Florida was notified on 10/17/2008 by the licensee's Medical Physicist regarding a dose of 3400 cGy (3400 rad) administered to unintended tissues during several breast cancer therapy treatments over the period of September 10 - 17, 2008. The apparent unintended dose was identified on 10/16/2008, when the patient reported to the licensee symptoms of erythma (skin reddening) to the breast not intended to be treated.
Specifically, the patient was being treated for breast cancer with an Ir-192 High Dose Rate (HDR) Afterloader unit (source strength, manufacturer, and model unknown at the time of this report). When the erythema was reported by the patient, the Medical Physicist reviewed the records and determined that the HDR Afterloader was mis-programmed such that the source stopped 10 centimeters short of the intended tumor bed in the right breast. As a result, the entire dose intended for the tumor bed was administered to the left breast that was not intended to be treated.
The Commonwealth of Florida did not currently have information regarding any potential long term effects for the patient due to this event. The Commonwealth of Florida will dispatch an inspector to the facility early next week to follow-up on this event. A written report of this event will be provide by the State at that time.