ENS 55468
From kanterella
Jump to navigation
Jump to search
ENS Event | |
|---|---|
07:00 Sep 13, 2021 | |
| Title | Radiography Device Malfunction |
| Event Description | The following information was received from the state of California via email:
On 9/14/2021, the licensee contacted the Radiation Health Branch (RHB) to report an incident that occurred on 9/13/2021 in which a radiography exposure device failed to retract to the fully shielded position. The licensee reported that after completing their second exposure shot, the radiography crew cranked the source to the fully shielded position which actuated the fully shielded source position indicator. The crew reverse cranked the source to confirm that the source was locked into the fully shielded position and discovered that the source was still exposed. The radiographers were able to lock the source into the final position after manipulating the locking mechanism manually. The licensee reported an exposure maximum of 33 millirem to one of the radiographers due to this incident. At this time, the licensee does not expect an overexposure to either radiographers nor members of the public. RHB will be investigating this incident further. |
| Where | |
|---|---|
| Mistras Group Richmond, California (NRC Region 4) | |
| License number: | 4886-48 |
| Organization: | California Radiation Control Prgm |
| Reporting | |
| Agreement State | |
| Time - Person (Reporting Time:+79.02 h3.293 days <br />0.47 weeks <br />0.108 months <br />) | |
| Opened: | K. Arunika Hewadikaram 14:01 Sep 16, 2021 |
| NRC Officer: | Brian Lin |
| Last Updated: | Sep 16, 2021 |
| 55468 - NRC Website
Loading map... {"minzoom":false,"maxzoom":false,"mappingservice":"leaflet","width":"350px","height":"250px","centre":false,"title":"","label":"","icon":"","lines":[],"polygons":[],"circles":[],"rectangles":[],"copycoords":false,"static":false,"zoom":6,"defzoom":14,"layers":["OpenStreetMap"],"image layers":[],"overlays":[],"resizable":false,"fullscreen":false,"scrollwheelzoom":true,"cluster":false,"clustermaxzoom":20,"clusterzoomonclick":true,"clustermaxradius":80,"clusterspiderfy":true,"geojson":"","clicktarget":"","imageLayers":[],"locations":[{"text":"\u003Cb\u003E\u003Cdiv class=\"mw-parser-output\"\u003E\u003Cp\u003E\u003Ca class=\"mw-selflink selflink\"\u003EENS 55468\u003C/a\u003E - \u003Ca href=\"/Mistras_Group\" title=\"Mistras Group\"\u003EMistras Group\u003C/a\u003E\n\u003C/p\u003E\u003C/div\u003E\u003C/b\u003E\u003Cdiv class=\"mw-parser-output\"\u003E\u003Cp\u003ERadiography Device Malfunction\n\u003C/p\u003E\u003C/div\u003E","title":"ENS 55468 - Mistras Group\n","link":"","lat":61.170877777777775,"lon":-149.87442222222222,"icon":""}],"imageoverlays":null} | |
Mistras Group with Agreement State | |
WEEKMONTHYEARENS 575662025-02-21T08:00:00021 February 2025 08:00:00
[Table view]Agreement State Damaged Radiography Equipment ENS 573572024-08-30T15:45:00030 August 2024 15:45:00 Agreement State Radiography Camera Failure ENS 554682021-09-13T07:00:00013 September 2021 07:00:00 Agreement State Radiography Device Malfunction ENS 542582019-08-30T16:30:00030 August 2019 16:30:00 Agreement State Agreement State - Radiography Camera Source Unable to Retract ENS 539702019-04-01T07:00:0001 April 2019 07:00:00 Agreement State Lost Then Found Category 2 Radiography Exposure Device ENS 475912012-01-04T15:30:0004 January 2012 15:30:00 Agreement State Agreement State Report - Radiography Camera Not Properly Secured 2025-02-21T08:00:00 | |