The following information was provided by the
Wisconsin Department of Health Services Radiation Protection Section (the Department) via email:
On September 11, 2025, at 1306 CDT, the Department received a telephone notification that the licensee's Jubilant DraxImage Inc. RUBY-FILL Rb-82 generator (Serial Number: GLO03160183) was unable to deliver a patient dose at approximately 1000 CDT on September 11, 2025.
The device provided an error during the infusion stating that the dosage could not be delivered to the patient. The generator stated that 0.2 millicuries of Rb-82 was delivered to the patient out of an intended 20 millicuries. At approximately 1300 CDT on the same day, the licensee completed the daily quality control tests on the generator and simulated a patient infusion with a shielded vial. The device delivered the intended dosage to the vial and appears to be in working order. The licensee has a scheduled visit with the manufacturer the week of September 15-19, 2025, to evaluate the cause of failure. No contamination resulted from the incident.
Event Report ID No.: WI250012