The following information was provided by the
South Carolina Department of Environmental Control (the Department) via email:
At approximately 2100 EST, on December 24, 2024, the licensee's staff noticed a tear in the Mylar window on one of their Thermo-Fisher FC-185 beta gauges, thus causing the system to be shut down. The system was put back into service at approximately 0000 on December 25, 2024. The licensee's radiation safety officer contacted the Department to report the incident at 1502 on December 25, 2024. The Thermo-Fisher Model TFC-185 contains 1250 mCi of Kr-85. The source model number is a Kr85-4, and the serial number is KA-2197. There were no radiation exposures resulting from the service of the gauge.
- * * UPDATE ON 03/06/2025 AT 0840 FROM ANDREW ROXBURGH TO BILLY NYTKO * * *
The licensee submitted its 30-day written report on January 16, 2025. The report indicated that the Mylar window was replaced at 2350 on December 24, 2025. The report also indicated that there were no increased radiation exposures to individuals during the repair. The Department considers this event closed.
Notified R1DO(Ford) and
NMSS Events (email)