BVY 14-006, Unsatisfactory Blind Quality Assurance Drug Testing Samples
| ML14042A198 | |
| Person / Time | |
|---|---|
| Site: | Vermont Yankee File:NorthStar Vermont Yankee icon.png |
| Issue date: | 02/05/2014 |
| From: | Chappell C Entergy Nuclear Operations |
| To: | Document Control Desk, Office of Nuclear Security and Incident Response |
| References | |
| BVY 14-006 | |
| Download: ML14042A198 (7) | |
Text
Entergy Entergy Nuclear Operations, Inc.
Vermont Yankee 320 Governor Hunt Road Vernon, VT 05354 Tel 802 257 7711 Coley C. Chappell Licensing Manager BVY 14-006 February 5, 2014 U.S. Nuclear Regulatory Commission Attn: Document Control Desk Washington, DC 20555
SUBJECT:
Unsatisfactory Blind Quality Assurance Drug Testing Samples Vermont Yankee Nuclear Power Station Docket No. 50-271 License No. DPR-28
Dear Sir:
On October 17, 2013 and again on October 24, 2013, Vermont Yankee Nuclear Power Station (VY) Fitness for Duty (FFD) Staff confirmed invalid test results from two blind Quality Assurance drug testing samples that were purchased from Professional Toxicology and sent to the Quest Diagnostics Laboratory in Norristown, Pennsylvania for analysis.
On January 7, 2014, the Arkansas Nuclear One FFD staff completed an investigation on behalf of Entergy pertaining to the blind sample errors. Pursuant to the report requirements of 10 CFR 26.719(c)(1), the investigation results and corrective actions are documented in the attachment to this letter.
There are no new regulatory commitments being made in this submittal.
Should you have any questions concerning this submittal, please contact Tina Emery-Howe, Supervisor, VY FFD at (802) 258-5804.
Sincerely,
[CCC/ble]
Attachment:
Blind QA Sample Error Investigation Report cc listing (next page)
BVY 14-006 / Page 2 of 2 cc:
Mr. William M. Dean Region 1 Administrator U.S. Nuclear Regulatory Commission 2100 Renaissance Blvd, Suite 100 King of Prussia, PA 19406-2713 Mr. Douglas Pickett, Project Manager Division of Operating Reactor Licensing Office of Nuclear Reactor Regulation U.S. Nuclear Regulatory Commission Mail Stop 08C2A Washington, DC 20555 USNRC Resident Inspector Vermont Yankee Nuclear Power Station 320 Governor Hunt Road Vernon, VT 05354 Mr. Christopher Recchia, Commissioner VT Department of Public Service 112 State Street, Drawer 20 Montpelier, VT 05620-2601
Docket No. 50-271 BVY 14-006 Attachment Blind QA Sample Error Investigation Report
Attachment to BVY 14-006 Page 1 of 4 Blind QA Sample Error Investigation Report This investigation report addresses errors related to invalid test results identified by Entergy at Arkansas Nuclear One (ANO) and Vermont Yankee (VY). On January 7, 2014 the ANO Fitness for Duty (FFD) staff completed an investigation on behalf of Entergy, as described below. The ANO report was submitted separately in Reference 1.
Identification of Error:
At ANO thirty FFD blind Quality Assurance (QA) specimens were purchased from Professional Toxicology (Pro Tox) on October 7, 2013. Within this group were four specimens from Lot 1307N-AMP that had been certified at Substance Abuse and Mental Health Services Administration (SAMSHA) / Health and Human Services (HHS) certified laboratories to be valid human urine specimens and positive for amphetamine. The pre-certified methamphetamine level in this lot was 1800 ng/ml.
All of the above specimens were maintained under refrigeration and/or were kept frozen until prepared for shipment to the Quest Diagnostics Laboratory in Lenexa, Kansas.
When prepared, all specimens were warmed to room temperature in manners traditionally used to prepare blind QA samples. Two of the four specimens from Lot 1307N-AMP identified as specimens 0058304 and 0058313 were shipped on October 28 and 29, 2013, respectively. Results of these specimens confirmed the presence of amphetamine at levels consistent with the Certification of Analysis obtained by Pro Tox (1742 ng/ml and 1751 ng/ml respectively). However, both specimens were determined to be invalid.
The invalid determination was based on the definition provided in 10 CFR 26.5 as:
result reported by an HHS-certified laboratory for a specimen that contains an unidentified adulterant, contains an unidentified interfering substance, has an abnormal physical characteristic, contains inconsistent physiological constituents, or has an endogenous substance at an abnormal concentration that prevents the laboratory from completing testing or obtaining a valid drug test result.
In this instance, the specimen results were reported as invalid based on inconsistent creatinine concentration and specific gravity results (10 CFR 26.161(f)(1)). Specifically, the creatinine concentrations were less than 2 mg/dl (reported creatinine levels of 1.5 mg/dl and 1.4 mg/dl respectively) and the specific gravity results were greater than 1.0010 but less than 1.0200 (reported 1.0155 and 1.0150 respectively).
VY also purchased thirty blind QA specimens from Pro Tox on September 23, 2013.
Within this group of specimens were the following two lot numbers Lot 1307N-AMP and Lot 1308-THC. The analysis of these specimens was performed at Quest Diagnostics Laboratory in Norristown, Pennsylvania and only the certification of the initial results was from Pro Tox.
Attachment to BVY 14-006 Page 2 of 4 Investigation:
Upon receipt of these results, the following actions were initiated:
- 1. The Medical Review Officer was informed of the error.
- 2. Pro Tox was notified and asked to initiate an investigation to determine the cause for the invalid results. When notified, the Pro Tox scientist, Dr. Stan Kammerer, expressed surprise and indicated that he had no plausible explanation for the results. He stated that the samples had originally been certified to be positive for amphetamine and valid with creatinine levels > 75 mg/dl. He further stated specimens from the same lot number had been distributed to other utilities and no other abnormal results had been reported. Dr. Kammerer inquired as to the method used by the Quest Lenexa lab to quantify creatinine. Since the method used to perform the initial certification was a colorimetric method, he questioned whether the same method was used for our confirmation.
- 3. Quest Diagnostics was notified and asked to validate the creatinine levels previously reported on specimens 0058304 and 0058313 and to confirm the method used to quantify creatinine. The creatinine levels were confirmed and the method used by Quest Diagnostics in the Lenexa lab was confirmed to be the colorimetric method, which was the same method used during the lot's original certification.
- 4. Entergy fleet plants were notified of the event and provided the lot number of the specimens in question. While other fleet plants had obtained samples from this lot, only Waterford and Vermont Yankee had processed these samples. Three specimens from this lot had been processed at the Quest Diagnostics laboratory in Norristown, Pennsylvania, by Vermont Yankee. Results of these specimens were consistent with the invalid results described above. Vermont Yankee specimens 8967960, 8967962, and 8967968 were submitted for testing on October 17 and 24, 2013. These specimen returned results that were positive for amphetamine but invalid based on creatinine levels of 1.6 mg/dl, 1.3 mg/dl, and 1.2 mg/dl. Condition Reports CR-VTY-2013-06412 and 06490 were written to document these errors. One specimen from lot 1307N-AMP was processed at Waterford 3. Although this specimen produced results that were positive for amphetamine and valid, the creatinine level was recorded to be 3.8 mg/dl. This value, although within acceptable validity parameters, represents a significant depletion of creatinine from the original certification level of > 75 mg/dl reported by Dr. Kammerer at Pro Tox.
Four specimens from Lot 1308-THC were also submitted by Vermont Yankee.
Specimens 8968033, 8968035, 8968036, and 8968037 had been certified as positive for THC and valid produced results that were positive for THC but also invalid due to low creatinine levels. All four of these specimens produced creatinine levels of 1.5 mg/dl. Although these specimens were assigned a different lot number than those described above, they are included in this investigation based on the similarity of the condition, and the common clean-urine volume and process from which they were prepared.
All Vermont Yankee specimens (amphetamine and THC) were processed at the Quest Diagnostics laboratory in Norristown, Pennsylvania.
Attachment to BVY 14-006 Page 3 of 4 Specimens from lot 1307N-AMP were also processed by non-Entergy licensees.
Specifically, Nine Mile Point and Dominion received invalid results from specimens in this lot. Ongoing dialogue with personnel at these facilities indicates that a cause for the depleted creatinine was not determined. Conversations with Dr. Stan Kammerer at Pro Tox over the course of the investigation failed to indicate a conclusive explanation for the depleted creatinine in the reference lot. According to Dr. Kammerer, he discussed the anomaly with representatives from SAMSHA who were also unable to offer an acceptable explanation. The following possible causes for the depleted creatinine levels were offered by Pro Tox:
- 1. A residual amount of bleach solution used to rinse the shipping containers could possibly have caused a chemical breakdown of the creatinine molecules after the lot is placed in shipping containers. This possibility was ruled out when the rinse solution was changed to a non-bleach solution and subsequent testing on specimens from a new lot, lot 1312N-AMP, continued to produce results with depleted creatinine.
- 2. One of the drug free urine providers used by Pro Tox had begun a drastic dietary regimen consisting of mega-vitamins, natural herbs and anti-oxidants. Pro Tox considered if possibly a constituent of their diet could be causing the creatinine to deteriorate. According to an email from Dr. Kammerer which is attached to this report, this scenario is still under investigation. However, two additional amphetamine specimens from lot 1312N-AMP were obtained from Pro Tox as part of our ongoing investigation. These specimens had been certified to be positive for amphetamine and valid with creatinine levels of 75 mg/dl according to Dr. Kammerer. This lot was supposed to have been prepared with urine obtained from new providers. Results from these two specimens indicated depleted creatinine levels of 14.9 mg/dl, and 15.0 mg/dl. These results were provided to Dr. Kammerer. He expressed continued surprise and offered no explanation for the depleted creatinine.
- 3. In an email correspondence between Jerry Woods, ANO FFD Supervisor, and Dr. Stan Kammerer, Director Professional Toxicology Services, Dr. Krammerer makes the following statement. "Several of our blind solutions have been reported 'invalid' due to the lack of creatinine. These blinds were prepared from drug-free urine that was tested at our certified lab and had results of at least 75 mg/dl of creatinine. The loss or degradation of creatinine is baffling to me and to the other scientists with whom I have discussed this situation. The only change in our procedure in the past seven years is that one of our donors of drug-free urine used to prepare our solutions recently changed his diet radically to a 'healthy' regimen including mega vitamins, natural herbs and anti-oxidants. I am hesitant to declare that this caused our creatinine problem, but we have had no problem since we stopped using that donors' urine. I will continue to monitor each donors' urine to determine whether their diet can affect the creatinine content."
Attachment to BVY 14-006 Page 4 of 4
==
Conclusion:==
Depleting levels of creatinine in blind QA specimens manufactured by Pro Tox has been identified in multiple lots of specimens and confirmed by two independent SAMSHA/HHS certified labs since July of 2013. Despite this recurring anomaly, the toxicologist at Pro Tox has been unable to provide a plausible explanation for the unexpected results.
Absent the ability of the blind QA sample vendor to provide reliable specimens to validate the integrity of the FFD testing program, the vendor in question will not be used until valid explanation(s) are provided and issues producing unreliable results are corrected.
Immediate Actions:
The Arkansas Nuclear One and Vermont Yankee Medical Review Officers were notified.
Fleet plants were notified of the discrepancy.
A Condition Report was initiated at all Entergy plants affected by this condition.
An investigation was conducted.
Additional Corrective Actions:
Based on the above findings, another provider of blind QA specimens will be utilized throughout the Entergy fleet until the issues at Pro Tox have been identified and corrected.
Reference