| Title | Medical Event Due to Potential Different Fractional Dose Delivered than Prescribed |
| Event Description | The licensee provided notification that a patient received 2 occurrences of a dose less than prescribed when delivering ten fractions of a treatment. Each of the underdoses were approximately 50% of the 340 Gray prescribed fractional dose. The patient will receive additional dose fractions in order to achieve the written directive total dose. The Radiation Oncologist has notified the patient and attending physician.
On January 17, 2012 the NRC Operations Center was verbally notified of two Therapeutic Underdose Occurrences discovered by the licensee on January 16 and 17, 2012. These occurrences involved a fractionated Breast High Dose Rate Afterloader (HDR) treatment with a SenoRx Contura multicatheter breast applicator. The first and third delivered treatment fractions were found to be less than 50% of the intended fractional dose. The entire course of the treatment in the written directive included ten equal-dose fractions of 3.4 Gray per fraction for a total dose of 34 Gray to the prescribed treatment site. To correct for the underdose occurrences, two additional treatment fractions were added and the treatment plan was modified to achieve the total dose specified in the written directive. The licensee now believes that this medical event has also caused an unintended dose to skin outside of the prescribed treatment site, requiring notification under 10CFR35.3045(a)(3). The licensee has performed computer simulation, calculations and physical measurements using TLDs simulating the treatment geometry to model the unintended skin dose. The event delivered an unintended skin dose exceeding at least the skin erythema threshold (2 Gy). The licensee is continuing to monitor the patient response to the skin dose and is working to refine the unintended skin dose estimates. An NRC reactive inspection team is on-site. Notified R4DO (Jeff Clark) and FSME (Greg Suber)
The licensee confirmed that they agree with their medical consultants' findings that the patient received approximately 2720 rads of unintended skin dose. Notified R4DO (Gaddy) and FSME (McIntosh). A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient. |
| Where | |
|---|---|
| Avera McKennan Hospital Sioux Falls, South Dakota (NRC Region 4) | |
| License number: | 4016571-01 |
| Organization: | Avera Mckennan Hospital |
| Reporting | |
| 10 CFR 35.3045(a)(1) | |
| Time - Person (Reporting Time:+14.07 h0.586 days <br />0.0837 weeks <br />0.0193 months <br />) | |
| Opened: | Richard Massoth 13:04 Jan 17, 2012 |
| NRC Officer: | Vince Klco |
| Last Updated: | Apr 12, 2012 |
| 47599 - NRC Website
Loading map... {"minzoom":false,"maxzoom":false,"mappingservice":"leaflet","width":"350px","height":"250px","centre":false,"title":"","label":"","icon":"","lines":[],"polygons":[],"circles":[],"rectangles":[],"copycoords":false,"static":false,"zoom":6,"defzoom":14,"layers":["OpenStreetMap"],"image layers":[],"overlays":[],"resizable":false,"fullscreen":false,"scrollwheelzoom":true,"cluster":false,"clustermaxzoom":20,"clusterzoomonclick":true,"clustermaxradius":80,"clusterspiderfy":true,"geojson":"","clicktarget":"","imageLayers":[],"locations":[{"text":"\u003Cb\u003E\u003Cdiv class=\"mw-parser-output\"\u003E\u003Cp\u003E\u003Ca class=\"mw-selflink selflink\"\u003EENS 47599\u003C/a\u003E - \u003Ca href=\"/Avera_McKennan_Hospital\" title=\"Avera McKennan Hospital\"\u003EAvera McKennan Hospital\u003C/a\u003E\n\u003C/p\u003E\u003C/div\u003E\u003C/b\u003E\u003Cdiv class=\"mw-parser-output\"\u003E\u003Cp\u003EMedical Event Due to Potential Different Fractional Dose Delivered than Prescribed\n\u003C/p\u003E\u003C/div\u003E","title":"ENS 47599 - Avera McKennan Hospital\n","link":"","lat":43.53422222222222,"lon":-96.71395,"icon":""}],"imageoverlays":null} | |
Avera McKennan Hospital with 10 CFR 35.3045(a)(1) | |
WEEKMONTHYEARENS 571712024-06-12T21:00:00012 June 2024 21:00:00
[Table view]10 CFR 35.3045(a)(1) Medical Event - LUTETIUM-177 Misadministration ENS 475992012-01-16T23:00:00016 January 2012 23:00:00 10 CFR 35.3045(a)(1) Medical Event Due to Potential Different Fractional Dose Delivered than Prescribed 2024-06-12T21:00:00 | |