10 CFR 35.457, Therapy-Related Computer Systems
From kanterella
Jump to navigation
Jump to search
Therapy-related computer systems
The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
Contents
a
(a) The source-specific input parameters required by the dose calculation algorithm;
b
(b) The accuracy of dose, dwell time, and treatment time calculations at representative points;
c
(c) The accuracy of isodose plots and graphic displays; and
d
(d) The accuracy of the software used to determine sealed source positions from radiographic images.