Calculation 98-ENG-02405D2, Revision 2, Justification for the Table Accuracy Value Used in the Minco RTD Sca Term, Attachment 8, Page 1 of 32
ML12227A522
Person / Time
Site: Millstone Dominion icon.png
Issue date: 08/09/2012
From: Nowlan N
Dominion Nuclear Connecticut
To:
Office of Nuclear Reactor Regulation
References
12-474A 98-ENG-02405D2, Rev 2
Download: ML12227A522 (33)


Text

Calculation 98-ENG-02405D2, Revision 2

ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PMPe Page I of 32 BWT Dominion-Engineering Technical Evaluation Cover Sheet P, #

/ a,~ 31 IZ S.

S S

  • .~

.I

1. Stations Doc Type:

Sub Type:

2. Document Number
3. Rev # 4. Decomissioning?

[Z MP E] SU 5] NA E] KW [] CO ETE 000 ETE-MP-2012-1135 0

5 Yes f] No (Note: If both SU and NA. then check CO) 5.

Title:

Justification for the Table Accuracy Value Used in the MINCo RTD SCA Term in Calculation No. 98-ENG-02405D2.

6. ETE Level
7. Unit(s) *
8. Quality Classification
9. FSRC Approval Req.?

[] Level El] Level 1 1 El Unit1 Ea Unit2 51 Unit3 l ISFSI El SR [] NS 5 NSQ 5

Yes [Z No 10.. Preparation, Review, and Approval Signatures (add or delete rows as needed)

Preparer : (Print)

Signature:

Date:

Digitally Signed By:

08/06/2012 Nancy Nowlan 12:03:43 pm Reviewer: (Print)

Signature:

Date:

Digitally Signed By:

08/06/2012 Konstantine Filippides 01:04:08 pm Engineering Supervisor / Designee: (Print)

Signature:

Date:

Digitally Signed By:

08/06/2012 Keith Deslandes Xý__

z*__

02:17:27 pm Standard Attachments Attachment

  1. of pages Reviewed/No Not Req.

Impact

11. Design Effects and Considerations (DNES-AA-GN-1003)

F-_

12. Document Impact Summary (DRUL) (DNES-AA-GN-1002)
13. Considerations and Conditions for Document Updates (check N/A F7 if no document updates are noted on the DRUL) 5 All Document updates noted on the DRUL can be initiated immediately 5 Document updates noted on the DRUL are delayed until the following documents/actions are completed:

(e.g., WO, CR etc.) (See DRUL Remarks section) 10CFR5O.59 Attachments Attachment

  1. of pages Not Req.
14. 10CFR50.59/72.48 applicability review forms (DNAP-3004 or CM-AA-400)

Wk

15. 10CFR50.59/72.48 screen form (DNAP-3004 or CM-AA-400)

I*

16. 1 OCFR50.59/72.48 evaluation form (DNAP-3004 or CM-AA-400)

I2

17. Additional Attachments Attachment
  1. of pages Description MET LAB RTD Cal UncertaintyCalc,,

1 MET LAB RTD Cal UncertaintyInforn I

CalibrationReport SN 1309.pdf 1

CalibrationReport SN 1269.pdf 1

CalibrationReport SN 1270.pdf 1

MINCoAccuracyStatement.pdf 2

SN_1269,_SN 1270 and SN 1309 C2 12 SN 1269 SN_1270 SN 1309_2009_1V 9

1 Aug 6, 2012 3:06 PM

ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 2 of 32 1'977X61-111460J7-9.

Engineering Technical Evaluation Cover Sheet Domninion-CMA'-T-0 ATTACHMENT

.a l2o 18.DDistribution Primary Recipient(s):

(Enter the Name I Dept or Location for EACH Primary Recipient in this block.)

Copy To? Other Recipient / Department or Location Copy To? 1Other Recipient/Department or Location

]J Preparer

[]J Reviewer F]

Supervisor El Site DCE

[]

iffected organization El Program Owners 2.0 Source Document Calculation No. 98-ENG-02405D2, "Millstone Unit 2 Service Water Inlet Temperature-Indicator Accuracy TI-6928, TI-6929 & TI-6930."

4.0 Purpose The purpose of this engineering technical evaluation is to provide justification for the MINCo RTD "Table Accuracy" term used within calculation no. 98-ENG-02405D2. Also, this engineering technical evaluation will contain the calibration tables for the calibrated RTDs TE-6928, TE-6929 and TE-6930.

Aug 6, 2012 3:06 PM

ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 3 of 32 A) *AcJi/I,/y/ 6Jr- *"

.u*,miionEngineering Technical Evaluation k(6 o r 3-ominion"Cover Sheet CMAA-T 0

ATTACHMENT

.a 3o 7.0 Discussion TE-6928, TE-6929 and TE-6930 are the Vital AC Switchgear Room Cooling Coil (Xl 81A and X811B) Service Water Inlet RTDs. Loops T-6928, T-6929 and T-6930 are used in technical specification surveillance SP 2619A, "Control Room Shift Checks," to record and verify that the "Ultimate Heat Sink" temperature is less than 74.5 degrees F.

TE-6928, TE-6929 and TE-6930 are MINCo. model no. XS7927PKIL76S RTDs, serial nos. S/N 1309, S/N 1270 and S/N 1269, respectively. RTDs S/N 1269 and S/N 1270 were certified by MINCo in 1993 and RTD S/N 1309 was certified by MINCo in 2000.

MINCo certified all three RTDs to a stated accuracy value of+/- 0.025 degrees C (refer to the attached MINCo "Report of Calibrations" and "MINCo Accuracy Statement"). This supported the SCA term used in calculation no. 98-ENG-02405D2, revision no. 0.

A subsequent calibration of these particular serial number RTDs was performed by the Millstone Metrology Lab in 2009. New R vs T tables were generated and used for the calibration of each of the temperature loops (refer to the attached "SN 1269, SN 1270 and SN 1309 Met Lab Calibration Curves"). Information provided by the Millstone Metrology Lab indicates that the standards used by the lab to calibrate these RTDs is +/- 0.0038 degrees C (refer to the attached "Met Lab RTD Cal Uncertainty Cales. And Met Lab RTD Cal Uncertainty Information). This uncertainty value bounds the A: 0.025 degrees C used in the SCA term in calculation no. 98-ENG-02405D2.

To ensure the Metrology Lab curves approximated those produced by MINCo, this ETE compared the values of the M]NCo curves to the Metrology Lab curves. The original R vs T tables provided by MINCo were not found. However, within MINCo's website, an RTD Table Generator was used to generate the R vs T tables using the coefficients of the original M1NCo RTD calibration reports. Using EXCEL, R vs T curves were plotted using resistance and temperature values over the 0 to 100 degrees F calibrated span of these temperature loops in 5 degrees F increments (21 data points). Again using EXCEL, R vs T curves were plotted using the 2009 Metrology Lab generated tables for these RTDs. The two curves generated for each serial no. RTD were compared. The worst case error noted and expressed as a percent of span (0 to 100 degrees F span) for S/N 1269, S/N 1270 and S/N 1309 was +/- 0.033 % (+/- 0.033 degrees F), +/-

0.01 % (- 0.01 degrees F) and +/- 0.066 % (+/- 0.066 degrees F), respectively. As stated earlier, S/N 1269 and S/N 1270 were certified by MINCO 16 years prior to the Metrology Lab calibration and S/N 1309 was certified by MINCo 9 years prior to the Metrology Lab calibraion. For S/N 1269 and S/N 1270, a MINCo stated 16 years drift term "root sum of the square" combined with the MINCo repeatability statement for this model no. RTD equates to +/- 0.36 % (0.36 degrees F). For S/N 1309, a MINCo stated 9 years drift term "root sum of the square" combined with the MINCo repeatability statement for this model no. RTD equates to +/- 0.27 % (0.27 degrees F). The worst case error in the curves generated by MINCo as compared to the 2009 Metrology Lab calibrations indicates that these particular serial no. RTDs did not drift very much over their respective 9 year and 16 year period and certainly much less than expected.

8.0 Conclusions The +/- 0.025 degrees C value used in the SCA term for RTD accuracy in calculation no. 98-ENG-2405D2 is valid. This value is used as "Table Accuracy" in revision no. 2 of this calculation.

Aug 6, 2012 3:06 PM

ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 4 of 32 A,7r~c/lA1e,~i-r g Engineering Technical Evaluation Cover Sheet

ý) Fominion-CMA-T-0 ATTCHEN 2

e4o 12.0 References All references are attached to this ETE.

1. Calibration Report SN 1269,
2. Calibration Report SN 1270,
3. Calibration Report SN 1309,
4. MET LAB RTD Cal Uncertainty Calcs,
5. MET LAB RTD Cal Uncertainty Information,
6. SN 1269 SN 1270 SN 1309 Met Lab Cal Curves,
7. SN 1269, SN 1270, SN 1309 Calibration Curve Comparisons (EXCEL spreadsheets and charts),
8. MINCo Accuracy Statement.

Aug 6, 2012 3:06 PM

Attachment #1 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 5 of 32

,4/7%6/n'n6A"r a 5-a-3 2--

MINCO PRODUCTS, INC.

REPORT No.

2013 7300 COMMERCE LANE JOB No.

03-30932

FRIDLEY, MN 55432 P.O.# 259053 TELEPHONE:

(612) 571-3121 FAX:

(612) 571-0927 REPORT OF CALIBRATION CERTIFICATE

    • T*

FOR PLATINUM RESISTANCE THERMOMETER MODEL XS7927PKIL76S(B)

SERIAL No.

1269 PREPARED BY MINCO PRODUCTS, INC.

FOR NORTHEAST NUCLEAR ENERGY CO THE THERMOMETER RESISTANCE WAS MEASURED AT FIXED TEMPERATURE POINTS DEFINED BY ITS-90 OR BY COMPARISON WITH A CALIBRATED PLATINUM RESISTANCE THERMOMETER TRACEABLE TO THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY.

RESISTANCES WERE MEASURED WITH AN EXCITATION CURRENT OF 1.OmA.

TEMPERATURE IS DERIVED FROM MEASURED RESISTANCE AS FOLLOWS:

FOR TEMPERATURES BELOW 273.16K:

Wref(T) = W(T) a4*[W(T)-I]

b4*(W(T)-I)lnW(T)

FOR TEMPERATURES AT AND ABOVE 273.16K:

Wref(T)

= W(T) a8*[W(T)-I]

- b8*[W(T)-I]-2 WHERE W(T)

R(T)/R(273.16K),

FOR T IN KELVIN; Wref(T) DEFINES TEMPERATURE ACCORDING TO THE ITS-90 REFERENCE FUNCTION (SEE FOLLOWING PAGE);

AND THE COEFFICIENTS a4, b4, a8, AND b8 ARE UNIQUE TO EACH INDIVIDUAL THERMOMETER.

a8 = -4.0758054D-04 b8 = +8.3733738D-05 a4 = -3.6445181D-04 b4 = -2.8576799D-06 R(OC) = 199.872 OHMS R(100C)/R(OC) = 1.3927 MEASUREMENT DATA TEMP (-C)

RESISTANCE (OHMS)

-189.34 43.2061

-38.83 168.742 0.010 199.880 100.00 278.358 231.928 378.273 MEASUREMENTS BY:

D.

JACOBS CERTIFIED BY:

G.

ENGELBRECHT DATE CERTIFIED:

10 JUNE 1993 N.I.S.T. TEST REPORT No.

811/250375 DATE OF N.I.S.T. TEST:

4 DECEMBER 1992

Attachment #2 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 6 of 32 A4 7rACI/,'f C,,

5--

6 a MINCO PRODUCTS, INC.

REPORT No.

2014 7300 COMMERCE LANE JOB No. '03-30932

FRIDLEY, MN 55432 P.O.# 259053 TELEPHONE:

(612) 571-3121 FAX:

(612) 571-0927 REPORT OF CALIBRATION C ERT I F I CATE FOR PLATINUM RESISTANCE THERMOMETER MODEL XS7927PK1L76S(B)

SERIAL No.

1270 PREPARED BY MINCO PRODUCTS, INC.

FOR NORTHEAST NUCLEAR ENERGY CO THE THERMOMETER RESISTANCE WAS MEASURED AT FIXED TEMPERATURE POINTS DEFINED BY ITS-90 OR BY COMPARISON WITH A CALIBRATED PLATINUM RESISTANCE THERMOMETER TRACEABLE TO THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY.

RESISTANCES WERE MEASURED WITH AN EXCITATION CURRENT OF 1.OmA.

TEMPERATURE IS DERIVED FROM MEASURED RESISTANCE AS FOLLOWS:

FOR TEMPERATURES BELOW 273.16K:

Wref(T) = W(T) a4*[W(T)-I]

b4*[W(T)-I]InW(T)

FOR TEMPERATURES AT AND ABOVE 273.16K:

Wref(T)

= W(T) a8*(W(T)-I]

b8*[W(T)-I]^2 WHERE W(T)

= R(T)/R(273.16K),

FOR T IN KELVIN; Wref(T) DEFINES TEMPERATURE ACCORDING TO THE ITS-90 REFERENCE FUNCTION (SEE FOLLOWING PAGE);

AND THE COEFFICIENTS a4, b4, a8, AND b8 ARE UNIQUE TO EACH INDIVIDUAL THERMOMETER.

a8 = -5.4864581D-04 b8 = +1.0006569D-04 a4 = -3.3185991D-04 b4 = -2.9061316D-06 R(OC) = 199.904 OHMS R(100C)/R(OC) = 1.3926 MEASUREMENT DATA TEMP (-C)

RESISTANCE (OHMS)

-189.34 43.2079

-38.83 168.768 0.010 199.912 100.00 278.392 231.928 378.311 MEASUREMENTS BY:

D.

JACOBS CERTIFIED BY:

G.

ENGELBRECHT DATE CERTIFIED:

10 JUNE 1993 N.I.S.T. TEST REPORT No.

811/250375 DATE OF N.I.S.T. TEST:

4 DECEMBER 1992

Attachment #3 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 7 of 32 MINCO Products, Inc.

Report No.

3306 7300 Commerce Lane Job No.

3-254185 Fridley, MN 55432 P.O.# 02140472 Tel: (612) 571-3121 t

FAX:

(612) 571-0927 Report of Calibration CERTIFICATE for Platinum Resistance Thermometer Model XS7927PKIL76S Serial No.

1309 Prepared by MINCO PRODUCTS, INC.

for NORTHEAST NUCLEAR ENERGY CO.

The thermometer resistance was measured at fixed temperature points defined by ITS-90 or by comparison with a calibrated Platinum Resistance Thermometer traceable to the National Institute of Standards and Technology.

Resistances were measured with an excitation current of 1.0mA.

Temperature is derived from measured resistance as follows:

for temperatures below 273.16K:

Wref(T)

= W(T) a4*[W(T)-I]

b4*[W(T)-IjlnW(T) for temperatures at and above 273.16K:

Wref(T) = W(T) a8*[W(T)-I]

b8*[W(T)-1I]2 1

where W(T)

= R(T)/R(273.16K),

for T in Kelvin; Wref(T) defines temperature according to the ITS-90 reference function (see following page);

and the coefficients a4, b4, a8, and b8 are unique to each individual thermometer.

a8 = -4.5088398D-04 b8 = +1.4870006D-04 a4 = -6.0597493D-04.

b4 = +3.1314950D-04 R(0 0 C)

= 200.008 ohms R(100 0 C)/R(00 C) = 1.3927 Measurement data Temp (OC)

Resistance (ohms)

-189.34 43.349

-38.83 168.866 0.010 200.016 100.00 278.546 231.928 378.533 Measurements by:

JES Certified by:

M. SWENSON Date certified:

14 July 2000 N.I.S.T. Test Report No.

14-HS876-1-1 Date of N.I.S.T. test:

6 May 1999

Attachment #4 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 8 of 32

,47YA!&46Ore-~

Hart Thermometry System Accuracies SPRT + Thermometer Uncertainties Device Temperature Accuracy Combined Uncertainty 1590

-200 to 6610C +/-0.00150C

+0.002500 5681

-200to661°C +/-0.0020C Bath Uncertainties Device Parameter Tem erature Specification

-40°C

+/-0.0020C 00C

+/-0.0010C 7037 Stability 250C

+/-0.00050C

+/-10000

+0.0020C Uniformity All

+/-0.0020C Stability 1000C

+/-0.0030C 6022 3000C

+/-0.0050C 6022 1000C

+/-0.0040C Uniformity 3000C

+/-0.0120C

<3000C

+/-0.0300C 9260 Stability

>3000C

+/-0.0500C Uniformity 50 to 6800C

+/-0.0200C Fixed Point Fixed Point Cell Temperature Uncertainty 5901 D-Q TPW 0.010C

+/-0.0001 0C 5946 MP Zn 429.5270C

+/-0.0020C Combined Uncertainties (Bath + 1590 + 5681)

Bath Temperature Combined Uncertainty

-400C

+/-0.0038°C 000C

+/-0.00340C 250C

+/-0.00320C 1000C

+/-0.00380C 1000C

+/-0.00560C 3000C

+/-0.01 320C 9260

<3000C

+/-0.03610C

>3000C

+0.05390C

0=>

S Nancy M Nowlan (Generation - 6) 0, From:

Robert J Byrnes (Generation - 4)

Sent:

Thursday, August 02, 2012 3:26 PM To:

Nancy M Nowlan (Generation - 6)

Subject:

Accuracy of Data Generated by Metrology Laboratory RTD Calibrations

Nancy, Measurement uncertainty for all standards used for calibration of RTDs by the Millstone Metrology Laboratory is +/-0.0038*C. This excludes the uncertainty of the Callendar Van Dusen equation, which generates the Temperature vs. Resistance table and coefficients. The CVD equation, used to generate the table and m

coefficients, contributes such an inconsequential uncertainty, that it is not worthy of consideration.

The RTD data provided by Minco has an uncertainty of 0.0250C. The calibrations performed by the Millstone Metrology Laboratory are 6.58 times more accurate.

-P bob byrnes Instrument Tech Ill C"

Tie Line: 8.759.6229 Phone: 860.447.1791 x6229 0

Fax: 860.444.5317 0

Pager: 860.504.4365 0D "The only stupid question is the one you don't ask."

0 cc O

Attachment #6 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 10 of 32

-Report for Callendar-Van Dusen Coefficients Model: XS7-927PK I Serial: 1309 t-Da~te-: August 04,2009 Rekfrence (OC)

-1 9:9433 0.0380 25.0211 50,0 176 75.0155 I00.0214 UUT (Ohms) 184.0662 199.9719 219.8137 239.5121 259.0611 278.4672

,5 5-.5 7 C) e7(3 Residual (QC) 0.0000 0.0000

-0.0016 0.0003 0.0012

-0.0007 Coefficients:

RO = 199.941595 ALPHA = 0.00392657 DELTA = 1.54203 BETA = -9.55746 P'age Iof 3

Attachment #8 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 11 of 32

,f,VrCftAlC' ce&r Model: XS7927PK1 Serial: 1309 Callendar-Van Dusen Temperature vs. Resistance Table OF

-4.00

-3.00

-2.00

-1.00 0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 16.00 17.00 18,00 19.00 20.00 21.00 22.00 23.00 24.00 25.00 26.00 27.00 28.00 29.00 30.00 31.00 32.00 33.00 34.00 35.00 36.00 37.00 38.00 39.00 40.00 41.00 42.00 43.00 44.00 45.00 46.00 47.00 48.00 49.00 50.00 51,00 52.00 63.00 Resistance 184.02137 184.46076 184.90046 185.34044 185.78069 186.22121 186.66197 187.10296 187.54418 187.98560 188.42723 188.86903 189.31101 189.75315 190.19543 190.63785 191.08040 191.52306 191,96583 192.40868 192.85162 193.29463 193.73770 194.18082 194.62397 195.06716 195.51037 195.95359 196.39682 196.84003 197.28323 197.72640 198.16954 198.61264 199.05569 199.49867 199.94159 200.38444 200.82721 201.26991 201.71253 202.15508 202.59756 203.03995 203.48228 203.92453 204.36670 204.80880 205.25083 205.69278 206.13465 206.57645 207.01818 207.45983 207.90141 208.34291 208.78433 209.22569 dr/dT OF 0.4392383 54.00 0.4395457 55.00 0.4398394 56.00 0.4401196 57.00 0,4403863 58.00 0,44063991 59.00 0,4408804 60.00 0.44110801 61.00 0.4413230 62.00 0.4415254 63.00 0.4417155 64.00 0.4418934 65.00 0,4420593 66.00 0.4422133 67.00 0.4423557 68.00 0.4424866 69.00 0.4426062 70.00 0.4427146 71.00 0.4428121 72.00 0.4428987 73.00 0.44297471 74.00 0.4430403 75.00 0.4430955 76.00 0.4431406 77.00 0.44317581 78.00 0.4432012!

79.00 0.4432170f 80:00 0.4432233, 81.00 0,4432204!

82.00 0.44320841 83.00 0.44318751 84.00 0.44315781 85.00 0.44311961 86.00 0.4430729 87.00 0.4430180 88.00 0.4429551 89.00 0.4428843 90.00 0.4428095 91.00 0.4427348 92.00 0.4426601 93.00 0.4425853 94.00 0.4425106 95.00 0.4424359 96.00 0.44236111 97.00 0.4422864 98.00 0,44221171 99.00 0.44213701 100.00 0.4420622 101.00 0.4419875 102.00 0.4419128 103.00 0.4418380 104.00 0.4417633 105.00 0.4416886 106.00 0,44161381 107.00 0.4415391 108.00 0.44146441 109.00 0.44138971 110.00 0.4413149 111.00 Resistance 209.66696 210.10817 210.54929 210.99035 211.43133 211.87223 212.31306 212.75381 213.19449 213.63510 214.07563 214.51608 214.95647 215.39677 215.83700 216.27716 216.71724 217.15725 217.59718 218.03704 218.47682 218.91653 219.35616 219.79572 220.23521 220.67462 221.11395 221,55321 221.99240 222.43151 222.87054 223.30950 223.74839 224.18720 224.62594 225.06460 225.50319 225.94170 226.38014 226.81850 227.25679 227.69500 228.13314 228,57120 229.00919 229.44711 229.88495 230,32271 230.76040 231.19802 231.63556 232.07303 232.51042 232.94774 233.38498 233.82215 234.25924 234.69626 dr/dT i

OF Resistance 0.4412402 112.00 235.13320 0.44116551 113.00 235.57007 0.44109071 114.00 236.00686 0.4410160 115.00 236.44358 0.44094131 116.00 236.88022 0.44086651 117.00 237.31679 0.4407918 118.00 237.75329 0.4407171 119.00 238.18971 0.4406424 120.00 238.62605 0.4405676 121.00 239.06233 0,4404929 122.00 239.49852 0.4404182 123.00 239.93464 0.4403434 124.00 240.37069 0.4402687 125.00 240.80666 0.4401940 126.00 241.24256 0.4401192 127.00 241.67838 0.4400445 128.00 242.11413 0.4399698 129.00 242.54980 0.4398951 130.00 242.98540 0.4398203 131.00 243.42092 0.4397456 132.00 243.85637 0.4396709 133.00 244.29174 0.4395961 134.00 244.72704 0.4395214 135.00 245.16227 0.4394467 ; 136.00 245.59742 0.4393719 137.00 246.03249 0.4392972 138.00 246.46749 0.4392225 139.00 246.90242 0.43914781 140.00 247.33727 0,4390730 141.00 247.77205 0,4389983 142.00 248.20675 0,4389236 143.00 248.64137 0.4388488 144.00 249.07592 0.4387741 145.00 249.51040 0,4386994 146.00 249.94480 0.4386246 147.00 250.37913 0.4385499 148.00 250.81338 0.4384752 149.00 251.24756 0.4384005 150.00 251,68167 0.4383257 151.00 252,11570 0,4382510 152.00 252.54965 0.4381763 153.00 252,98353 0.4381015 154.00 253.41733 0.4380268 155.00 253.85106 0.4379521 156.00 254.28472 0.4378773 157.00 254,71830 0.4378026 158.00 255.15180 0.4377279 159.00 255.58524 0.4376532 160.00 256,01859 0.4375784 161.00 256,45187 0.4375037 162.00 256.88508 0.4374290 163.00 257.31821 0.4373542 164.00 257.75127 0.4372795 165.00 258.18425 0.4372048 166.00 258.61716 0.4371300 167.00 259.04999 0.4370553 168.00 259.48275 0.4369806 169,00 259.91543 drldT 0.4369059 0.4368311 0,4367564 0.4366817 0.4366069 0.4365322 0.4364575 0.4363827 0.4363080 0.4362333 0.4361586 0.4360838 0.4360091 0,4359344 0,4358596 0.4357849 0.4357102 0.4356354 0.4355607 0.4354860 0.4354113 0.4353365 0.4352618 0.4351871 0.4351123 0.4350376 0.4349629 0.4348881 0.4348134 0.4347387 0.4346640 0.4345892 0,4345145 0.4344398 0.4343650 0.4342903 0.4342156 0,4341408 0,4340661 0.4339914 0.4339167 0.4338419 0.4337672 0.4336925 0.4336177 0,4335430 0.4334683 0.4333935 0.4333188 0.4332441 0.4331694 0.4330946 0.4330199 0.4329452 0.4328704 0.4327957 0.4327210 0.4326462 Page 2 of 3 Date: August 04,2009

Attachment #6 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 12 of 32 Model: XS7927PKI Serial: 1309 Callendar-Van Dusen Temperature vs. Resistance Table OF Resistance drldT 170.00 260.34804 0.4325715 171.00 260.78058 0.4324968 172.00 261.21304 0.4324221 173.00 261.64542 0.4323473 174.00 262.07773 0.4322726 175.00 262.50997 0.4321979 176.00 262.94213 0.4321231 177.00 263.37421 0.4320484 178.00 263.80622 0.4319737 179.00 264.23816 0.4318989 180.00 264.67002 0.4318242 181.00 265.10181 0.4317495 182.00 265.53352 0.4316748 183.00 265.96516 0.4316000 184.00 266.39672 0.4315253.

185.00 266.82821 0.4314506*

186.00 267.25962 0.4313758!

187.00 267.69096 0,4313011!

188.00 268.12222 0.4312264; 189.00 268.55341 0.4311516.

190.00 268,98453 0.4310769-191.00 269.41557 0.4310022:

192.00 269.84653 0.4309275:

193.00 270.27742 0,4308527:

194.00 270,70824 0.4307780 195.00 271.13898 0.4307033, 196.00 271.56964 0.4306285 197.00 272.00023 0.4305538 198.00 272.43075 0.4304791 199.00 272.86119 0.4304044 200.00 273.29156 0.4303296 201.00 273.72185 0.4302549 202.00 274.15207 0.4301802 203.00 274.58221 0.4301054 204.00 275.01228 0.4300307 205.00 275.44227 0.4299560 206.00 275.87219 0.4298812 207.00 276.30204 0.4298065 208.00 276.73181 0.4297318 209.00 277.16150 0.4296571 210.00 277.59112 0.4295823 211.00 278.02066 0.4295076 212.00 278.45014 0.4294329 Page 3 of 3 Date: August 04,2009

Attachment #6 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 13 of 32

,rA(#C e A17-Et Report No: RU200908073-001 Page 1 of 3 Report of Calibration Dominion Rope Ferry Road Bldg 433, Metrology Laboratory Waterford, CT 06385 Model:

XS7927PK1 Customer:

Dominion Serial:

1270

, 1'o Rope Ferry Road

==

Description:==

RTD PLANT INSTALLED Bldg 433, Metrology Laboratory Waterford, CT 06385 Calibration Range:

Limited Received Condition:

Recalibration Current:

1.0 mA Procedure:

MTE 1142E Actual Value (Reference)

(C)

-19.92670 0.04100 25.02490 50.01590 75.01290 100.01760 UUT (Test Sensor)

(Ohms) 183.96920 199.93170 219.76970 239.46400 259.01160 278.41470 Residuals (C) 0.0000 0.0000 0.0005

-0.0003

-0.0001 0.0001 Method of Realization COMP COMP COMP COMP COMP COMP

'Callendar-Van Dusen Coefficients:

RO = 199.89901 ALPHA= 0.00392709 DELTA= 1.53452 BETA = -0.12847 Test Equipment Manufacturer Model Description Serial Number Recall Date Hart Scientific 1590 Super Thermometer I1 A68350 09/03/2009 Hart Scientific 5681 SPRT, 25 ohm 1466 02/25/2010 Hart Scientific 7037 Bath, Deep Low-Temp 93046 NCR Notes: Initial calibration to establish baseline data.

Acceptance Criteria: RO = 199,8000 to 200.2000 ohms Prerequisitesp/et: Initial:

Usage Review Required:

Yes __

CR#:

111

'inaf Calibration Acceptable: Yes %_ No__

Technician:

-17 Calibration Date:

Recall Date:

Temperature:

Humidity:

Customer Order:

08/17/2009 08/12/2029 73.OF 45%

N/A R61~rerI3 Byrne~:

Approved By:

This report shahl not be reproduced except in full without written approval ofDondnion.

Attachment #6 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 14 of 32 A(f1~416-A'r 08/18/2009 Temperature vs.'-% tance Table Repoi

.o : RJ200908073.

F 0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 16.00 17.00 18.00 19.00 20.00 21.00 22.00 23.00 24.00 25.00 26.00 27.00 28.00 29.00 30.00 31.00 32.00 33.00 34,00 35,00 36.00 37.00 38.00 39.00 40.00 41.00 42.00 43.00 44.00 45.00 46.00 47.00 48.00 49.00 50.00 51.00 52.00 53.00 54.00 55.00

" 56.00 57.00 58.00 Resistance 185.69151 186.13661 186.58163 187.02658 187.47147 187.91628 188.36102 188.80570 189.25030 189.69482 190.13928 190,58367 191.02798 191.47223 191.91640 192.36050 192.80452 193.24848 193.69236 194.13617 194.57990 195.02357 195.46716 195,91067 196.35412 196,79749 197.24078 197.68401 198.12715 198.57023 199.01323 199.45616 199.89901 200.34179 200.78449 201.22712 201.66968 202,11216 202.55456 202.99690 203.43915 203.88134 204.32345 204.76548 205,20744 205,64933 206.09114 206.53288 206.97454 207.41613 207.85764 208.29908 208.74045 209.18174 209.62295 210.06410 210.50516 210.94616 211.38708 dr/dT 0.4451292 0.4450593 0.4449891 0.4449188 0.4448484 0.4447777 0.4447069 0.4446360 0.4445648 0.4444936 0.4444221 0.4443505 0.4442788 0.4442069 0,4441348 0.4440626 0.4439903 0.4439178 0.4438452 0.4437725 0.4436996 0.4436266 0.4435534 0.4434802 0.4434068 0.4433332 0.4432596 0,4431859 0.4431120 0.4430380 0.4429639 0.4428897 0.4428154 0.4427410 0.4426667 0.4425923 0.4425180 0.4424436 0.4423692 0.4422949 0.4422205 0.4421462 0.4420718 0.4419974 0.4419231 0.4418487 0.4417744 0.4417000 0.4416256 0.4415513 0,4414769 0.4414026 0.4413282 0.4412538 0.4411795 0.4411051 0.4410308 0.4409564 0.4408820 6

6 6

6 6

6.

6' 6

67 6*

7.

71 7E 7

71 75 76 77 78 79 80 8t 82, 83 F

59.00 0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00

).00 0.00 t.00 2.00 3.00 84.00 85.00 86.00 87.00 88.00 89.00 90.00 91.00 92.00 93.00 94.00 95.00 96.00 97.00 98.00 99.00 100.00 101.00 102.00 103.00 104.00 105.00 106.00 107.00 108.00 109.00 110.00 111.00 112.00 113.00 114.00 115.00 116.00 117.00 Resistance 211.82792 212.26869 212.70939 213.15001 213.59056 214.03103 214.47143 214.91175 215.35200 215.79218 216.23228 216.67231 217.11226 217.55214 217.99194 218.43167 218.87133 219.31091 219.75041 220.18985 220.62920 221.06849 221,50770 221.94683 222.38589 222.82488 223.26379 223.70262 224.14139 224.58008 225.01869 225.45723 225.89569 226.33409 226.77240 227.21064 227.64881 228.08691 228.52493 228.96287 229.40074 229.83854 230.27626 230.71391 231.15148 231.58898 232.02640 232.46375 232.90103 233.33823 233.77536 234.21241 234.64939 235.08629 235.52312 235.95987 236.39656 236.83316 237.26969 dr/dT 0.4408077 0.4407333 0.4406590 0.4405846 0.4405102 0.4404359 0.4403615 0.4402872 0.4402128 0.4401384 0.4400641 0.4399897 0.4399154 0.4398410 0.4397666 0.4396923 0.4396179 0.4395436 0.4394692 0.4393948 0.4093205 0.4392461 0.4391718 0,4390974 0.4390230 0.4389487 0.4388743 0.4388000 0,4387256 0.4386512 0.4385769 0.4385025 0.4384282 0.4383538 0.4382794 0.4382051 0.4381307 0.4380564 0.4379820 0.4379076 0.4378333 0.4377589 0.4376846 0.4376102 0.4375358 0.4374615 0.4373871 0.4373128 0.4372384 0,4371640 0.4370897 0.4370153 0.4369410 0.4368666 0.4367922 0.4367179 0.4366435 0.4365692 0.4364948 139.00 140.00 141.00 142.00 143.00 144.00 145.00 146.00 147.00 148.00 149.00 150.00 151.00 152.00 153.00 154.00 155.00 156.00 157.00 158.00 159.00 160.00 161.00 162.00 163.00 164.00 165.00 166.00 167.00 168.00 169.00 170.00 171.00 172.00 173.00 174.00 175.00 176.00 F

118.00 119.00 120.00 121.00 122.00 123.00 124.00 125.00 126.00 127.00 128.00 129.00 130.00 131.00 132.00 133,00 134.00 135.00 136.00 137.00 138.00 Resistance 237.70615 238.14253 238.57884 239.01508 239.45124 239.88732 240.32334 240.75927 241.19514 241.63092 242.06664 242.50228 242.93784 243.37333 243.80875 244.24409 244.67936 245.11455 245.54967 245.98472 246.41969 246.85458 247.28941 247.72415 248,15883 248,59342 249.02795 249.46240 249.89677 250.33108 250.76530 251.19945 251.63353 252.06754 252.50147 252.93532 253.36910 253.80281 254.23644 254.67000 255.10348 255.53689 255.97022 256.40348 256.83667 257.26978 257.70282 258.13578 258.56867 259.00148 259.43422 259.86689 260.29948 260.73200 261.16444 261.59681 262.02910 262.46132 262.89346

-001 Page2of3 dr/dT 0.4364204 0.4363461 0.4362717 0.4361974 0.4361230 0.4360486 0.4359743 0.4358999 0.4358256 0.4357512 0.4356768 0.4356025 0.4355281 0.4354538 0.4353794 0.4353050 0.4352307 0.4351563 0.4350820 0.4350076 0.4349332 0.4348589 0.4347845 0.4347102 0.4346358 0.4345614 0.4344871 0.4344127 0.4343384 0.4342640 0.4341896 0.4341153 0.4340409 0.4339666 0.4338922 0.4338178 0.4337435 0.4336691 0.4335948 0.4335204 0.4334460 0.4333717 0.4332973 0.4332230 0.4331486 0.4330742 0.4329999 0.4329255 0.4328512 0.4327768 0.4327024 0.4326281 0.4325537 0.4324794 0.4324050 0.4323306 0.4322563 0.4321819 0.4321076

Attachment #6 ETE-MP-2012-1135 Report Generated on 816/2012 3:06:09 PM

.,, 70CHW1 S

Page 15 of 32 08/18/2009 Temperature vs. RL,tance Table F

Resistance dr/dT 177.00 263.32553 0.4320332 178.00 263.75753 0,4319588 179.00 264.18945 0.4318845 180.00 264.62130 0.4318101 181.00 265.05307 0.4317358 182.00 265.48477 0.4316614 183.00 265.91639 0.4315870 184.00 266.34794 0.4315127 185.00 266.77942 0.4314383 186.00 267.21082 0.4313640 187.00 267.64215 0.4312896 188.00 268.07340 0.4312152 189.00 268.50458 0.4311409 190.00 268.93568 0.4310665 191.00 269.36671 0.4309922 192.00 269.79767 0.4309178 193.00 270.22855 0.4308434 194.00 270.65935 0.4307691 195.00 271.09008 0.4306947 196.00 271.52074 0.4306204 197.00 271.95133 0.4305460 198.00 272.38183 0.4304716 199.00 272.81227 0.4303973 200.00 273.24263 0.4303229 201.00 273.67291 0.4302486 202.00 274.10313 0.4301742 203.00 274.53326 0.4300998 204.00 274.96333 0.4300255 205.00 275.39331 0.4299511 206.00 275.82323 0.4298768 207.00 276.25307 0.4298024 208.00 276.68283 0.4297280 209.00 277.11252 0.4296537 210.00 277.54214 0.4295793 211.00 277.97168 0.4295050 212.00 278.40115 0.4294306 R.9 0-/1" Pag of5 Z--

Re poi,. o: RJ200908073-001 Page 3 of 3

Attachment #6 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 16 of 32 ff,4 c:/

Ito a//"o Report No: RJ200908072-001 Page 1 of 3 Report of Calibration Dominion Nuclear Connecticut Rope Fenry Road Bldg 433, Metrology Laboratory Waterford, CT 06385

-31n.,OD-2 76~3a Model:

XS7927PK1 Customer:

Dominion Nuclear Connecticut Serial:

1269 6* 730 Rope Ferry Road

==

Description:==

RTD PLANT INSTALLED Bldg 433, Metrology Laboratory Waterford, CT 06385 Calibration Range:

Limited Received Condition:

New Current:

1.0 mA Procedure:

MTE 1142E Actual Value (Reference)

(C)

-19.92670 0.04100 25.02480 50.01740 75.01330 100.01830 UUT (Test Sensor)

(Ohms) 183.91050 199.86790 219.69950 239.38870 258.92880 278.32550 Residuals (C) 0.0000 0.0000 0.0006

-0.0004 0.0000 0.0001 Method of Realization COMP COMP COMP COMP COMP COMP Callendar-Van Dusen Coefficients:

RO = 199.83522 ALPHA = 0,00392705 DELTA'= 1.53631 BETA = -0.14945 Test Equipment Manufacturer Model Description Serial Number Recall Date Hart Scientific 1590 Super Thermometer II A68350 09/03/2009 Hart Scientific 5681 SPRT 1466 02/25/2010 Hart Scientific 7037 Bath, Deep Low-Temp 93046 NCR Notes: Initial calibration to establish baseline data.

Acceptance Criteria: RO = 19 9.8 p 200.2 ohms Prerequisites rpet: Initial:

Usage Review Required:

Yes CR#:

Pinal'Cahbration Acceptable: Yes/x_.

No _

Calibration Date:

Recall Date:

Temperature:

Humidity:

Customer Order:

08/06/2009 08/01/2029 73.OF 38%

53M20707631 Technician:

Approved By:

This report shall not be reproduced except in.fidl without written approval of Dominion Nuclear Connecticut.

Attachment #6 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 17 of 32 1.-rg.C/Y1IhW7-o' 08/07/2009 Temperature vs. Resihance Table Report

,: RJ200908072-001 Page Z of 3 F

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 16.00 17.00 18.00 19.00 20.00 21.00 22.00 23.00 24.00 25.00 26.00 27.00 28.00 29.00 30.00 31.00 32.00 33.00 34.00 35.00 36.00 37.00 38.00 39.00 40.00 41.00 42.00 43.00 44.00 45.00 46.00 47.00 48.00 49.00 50.00 51.00 52.00 53.00 54.00 55.00 56.00 57.00 58.00 Resistance drldT 185.63222 0.4449825 186.07716 0.4449132 186.52204 0.4448437 186.96685 0.4447740 187.41159 0.4447041 187.85626 0.4446340 188.30086 0.4445637 188.74539 0.4444932 189,18984 0.4444226 189.63423 0.4443517 190.07855 0.4442807 190.52279 0.4442095 190.96697 0.4441381 191.41107 0.4440666 191.85510 0.4439948 192.29906 0.4439229 192.74295 0.4438509 193.18676 0.4437786 193.63050 0.4437062 194.07417 0.4436337 194.51777 0.4435610 194.96129 0.4434881 195.40475 0.4434151 195.84812 0.4433420 196.29143 0.4432687 196.73466 0.4431952 197.17782 0.4431216 197.62091 0.4430479

.198.06392 0.4429741 198.50685 0,4429001 198.94972 0.4428260 199.39251 0.4427517 199.83522 0.4426774 200.27786 0,4426030 200.72043 0.4425285 201.16292 0.4424541 201.60533 0.4423797 202.04768 0.4423053 202.48994 0.4422309 202.93214 0.4421564 203.37426 0.4420820 203.81630 0.4420076 204.25827 0.4419332 204.70017 0.4418587 205.14199 0.4417843 205.58374 0.4417099 206.02541 0.4416355 206.46701 0.4415611 206.90853 0.4414866 207.34998 0.4414122 207.79136 0.4413378 208.23266 0.4412634 208.67388 0.4411889 209,11503 0.4411145 209.55611 0.4410401 209.99712 0.4409657 210.43804 0.4408913 210.87890 0.4408168 211.31968 0.4407424 F

Resistance drldT 59.00 60.00 61.00 62.00 63.00 64.00 65.00 66.00 67.00 68.00 69.00 70.00 71.00 72.00 73.00 74.00 75.00 76.60 77.00 78.00 79.00 80.00 81.00 82.00 83.00 84.00 85.00 86.00 87.00 88.00 89.00 90.00 91.00 92.00 93.00 94.00 95.00 96.00 97.00 98.00 99.00 100.00 101.00 102.00 103.00 104.00 105.00 106.00 107.00 108.00 109.00 110.00 111.00 112.00 113.00 114.00 115.00 116.00 117.00 211.76038 0.4406680 212.20101 0,4405936 212.64157 0.4405191 213.08205 0.4404447 213.52246 0.4403703 213.96279 0.4402959 214.40305 0.4402215 214.84324 0.4401470 215.28334 0.4400726 215.72338 0.4399982 216.16334 0.4399238 216.60323 0.4398493 217.04304 0.4397749 217.48278 0.4397005 217.92244 0.4396261 218.36203 0.4395517 218.80154 0.4394772 219'.24098 0,4394028 219.68035 0.4393284 220.11964 0.4392540 220.55886 0.4391795 220.99800 0.4391051 221.43707 0.4390307 221.87606 0,4389563 222.31498 0.4388819 222,75383 0.4388074 223.19260 0.4387330 223,63129 0.4386586 224.06991 0.4385842 224.50846 0.4385097 224.94693 0.4384353 225.38533 0,4383609 225.82365 0.4382865 226.26190 0.4382121 226.70008 0.4381376 227.13818 0.4380632 227.57620 0,4379888 228.01416 0.4379144 228.45203 0.4378399 228.88984 0.4377655 229.32756 0.4376911 229,76522 0.4376167 230,20280 0.4375423 230.64030 0.4374678 231.07773 0.4373934 231.51509 0.4373190 231.95237 0.4372446 232.38958 0.4371701 232.82671 0.4370957 233.26377 0.4370213 233.70075 0.4369469 234.13766 0.4368725 234.57450 0.4367980 235.01126 0,4367236 235.44795 0.4366492 235.88456 0.4365748 236.32110 0.4365003 236.75756 0.4364259 237.19395 0.4363515 F

Resistance drldT 118.00 237.63026 0.4362771 119.00 238.06650 0.4362027 120.00 238.50267 0.4361282 121.00 238.93876 0.4360538 122.00 239.37477 0.4359794 123.00 239.81072 0.4359050 124.00 240.24658 0.4358305 125.00 240,68238 0.4357561 126.00 241.11810 0.4356817 127.00 241.55374 0.4356073 128.00 241.98931 0.4355329 129.00 242.42481 0.4354584 130.00 242.86023 0.4353840 131.00 243.29558 0.4353096 132.00 243.73085 0.4352352 133.00 244.16605 0.4351607 134.00 244.60117 0.4350863 135.00 245,03622 0.4350119 136.00 245.47119 0.4349375 137.00 245.90609 0.4348631 138.00 246.34092 0.4347886 139.00 246.77567 0.4347142 140.00 247.21035 0.4346398 141.00 247.64495 0.4345654 142.00 248.07948 0.4344909 143.00 248.51393 0.4344165 144.00 248.94831 0.4343421 145.00 249.38262 0.4342677 146.00 249.81685 0.4341933 147.00 250.25100 0.4341188 148.00

  • 250.68508 0.4340444 149.00 251.11909 0.4339700 150.00 251.55302 0.4338956 151.00 251.98688 0.4338211 152.00 252.42067 0.4337467 153.00 252.85438 0.4336723 154.00 253.28801 0.4335979 155.00 253.72157 0.4335235 156.00 254.15506 0.4334490 157.00 254.58847 0.4333746 158.00 255.02181 0.4333002 159.00 255.45507 0.4332258 160.00 255.88826 0.4331513 161.00 256.32137 0.4330769 162.00 256.75441 0.4330025 163.00 257.18738 0.4329281 164.00 257.62027 0.4328537 165.00 258.05309 0.4327792 166.00 258.48583 0.4327048 167.00 258.91849 0.4326304 168.00 259.35109 0.4325560 169.00 259.78361 0.4324815 170.00 260.21605 0.4324071 171.00 260.64842 0.4323327 172.00 261.08072 0.4322583 173.00 261.51294 0.4321839 174.00 261.94508 0.4321094 175.00 262.37716 0.4320350 176.00 262.80915 0.4319606

Attachment #6 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM 4

m7-e Page 18 of 32 08/07/2009 Temperature vs. Resx-tance Table

.F Resistance drldT 177.00 263.24108 0.4318862 178.00 263.67293 0.4318117 179.00 264.10470 0.4317373 180.00 264.53640 0.4316629 181.00 264.96803 0.4315885 182.00 265.39958 0.4315141 183.00 265.83106 0.4314396 184.00 266.26246 0.4313652 185.00 266.69379 0.4312908 186.00 267.12504 0.4312164 187.00 267.55622 0.4311419 188.00 267.98732 0.4310675 189.00 268.41835 0.4309931 190.00 268.84931 0.4309187 191.00 269.28019 0.4308443 192.00 269.71100 0.4307698 193.00 270.14173 0.4306954 194.00 270.57239 0.4306210 195.00 271.00297 0.4305466 196.00 271.43348 0.4304721 197.00 271.86392 0.4303977 198.00 272.29428 0.4303233 199.00 272.72456 0.4302489 200.00 273.15477 0.4301745 201.00 273.58491 0,4301000 202.00 274.01498 0.4300256 203.00 274.44496 0.4299512 204.00 274.87488 0.4298768 205.00 275.30472 0.4298023 206.00 275.73448 0.4297279 207.00 276.16417 0.4296535 208.00 276.59379 0.4295791 209.00 277.02333 0.4295047 210.00 277.45280 0.4294302 211.00 277.88219 0.4293558 212.00 278.31151 0.4292814 Report j: RJ200908072-001 Page 3 of 3

Attachment #7 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 19 of 32 Sensor Calculator - Resistance table generator 9,'f'466Il

/9 6

Page 1 of 1 PJ c~

R WE Search MINCO Home Products Applications Resources About Us Table Options Select Element Start Over Interchangeability Quick Calc Gen Temperature Vs. Resistance PLATINUM: PIA TCR = 3.93e-3 Ro = 199.872 A4 =-0.00036445181 B4 =-2.8576799E-06 A8 =-0.00040758054 B8 = 8.3733738E-05 Values are in Ohms Temp

+0

+1

+2

+3

+4

+5

+6

+7

+8

+9

-F 0

185.6584186.1035186.548 186.9934187.4383187.8831 188.3278188.772 189.2170189.661.

10 190.1059 190.5502190.9945 191.4387 191.882 192.3268192.7708193.2147 193.658 194.102 20 194.5459 194.9895195.4330 195.8765 196.3198196.7631197.2063 197.6495 198.092 198.5355 30 198.9784.199.4213199.8640 200.3067 200.7493 201.1918 201.6343 202.0766 202.5189 202.9611 40 203.40331203.8454 204.2873 204.72931205.1711 205.612 206.0546206.496 206.9377 207.3792 50 207.82061208.2619 208.703 209.14431209.585 210.0265 210.467 210.908 211.3491211.7898 60 212.230 212.6710213.1115213.5520213.9923214.4326214.8728215.312 215.7530216.1929 70 216.63281217.0727 217.5124 217.9521 218.3917 218.8312219.2707 219.7101 220.1494 220.5886 80 -1221.02781221.46681221.905 222.3448222.7837223.222 223.6611224.0998224.5383224.9768 1.90 1225.41521225.8536 226.291

_226.7300 227.1681227.6062 228.0442228.4821 228.919 229.3576 Temp

+0

+1

+2

+3

+4

+6

+9 100 229.7953 1

1 1

© 2011 Minco, All Rights Reserved Flexible Heaters I Flex Circuits J Sensors I Instruments I Careers I Privacy Policy I http://www.minco.com/tools/sensorcalc/rtd/display.aspx?action=options 08/06/2012

Attachment #7 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 20 of 32 19 1-M1/14ý0i &A17 e

p4wr&z i 01,3--

TEMP F 0

5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 ITS-90 R (ohms) 185.6584 187.8831 190.1059 192.3268 194.5459 196.7631 198.9784 201.1918 203.4033 205.6129 207.8206 210.0265 212.2305 214.4326 216.6328 218.8312 221.0278 223.2224 225.4152 227.6062 229.7953 Met Lab R (ohms) 185.63222 187.85626 190.07855 192.29906 194.51777 196.73466 198.94972 201.16292 203.37426 205.58374 207.79136 209.99712 212.20101 214.40305 216.60323 218.80154 220.998 223.1926 225.38533 227.5762 229.76522 Serial # 1269 Error(ohms) 0.02618 0.02684 0.02735 0.02774 0.02813 0.02844 0.02868 0.02888 0.02904 0.02916 0.02924 0.02938 0.02949 0.02955 0.02957 0.02966 0.0298 0.0298 0.02987 0.03 0.03008 Error

% Reading 0.0141 0.0143 0.0144 0.0144 0.0145 0.0145 0.0144 0.0144 0.0143 0.0142 0.0141 0.0140 0.0139 0.0138 0.0136 0.0136 0.0135 0.0133 0.0133 0.0132 0.0131 Error

% Span 0.032 0.032 0.033 0.033 0.033 0.033 0.033 0.033 0.032 0.032 0.032 0.032 0.031 0.031 0.031 0.031 0.031 0.030 0.030 0.030 0.030 span 44.1369 Certified by Minco in 1993 Met Lab Cal performed in 2009 16 years between Minco and Met Lab Reports repeatability Drift 0.01%.05deg C/yr 16 year drift RSS drift & repeatability 0.36 0.3604

3 Serial # 1269 ITS-90 VS MET LAB 230 -

225 -

m iq 220

-a IZ 215 E

210 0

C 0D

-ITS TOCaliData0 205

-M LAB Cal Data C

200 195 190 0

10 20 30 40 50 60 70 80 90 100 Temperature (Deg F)

CD 0

(-3 N)

3 Serial # 1269 ITS-90 VS MET LAB 0)

.220 219.8 219.6 219.4 219.2 219 218.8 218.6 218.4 218.2 218

  • ~ITS9!0 Cal Data

-M LAB Cal Data m

--A CA, 0CD 0

P.

(C, 74 74.2 74.4 74.6 74.8 75 Temperature (Deg F) 75.2 75.4 75.6 75.8 76 QI-V CD to0 (0

w N3

Attachment #7 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 23 of 32

,,# jrf9-6'41&A/F f

Sensor Calculator - Resistance table generator Sý,.,r CNO\\

1-L-)

Page 1 of 1 Search I If MINCO Home Products Applications i Resources i About Us Table Options Select Element Start Over Interchangeability Quick Calc Gen.

Temperature Vs. Resistance PLATINUM: PIA TCR = 3.93e-3 R0 = 199.904 A4 = -0.00033185991 B4 = -2.9061316E-06 A8 = -0.00054864581 B8 = 0.00010006569 Values are in Ohms Temp

+0

+1

+2

+3

+4

+5

+6

+7

+8

+9 oF I

0 185.6876186.132 186,577 187,0229187.4679187.9128188.3576188.8023189.2469189.6915 10 190.1360190.5804 191.0248191.4691 191.9133192.3574 192.8014 193.2454193.6893194.1331 20 194.576 195.020 195.4642195.9077196.3511196.7945197.237 197.6810198.1242 198.5673 30 199.0103199.453 199,8960200.3387 200.7813201.2238201.6663202.1087202.5510202.9932 40 203.4354 203.8774 204.3194 204.7614205.2032205.6450206.086 206.5283 206.9699207.4114 50 207.8528 208.2941 208.735 4209.1765 209.6176210.0587 210.4996(210.9405211.3813211.8220 60 212.2627 212.7033 213.1438213.5842214.0246214.4649214,9051215.3452215.7853216.2253 70 216.6652217.105 217.5448217.9845 218.4241218.863 219.30311219.7425220.1818220.6210 80 221.060 221.4993,221.9383 222.3773222.8161223.2549223.6936 224.1323 224.5709 225.0094 90 225.4478 225,8861 *226.3244 226.7626 227.2007 227.63881228.0768 228.5141228.9525 229.3903 Temp

+0

+I1

+2

+3

+4

+5

+6

+7

+8

+9 Fo 100 1229.8279

© 2011 Minco, All Rights Reserved Flexible Heaters I Flex Circuits I Sensors I Instruments I Careers I Privacy Policy I http://www.niinco.com/tools/sensorcalc/rtd/display.aspx?action=options 08/06/2012

Attachment #7 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM

-Page 24 of 32 TEMP F 0

5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 ITS-90 R (ohms) 185.6876 187.9128 190.1360 192.3574 194.5769 196.7945 199.0103 201.2238 203.4354 205.6450 207.8528 210.0587 212.2627 214.4649 216.6652 218.8636 221.0602 223.2549 225.4478 227.6388 229.8279 Met Lab R (ohms) 185.69151 187.91628 190.13928 192.36050 194.57990 196.79749 199.01323 201.22712 203.43915 205.64933 207.85764 210.06410 212.26869 214.47143 216.67231 218.87133 221.06849 223.26379 225.45723 227.64881 229.83854 Serial # 1270 Error(ohms)

-0.00391

-0.00348

-0.00328

-0.0031

-0.003

-0.00299

-0.00293

-0.00332

-0.00375

-0.00433

-0.00484

-0.0054

-0.00599

-0.00653

-0.00711

-0.00773

-0.00829

-0.00889

-0.00943

-0.01001

-0.01064 Error

% Reading

-0.0021

-0.0019

-0.0017

-0.0016

-0.0015

-0.0015

-0.0015

-0.0016

-0.0018

-0.0021

-0.0023

-0.0026

-0.0028

-0.0030

-0.0033

-0.0035

-0.0038

-0.0040

-0.0042

-0.0044

-0.0046 Error

% Span

-0.005

-0.004

-0.004

-0.004

-0.003

-0.003

-0.003

-0.004

-0.004

-0.005

-0.005

-0.006

-0.006

-0.007

-0.007

-0.008

-0.008

-0.009

-0.009

-0.010

-0.010 span 44.1403 Certified by Minco in 1993 Met Lab Cal performed in 2009 16 years between Minco and Met Lab Reports repeatability Drift 0.01%.05deg C/yr 16 year drift RSS drift & repeatability 0.36 0.3604

C)

CD Serial # 1270 ITS-90 VS MET LAB 230.0000 225.0000 220.0000 215.0000 E

210.0000 205.0000 200.0000 195.0000 190.0000 185.0000 ITS O Cal Data M ETLAB Cal Data M

-i CD C)

C)

CD 0

3 0

0)

N 0

10 20 30 40 50 Temperature (Deg F) 60 70 80 90 100 ID C) 0 C.)

N)

Serial 1270 ITS-90 VS MET LAB 0

0 0

a, U

(0 a) 220.0000 219.8000 219.6000 219.4000 219.2000 219.0000 218.8000 218.6000 218.4000 218.2000 218.0000 Serial # 1270 ITS-90 VS MET LAB ITS! 10 Cal Data so-METLAB Cal Data m

-Im

-u N,

0 N,

C71 0

0 0

0 0

00.

0 co N,

0 N,

Co 00) 6 CD

  • 4, 74 74.2 74.4 74.6 74.8 75 Temperature (Deg F) 75.2 75.4 75.6 75.8 76 (D

0 Co N,

Attachment #7 ETE-MP-2012-1135 Report Generated on 816/2012 3:06:09 PM Page 27 of 32 Sensor Calculator - Resistance table generator Page 1 of 1

/'A 166 2 -,3Zý Faý Search~

MINCO Home Products I Applications Resources 1 About Us C

Table Options Select Element Start Over Interchangeability Quick Calc Genej Temperature Vs. Resistance PLATINUM, PIA TCR = 3.93e-3 Ro = 200.008 A4 =-0.00060597493 B4 = 0.0003131495 AB =-0.00045088398 Ba =0.00014870006 Values are in Ohms emp

+0

+1

+2

+3

+4

+5

+6

+7

+8

+9 0F 0

185.7885186.2337186.6789 187.1241187.5691188.0141188.4590188.9038189.3485189.7932 10 190.2378190.6823191.1267191.5711192.015 192.459 192.9038193.3478193.7918194.2358 20 194.6796195.1234195.5671196.0107 196.4543 196.8978 197.3412 197.7845 198.2277 198.6709 30 199.1140,199.55711200.0000 200.4430.200.8859 201.3287.201.7714 202.2140 202.6566 203.0991 40 203.5415 203.9839 204.42621204.8684 205.3105 205.7526 206.1945 206.6364 207.0783 207.5200 50 207.9617 208.40331208.8448 209.28631209.7277 210.1690 210.610 211.0514 211.4924 211.9334 60 212.3744 212.8152 213.2560213.6967 214.1374 214.5779 215.0184 215.4588 215.8992 216.339 70 216.7796217.2197 217.65981218.0997 218.5396 218.9794 219.4192219.8588220.298 220.737 80 221.177 221.6168 222.05611222.49532 22.9344 223.3735 223.8125 224.2514 224.6903 225.129 90 225.5677 226.0064 226.444 226.8834 227.3218 227.7601 228.198 228.6366 229.0747 229.512 Temp

+0

+1

+2

+3

+4

+5

+6

+7

+8

+9 100 1 229.9507

+1 1

171

@ 2011 Minco, All Rights Reserved Flexible Heaters I Flex Circuits I Sensors I Instruments I Careers I Privacy Policy J Sit http://www.minco.com/tools/sensorcalc/rtd/display.aspx?action-options 08/06/2012

Attachment #7 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 28 of 32 J9 6Z8 ot 31.

Serial # 1309 ITS-90 Met Lab Error Error TEMP F R (ohms)

R (ohms)

Error(ohms)

% Reading

% Span 0

185.7885 185.78069 0.00781 0.0042 0.010 5

188.0141 187.98560 0.0285 0.0152 0.034 10 190.2378 190.19543 0.04237 0.0223 0.050 15 192.4596 192.40868 0.05092 0.0265 0.060 20 194.6796 194.62397 0.05563 0.0286 0.065 25 196.8978 196.84003 0.05777 0.0293 0.066 30 199.1140 199.05569 0.05831 0.0293 0.066 35 201.3287 201.26991 0.05879 0.0292 0.066 40 203.5415 203.48228 0.05922 0.0291 0.066 45 205.7526 205.69278 0.05982 0.0291 0.066 50 207.9617 207.90141 0.06029 0.0290 0.066 55 210.1690 210.10817 0.06083 0.0289 0.066 60 212.3744 212.31306 0.06134 0.0289 0.065 65 214.5779 214.51608 0.06182 0.0288 0.065 70 216.7796 216.71724 0.06236 0.0288 0.065 75 218.9794 218.91653 0.06287 0.0287 0.065 80 221.1774 221.11395 0.06345 0.0287 0.065 85 223.3735 223.30950 0.064 0.0287 0.065 90 225.5677 225.50319 0.06451 0.0286 0.065 95 227.7601 227.69500 0.0651 0.0286 0.065 100 229.9507 229.88495 0.06575 0.0286 0.065 span 44.1622 Certified by Minco in 2000 Met Lab Cal performed in 2009 16 years between Minco and Met Lab Reports repeatability Drift 9 year drift RSS drift & repeatability 0.01%.05deg C/yr 0.27 0.2706

0

7 230.0000 225.0000 220.0000 215.0000 E

0 210.0000 cz 205.0000 200.0000 195.0000 190.0000 185-0000 Serial # 1309 ITS-90 VS MET LAB

-ITS 0 Cal Data

=ME" LAB Cal Data CD

-t

,m 0G)

Co M

CD 0

0 C.

0L 0

10 20

.30 40 50 Temperature (Deg F) 60 70 80 90 100 "0

(0 0

0 3

Serial # 1309 ITS-90 VS MET LAB 220.0000 219.8000 219.6000 219.4000 2.

o 219.2000 U

S219.0000 218.8000 218.6000 218.4000 MT "0 Cal Data

~~~~~~~

~

~

IVE

____~~LAB CalI Data 0

0b 0

C-5)

A 218.2000 218.0000 I

+

I

+

74 74.2 74.4 74.6 74.8 75 75.2 75.4 75.6 75.8 76 Temperature (Deg F)

(.0 0

0 CD Nancy M Nowlan (Generation - 6)

From:

Anderson, Tereasa [tereasa.anderson@minco.com]

Sent:

Monday, July 30, 2012 6:25 PM To:

Gus J Filippides (Generation - 6)

Cc:

Nancy M Nowlan (Generation - 6)

Subject:

Calibrated info reference REQ acc rating to old PO# 932995, Gus from DOM!

Attachments:

TS102(N)_L-6.pdf; S7927.pdf

Gus, I just now got it from engineering.

"Somewhere, I seem to have missed the original request for a 0.025°C accuracy guarantee. I have attached an old catalog page that describes the different m

calibration tables, and their associated accuracy. Please note that their probe model (57927) is not listed in this table (it had already been discontinued whenT the catalog was published). However, calibration table RT01 is listed as being accurate to +/-0.025°C. I have also attached the S7927 specification drawing that provides repeatability and stability for this model.

You may notice that the model number listed on the calibration certs is "XS7927PK ". The "X" prefix designates an alteration of some sort from the standard

0 design. In this case, "X" stands for a 200-ohm platinum element. That does not affect table calibration accuracy, nor repeatability, nor stability."

CD Thank you!

Tereasa Anderson I Customer Service Representative I Minco Products Tel: 763.586.2920 I Fax: 763-571-0927 Tereasa.anderson@minco.com 0

7300 Commerce Lane I Minneapolis, MN 55432 wwW.rn nc0, corn co From: Gus J Filippides [1]

Sent: Monday, July 30, 2012 10:28 AM To: Anderson, Tereasa Cc: Nancy M Nowlan

Subject:

FW: Calibrated info reference to old PO# 932995, Gus from DOM!

Hello Tereasa, Did you get a response to my follow up question that I sent last week. The calibration reports do not cite any acceptance criteria for the RTDs, ie 0.025 degrees F uncertainty.

What is the accuracy of the RTDs in question that were supplied to us?

CD 1

0C.,

Attachment #8 ETE-MP-2012-1135 Report Generated on 8/6/2012 3:06:09 PM Page 32 of 32 Ref&nc~1Thiiriieters Calibration Minco's fully equipped standards lab can calibrate and furnish Individual tables for any thermometer in this bulletin. All temperatures are based on ITS-90. There are two types of calibration:

Fixed point calibrations furnishes the actual fixed point val-ues and the resulting coefficients.

Comparison calibrations Used for oil filled thermometers, comparison calibrations match the thermometer to be measured against an NIST traceable standard, rather than fixed points.

Absolute vs ratio tables reads 1.77328 at 2001C. A ratio like this tends to remain more stable over time than an absolute resistance.

You can periodically recheck Rooc in an ice bath and use the latest value for ratio calculation.

RTD calculations These calibrations use measurements taken at points derived from physical constants, such as the triple point of water or the freezing point of tin.

The observed values are used to cal-culate coefficients for the interpolat-ing equations of ITS-90. In addition to the comniete R vs T table. Minco Tables may list either absolute resis-tance values (RT vs T) or resistance ratios (R-t/Ro vs T) indexed to 1. For example, if the absolute values of a thermometer are 177.47 Q at 2000C and 100.08 9 at 01C, the ratio table Minco offers two options to create and print RTD tables. You can download the DOS PC program "RTD Cailc" for use on your com-puter. We also offer "SensorCalc", an on-line tool for creating, printing and saving RTD tables. Both are available at www mincn.com/support.

Thermometer model (accuracy specifications may vary for other models)

S9688-S9691

S7924, S7928

$7929 S9762 S1059 S7932-S7933 S79253 Type of calibration Fixed point Fixed point Fixed point Fixed point Comparison Fixed point

-189 to

-189to

-189 to

-269 to

-189 to

-189 to Temperature range 500oM 500OC 660o0 260°0 2600C

-100 to 25000 64000 Ordering number for tables RT vs T, 10C Increments RT05 RT05 RT16 RT01 RT09 RT12 ir/R0 vs T, 10C Increments RT06 RT06 RT17 RT02 RTlO RT13 RT vs T, O.10C Increments RT07 RT18 RT11 1 RT03 RT14 Rr/Ro vs T, 0.1 *C Increments RT08 RT19 RT04 RTI5 Table accuracy

+/-0.0250C

+/-0,02500

+/-0.0250C

+/-0.025

+/-0.0250C

+/-0.03500

+/-0.0250C Data points supplied

-268.95*C (4.2K): BP helium*

-189.340C : TP of argon2

-50"C, -25°C, 175"*3

-38.830C : TP of mercury2 0.01°C :Triple Point of water 100'C: BP water2 231.928°C : FP tin 419.5270C: FP zinc 660.323°C : FP aluminum

(+0.05-0) 1 Cryogenic calibrations available on special order. 2 Comparison calibration measurement.

3 S7924, S7925 calibrations based on IEC 751, not ITS-90 The Minco strain-free element Minco precision thermometers rely on a unique element de-sign for superior precision. It consists of a helical coil of plati-num wire, spiraled onto a grooved ceramic bobbin. This con-struction gives more support to the coils than "birdcage" designs, but does not stress the element during thermal ex-pansion. The result is stable and repeatable measurements under a variety of conditions.

Page L-6 Mmcv Sales: 1-763-571-3121 Fax: 1-763-571-0927 Web: www.minoo.com Bulletin TS-102 Page L-6 Minco Sales: 1-763-571-3121 Fax. 1-763-571-0927 Web: www.minco.com Bulletin TS-102