10 CFR 35.457, Therapy-Related Computer Systems

From kanterella
Jump to navigation Jump to search

Therapy-related computer systems The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:

a

(a) The source-specific input parameters required by the dose calculation algorithm;

b

(b) The accuracy of dose, dwell time, and treatment time calculations at representative points;

c

(c) The accuracy of isodose plots and graphic displays; and

d

(d) The accuracy of the software used to determine sealed source positions from radiographic images.


External Links

NRC website 10 CFR 35.457