10 CFR 35.61, Calibration of Survey Instruments

From kanterella
(Redirected from 10 CFR 35.61)
Jump to navigation Jump to search

Calibration of survey instruments

a

(a) A licensee shall calibrate the survey instruments used to show compliance with this part and 10 CFR Part 20 before first use, annually, and following a repair that affects the calibration. A licensee shall--

(1) Calibrate all scales with readings up to 10 mSv (1000 mrem) per hour with a radiation source;
(2) Calibrate two separated readings on each scale or decade that will be used to show compliance; and
(3) Conspicuously note on the instrument the date of calibration.

b

(b) A licensee may not use survey instruments if the difference between the indicated exposure rate and the calculated exposure rate is more than 20 percent.

c

(c) A licensee shall retain a record of each survey instrument calibration in accordance with § 35.2061.


External Links

NRC website 10 CFR 35.61