pathway tailored to 13 specific radiopharmaceuticals.
14 Radionuclide therapy possesses the 15 highest risk and the highest impact of all nuclear 16 medicine procedures. And if dosages are not properly 17 handled or administered, these therapies can cause 18 unintentional, serious organ or tissue injury. The 19 newer therapeutic radionuclides have become 20 increasingly more complex administrations. And with 21 the potential for multi-organ or tissue toxicities, 22 and, hence, this requires a basic competency in 23 radiation therapy and radiation safety.
24 A potential limited scope AU pathway for 25 radionuclide therapy must ensure that the basic NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
21 1 knowledge topics in 10 CFR 35.390 are obtained thereby 2 obtaining an equivalent level of therapeutic 3 competency and competency in radiation safety.
4 When investigating the feasibility of a 5 limited scope AU pathway for 10 CFR 35.390, the NRC 6 staff, with ACMUI input, proposed a list of required 7 basic knowledge topics for AUs involved in 8 radionuclide therapy. The proposed curriculum began 9 with the knowledge topics of 10 CFR 35.390.
10 But due to the complexity and overlap of 11 these basic knowledge topics, the Subcommittee 12 concluded that it is not feasible to tailor the T&E 13 requirements for a limited scope authorized user for 14 each specific radiopharmaceutical, nor is it feasible 15 to create categories for specific therapeutic 16 individual radionuclides because each such category 17 would encompass nearly all of the knowledge topics in 18 10 CFR 35.390.
19 The NRC staff, with external stakeholder 20 input from the medical community, and the 21 Subcommittee, agreed that the knowledge topics in 10 22 CFR 35.390 are the basic minimum knowledge required 23 for any radionuclide therapy.
24 In considering the above, the 25 subcommittee does not recommend a limited scope AU NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
22 1 pathway for radionuclide therapy requiring a written 2 directive. Unlike the iodine I-131 sodium iodide, 3 limited scope authorized user, under 10 CFR 35.392 4 and 35.394, which are for endocrinologists, the 5 emerging radionuclide therapies have multiple 6 contraindications and more toxicities versus the I-7 131 sodium iodide, which is specific to one organ, 8 the thyroid gland for thyroid therapy.
9 It would be too cumbersome to develop and 10 provide oversight for specific training and 11 experience requirements within the regulations to fit 12 each radionuclide therapy. All of the classroom 13 laboratory training areas and work experience topics 14 contained in 35.390 are applicable to any 15 radionuclide therapy and are essential for radiation 16 safety of the patient, personnel, and public.
17 It would be difficult in defining a 18 limited scope authorization and what radionuclides or 19 radiopharmaceuticals are to be included. Each 20 therapeutic radiopharmaceutical has unique radiation 21 safety issues, which require a comprehensive 22 understanding of all of the training and experience 23 topics in 35.390, regardless of the types of radiation 24 emissions, chemical properties, or mode of 25 administration.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
23 1 During a public NRC meeting held on 2 December 11, 2018, a novel team approach was proposed 3 where an onsite authorized nuclear pharmacist, or 4 ANP, would prepare the radionuclide for therapy and 5 handled the radiation safety components, while the 6 limited scope authorized user would administer the 7 patient-ready dose and manage patient care.
8 The perceived benefits of an AU 9 partnership should be carefully reviewed. Although 10 well intended, a fragmented approach to a therapeutic 11 procedure can have the unintended consequence of 12 making things worse. Furthermore, if an onsite ANP 13 or authorized nuclear pharmacist is available, a 14 fully trained authorized user is also likely 15 available for the entire radionuclide therapies, 16 which are generally not on an emergent basis.
17 There are also fewer authorized nuclear 18 pharmacists than authorized users, and authorized 19 nuclear pharmacists are generally concentrated, as 20 are authorized users, in urban and not-rural areas.
21 The safe and effective administration of 22 radionuclide therapy is best accomplished by a 23 comprehensively trained physician authorized user who 24 is responsible for the entire therapeutic procedure 25 and who has a thorough knowledge and understanding of NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
24 1 the therapy, to include the various factors and 2 potential toxicities and serious hazards that can 3 occur to the patient, the personnel, and the public.
4 Topic 3, competency assessment for the 5 limited scope authorized user pathway. In the initial 6 limited stakeholder outreach, the majority of 7 respondents favored using an examination to confirm 8 the successful acquisition of 10 CFR 35.390 outreach 9 topics and to confirm the individuals competency to 10 independently function as comprehensive or limited 11 scope authorized user under 10 CFR 35.390. It is 12 also critical to validate that the proposed 13 curriculum was successfully attained.
14 For this confirmation and proficiency, 15 the NRC staff and the subcommittee agree that a 16 competency assessment is necessary. This assessment 17 should not be based on hours or preceptor 18 attestations, but, rather, on an initial and 19 continued competency evaluation over time.
20 The subcommittee supports broader input 21 from the medical community to create an AU competency 22 assessment with final approval by the NRC. The 23 subcommittee further supports the periodic 24 reassessment of authorized user competency, 25 particularly in relation to the frequent or NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
25 1 infrequent performance of radionuclide therapy. This 2 infrequency concept has raised similar concerns in 3 board certification/ recertification process, and the 4 recertification process is generally seven to ten 5 years between recertification exams.
6 The American Board of Medical 7 Specialties, or the ABMS, is a nonprofit organization 8 of 24 medical specialty boards which serves the public 9 in quality health care through professional and 10 educational standards. And the ABMS has supported a 11 program of continuing professional certifications for 12 physician lifelong learning and self-assessment.
13 The American Boards of Radiology and 14 Nuclear Medicine are ABMS member boards. And to 15 promote continued professional competency for their 16 diplomats, the ABR and ABNM have transitioned to or 17 are transitioning to this continuous longitudinal 18 assessment.
19 In regards to radionuclide therapy, this 20 infrequency concept and procedure performance was 21 also reviewed by the subcommittee. Because of the 22 ability to eliminate and destroy tissue, therapeutic 23 radionuclide procedures pose a much higher risk to 24 the patient, personnel, and public than do diagnostic 25 procedures. The potential for limited scope AUs, and NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
26 1 the higher likelihood that the infrequent performance 2 for radionuclide therapy in rural areas would make it 3 difficult for physicians to retain basic AU 4 competency in radionuclide therapy.
5 To attest to the successful acquisition 6 of the authorized user knowledge topics in 10 CFR 7 35.390, the overall limited stakeholder input and the 8 subcommittee support a formal competency 9 certification and a continuous certification process.
10 The goal of certification is to validate 11 that an authorized user candidate has achieved a 12 predetermined level of competence, and the current -
13 - this certification is to confirm the acquisition of 14 a basic knowledge curriculum and the ability to 15 independently function as an authorized user for 16 specific radionuclide therapy or therapies.
17 Although the subcommittee does not 18 recommend adoption of a limited scope AU pathway for 19 therapy, if the NRC pursues such a pathway, the 20 subcommittee strongly recommends an initial formal 21 competency assessment and competency reassessment 22 with ongoing longitudinal reassessment with specific 23 emphasis on radiation safety.
24 The entity or entities that will 25 administer this formal competency assessment and NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
27 1 reassessments must develop a methodology that ensures 2 that passing these examinations is empirically 3 determined. This latter aspect is the Angoff Method, 4 which is a widely used standard in test development 5 and creates a test that will be legally defensible 6 and meet the standards for educational and 7 psychological testing.
8 So, in summary, the ACMUI Subcommittee on 9 Training and Experience Requirements for All 10 Modalities addressed the NRC staff request to assess 11 the feasibility of a limited scope authorized user 12 pathway for 10 CFR 35.390, which was initially 13 predicated on the concern about a potential future 14 shortage of authorized users.
15 At the present time, there are no 16 objective data to support an authorized user 17 shortage. The subcommittee does not recommend the 18 development of a limited scope AU pathway for the 19 administration of unsealed byproduct material where 20 a written directive is required.
21 If the NRC moves forward in pursuing an 22 alternative limited scope authorized user pathway, 23 the subcommittee strongly recommends that the limited 24 scope authorized user must successfully acquire the 25 knowledge topics in 10 CFR 35.390, which would be a NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
28 1 minimum requirement for all authorized users involved 2 in radionuclide therapy.
3 The subcommittee also conclude that due 4 to the complexity and overlap in these basic knowledge 5 topics, it would be difficult to safely and 6 practically create specific categories for 7 therapeutic radiopharmaceuticals.
8 Despite the ACMUI Subcommittee 9 recommendation against this action, if the NRC 10 chooses to pursue the creation of a limited scope 11 authorized user pathway for unsealed byproduct 12 material where a written directive is required, the 13 subcommittee strongly recommends that the authorized 14 user candidate must acquire the basic knowledge 15 topics in 10 CFR 35.390 and satisfactorily complete 16 a formal competency assessment.
17 Furthermore, the individuals continued 18 status as a limited scope authorized user is dependent 19 on successfully maintaining a formal periodic 20 reassessment of competency. This final and most 21 critical component, and the attainment and 22 maintenance of any authorized user status, will 23 optimize patient care while ensuring the protection 24 of the publics health and safety.
25 So the Subcommittee position and four NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
29 1 recommendations are the following. First, the 2 subcommittee strongly supports and reaffirms the 3 committees 2016 position on maintaining the current 4 and existing authorized user pathway, which are the 5 board certification and alternate pathways as 6 codified in the regulations, which are adequate for 7 protecting public health and safety. Radionuclide 8 therapy poses the highest risk and the highest impact 9 of all nuclear medicine procedures.
10 Two, the Subcommittee concludes that 11 there is no objective data to confirm an authorized 12 user shortage.
13 Three, the Subcommittee does not 14 recommend a limited scope authorized user pathway for 15 unsealed byproduct material for which a written 16 directive is required.
17 And, four, the Subcommittee unanimously 18 agrees that in order to ensure the safety of patients, 19 personnel, and the public, if the NRC chooses to 20 pursue the creation of a limited scope authorized 21 user pathway for unsealed byproduct material, where 22 a written directive is required, the authorized user 23 candidate must acquire the basic knowledge topics in 24 10 CFR 390 and satisfactorily complete a formal 25 competency assessment.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
30 1 Furthermore, the individuals continued 2 status as a limited scope authorized user is dependent 3 on successfully maintaining a formal periodic 4 reassessment of competency.
5 And that is the end of my report, and, 6 Dr. Palestro, I turn the meeting over to you.
7 CHAIR PALESTRO: Thank you, Dr. Metter.
8 This is Dr. Palestro again. I now open this report 9 to comments by members of the Subcommittee.
10 MEMBER DILSIZIAN: Vasken here. Great 11 presentation, Dr. Metter. I guess, to me, Im just 12 going to summarize, if I get this correctly, what we 13 are proposing is to maintain the current AU pathway 14 training for 35.390. And not only maintain it, but 15 add not just training based on hours and attestation 16 alone, but additional competency assessment in this 17 formal test that would be concluding in a certificate, 18 which would be maintained subsequently by some number 19 of years where you have to be recertified. Is that 20 -- would that be the conclusion then?
21 VICE CHAIR METTER: That would be a 22 recommendation right now. The subcommittee was 23 looking at the feasibility of a limited scope 24 authorized user pathway. And these are some of the 25 recommendations we would make, and that would have to NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
31 1 be for a future subcommittee investigation.
2 MEMBER DILSIZIAN: Thank you.
3 CHAIR PALESTRO: Any other comments from 4 members of the subcommittee?
5 MEMBER SHEETZ: This is Mike Sheetz. I 6 have a comment.
7 CHAIR PALESTRO: Go ahead, Mr. Sheetz.
8 MEMBER SHEETZ: I want to thank Dr. Metter 9 for pulling all of this together. I have a couple of 10 comments. One, in consideration of the limited scope 11 AU/ANP partnership, while the ANP could help as an 12 RSO, we feel that the AU must have a comprehensive 13 knowledge and understanding of the entire therapeutic 14 procedure.
15 This includes all of the radiation safety 16 issues associated with the procedure from package 17 receipt, dose assay surveys, radioactive waste 18 disposal, instrument QA, radiation safety training, 19 personal monitoring, and others.
20 In the United Pharmacy Partners AU/ANP 21 partnership proposal, there was no delineation of 22 tasks of who would be responsible for the aspects of 23 this therapy. You know, would the ANP be physically 24 present during the administration of the procedure?
25 Would they be onsite for a person to do special NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
32 1 surveys and waste disposal?
2 So the details of the shared 3 responsibility have not been addressed, so its 4 difficult to make a determination of whether it is 5 really feasible. While there was a team approach 6 with other types of radiation therapy and medical 7 uses, such as the Y-90 microsphere therapy, gamma 8 radiosurgery, radioactive seed localization, the AU 9 really is knowledgeable in all of the areas and 10 supervises the other team members.
11 In one of the United Pharmacy Partners 12 letters, they say that this limited scope AU status 13 is only being requested for alpha and beta therapies 14 and not high-risk materials. I feel this, in itself, 15 is just a limited understanding of the radiation 16 physics and radiation safety associated with all 17 radiopharmaceuticals used for therapy.
18 Most of these radionuclides also emit 19 photons, gamma rays, or X-rays, you know, very strong, 20 and the alpha and/or beta particles also present a 21 radiation risk, you know, if not handled or 22 administered properly.
23 In my personal experience here at the 24 University of Pittsburgh Medical Center, before the 25 FDA approval of Lutathera, we were one of the only NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
33 1 sites in the western PA region who offered this 2 therapy. At that time, we had quite a few referrals 3 that came from distant locations, roughly 40 percent.
4 However, now a year after FDA approval, we dont get 5 those referrals. Almost all of our patients are from 6 the local area.
7 Also, our volumes have plateaued and 8 slowly dipped as many patients have been -- who have 9 been waiting for the therapy have received their 10 treatments. The point is, patients travel to get 11 medical care. Another important factor that needs to 12 be considered is that the lack of Lutathera 13 availability in many places is not because of an end 14 use shortage but the reluctance of many hospitals or 15 clinics to do Lutathera therapy because of its upfront 16 cost, which is upwards of $200,000 per patient in 17 drug cost for the four treatment, which obviously can 18 easily break the bank for smaller community entities, 19 even with one or two insurance ties.
20 In addition, the manufacturer of 21 Lutathera, AAA, if necessary with their applications, 22 if selective and a good institution, it will help set 23 up a program based somewhat on the expected volume of 24 patients.
25 So I believe the core program on NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
34 1 availability is really the extremely high cost of 2 these drugs and not the availability of AUs.
3 Thank you.
4 CHAIR PALESTRO: Thank you, Mr. Sheetz.
5 Any other comments from members of the 6 Subcommittee? Ron?
7 MS. HOLIDAY: Dr. Ennis, before you speak 8 -- this is Sophie Holiday -- for everybody else on 9 the phone, I just want to let you know that the 10 information that Mr. Sheetz was referring to was a 11 letter submitted to the NRC on February 20th from 12 UPPI, and that comment letter will be made publicly 13 available when it is appended to the meeting 14 transcript as part of the official record.
15 Thanks. Ill turn it back to you now, 16 Dr. Ennis.
17 MEMBER ENNIS: Thank you. Just kind of 18 a reflection. In these recommendations that people 19 have put forward, I see a very lack of the 20 understanding of the biological issues surrounding 21 radiotherapy and the complications thereof.
22 There seems to be some kind of implicit 23 understanding or -- that these are trivial drugs that 24 anyone can learn how to do, and its just a technical 25 application and a technical issue of this injection NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
35 1 rather than understanding the complex biology and how 2 radiation interacts with a variety of tissues.
3 Its interaction with each tissue is 4 unique. The complications that can cause -- that it 5 can cause, both acutely and chronically, are unique.
6 Its a level of subspecialization that [inaudible],
7 the full education, has listed in the listing of 8 topics. To imply otherwise would be somewhat similar 9 to suggesting that a primary care doctor could do 10 chemotherapy. The analogy is really quite similar, 11 but obviously possible.
12 Thank you.
13 CHAIR PALESTRO: Thank you, Dr. Ennis.
14 Any other comments from members of the 15 subcommittee?
16 MEMBER WEIL: Yes, please. This is Laura 17 Weil.
18 CHAIR PALESTRO: Go ahead, Ms. Weil.
19 MEMBER WEIL: The primary role of 20 regulation is to ensure safety. And its true that 21 the -- that regulation can limit access to some people 22 who cant travel for care, and that is a legitimate 23 barrier. However, its not a question of whether 24 those barriers exist. Its more a question of whether 25 the regulatory standards that create those barriers NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
36 1 are unnecessarily restrictive or ethically 2 responsible -- or nor is it the role of the regulator 3 to attempt to increase access by compromising 4 necessary safeguards that protect patients, their 5 families, health care providers, and the public.
6 Thank you.
7 CHAIR PALESTRO: Thank you, Ms. Weil.
8 Any other comments from other members of 9 the Subcommittee? Comments from members of the ACMUI?
10 MEMBER MARTIN: This is Melissa Martin.
11 Again, I would like to refer to the letter that was 12 submitted from the UPPI. A couple of comments after 13 reviewing that letter. As presented earlier, one of 14 the requests was that we look at the distribution of 15 authorized -- number of authorized users and the 16 geographic distribution.
17 In the letter from the UPPI, they 18 suggested that there would be a better geographic 19 distribution of nuclear pharmacists than there would 20 be of authorized users. Again, there has been no 21 data submitted to support that statement. I think we 22 would need to see some kind of data that shows where 23 the nuclear pharmacist would be distributed.
24 I know the NRC is working on collecting 25 the data for where the authorized users are located.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
37 1 But, again, if we have 4,000 authorized users, I think 2 there is a better geographic distribution already. I 3 think -- my impression is most of those authorized 4 users are connected to relatively moderate to large 5 size cities, and the patients -- again, that was 6 stated earlier -- the ones that live in the rural 7 areas are used to traveling and getting 8 accommodations to get their therapy.
9 So I dont see that that has been a big 10 -- I dont think that is the impediment to receiving 11 some of these treatments.
12 The proposal is that it would increase 13 access to the rural access -- rural areas, but there 14 has been no proposed limitation on these procedures 15 being performed by the team approach, limited to the 16 rural areas. So it would assume that a lot of these 17 procedures would be performed in the same geographic 18 mid-to-large sized cities that is currently being 19 performed.
20 So I dont see that that would -- I dont 21 understand how that would be restricted to rural 22 access.
23 One of the other questions I had was, 24 what is meant by minimally trained physician that 25 would be administering the isotope in the proposed NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
38 1 team approach? Thank you very much.
2 CHAIR PALESTRO: Thank you, Ms. Martin.
3 Any other comments from the ACMUI?
4 MEMBER GREEN: Dr. Palestro?
5 CHAIR PALESTRO: Yes.
6 MEMBER GREEN: This is Richard Green.
7 CHAIR PALESTRO: Yes, Mr. Green. Go 8 ahead.
9 MEMBER GREEN: The Subcommittees 10 position and recommendations are four bullets on the 11 document. The last bullet says that the Subcommittee 12 unanimously agrees, in that fourth bullet. Were the 13 three prior bullets unanimous decisions?
14 VICE CHAIR METTER: This is Darlene. Yes, 15 the Subcommittee -- actually, every member was given 16 the document and they all agreed and we could actually 17 add that, too, unless there is -- I misunderstood the 18 Subcommittee, and that when they submitted their 19 comments they did not agree, but I -- they did not 20 relate that, but we wanted to make an emphasis on 21 that last point because that is a very important 22 point.
23 I mean, theyre all important, but that 24 was a key component in assessing the competency of 25 the individual who is going to be an authorized user.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
39 1 And can I just ask the Subcommittee, was 2 that correct or was it -- I mean, is that a correct 3 reflection of our decision?
4 PARTICIPANT: Absolutely.
5 MEMBER SCHLEIPMAN: Robert Schleipman. I 6 totally agree. I think we all -- including for 7 myself, I fully agree with the recommendations.
8 PARTICIPANT: Yes, it was unanimous from 9 our discussions.
10 VICE CHAIR METTER: Thank you.
11 MEMBER GREEN: Thank you, Dr. Metter.
12 Another question I have. I agree completely that the 13 requirements for complete knowledge of all aspects of 14 10 CFR 35.390 are required for any physician who 15 administers radionuclide therapy. I dont think 16 there is any ability to skimp on that full, complete 17 knowledge.
18 It is my understanding that during the 19 course of training through one of the approved 20 pathways to become a diplomate in nuclear medicine, 21 nuclear radiology, radiation oncology, or diagnostic 22 radiology, that that training period encompasses more 23 than the 10 CFR 35.390.
24 Can you give an estimate of what time is 25 spent on 35.200 and other nuclear medicine NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
40 1 techniques, imaging, uptake, dilution, and excretion, 2 other than radionuclide therapy?
3 VICE CHAIR METTER: Well, they have -- as 4 far as its all incorporated into 48 months of 5 diagnostic radiology. But looking at the authorized 6 user status, we are really mainly just looking at the 7 equivalent training and experience requirements for 8 35.390, which is for the use of unsealed sources for 9 therapy and not for imaging and localization or 10 dilution and excretion.
11 MEMBER GREEN: Yeah. I thought the 12 charter --
13 VICE CHAIR METTER: Yeah. Im sorry.
14 The other thing, too, were not looking at hours. As 15 I mentioned in the report, were not looking at hours.
16 Were looking at level of competency and ability to 17 acquire that knowledge and use that knowledge safely 18 for our patients.
19 MEMBER GREEN: All right. Im just trying 20 to -- you know, I thought the challenge to the 21 Subcommittee was to discern in the training and 22 experience for all modalities. And so this is whats 23 required to successfully do 35.200, 35.100, 35.300 24 applications for radiopharmaceuticals.
25 And I -- I think I am seeing a resounding NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
41 1 support of the existing pathway, but I dont really 2 think I see a dissection of really what training and 3 education is required for this 35.300 versus 35.200 4 versus 35, you know 100. And thats what I thought 5 was the charge to the Committee -- to the 6 Subcommittee.
7 VICE CHAIR METTER: Mr. Green, the 8 Subcommittee is supposed to look at 35.300, but this 9 specific report addresses 35.390 and not the entire 10 35.300.
11 MEMBER GREEN: Thank you.
12 VICE CHAIR METTER: And that is 13 throughout the report, by the way.
14 MEMBER DILSIZIAN: Vasken here. Can I 15 have a comment?
16 VICE CHAIR METTER: Yes.
17 MEMBER DILSIZIAN: So, Dr. Metter, you 18 know, I think I know what youre intending to say, 19 but looking at this slide, and then in your statement 20 when you say, We are not looking for our self-21 competency. I think what you wanted to say is, were 22 not just looking for hours and attestation alone, but 23 an added competency assessment.
24 I think that its important to just say 25 -- because the way it comes out both in the slide and NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
42 1 your verbal statement, that if there is no minimum 2 hours2.314815e-5 days <br />5.555556e-4 hours <br />3.306878e-6 weeks <br />7.61e-7 months <br />, I think we -- we will understand there has got 3 to be some number of hours of training, and then it 4 falls with the competency test. Is that correct, 5 what Im hearing, because thats not very clear.
6 VICE CHAIR METTER: No. Exactly what I 7 said, that we are not looking at hours because hours 8 can be - its different for different individuals.
9 Were looking at the final result.
10 Its sort of like -- I believe I made the 11 same analogy during our meeting last year in the sense 12 of if you -- if you are taking your drivers license 13 -- you wanted to obtain a drivers license, and the 14 way you obtain it is -- the final result is going to 15 be the drivers license.
16 And whether you obtain it by going to a 17 driving school or going to a course with your parents 18 or with a different individual, if you -- if you learn 19 the basic requirements and the basic fundamental 20 knowledge and experience that you need to pass an 21 exam to get your certification, is what this is 22 implying. This is a new pathway in the 23 sense, if its a limited scope, its not like a 24 traditional one that weve done before. So I think 25 we need to really look at it very seriously because NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
43 1 hours1.157407e-5 days <br />2.777778e-4 hours <br />1.653439e-6 weeks <br />3.805e-7 months <br /> can be, you know, different for your number of 2 training hours. You can learn quicker than I can, 3 but I want to know the final result, that you are a 4 competent individual for the administration of the 5 radiopharmaceutical therapy for the patient and for 6 those involved with -- and the public.
7 And so, to me, the final result is that 8 you are able to do what is needed for our patients.
9 MEMBER DILSIZIAN: Well, the problem I 10 guess I have is that any ACGME training program, where 11 youre doing internal medicine, youre doing 12 radiology, or radiation oncology, has, first, a 13 minimum number of years of training or hours of 14 training for/by board certification.
15 You cant possibly ignore the 700 hours0.0081 days <br />0.194 hours <br />0.00116 weeks <br />2.6635e-4 months <br /> 16 requirement, plus competency. This is what you -- or 17 what Im hearing you say is that lets forget about 18 the 700 hours0.0081 days <br />0.194 hours <br />0.00116 weeks <br />2.6635e-4 months <br />. Some people can do it in 50 hours5.787037e-4 days <br />0.0139 hours <br />8.267196e-5 weeks <br />1.9025e-5 months <br />, as 19 long as they pass the test, and Im having trouble 20 with that. I think my recommendation would be to 21 maintain a certain number of hours, years, or whatever 22 it takes for an average person to be trained for/by 23 a competency test. Otherwise, I think that 700 --
24 what you are proposing, 700 hours, doesnt mean 25 anything. Maybe someone can do it in 50 hours5.787037e-4 days <br />0.0139 hours <br />8.267196e-5 weeks <br />1.9025e-5 months <br />.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
44 1 VICE CHAIR METTER: I understand what 2 youre saying.
3 CHAIR PALESTRO: Dr. - Im sorry, Dr.
4 Dilsizian?
5 MEMBER DILSIZIAN: Yes.
6 CHAIR PALESTRO: Yeah. Hi. Dr. Palestro 7 speaking now. At this point, your comments are sort 8 of moot because the Subcommittee has recommended 9 against the creation of the limited scope authorized 10 user status.
11 But going back a bit -- and do you 12 remember this -- this committee and these discussions 13 have been evolving over time. The concept was to --
14 if and when a program is -- a limited AU program is 15 in fact developed, to develop and to create a 16 curriculum, if you will, with all of the competencies 17 that need to be met, and then after that to go back 18 and determine hours that would be required to complete 19 it, because as I think we all agree at this point, we 20 have no idea how the 700 hours0.0081 days <br />0.194 hours <br />0.00116 weeks <br />2.6635e-4 months <br /> was arrived at.
21 And so rather than focusing on hours, the 22 concept would have been develop a program, make sure 23 it has all of the elements necessary to educate and 24 to develop a competent AU, and then after that figure 25 out how to -- how to translate that into hours.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
45 1 MEMBER DILSIZIAN: So, again, I guess I 2 just wanted to -- I think Mr. Green said this before 3 me. I guess I would have expected the subcommittee 4 to actually designate some hours to the current 5 curriculum that would sum up to some number that may 6 be 700 or 500 or 550, so that we can be reassured, as 7 would be the NRC staff, that these numbers are in the 8 cost ballpark of what it would take for someone to be 9 adequately trained.
10 CHAIR PALESTRO: Dr. Metter, would you 11 like to respond to that?
12 VICE CHAIR METTER: I would defer to you.
13 CHAIR PALESTRO: Thank you, Mr. Metter.
14 VICE CHAIR METTER: Anytime.
15 CHAIR PALESTRO: The answer is, again, I 16 think right now for the moment it is a moot point, 17 since the subcommittee has recommended against it.
18 Should the NRC decide to go forward with it, as the 19 subcommittee has indicated, they clearly would like 20 the opportunity to actively participate in the 21 development of such a program.
22 MEMBER GREEN: Dr. Palestro?
23 CHAIR PALESTRO: Yes.
24 MEMBER GREEN: This is Mr. Green again.
25 So with the recommendation not at this time to do a NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
46 1 limited AU status, will the subcommittee, or the 2 committee as a whole, take the challenge, as Dr.
3 Dilsizian has asked, to actually dive into the concept 4 and figure out how many hours -- what does it take to 5 learn this concept, that concept, this fact, this 6 methodology, so that we can actually better 7 substantiate that engraved-in-stone sort of -- we 8 dont know where it came from, but if we are asked by 9 staff or by the Commissioners to develop other 10 pathways or to validate the existing pathway, are we 11 going to be going through the exercise?
12 CHAIR PALESTRO: The answer is, when the 13 subcommittee was originally formed -- and, in fact, 14 if you look at the title of the subcommittee -- it is 15 training and experience for all modalities. And the 16 concept was -- and as Dr. Metter enumerated in her 17 report, we developed a template. We looked at 100, 18 and our plan was to go to 200, and then 300, but we 19 were -- or the subcommittee, I should say, was 20 redirected to focus on 390.
21 So the answer is, the concept for the 22 subcommittee is at some point to return to its 23 original charge and begin to look at the various 24 categories. And I would presume that the subcommittee 25 will take into account hours and necessary, for lack NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
47 1 of a better term, items that individuals need to be 2 knowledgeable in.
3 MEMBER GREEN: Dr. Metter, you are 4 currently the chair of the subcommittee. Would you 5 agree with that?
6 VICE CHAIR METTER: Yes, Dr. Palestro, I 7 do. And I also think, as far as if we do look back 8 at how these 700 hours0.0081 days <br />0.194 hours <br />0.00116 weeks <br />2.6635e-4 months <br /> were obtained, it may actually, 9 with the increasing complexities of the radionuclides 10 that are coming out for therapy, that the - were 11 assuming that the hours are going to decrease.
12 The hours may -- we dont know what the 13 hours1.50463e-4 days <br />0.00361 hours <br />2.149471e-5 weeks <br />4.9465e-6 months <br /> will do. They may actually increase because 14 you now have to know more about what -- about the 15 newer therapies and their toxicities and adverse 16 effects and effects on the patient and their -- and 17 the long-term consequences.
18 So that would be something wed have to 19 look at. But at this point in time, I dont think we 20 are looking at that because, as you said, we did not 21 support a limited scope AU pathway.
22 CHAIR PALESTRO: Thank you, Dr. Metter.
23 Any other comments from the ACMUI?
24 MEMBER DILSIZIAN: Vasken here. I just 25 want to go on the record just to kind of let us know NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
48 1 that despite -- I agree with all of what was stated 2 so far. It would have been nice to have an 3 approximation of the hours dedicated to the current 4 required training -- current, not the -- not the 5 shortened pathway.
6 That would be close to 700 hours0.0081 days <br />0.194 hours <br />0.00116 weeks <br />2.6635e-4 months <br />, and so 7 it would then reassure all of us on this call, as 8 well as the NRC Commissioners and the staff, that we 9 are not that far apart from what is currently 10 required.
11 CHAIR PALESTRO: Thank you, Dr.
12 Dilsizian. But, again, I just want to point out that 13 was not the charge or the subcharge that was given to 14 the subcommittee.
15 Any other comments/questions from the 16 ACMUI?
17 MEMBER OUHIB: Yes. This is Zoubir Ouhib.
18 My comment is really we seem to be focusing a little 19 bit on the numbers, and Im a little bit concerned 20 about that because, really, going back and seeing 21 what it takes in -- you know, on the big picture, you 22 know, in terms of patient evaluation, pre-treatment, 23 and so on and so forth, and then the treatment itself, 24 its not just an injection. There is a lot more that 25 goes into it than that, and then the patient NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
49 1 management after the fact, and so on and so forth.
2 So I think looking at what an authorized 3 user has to go through in terms of education, 4 training, and so on and so forth, and exams and so 5 on, I think that number is -- well, is irrelevant in 6 my opinion because its very easy to look at that and 7 figure out that it takes a lot more than the 700 8 perhaps.
9 CHAIR PALESTRO: Thank you, Mr. Ouhib.
10 Any other comments or questions from the 11 ACMUI? Dr. Metter, I have a question for you. The 12 subcommittee concluded that there are no objective 13 data to support an AU shortage at the present time.
14 We have a list of data, table of data, in which we 15 have about 270 new graduates per year.
16 How did you come to that conclusion? In 17 other words, what data did you use? Did you attempt 18 to determine the number of AUs throughout the United 19 States and then look at the number per 100,000 people 20 and determine thats sufficient or insufficient?
21 Exactly how did you arrive at that conclusion?
22 VICE CHAIR METTER: Well, there is -- you 23 know, those are the numbers that we have. And as we 24 had previously stated, most of these are concentrated 25 in urban areas and where the facilities and resources NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
50 1 are available for the safe use of the radionuclides.
2 And there is really no objective data for 3 that, and it was -- it was -- there is no objective 4 for that -- objective data regarding your question.
5 CHAIR PALESTRO: Okay. So this was sort 6 of I guess an intuitive approach or an intuitive 7 conclusion based on discussions among members of the 8 subcommittee?
9 Types of things that I would be 10 interested in that would make me feel more comfortable 11 with this sort of conclusion would be the following.
12 What are the trends over time if we look at the 13 specialties? Have the number of individuals being 14 certified, are they increasing, decreasing? Are they 15 remaining the same?
16 You mentioned briefly individuals 17 retiring, but did you do any -- any sort of 18 calculations, for lack of a better word, to see 19 whether or not the number of new graduates or newly 20 certified individuals are compensating for retirees, 21 that sort of thing?
22 And then the other issue, of course, that comes 23 up, and we have talked about lutetium-177, and there 24 has also been another agent that was approved last 25 July, the I-131 MIBG, what about the numbers of these NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
51 1 unsealed source radiopharmaceutical therapies?
2 Where are they in terms of total numbers?
3 Are they decreasing? Are they plateauing? Are they 4 increasing? Those are the sorts of data that would 5 make me feel more comfortable about accepting the 6 conclusion that there is nothing to suggest an 7 authorized user shortage at this time.
8 VICE CHAIR METTER: Those are very good 9 points, Dr. Palestro. I think looking at numbers can 10 be for or against an argument. And if you say you 11 have one authorized user for X number of patients, 12 that would be assuming that the distribution of 13 patients is equal and homogeneous throughout that 14 territory, which of course, as I mentioned earlier, 15 its not because theyre generally concentrated in 16 urban areas.
17 So, in itself, the distribution of the 18 population is not equal. And as far as the number of 19 retirees, that is not available, and I dont know how 20 that could be available. But as far as we know, I 21 can give you the number of practicing authorized users 22 by the American Board of Nuclear Medicine that they 23 obtained -- that I obtained from them, and we can 24 look at the number of authorized user graduates from 25 these programs.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
52 1 We do not know -- and I dont believe the 2 NRC collects data on exactly what authorized users do 3 what therapies. And, you know, that, unfortunately, 4 is not available. It would be my desire to have that 5 available, so we can be -- make a more educated 6 summary of what your questions pose. But at this 7 point in time, the data is not available.
8 CHAIR PALESTRO: Okay. That doesnt 9 surprise me because we -- this has come up on numerous 10 occasions in the past. And that is one of the reasons 11 that I am concerned, because as we all know, there is 12 -- it takes a long time to effect changes in rules 13 and regulations, and so forth.
14 And my concern has been, continues to be, 15 that I would much prefer to be proactive rather than 16 reactive. And by that I mean I would not like to 17 find out two or three or four years from now that, in 18 fact, there is a shortage, and now we need to do 19 something about it, rather than potentially being 20 proactive and working on developing a plan to prevent 21 that, if you will, a preemptive plan.
22 So thats my concern about your 23 conclusion, and it may be that there is no way to get 24 more substantive data, and thats the way it is. But 25 thank you.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
53 1 VICE CHAIR METTER: So as far as numbers 2 go, there are an increasing interest for -- I know 3 for at least radiology residents to enter the pathway, 4 the redesigned pathway into nuclear radiology 5 positions, because of the increasing therapy.
6 The other thing is I believe your data, 7 you know, is something that we need to look at. But, 8 again, its unavailable, and I think that if you look 9 at the number of authorized users that are going to 10 be coming out of the next four years, there will 11 actually be another 1,000 at least. And so that is 12 going to increase it by, you know, another 1,000 for 13 the 3,600 currently available.
14 And is there -- there is no objective 15 data, like you said, that there is a potential --
16 that there is a current authorized user shortage.
17 And we do have to remember our patients.
18 As Ms. Weil mentioned that, you know, there is, 19 unfortunately, a perhaps geographic barrier regarding 20 the accessibility of certain therapies and 21 treatments, but thats not the specifics for nuclear 22 medicine. Its for any medical and health therapy.
23 And, you know, its our responsibility as 24 regulators to protect the public, and I think 25 protecting the public is to have a confident NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
54 1 individual who understands the entire therapy to be 2 the one that is going to be responsible for the 3 patient.
4 CHAIR PALESTRO: Well, I certainly agree 5 with that and your comment that over the next four 6 years there is going to be an increase of 1,000 7 authorized users, give or take. Again, that comment 8 fails to take into account retirees. There are 9 clearly people who are retiring, and I have no idea 10 what that number is. So while it may remain the same, 11 it may increase by a couple hundred, it may decrease 12 by a couple hundred. Im not sure, but, again, 13 without those data, I would not assume that the number 14 is going to increase by 1,000 over the next four 15 years.
16 VICE CHAIR METTER: Well, all that you 17 know is that the number of authorized users coming 18 out will be another 1,000. We dont know, like you 19 said, how many are going to retire. But at least the 20 number -- you know, there is an increasing number of 21 available authorized users.
22 CHAIR PALESTRO: Okay. Any other 23 comments from the ACMUI?
24 MEMBER ENNIS: Yes. Dr. Palestro, this 25 is Ron.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
55 1 CHAIR PALESTRO: Yes, Dr. Ennis.
2 MEMBER ENNIS: So just -- I have a little 3 more specific information about the radiation 4 oncology piece in terms of distribution, and ASTRO is 5 actually putting together a heat map of distribution 6 across the country.
7 And for the part of our radiation 8 oncology practices that involves the external 9 radiation and treatments, those are, as you might 10 imagine, even more challenging in the sense the 11 patients have to come for daily treatments as opposed 12 to a single or a periodic injection, and yet we are 13 not seeing, you know, any groundswell of concern of, 14 you know, tremendous shortages in the rural areas.
15 Although there are challenges to 16 practices in radiation oncology in rural areas, there 17 is no overwhelming sense of lack of access, so that 18 that would suggest that there are AUs even in the 19 more rural parts available for the nuclear medicine 20 aspects of radiotherapy as well.
21 CHAIR PALESTRO: Thank you, Dr. Ennis.
22 Any other comments from the ACMUI?
23 MEMBER GREEN: Dr. Palestro, this is 24 Richard Green. I just want to, again, repeat your 25 concerns about being proactive versus reactive.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
56 1 Personally, I am involved in seven radionuclide 2 therapy investigational agents that are across the 3 gambit from AML to two for prostate cancer, non-4 Hodgkins lymphoma, and there is -- again, I am 5 dealing with seven.
6 So there is a high likelihood that were 7 going to see multiple radionuclide therapies coming 8 out for a multitude of different targets, cancers, in 9 the next two or three years.
10 Thank you.
11 CHAIR PALESTRO: Thank you, Mr. Green.
12 Any other comments from the ACMUI?
13 MEMBER OUHIB: Yes. Hi. This is Zoubir 14 Ouhib. Just a general comment looking at, you know, 15 can you imagine having 1,000 users doing one case 16 every six months versus having, you know, 500 17 qualified authorized users that will do multiple 18 cases regularly? Their expertise, their skills, and 19 all of that, will be much, much better, and, sure 20 enough, will most likely have less issues and 21 complications. So thats my take.
22 CHAIR PALESTRO: Thank you, Mr. Ouhib.
23 Any other comments from the ACMUI?
24 MEMBER SHEETZ: This is Mike Sheetz.
25 CHAIR PALESTRO: Yes, Mr. Sheetz.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
57 1 MEMBER SHEETZ: The availability of AUs 2 is one issue. But, really, the issue is the number 3 of institutions or licensees, and so one of the 4 questions would be, is the current number of 5 institutions or licensees, you know, insufficient to 6 provide these therapies?
7 And, again, Ill go back to a comment 8 made earlier. These new targeted therapies are very 9 expensive. They are complex. They require a program 10 to be developed and set up, you know, usually with 11 the manufacturer. And so its not something that can 12 be set up at every clinic. Its going to be set up 13 at existing hospitals that have nuclear medicine and 14 radiation oncology programs, for which I assume there 15 will be a sufficient number of AUs, you know, who 16 will meet the requirements.
17 Thank you.
18 CHAIR PALESTRO: Thank you, Mr. Sheetz.
19 Any other comments from the ACMUI?
20 Hearing none, at this point, I would like to open up 21 the phone lines to the public.
22 MS. AYOADE: Dr. Palestro?
23 CHAIR PALESTRO: Yes.
24 MS. AYOADE: This is Maryann Ayoade from 25 NRC. I just wanted to clarify, and I would put up a NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
58 1 slide with regards to the NRC specialty boards that 2 are recognized, as I know you are aware. But I just 3 wanted to clarify that the current certification 4 boards that recognize the NRC for 390 training are 5 the American Board of Nuclear Medicine Certification, 6 the American Board of Radiology for Radiation 7 Oncology, and the American Osteopathic Board of 8 Radiology for Radiation Oncology.
9 With that being said, with regards to 10 this next slide, there is -- there are a number of 11 authorized users in training. We would be looking at 12 -- currently, the ones that recognize the NRC would 13 be the nuclear medicine and the radiation oncology.
14 CHAIR PALESTRO: Correct.
15 MS. AYOADE: For the nuclear radiology, 16 they still have an application with it, so they are 17 not -- they are still not recognized.
18 And then, like Dr. Metter mentioned, the 19 redesigned DR is still in the works. I just wanted 20 to clarify that.
21 CHAIR PALESTRO: Okay. Thank you.
22 OPERATOR: Okay. Thank you. We will now 23 begin the question and answer session. If you would 24 like to ask a question, please press star followed by 25 one, unmute your phone, and record your name clearly.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
59 1 If you need to withdraw your question, press star 2 followed by two.
3 Again, to ask a question, you would press 4 star followed by one. It will take a moment for the 5 questions to come through. Please stand by.
6 MS. HOLIDAY: Before we turn it over to 7 the public comment -- this is Sophie from the NRC, 8 and Id like to ask a question for clarification.
9 OPERATOR: Sure.
10 MS. HOLIDAY: Earlier, Dr. Dilsizian, you 11 made a comment regarding if the subcommittees 12 recommendation was that there would be a number of 13 training and experience hours in addition to 14 competency, and Dr. Palestro stated that the 15 subcommittees recommendation was that we would 16 maintain the current and existing AU pathways, but 17 that if the NRC decided to move forward with the 18 limited scope AU pathways, then the subcommittee 19 would recommend that there be a competency 20 assessment.
21 However, Dr. Palestro, he went forward in 22 saying that the subcommittees recommendation was 23 that there would be a curriculum developed, and then 24 based on the topics being covered in the curriculum, 25 a number of hours would be determined as a result of NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
60 1 that curriculum.
2 So my question is, if that is indeed the 3 recommendation from the Subcommittee, that is not 4 reflected in the subcommittee report. What the report 5 does say is that if -- you know, like as I said, NRC 6 decides to move forward with this, there should be 7 competency assessments, which should not be based on 8 hours9.259259e-5 days <br />0.00222 hours <br />1.322751e-5 weeks <br />3.044e-6 months <br /> or separate attestations.
9 So I think maybe this goes to what Dr.
10 Dilsizian was saying, if this is in fact what the 11 subcommittee is recommending, that there be a 12 curriculum, and then from the curriculum a 13 delineation of hours, then the report would need to 14 be revised to reflect that.
15 Is that, in fact, the Subcommittees 16 position?
17 CHAIR PALESTRO: Sophie, this is Dr.
18 Palestro. Thank you for pointing that out to me.
19 And, Dr. Metter, really this question is 20 for you and the subcommittee.
21 VICE CHAIR METTER: Well, that part of it 22 has not been discussed with the subcommittee. And 23 right now, this report is that right now we dont 24 recommend hours. But we do recommend that the 25 curriculum be established, and right now we do not NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
61 1 recommend the limited scope AU pathway.
2 We would be happy to work with NRC staff 3 if the NRC decides to go on a limited scope pathway, 4 and that will be determined at that time.
5 MS. HOLIDAY: So my understanding is that 6 you are saying, again, that the subcommittee does not 7 recommend the development of the limited AU pathway.
8 However, should the NRC move forward, then the ACMUI 9 would like to work with the NRC to develop such a 10 curriculum?
11 VICE CHAIR METTER: Yes.
12 MS. HOLIDAY: Okay. Thank you.
13 MEMBER DILSIZIAN: Vasken here. Just --
14 can I just add one comment? A curriculum has been 15 forwarded I think to us -- by the Society of Nuclear 16 Medicine and Molecular Imaging of what entails to 17 have the training to be an authorized user, correct, 18 a 700-hour pathway? Its just that the curriculum 19 hasnt been assigned hours; is that correct? We do 20 have a curriculum in hand.
21 VICE CHAIR METTER: The subcommittee will 22 have to look at that curriculum and review it.
23 MEMBER DILSIZIAN: Okay.
24 CHAIR PALESTRO: All right. There being 25 no other comments, if we can move ahead to comments NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
62 1 from the public?
2 OPERATOR: Our first question is coming 3 from Carol Marcus. Your line is now open. You may 4 begin.
5 DR. MARCUS: Thank you very much. I am 6 a nuclear medicine physician from UCLA. I have two 7 comments. One is, since the SNMMI, the ACNM, the 8 ACR, ASTRO, and the ACMUI have all recommended against 9 this NRC rulemaking, it would be very strange to me 10 if the NRC, which has no medical competence 11 whatsoever, would go against the unanimous opinion of 12 the medically competent groups involved. Thats my 13 first comment.
14 My second comment is that most of these 15 therapies require sophisticated imaging studies ahead 16 of time to ascertain whether the patients are good 17 candidates for the therapies. And this kind of 18 sophisticated imaging is found in urban areas. It is 19 not found in rural areas either. And since the 20 patients are going to have to have sophisticated 21 imaging procedures in urban areas, they certainly 22 might as well get their therapy there.
23 Nobody has talked about how they would 24 get these sophisticated imaging studies in a rural 25 area with some limited-competence doctor giving the NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
63 1 therapy. And it doesnt make sense. They have to 2 have the diagnostic tests first, in most cases.
3 Thats the end of my comment.
4 CHAIR PALESTRO: Thank you, Dr. Marcus.
5 Other comments from the public?
6 OPERATOR: Yes, sir. Our next question 7 is going to come from Dr. Paul Wallner. Your line is 8 now open.
9 DR. WALLNER: Thank you. My name is Dr.
10 Paul Wallner. Im a practicing radiation oncologist.
11 Im speaking today on behalf of the American College 12 of Radiology.
13 The ACR represents approximately 38,000 14 members, including diagnostic and interventional 15 radiologists, nuclear medicine physicians, radiation 16 oncologists, and medical physicists, many of whom are 17 authorized users, authorized medical physicists, or 18 radiation safety officers.
19 We commend the ACMUI Subcommittee for its 20 thoughtful review of this controversial topic, and we 21 generally concur with the subcommittees analysis and 22 recommendations.
23 I would like to make three comments 24 related to some of the discussion earlier today. One 25 is that there is definitely an increase in the number NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
64 1 of nuclear radiologists, diagnostic radiologists, who 2 are authorized user-eligible, and radiation 3 oncologists. And that number is based on increasing 4 actual real numbers and increasing number of training 5 programs. Those numbers do exceed the number of 6 retirees.
7 Secondly, related to the issue of hours, 8 I would urge the Commission to avoid placing specific 9 hours1.041667e-4 days <br />0.0025 hours <br />1.488095e-5 weeks <br />3.4245e-6 months <br /> on any of these core competencies. The ACGME 10 requires that all of its training programs review 11 core competency of all trainees at least twice a year.
12 Residents cannot advance to the next level of training 13 unless they have fulfilled all of those competencies.
14 And a third issue related to the new --
15 potential new agents in the pipeline, I would suggest 16 that history would suggest that many of these agents 17 will never reach the marketplace, number one. And, 18 number two, as we are seeing over the last 30 or 40 19 years, except for iodine-131, many of these agents 20 enter the marketplace and then are quickly changed -
21 - it is quickly eliminated by disruptive technologies 22 or interventions.
23 With regard to topic 3 of the 24 subcommittee report, I would note that the American 25 Board of Radiology and the American Board of Nuclear NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
65 1 Medicine currently employ maintenance and 2 certification programs to facilitate career-long 3 assessments for its diplomates, and their previous 4 issues are evaluated and assessed on an ongoing basis.
5 We strongly agree with the subcommittee 6 recommendation against adoption of a limited scope AU 7 pathway mechanism that we believe would fail to 8 provide reasonable assurance of the adequate 9 protection of health and safety. Moreover, any such 10 regulatory changes would greatly increase risk and 11 introduce unintended consequences, conflicts, and 12 burdens that would far outweigh the theoretical and 13 unrealistic rural access improvements suggested by 14 drug manufacturers.
15 We hope that the NRC staff future 16 recommendations will prioritize health and safety in 17 radiopharmaceutical therapy, and follow the advice of 18 the ACMUI members and patient care experts.
19 Thank you for this opportunity.
20 CHAIR PALESTRO: Thank you. Other 21 comments?
22 OPERATOR: Yes, there is. Our next 23 comment is coming from Bennett Greenspan. Your line 24 is now open.
25 DR. GREENSPAN: Thank you. Can you hear NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
66 1 me?
2 CHAIR PALESTRO: Yes.
3 DR. GREENSPAN: I just want to make a few 4 comments. First of all, I agree with the assessment 5 from the subcommittee, and I agree with their 6 recommendations. I also agree that a competency 7 assessment is a far better approach than number of 8 hours9.259259e-5 days <br />0.00222 hours <br />1.322751e-5 weeks <br />3.044e-6 months <br />. I also agree that the NRC is charged with 9 protecting the public, and thats good, and thats 10 why in rural areas with people with limited abilities, 11 thats really not sufficient, and I think patients 12 would not be protected being treated by physicians 13 who dont really understand what they are dealing 14 with.
15 As far as a curriculum, I am not 16 representing the Society here, but I did formulate a 17 curriculum for the Society of Nuclear Medicine and 18 Molecular Imaging, which was submitted last summer, 19 and that is one -- one example of a curriculum. I 20 think the ACMUI may be developed one also.
21 Numerous societies are against having a 22 limited scope pathway. As Carol Marcus suggested, 23 that includes now ACMUI and the board certifications 24 ABNM and ABR, but also SNMMI, ACNM, ACR, ASTRO, which 25 you all mentioned previously, and also AAPM, the NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
67 1 American Association of Physicists in Medicine.
2 Im not specifically representing them 3 here either, but all of these societies recommend 4 against a limited scope authorized user pathway, and 5 I agree with that as well.
6 Thank you very much.
7 CHAIR PALESTRO: Thank you. Other 8 comments?
9 OPERATOR: We do have another comment.
10 This next participant, we do not have your name 11 recorded, but your line is open.
12 DR. RAZMARIA: Hello? Can you hear me?
13 CHAIR PALESTRO: Yes.
14 DR. RAZMARIA: Hi. Yeah. My name is 15 Aria Razmaria. I am a senior resident training in 16 nuclear medicine. As training physicians, we 17 appreciate the efforts by the ACMUI to provide insight 18 into the concerns of shortage of authorized users and 19 bring data into the discussion rather than just 20 assumption.
21 As to just physicians, our primary 22 obligation is toward the patients. Radionuclide 23 therapies are highly proven therapies with high 24 potential for serious side effects. We appreciate a 25 strong recommendation towards maintaining the NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
68 1 training and experience requirements, but are puzzled 2 by discussion of the option of a possibility of NRC 3 -- (telephone interference).
4 NRC is charged with radiation safety of 5 the public. As physicians, we are tasked with a 6 responsibility towards each and every patient, and 7 high quality of care. We have undermined the 8 authority and it causes confusion in the 9 recommendation. If there is a clause mentioned in 10 the staff recommendation, it is not solid.
11 As physicians, our primary concern is 12 towards patients, and above all to maintain the 13 patient interest, above all -- despite all economic 14 pressures and insights. We strongly maintain that 15 radionuclide therapies are highly individualized 16 therapeutic modality and require thorough and in-17 depth trainings that are already established by 18 specialty boards. Creating new competency-assessing 19 tools would be a significant undertaking, going above 20 NRCs purview and undermine existing competency 21 assessment mechanisms in place by specialty boards.
22 In summary, the subcommittee report 23 sufficiently underlines the fact that there is no 24 shortage of authorized users, which are the original 25 arguments of creating alternate pathways.
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
69 1 If NRC decides to pursue the creation of 2 new pathways with limited training requirements, as 3 physicians we cannot share responsibility for any 4 consequences a physician might have, the health of 5 the patients, and NRC has to be the sole bearer of 6 responsibility to the public for any consequences 7 this rulemaking might have.
8 Thank you for your attention.
9 CHAIR PALESTRO: Thank you. Other 10 comments?
11 OPERATOR: Yes. Our next comment comes 12 from George Segall. Your line is now open.
13 DR. SEGALL: Thank you very much. I am 14 the Executive Director of the American Board of 15 Nuclear Medicine, as well as a practicing physician, 16 and Id like to state that the Board strongly 17 recommends -- strongly supports the recommendation of 18 the Training and Education Subcommittee.
19 A few interesting questions came up 20 during the conference today. There was one question 21 regarding the distribution of authorized users in the 22 United States. I think the NRC is best positioned to 23 answer this question. However, in the comment letter 24 submitted by the American Board of Nuclear Medicine, 25 we did give you the distribution data of our NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
70 1 diplomates who are the largest block of diplomates 2 who are authorized users. And the numbers are in the 3 letter, but I can say that we have authorized users 4 in all 50 states and territories.
5 Dr. Palestro, during his public 6 conference call, queried about the number of retiring 7 diplomates and whether the pipeline would more than 8 offset the number of retired diplomates. I think all 9 specialty boards keep track of this information. The 10 ABNM can certainly supply it.
11 I can tell you, just as I was looking up 12 while listening to these proceedings, that in the 13 past four years the ABNM has had an average number of 14 retirees of 48 per year, which is below the average 15 number of new diplomates for this period of time, 16 which is 63.
17 So we do not see any sort of attrition in 18 the total number of diplomates, and has -- and as has 19 been brought up by other participants in this call, 20 there are new training pathways that are encouraging 21 increasing numbers of young professionals to seek the 22 training necessary to become authorized users.
23 During this conference call, there was an 24 interesting proposal that the limited authorized user 25 pathway would be predicated on having an authorized NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
71 1 nuclear pharmacist to be responsible for the nuclear 2 safety aspects of treatment. This just wouldnt work.
3 The authorized nuclear pharmacist would not be onsite 4 and could not provide the level of coordination 5 necessary of all of the personnel that is required in 6 handling a spill of radioactive materials that do 7 occur.
8 And the coordination of these personnel 9 involve physicians, technologists, as well as 10 physicists. So the authorized user responsible for 11 the radiation safety must be actually physically 12 present in the department.
13 Then the other question came out about 14 de minimis standards for the duration of training, 15 and lets say all education met de minimis standards 16 in terms of duration, which is a practical necessity.
17 In terms of radiation safety, 18 particularly with skills, these are unexpected events 19 that are relatively uncommon. The best training 20 occurs on -- when an event like this happens and the 21 training is present. The likelihood of acquiring 22 this training is directly proportional to the 23 duration of training.
24 Similar events involving radioactive 25 materials used in -- used for diagnosis occur more NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
72 1 frequently, and though they are still uncommon. The 2 likelihood of gaining the necessary experience to 3 fully and effectively respond to these situations 4 cannot be gained in a short period of time.
5 To directly answer the question of one of 6 the committee members regarding how much time is 7 devoted to 35.390 training, in the ACGME-accredited 8 residency training programs that lead to 9 certification by boards who have deemed-status from 10 the NRC, the minimum training -- the minimum -- is 16 11 months, and as has been noted earlier, up to three 12 years.
13 And training is gaining the experience to 14 handle spills and other unexpected events throughout 15 their training period. It is a continuous educational 16 process. And as a practical matter, there must be 17 de minimis standards, and my expert opinion is that 18 700 hours0.0081 days <br />0.194 hours <br />0.00116 weeks <br />2.6635e-4 months <br /> is really a minimum requirement to obtain 19 the experience necessary for competency.
20 Thank you.
21 CHAIR PALESTRO: Thank you. Other 22 comments?
23 OPERATOR: We show no other comments at 24 this time. As a reminder, if you would like to make 25 a comment, or if you have a question, please press NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
73 1 star followed by one.
2 CHAIR PALESTRO: Ms. Holiday?
3 MS. HOLIDAY: Yes, Dr. Palestro.
4 CHAIR PALESTRO: Yes. Next, the ACMUI 5 needs to vote on the draft report; am I correct?
6 MS. HOLIDAY: That is correct. If there 7 are no additional comments from members of the public, 8 you are free to go ahead and make a motion to vote on 9 the report.
10 CHAIR PALESTRO: The report itself, am I 11 correct, isnt that the motion and we need a second?
12 MS. HOLIDAY: Correct. And that would be 13 the report including all of the recommendations 14 within the report.
15 CHAIR PALESTRO: Yes. All right. So do 16 we have a second for the subcommittees report?
17 MS. HOLIDAY: Well, is there a motion?
18 CHAIR PALESTRO: Well, the report itself 19 is the motion; is it not?
20 MS. HOLIDAY: No. Generally, a member on 21 the committee --
22 CHAIR PALESTRO: Im sorry. I 23 misunderstood.
24 MS. HOLIDAY: Thats okay.
25 CHAIR PALESTRO: All right. Can we have NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
74 1 a motion?
2 MEMBER ENNIS: Motion to approve the 3 report.
4 MEMBER OHARA: Second.
5 CHAIR PALESTRO: Second?
6 MEMBER OHARA: Yes. Second.
7 CHAIR PALESTRO: All right. Dr. Ennis 8 made the original motion; am I correct?
9 MEMBER ENNIS: Yes.
10 CHAIR PALESTRO: And who made the second, 11 for the transcriptionist?
12 MEMBER OHARA: Michael OHara.
13 CHAIR PALESTRO: Dr. OHara seconded.
14 Any discussion?
15 VICE CHAIR METTER: This is Darlene. I 16 would like to -- Sophie, you had made a comment about 17 developing the curriculum. Would that need to be in 18 this report?
19 MS. HOLIDAY: If that is a recommendation 20 from the Subcommittee, at this time, you would need 21 to make a vote on that before the overall vote on the 22 report. So my suggestion would be that a motion be 23 put forward for the subcommittee report to be amended 24 to include language reflecting that the subcommittee 25 recommends or maintains its recommendation that it NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
75 1 does not support a limited scope AU pathway. However, 2 should the NRC move forward with developing such a 3 pathway, that the NRC staff and the ACMUI work 4 together to develop the core topics in the curriculum.
5 Is that the recommendation?
6 VICE CHAIR METTER: That is exactly the 7 recommendation. I move that that be introduced as 8 the motion, to make that recommendation.
9 CHAIR PALESTRO: Do we have a second?
10 MEMBER SCHLEIPMAN: Robert Schleipman. I 11 second that.
12 CHAIR PALESTRO: All right. All in favor?
13 (Chorus of ayes.)
14 CHAIR PALESTRO: Any opposed?
15 All right. Now we will go back and vote 16 on the original motion.
17 MS. HOLIDAY: Dr. Palestro --
18 CHAIR PALESTRO: Yes.
19 MS. HOLIDAY: -- if I may, were there any 20 abstentions?
21 CHAIR PALESTRO: Im sorry. Yes. Any 22 abstentions?
23 MS. HOLIDAY: Thank you.
24 CHAIR PALESTRO: Now we will go back to 25 vote on the original motion. All in favor?
NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
76 1 (Chorus of ayes.)
2 CHAIR PALESTRO: Any opposed? I am going 3 to oppose it for the reasons that I expressed 4 initially. My concerns -- and I do thank Dr. Segall 5 for his input on retirees. That was certainly very 6 helpful. But before I am comfortable with saying 7 there is no shortage, or there is no potential for a 8 shortage, I would like to see more substantive data.
9 And, again, my biggest concern is the 10 long lag time between initiation and completion of 11 rules changes, and so forth. So I am going to go on 12 the record as being opposed.
13 Any abstentions? All right.
14 MS. HOLIDAY: Okay. Thank you. So I 15 have that the motion passes to approve the report 16 with the amendment regarding the development of the 17 curriculum between the NRC staff and the ACMUI, and 18 that the ACMUI approved the report with one opposing 19 vote.
20 VICE CHAIR METTER: Correct.
21 CHAIR PALESTRO: Correct.
22 MS. HOLIDAY: Okay.
23 CHAIR PALESTRO: We can adjourn the 24 meeting, Ms. Holiday?
25 MS. HOLIDAY: That is correct. So, at NEAL R. GROSS COURT REPORTERS AND TRANSCRIBERS 1323 RHODE ISLAND AVE., N.W.
(202) 234-4433 WASHINGTON, D.C. 20005-3701 (202) 234-4433
77 1 this time, I would like to thank all of the ACMUI 2 members, particularly the ACMUI Subcommittee members, 3 for developing this subcommittee report and for 4 discussing it today.
5 Thank you to the Committee members, NRC 6 staff, and members of the public who have participated 7 and provided thoughtful comment and feedback during 8 this time.
9 As I said earlier in this meeting, the 10 letter that was submitted to the NRC staff from UPPI 11 will be appended to the meeting transcript as part of 12 the official record.
13 If you joined this call after the call 14 started, please make sure that you send an email to 15 Kellee Jamerson, K-E-L-L-E-E dot J-A-M-E-R-S-O-N at 16 nrc.gov, and myself, Sophie Holiday, S-O-P-H-I-E dot 17 H-O-L-I-D-A-Y at nrc.gov, so that we may 18 appropriately capture you as a participant of the 19 meeting.
20 With that being said, thank you for your 21 time, and the meeting is adjourned.
22 CHAIR PALESTRO: Thank you.
23 (Whereupon, the above-entitled matter 24 went off the record at 11:49 a.m.)