ANALYTICAL REPORT TestAmerica Laboratories, Inc.

TestAmerica St. Louis 13715 Rider Trail North Earth City, MO 63045 Tel: (314)298-8566 TestAmerica Job ID: 160-11841-1 Client Project/Site: HDP For:

Perma-Fix Environmental Services Inc.

2800 Solway Road Knoxville, Tennessee 37931 Attn: Brian Miller Authorized for release by:

5/26/2015 5:10:37 PM Elizabeth Hoerchler, Project Mgmt. Assistant elizabeth.hoerchler@testamericainc.com Designee for Ivan Vania, Project Manager II (314)298-8566 ivan.vania@testamericainc.com This report has been electronically signed and authorized by the signatory. Electronic signature is intended to be the legally binding equivalent of a traditionally handwritten signature.

Results relate only to the items tested and the sample(s) as received by the laboratory.

1 2

3 4

5 6

7 8

9 10 11 12 13

Table of Contents Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP TestAmerica Job ID: 160-11841-1 Page 2 of 19 TestAmerica St. Louis 5/26/2015 Cover Page..............................................

1 Table of Contents.........................................

2 Case Narrative...........................................

3 Chain of Custody..........................................

5 Receipt Checklists.........................................

6 Definitions/Glossary.......................................

7 Method Summary.........................................

8 Sample Summary.........................................

9 Detection Summary........................................

10 Client Sample Results......................................

11 QC Sample Results........................................

14 QC Association Summary...................................

17 Tracer Carrier Summary....................................

19 1

2 3

4 5

6 7

8 9

10 11 12 13

Case Narrative Client: Perma-Fix Environmental Services Inc.

TestAmerica Job ID: 160-11841-1 Project/Site: HDP Job ID: 160-11841-1 Laboratory: TestAmerica St. Louis Narrative CASE NARRATIVE Client: Perma-Fix Environmental Services Inc.

Project: HDP Report Number: 160-11841-1 With the exceptions noted as flags or footnotes, standard analytical protocols were followed in the analysis of the samples and no problems were encountered or anomalies observed. In addition all laboratory quality control samples were within established control limits, with any exceptions noted below. Each sample was analyzed to achieve the lowest possible reporting limit within the constraints of the method. In some cases, due to interference or analytes present at high concentrations, samples were diluted. For diluted samples, the reporting limits are adjusted relative to the dilution required.

TestAmerica St. Louis attests to the validity of the laboratory data generated by TestAmerica facilities reported herein. All analyses performed by TestAmerica facilities were done using established laboratory SOPs that incorporate QA/QC procedures described in the application methods. TestAmericas operations groups have reviewed the data for compliance with the laboratory QA/QC plan, and data have been found to be compliant with laboratory protocols unless otherwise noted below.

The test results in this report meet all NELAP requirements for parameters for which accreditation is required or available. Any exceptions to NELAP requirements are noted in this report. Pursuant to NELAP, this report may not be reproduced, except in full, without the written approval of the laboratory.

Calculations are performed before rounding to avoid round-off errors in calculated results.

All holding times were met and proper preservation noted for the methods performed on these samples, unless otherwise detailed in the individual sections below.

All solid sample results for Chemistry analyses are reported on an "as received" basis unless otherwise indicated by the presence of a %

solids value in the method header. All soil/sediment sample results for radiochemistry analyses are based upon sample as dried and disaggregated with the exception of tritium, carbon-14, and iodine-129 by gamma spectroscopy unless requested as wet weight by the client.

This laboratory report is confidential and is intended for the sole use of TestAmerica and its client.

RECEIPT The samples were received on 05/14/2015; the samples arrived in good condition, properly preserved. The temperature of the coolers at receipt was 20.0° C.

TECHNETIUM-99 (ICPMS)

Samples L10-13-14-B-E-B-00 (160-11841-1), L10-13-15-B-E-B-00 (160-11841-2), L10-14-13-B-E-B-00 (160-11841-3) and L10-14-14-B-E-B-00 (160-11841-4) were analyzed for Technetium-99 (ICPMS) in accordance with EPA SW-846 Method 6020A. The samples were prepared on 05/18/2015 and analyzed on 05/19/2015.

No analytical or quality issues were noted, other than those described above or in the Definitions/Glossary page.

TECHNETIUM-99 (ICPMS)

Samples L10-13-14-B-E-B-00 (160-11841-1), L10-13-15-B-E-B-00 (160-11841-2), L10-14-13-B-E-B-00 (160-11841-3) and L10-14-14-B-E-B-00 (160-11841-4) were analyzed for Technetium-99 (ICPMS) in accordance with EPA SW-846 Method 6020A. The samples were prepared on 05/18/2015 and analyzed on 05/19/2015.

TestAmerica St. Louis Page 3 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

Case Narrative Client: Perma-Fix Environmental Services Inc.

TestAmerica Job ID: 160-11841-1 Project/Site: HDP Job ID: 160-11841-1 (Continued)

Laboratory: TestAmerica St. Louis (Continued)

No analytical or quality issues were noted, other than those described above or in the Definitions/Glossary page.

PERCENT SOLIDS Samples L10-13-14-B-E-B-00 (160-11841-1), L10-13-15-B-E-B-00 (160-11841-2), L10-14-13-B-E-B-00 (160-11841-3) and L10-14-14-B-E-B-00 (160-11841-4) were analyzed for percent solids in accordance with EPA Method 160.3 MOD. The samples were analyzed on 05/16/2015.

No analytical or quality issues were noted, other than those described above or in the Definitions/Glossary page.

CESIUM-137 & OTHER GAMMA EMITTERS (GS)

Samples L10-13-14-B-E-B-00 (160-11841-1), L10-13-15-B-E-B-00 (160-11841-2), L10-14-13-B-E-B-00 (160-11841-3) and L10-14-14-B-E-B-00 (160-11841-4) were analyzed for Cesium-137 & Other Gamma Emitters (GS) in accordance with DOE GA-01-R. The samples were leached on 05/18/2015, and prepared and analyzed on 05/20/2015.

Prep batch: 191719 Insufficient volume was provided for the following sample to fill the client requested marn soil geometry: L10-13-14-B-E-B-00 (160-11841-1), L10-13-15-B-E-B-00 (160-11841-2), L10-14-13-B-E-B-00 (160-11841-3), L10-14-14-B-E-B-00 (160-11841-4) and (160-11841-A-1-A DU) Samples were put into a tuna can instead.

No additional analytical or quality issues were noted, other than those described above or in the Definitions/Glossary page.

TestAmerica St. Louis Page 4 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

Page 5 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13 TestAmerica St. Louis 13715 Rider Trail North Earth City, MO 63045 phone 314.298.8566 fax Client Contact Perrna-Fix Environmental Services 2800 Solway Road Knoxville, TN 37931 (865) 690-0501 Phone (865) 539-9868 FAX Chain of Custody Record Regulatory Program: D DW D NPDES O RCRA 0 Other:

Project Manager: Brian Miller TeUFax: (314) 810-3333 Analysis Turnaround Time 0 CALENDAR DAYS O WORKING DAYS TAT if differentfrom Below ___ _

D 2wee1-

~

e f

0, Date: 05/14/2015 Carrier: Courier 1

1 ° r

C

-Cl)

O 2days

.!!1:;;

D i i"amp.e I 1!

~

i:o E Project Name: Hematite Decommissioning Project Site: Westinghouse Hematite, MO P 0#832953

.s

~

E..

c.,

a,::,

Q. 0 Cl).c:

.. 0, E.!:

e"lii o,"lt 0,

CD O

3; !:::.

~

Sample Identification Sample I Sample Date Time L 10-13-14-B-E-B-OO 5/14/15 1415 L 10-13-15-B-E-B-OO 5/14/15 1425 L 10-14-13-B-E-B-OO 5/14/15 1430 L 10-14-14-B-E-B-OO 5/14115 1435 I ~ -----

Type (C=Comp, I I # of G=Grab)

Matrix Cont.

*~

(I)

u. °"

G Soil 1

IN G

Soil 1 IN G

Soil 1 IN G

Soil 1

IN C) 3:

~ g,j Q.

< ~ ~.

X XI X X

XIX X

XIX X

XI X Test America THE LEADER IN ENVIRONMENTAL TESTING TestAmerica Laboratories, Inc.

COC No: 051415-FSS2 1

of 1

COCs Sampler:

For Lab Use Only:

Walk-in Client:

I I

Lab Sampling:

Job I SDG No.:

Sample Specific Notes:

Matt Cushman Matt Cushman Matt Cushman Matt Cushman

___---7

~ ~

\\I\\\\\\\\\\\\\\I\\\\\\\\\\\\\\\\\\\\\\\\\\\\I\\\\\\\\I 160-11841 Chain of Custody

--r---+-_

N I

~

5 A

l

  • -- -~!,--+--+-1--1--

r

~

er~Ntttl'PniW,s$ "'%Jt~'J'.~.~~ 1a:~~tl~ ~~~t~!fJl{.Jr'-~1if?~ffi:i...~ll-~R5fC!ft\\e"f~~n~;1:?1'I1,r1l;;ti~~'~*1~1Pll~,;{J1?1A¥;

Possible Hazard Identification:

Sample Disposal ( A fee may be assessed if samples are retained longer than 1 month)

Are any samples from a listed EPA Hazardous Waste? Please List any EPA Waste Codes for the sample in the

,,comments Section if the lab is to dispose of the sample.

0 Non-Hazard O Flammable O Skin Irritant O Poison B O Unknown 0 RebJm to Client 0 Disposal by Lab 0 Archive for Months Special Instructions/QC Requirements & Comments: Potential for Radiological Contamination Custody Seals Intact:

0 Yes D No Corr'd: ___

Therm ID No.:

Company:

T4 I t.,5"L 1~~zv~

Relinquished by:-

Company:

Relinquished by:

Company:

Date/Time:

Received in Laboratory by:

Company:

Date/Time:

I Hemallie Decommronmg P roject I

Pe-mafu< (for Westinghoose) Non-Propaetar.y O ass 3 ! -

reJi!JP.t, W A-QGS. Cham cfCuswdy j.....

j

Login Sample Receipt Checklist Client: Perma-Fix Environmental Services Inc.

Job Number: 160-11841-1 Login Number: 11841 Question Answer Comment Creator: Daniels, Brian J List Source: TestAmerica St. Louis List Number: 1 True Radioactivity wasn't checked or is </= background as measured by a survey meter.

True The cooler's custody seal, if present, is intact.

N/A Sample custody seals, if present, are intact.

True The cooler or samples do not appear to have been compromised or tampered with.

N/A Samples were received on ice.

True Cooler Temperature is acceptable.

True Cooler Temperature is recorded.

True COC is present.

True COC is filled out in ink and legible.

True COC is filled out with all pertinent information.

N/A Is the Field Sampler's name present on COC?

True There are no discrepancies between the containers received and the COC.

True Samples are received within Holding Time.

True Sample containers have legible labels.

True Containers are not broken or leaking.

True Sample collection date/times are provided.

True Appropriate sample containers are used.

True Sample bottles are completely filled.

True Sample Preservation Verified.

True There is sufficient vol. for all requested analyses, incl. any requested MS/MSDs N/A Containers requiring zero headspace have no headspace or bubble is

<6mm (1/4").

True Multiphasic samples are not present.

True Samples do not require splitting or compositing.

N/A Residual Chlorine Checked.

TestAmerica St. Louis Page 6 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

Definitions/Glossary TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Qualifiers Metals Qualifier Description J

Result is less than the RL but greater than or equal to the MDL and the concentration is an approximate value.

Qualifier Rad Qualifier Description U

Result is less than the sample detection limit.

Qualifier Glossary These commonly used abbreviations may or may not be present in this report.

Listed under the "D" column to designate that the result is reported on a dry weight basis Abbreviation

%R Percent Recovery CFL Contains Free Liquid CNF Contains no Free Liquid DER Duplicate error ratio (normalized absolute difference)

Dil Fac Dilution Factor DL, RA, RE, IN Indicates a Dilution, Re-analysis, Re-extraction, or additional Initial metals/anion analysis of the sample DLC Decision level concentration MDA Minimum detectable activity EDL Estimated Detection Limit MDC Minimum detectable concentration MDL Method Detection Limit ML Minimum Level (Dioxin)

NC Not Calculated ND Not detected at the reporting limit (or MDL or EDL if shown)

PQL Practical Quantitation Limit QC Quality Control RER Relative error ratio RL Reporting Limit or Requested Limit (Radiochemistry)

RPD Relative Percent Difference, a measure of the relative difference between two points TEF Toxicity Equivalent Factor (Dioxin)

TEQ Toxicity Equivalent Quotient (Dioxin)

TestAmerica St. Louis Page 7 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

Method Summary TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Method Method Description Laboratory Protocol SW846 6020A Metals (ICP/MS), Tc-99 TAL SL EPA Moisture Percent Moisture TAL SL SW846 6020A Metals (ICP/MS), Tc-99 in Activity TAL SL DOE GA-01-R Cesium-137 & Other Gamma Emitters (GS)

TAL SL Protocol

References:

DOE = U.S. Department of Energy EPA = US Environmental Protection Agency SW846 = "Test Methods For Evaluating Solid Waste, Physical/Chemical Methods", Third Edition, November 1986 And Its Updates.

Laboratory

References:

TAL SL = TestAmerica St. Louis, 13715 Rider Trail North, Earth City, MO 63045, TEL (314)298-8566 TestAmerica St. Louis Page 8 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

Sample Summary TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Lab Sample ID Client Sample ID Received Collected Matrix 160-11841-1 L10-13-14-B-E-B-00 Solid 05/14/15 14:15 05/14/15 16:54 160-11841-2 L10-13-15-B-E-B-00 Solid 05/14/15 14:25 05/14/15 16:54 160-11841-3 L10-14-13-B-E-B-00 Solid 05/14/15 14:30 05/14/15 16:54 160-11841-4 L10-14-14-B-E-B-00 Solid 05/14/15 14:35 05/14/15 16:54 TestAmerica St. Louis Page 9 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

Detection Summary TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Client Sample ID: L10-13-14-B-E-B-00 Lab Sample ID: 160-11841-1

Technetium 99 RL 0.000082 mg/Kg MDL 0.000025 Analyte Result Qualifier Unit Dil Fac D Method Prep Type Total/NA 1

J 0.000031 6020A Client Sample ID: L10-13-15-B-E-B-00 Lab Sample ID: 160-11841-2 No Detections.

Client Sample ID: L10-14-13-B-E-B-00 Lab Sample ID: 160-11841-3 No Detections.

Client Sample ID: L10-14-14-B-E-B-00 Lab Sample ID: 160-11841-4 No Detections.

TestAmerica St. Louis This Detection Summary does not include radiochemical test results.

Page 10 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

Client Sample Results TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Lab Sample ID: 160-11841-1 Client Sample ID: L10-13-14-B-E-B-00 Matrix: Solid Date Collected: 05/14/15 14:15 Percent Solids: 71.3 Date Received: 05/14/15 16:54 Method: 6020A - Metals (ICP/MS), Tc-99 RL MDL Technetium 99 0.000031 J 0.000082 0.000025 mg/Kg

05/18/15 09:50 05/19/15 13:01 1

Analyte Dil Fac Analyzed Prepared Unit D

Result Qualifier Method: 6020A - Metals (ICP/MS), Tc-99 in Activity Analyte Technetium 99 0.528 (2+/-)

0.171 (2+/-)

1 05/19/15 13:01 05/18/15 09:50 pCi/g 0.280 1.64 RL MDC 0.163 Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Uncert.

Uncert.

Re 30 - 110 Carrier 05/18/15 09:50 05/19/15 13:01 1

Dil Fac Analyzed Prepared Qualifier Limits

%Yield 86 Method: GA-01-R - Cesium-137 & Other Gamma Emitters (GS)

Analyte Actinium 228 1.36 (2+/-)

0.197 (2+/-)

1 05/20/15 17:43 05/20/15 14:46 pCi/g 0.103 RL MDC 0.140 Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Uncert.

Uncert.

pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.0971 0.0586 0.0585 U

0.0301 Americium 241 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.393 0.448 0.425 1.39 Bismuth 212 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.0650 0.100 0.0719 0.673 Bismuth 214 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.0525 0.177 0.0610 1.28 Lead 212 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.0621 0.110 0.0706 0.812 Lead 214 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.344 2.00 0.807 17.9 Potassium 40 pCi/g 05/20/15 14:46 05/20/15 17:43 1

3.86 2.33 2.32 U

1.58 Protactinium 234m pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.983 0.259 0.251 U

0.572 Protactinium 231 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.0650 0.100 0.0719 0.673 Radium 226 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.840 0.548 0.530 1.31 Thorium 234 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.180 0.223 0.144 1.68 Uranium 235 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.103 0.197 0.140 1.36 Thorium 232 Tl-208 0.467 0.0460 0.0668 0.0333 pCi/g 05/20/15 14:46 05/20/15 17:43 1

(2+/-)

(2+/-)

MDC RL Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Other Detected Radionuclides Uncert.

Uncert.

Lab Sample ID: 160-11841-2 Client Sample ID: L10-13-15-B-E-B-00 Matrix: Solid Date Collected: 05/14/15 14:25 Percent Solids: 75.9 Date Received: 05/14/15 16:54 Method: 6020A - Metals (ICP/MS), Tc-99 RL MDL Technetium 99 ND 0.000076 0.000023 mg/Kg

05/18/15 09:50 05/19/15 13:05 1

Analyte Dil Fac Analyzed Prepared Unit D

Result Qualifier Method: 6020A - Metals (ICP/MS), Tc-99 in Activity Analyte Technetium 99 0.0548 U (2+/-)

0.0761 (2+/-)

1 05/19/15 13:05 05/18/15 09:50 pCi/g 0.260 1.52 RL MDC 0.0760 Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Uncert.

Uncert.

Re 30 - 110 Carrier 05/18/15 09:50 05/19/15 13:05 1

Dil Fac Analyzed Prepared Qualifier Limits

%Yield 87 TestAmerica St. Louis Page 11 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

Client Sample Results TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Lab Sample ID: 160-11841-2 Client Sample ID: L10-13-15-B-E-B-00 Matrix: Solid Date Collected: 05/14/15 14:25 Date Received: 05/14/15 16:54 Method: GA-01-R - Cesium-137 & Other Gamma Emitters (GS)

Analyte Actinium 228 1.43 (2+/-)

0.211 (2+/-)

1 05/20/15 17:48 05/20/15 14:46 pCi/g 0.145 RL MDC 0.153 Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Uncert.

Uncert.

pCi/g 05/20/15 14:46 05/20/15 17:48 1

0.117 0.0822 0.0822 U

0.0253 Americium 241 pCi/g 05/20/15 14:46 05/20/15 17:48 1

0.565 0.638 0.611 1.78 Bismuth 212 pCi/g 05/20/15 14:46 05/20/15 17:48 1

0.0807 0.155 0.111 1.04 Bismuth 214 pCi/g 05/20/15 14:46 05/20/15 17:48 1

0.0834 0.188 0.0822 1.30 Lead 212 pCi/g 05/20/15 14:46 05/20/15 17:48 1

0.0784 0.141 0.0871 1.06 Lead 214 pCi/g 05/20/15 14:46 05/20/15 17:48 1

0.342 2.16 1.07 18.3 Potassium 40 pCi/g 05/20/15 14:46 05/20/15 17:48 1

6.37 2.59 2.55 U

4.23 Protactinium 234m pCi/g 05/20/15 14:46 05/20/15 17:48 1

1.45 0.271 0.266 U

0.491 Protactinium 231 pCi/g 05/20/15 14:46 05/20/15 17:48 1

0.0807 0.155 0.111 1.04 Radium 226 pCi/g 05/20/15 14:46 05/20/15 17:48 1

1.28 1.03 1.02 U

1.23 Thorium 234 pCi/g 05/20/15 14:46 05/20/15 17:48 1

0.302 0.000467 0.000466 U

-0.0000804 Uranium 235 pCi/g 05/20/15 14:46 05/20/15 17:48 1

0.145 0.211 0.153 1.43 Thorium 232 Th-227 0.685 0.0997 0.134 0.0772 pCi/g 05/20/15 14:46 05/20/15 17:48 1

(2+/-)

(2+/-)

MDC RL Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Other Detected Radionuclides Uncert.

Uncert.

Tl-208 0.421 0.0503 0.0666 0.0420 pCi/g 05/20/15 14:46 05/20/15 17:48 1

Lab Sample ID: 160-11841-3 Client Sample ID: L10-14-13-B-E-B-00 Matrix: Solid Date Collected: 05/14/15 14:30 Percent Solids: 80.2 Date Received: 05/14/15 16:54 Method: 6020A - Metals (ICP/MS), Tc-99 RL MDL Technetium 99 ND 0.000070 0.000021 mg/Kg

05/18/15 09:50 05/19/15 13:09 1

Analyte Dil Fac Analyzed Prepared Unit D

Result Qualifier Method: 6020A - Metals (ICP/MS), Tc-99 in Activity Analyte Technetium 99 0.286 (2+/-)

0.0509 (2+/-)

1 05/19/15 13:09 05/18/15 09:50 pCi/g 0.240 1.40 RL MDC 0.0435 Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Uncert.

Uncert.

Re 30 - 110 Carrier 05/18/15 09:50 05/19/15 13:09 1

Dil Fac Analyzed Prepared Qualifier Limits

%Yield 89 Method: GA-01-R - Cesium-137 & Other Gamma Emitters (GS)

Analyte Actinium 228 1.27 (2+/-)

0.179 (2+/-)

1 05/20/15 17:43 05/20/15 14:46 pCi/g 0.127 RL MDC 0.123 Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Uncert.

Uncert.

pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.131 0.0778 0.0778 U

-0.00351 Americium 241 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.471 0.552 0.527 1.60 Bismuth 212 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.102 0.136 0.106 0.824 Bismuth 214 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.0829 0.177 0.0787 1.22 Lead 212 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.104 0.124 0.0861 0.853 Lead 214 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.443 2.35 1.07 20.4 Potassium 40 pCi/g 05/20/15 14:46 05/20/15 17:43 1

6.00 3.44 3.44 U

2.08 Protactinium 234m TestAmerica St. Louis Page 12 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

Client Sample Results TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Lab Sample ID: 160-11841-3 Client Sample ID: L10-14-13-B-E-B-00 Matrix: Solid Date Collected: 05/14/15 14:30 Date Received: 05/14/15 16:54 Method: GA-01-R - Cesium-137 & Other Gamma Emitters (GS) (Continued)

Analyte Protactinium 231 0.432 U (2+/-)

0.251 (2+/-)

1 05/20/15 17:43 05/20/15 14:46 pCi/g 1.34 RL MDC 0.246 Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Uncert.

Uncert.

pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.102 0.136 0.106 0.824 Radium 226 pCi/g 05/20/15 14:46 05/20/15 17:43 1

1.23 0.921 0.912 U

1.20 Thorium 234 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.184 0.136 0.135 0.187 Uranium 235 pCi/g 05/20/15 14:46 05/20/15 17:43 1

0.127 0.179 0.123 1.27 Thorium 232 Tl-208 0.416 0.0465 0.0635 0.0390 pCi/g 05/20/15 14:46 05/20/15 17:43 1

(2+/-)

(2+/-)

MDC RL Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Other Detected Radionuclides Uncert.

Uncert.

Lab Sample ID: 160-11841-4 Client Sample ID: L10-14-14-B-E-B-00 Matrix: Solid Date Collected: 05/14/15 14:35 Percent Solids: 78.5 Date Received: 05/14/15 16:54 Method: 6020A - Metals (ICP/MS), Tc-99 RL MDL Technetium 99 ND 0.000070 0.000021 mg/Kg

05/18/15 09:50 05/19/15 13:12 1

Analyte Dil Fac Analyzed Prepared Unit D

Result Qualifier Method: 6020A - Metals (ICP/MS), Tc-99 in Activity Analyte Technetium 99 0.281 (2+/-)

0.0925 (2+/-)

1 05/19/15 13:12 05/18/15 09:50 pCi/g 0.240 1.40 RL MDC 0.0888 Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Uncert.

Uncert.

Re 30 - 110 Carrier 05/18/15 09:50 05/19/15 13:12 1

Dil Fac Analyzed Prepared Qualifier Limits

%Yield 91 Method: GA-01-R - Cesium-137 & Other Gamma Emitters (GS)

Analyte Actinium 228 1.22 (2+/-)

0.171 (2+/-)

1 05/20/15 20:44 05/20/15 14:46 pCi/g 0.107 RL MDC 0.117 Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Uncert.

Uncert.

pCi/g 05/20/15 14:46 05/20/15 20:44 1

0.114 0.0692 0.0691 U

0.0397 Americium 241 pCi/g 05/20/15 14:46 05/20/15 20:44 1

0.407 0.451 0.427 1.40 Bismuth 212 pCi/g 05/20/15 14:46 05/20/15 20:44 1

0.0746 0.119 0.0844 0.804 Bismuth 214 pCi/g 05/20/15 14:46 05/20/15 20:44 1

0.0577 0.179 0.0666 1.29 Lead 212 pCi/g 05/20/15 14:46 05/20/15 20:44 1

0.0846 0.146 0.102 1.00 Lead 214 pCi/g 05/20/15 14:46 05/20/15 20:44 1

0.449 2.44 1.00 21.8 Potassium 40 pCi/g 05/20/15 14:46 05/20/15 20:44 1

3.81 2.46 2.44 U

3.45 Protactinium 234m pCi/g 05/20/15 14:46 05/20/15 20:44 1

1.12 0.375 0.364 U

0.850 Protactinium 231 pCi/g 05/20/15 14:46 05/20/15 20:44 1

0.0746 0.119 0.0844 0.804 Radium 226 pCi/g 05/20/15 14:46 05/20/15 20:44 1

1.10 0.360 0.335 1.24 Thorium 234 pCi/g 05/20/15 14:46 05/20/15 20:44 1

0.247 0.149 0.149 U

0.0770 Uranium 235 pCi/g 05/20/15 14:46 05/20/15 20:44 1

0.107 0.171 0.117 1.22 Thorium 232 Tl-208 0.420 0.0549 0.0701 0.0414 pCi/g 05/20/15 14:46 05/20/15 20:44 1

(2+/-)

(2+/-)

MDC RL Dil Fac Analyzed Prepared Unit Result Qualifier Count Total Other Detected Radionuclides Uncert.

Uncert.

TestAmerica St. Louis Page 13 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

QC Sample Results TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Method: 6020A - Metals (ICP/MS), Tc-99 Client Sample ID: Method Blank Lab Sample ID: MB 160-191187/1-A Matrix: Solid Prep Type: Total/NA Analysis Batch: 191454 Prep Batch: 191187 RL MDL Technetium 99 ND 0.000053 0.000016 mg/Kg 05/18/15 09:50 05/19/15 11:43 1

MB MB Analyte Dil Fac Analyzed Prepared D

Unit Result Qualifier Client Sample ID: Lab Control Sample Lab Sample ID: LCS 160-191187/2-A Matrix: Solid Prep Type: Total/NA Analysis Batch: 191454 Prep Batch: 191187 Technetium 99 0.00124 0.00120 mg/Kg 97 80 - 120 Analyte LCS LCS D

Unit Result Qualifier

%Rec Spike Added

%Rec.

Limits Client Sample ID: Matrix Spike Lab Sample ID: 160-11861-A-15-B MS Matrix: Solid Prep Type: Total/NA Analysis Batch: 191454 Prep Batch: 191187 Technetium 99 0.000037 J 0.00155 0.00156 mg/Kg 98 75 - 125

Analyte MS MS D

Unit Result Qualifier

%Rec Spike Added Sample Result Sample Qualifier

%Rec.

Limits Client Sample ID: Matrix Spike Duplicate Lab Sample ID: 160-11861-A-15-C MSD Matrix: Solid Prep Type: Total/NA Analysis Batch: 191454 Prep Batch: 191187 Technetium 99 0.000037 J 0.00156 0.00154 mg/Kg 96 75 - 125 1

30

Analyte MSD MSD D

Unit Result Qualifier

%Rec Spike Added Sample Result Sample Qualifier

%Rec.

Limits Limit RPD RPD Method: 6020A - Metals (ICP/MS), Tc-99 in Activity Client Sample ID: Method Blank Lab Sample ID: MB 160-191187/1-A Matrix: Solid Prep Type: Total/NA Analysis Batch: 191455 Prep Batch: 191187 Technetium 99 Analyte U

1 05/19/15 11:43 05/18/15 09:50 pCi/g 0.183 MDC 1.07 RL 0.0446 0.0446 (2+/-)

(2+/-)

MB 0.009865 MB Dil Fac Prepared Analyzed Unit Result Qualifier Uncert.

Count Uncert.

Total Carrier Re 30 - 110 05/18/15 09:50 05/19/15 11:43 1

Dil Fac Prepared Analyzed Qualifier Limits

%Yield 94 MB MB Client Sample ID: Lab Control Sample Lab Sample ID: LCS 160-191187/2-A Matrix: Solid Prep Type: Total/NA Analysis Batch: 191455 Prep Batch: 191187 Technetium 99 Analyte 120 80 97 20.63 21.2 2.21 1.08 0.185 RL MDC (2+/-)

LCS LCS pCi/g Unit Result Qual

%Rec Added Spike Limits

%Rec.

Uncert.

Total Re Carrier 30 - 110 LCS Qualifier Limits

%Yield 93 LCS TestAmerica St. Louis Page 14 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

QC Sample Results TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Method: 6020A - Metals (ICP/MS), Tc-99 in Activity (Continued)

Client Sample ID: Matrix Spike Lab Sample ID: 160-11861-A-15-B MS Matrix: Solid Prep Type: Total/NA Analysis Batch: 191455 Prep Batch: 191187 Technetium 99 Analyte 125 75 98 26.71 26.6 0.635 2.54 1.42 0.243 RL MDC (2+/-)

MS MS pCi/g Unit Result Qual

%Rec Added Spike Result Sample Sample Qual Limits

%Rec.

Uncert.

Total Re Carrier 30 - 110 MS Qualifier Limits

%Yield 89 MS Client Sample ID: Matrix Spike Duplicate Lab Sample ID: 160-11861-A-15-C MSD Matrix: Solid Prep Type: Total/NA Analysis Batch: 191455 Prep Batch: 191187 Technetium 99 Analyte 1

0.05 125 75 96 26.45 26.8 0.635 2.52 1.44 0.246 RL MDC (2+/-)

MSD MSD pCi/g Unit Result Qual

%Rec Limit Added Spike Result Sample Sample Qual Limits

%Rec.

Uncert.

Total RER RER Re Carrier 30 - 110 MSD Qualifier Limits

%Yield 88 MSD Method: GA-01-R - Cesium-137 & Other Gamma Emitters (GS)

Client Sample ID: Method Blank Lab Sample ID: MB 160-191719/1-A Matrix: Solid Prep Type: Total/NA Analysis Batch: 191651 Prep Batch: 191719 Actinium 228 Analyte U

1 05/20/15 17:47 05/20/15 14:46 pCi/g 0.0758 MDC RL 0.0350 0.0345 (2+/-)

(2+/-)

MB 0.05668 MB Dil Fac Prepared Analyzed Unit Result Qualifier Uncert.

Count Uncert.

Total 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.0495 0.0288 0.0288 U

0.0002267 Americium 241 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.285 0.174 0.173 U

0.1387 Bismuth 212 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.0647 0.0558 0.0558 U

-0.01474 Bismuth 214 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.0490 0.0337 0.0337 U

-0.004958 Lead 212 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.0624 0.0336 0.0336 U

0.006167 Lead 214 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.496 1.54 1.54 U

-0.4771 Potassium 40 1

05/20/15 17:47 05/20/15 14:46 pCi/g 2.76 1.59 1.59 U

0.5271 Protactinium 234m 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.627 0.181 0.181 U

0.07186 Protactinium 231 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.0647 0.0558 0.0558 U

-0.01474 Radium 226 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.529 0.329 0.329 U

-0.06993 Thorium 234 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.111 0.0747 0.0747 U

0.01890 Uranium 235 1

05/20/15 17:47 05/20/15 14:46 pCi/g 0.0758 0.0350 0.0345 U

0.05668 Thorium 232 Other Detected Radionuclide None MB (2+/-)

(2+/-)

MDC RL pCi/g 05/20/15 14:46 05/20/15 17:47 1

MB Dil Fac Prepared Analyzed Unit Result Qualifier Uncert.

Count Uncert.

Total Radionuclides Other Detected TestAmerica St. Louis Page 15 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

QC Sample Results TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Method: GA-01-R - Cesium-137 & Other Gamma Emitters (GS) (Continued)

Client Sample ID: Lab Control Sample Lab Sample ID: LCS 160-191719/2-A Matrix: Solid Prep Type: Total/NA Analysis Batch: 191652 Prep Batch: 191719 Americium 241 Analyte 116 87 103 99.89 97.3 10.4 0.549 RL MDC (2+/-)

LCS LCS pCi/g Unit Result Qual

%Rec Added Spike Limits

%Rec.

Uncert.

Total Cesium 137 30.4 30.50 3.19 0.200 0.141 pCi/g 100 87 - 120 Cobalt 60 19.7 19.57 1.98 0.0579 pCi/g 99 87 - 115 Client Sample ID: L10-13-14-B-E-B-00 Lab Sample ID: 160-11841-1 DU Matrix: Solid Prep Type: Total/NA Analysis Batch: 191652 Prep Batch: 191719 Actinium 228 Analyte 1

0.07 1.382 1.36 0.206 0.138 RL MDC (2+/-)

DU DU pCi/g Unit Result Qual Limit Result Sample Sample Qual Uncert.

Total RER RER Americium 241 0.0301 U

-0.02140 U 38.6 0.139 pCi/g 0

1 Bismuth 212 1.39 1.545 0.410 0.453 pCi/g 0.18 1

Bismuth 214 0.673 0.8606 0.149 0.0970 pCi/g 0.75 1

Lead 212 1.28 1.350 0.191 0.0661 pCi/g 0.18 1

Lead 214 0.812 0.8098 0.117 0.0868 pCi/g 0.01 1

Potassium 40 17.9 17.82 2.14 0.484 pCi/g 0.01 1

Protactinium 234m 1.58 U

-1.162 U 3.81 6.52 pCi/g 0.45 1

Protactinium 231 0.572 U 0.5670 U 0.310 1.34 pCi/g 0.01 1

Radium 226 0.673 0.8606 0.149 0.0970 pCi/g 0.75 1

Thorium 234 1.31 1.008 U 0.394 1.10 pCi/g 0.32 1

Uranium 235 1.68 1.955 0.308 0.269 pCi/g 0.52 1

Thorium 232 1.36 1.382 0.206 0.138 pCi/g 0.07 1

Tl-208 1

0.34 0.4207 0.467 0.0674 0.0443 RL MDC (2+/-)

DU DU pCi/g Unit Result Qual Limit Result Sample Sample Qual Uncert.

RER RER Total Other Detected Radionuclides TestAmerica St. Louis Page 16 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

QC Association Summary TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Metals Prep Batch: 191187 Lab Sample ID Client Sample ID Prep Type Matrix Method Prep Batch Solid None 160-11841-1 L10-13-14-B-E-B-00 Total/NA Solid None 160-11841-2 L10-13-15-B-E-B-00 Total/NA Solid None 160-11841-3 L10-14-13-B-E-B-00 Total/NA Solid None 160-11841-4 L10-14-14-B-E-B-00 Total/NA Solid None 160-11861-A-15-B MS Matrix Spike Total/NA Solid None 160-11861-A-15-C MSD Matrix Spike Duplicate Total/NA Solid None LCS 160-191187/2-A Lab Control Sample Total/NA Solid None MB 160-191187/1-A Method Blank Total/NA Analysis Batch: 191454 Lab Sample ID Client Sample ID Prep Type Matrix Method Prep Batch Solid 6020A 191187 160-11841-1 L10-13-14-B-E-B-00 Total/NA Solid 6020A 191187 160-11841-2 L10-13-15-B-E-B-00 Total/NA Solid 6020A 191187 160-11841-3 L10-14-13-B-E-B-00 Total/NA Solid 6020A 191187 160-11841-4 L10-14-14-B-E-B-00 Total/NA Solid 6020A 191187 160-11861-A-15-B MS Matrix Spike Total/NA Solid 6020A 191187 160-11861-A-15-C MSD Matrix Spike Duplicate Total/NA Solid 6020A 191187 LCS 160-191187/2-A Lab Control Sample Total/NA Solid 6020A 191187 MB 160-191187/1-A Method Blank Total/NA General Chemistry Analysis Batch: 190916 Lab Sample ID Client Sample ID Prep Type Matrix Method Prep Batch Solid Moisture 160-11841-1 L10-13-14-B-E-B-00 Total/NA Solid Moisture 160-11841-2 L10-13-15-B-E-B-00 Total/NA Solid Moisture 160-11841-3 L10-14-13-B-E-B-00 Total/NA Solid Moisture 160-11841-4 L10-14-14-B-E-B-00 Total/NA Solid Moisture 160-11841-4 DU L10-14-14-B-E-B-00 Total/NA Rad Prep Batch: 191187 Lab Sample ID Client Sample ID Prep Type Matrix Method Prep Batch Solid None 160-11841-1 L10-13-14-B-E-B-00 Total/NA Solid None 160-11841-2 L10-13-15-B-E-B-00 Total/NA Solid None 160-11841-3 L10-14-13-B-E-B-00 Total/NA Solid None 160-11841-4 L10-14-14-B-E-B-00 Total/NA Solid None 160-11861-A-15-B MS Matrix Spike Total/NA Solid None 160-11861-A-15-C MSD Matrix Spike Duplicate Total/NA Solid None LCS 160-191187/2-A Lab Control Sample Total/NA Solid None MB 160-191187/1-A Method Blank Total/NA Leach Batch: 191189 Lab Sample ID Client Sample ID Prep Type Matrix Method Prep Batch Solid Dry and Grind 160-11841-1 L10-13-14-B-E-B-00 Total/NA Solid Dry and Grind 160-11841-1 DU L10-13-14-B-E-B-00 Total/NA Solid Dry and Grind 160-11841-2 L10-13-15-B-E-B-00 Total/NA Solid Dry and Grind 160-11841-3 L10-14-13-B-E-B-00 Total/NA Solid Dry and Grind 160-11841-4 L10-14-14-B-E-B-00 Total/NA TestAmerica St. Louis Page 17 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

QC Association Summary TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Rad (Continued)

Prep Batch: 191719 Lab Sample ID Client Sample ID Prep Type Matrix Method Prep Batch Solid Fill_Geo-0 191189 160-11841-1 L10-13-14-B-E-B-00 Total/NA Solid Fill_Geo-0 191189 160-11841-1 DU L10-13-14-B-E-B-00 Total/NA Solid Fill_Geo-0 191189 160-11841-2 L10-13-15-B-E-B-00 Total/NA Solid Fill_Geo-0 191189 160-11841-3 L10-14-13-B-E-B-00 Total/NA Solid Fill_Geo-0 191189 160-11841-4 L10-14-14-B-E-B-00 Total/NA Solid Fill_Geo-0 LCS 160-191719/2-A Lab Control Sample Total/NA Solid Fill_Geo-0 MB 160-191719/1-A Method Blank Total/NA TestAmerica St. Louis Page 18 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13

Tracer/Carrier Summary TestAmerica Job ID: 160-11841-1 Client: Perma-Fix Environmental Services Inc.

Project/Site: HDP Method: 6020A - Metals (ICP/MS), Tc-99 in Activity Prep Type: Total/NA Matrix: Solid Lab Sample ID Client Sample ID (30-110)

Re 86 160-11841-1 Percent Yield (Acceptance Limits)

L10-13-14-B-E-B-00 87 160-11841-2 L10-13-15-B-E-B-00 89 160-11841-3 L10-14-13-B-E-B-00 91 160-11841-4 L10-14-14-B-E-B-00 89 160-11861-A-15-B MS Matrix Spike 88 160-11861-A-15-C MSD Matrix Spike Duplicate 93 LCS 160-191187/2-A Lab Control Sample 94 MB 160-191187/1-A Method Blank Tracer/Carrier Legend Re = Re TestAmerica St. Louis Page 19 of 19 5/26/2015 1

2 3

4 5

6 7

8 9

10 11 12 13