Materials Inspection Report U.S. NUCLEAR REGULATORY COMMISSION Page 1 of 1
Add Continuation Page Licensee/Location Inspected:
Premier Medical Clinics
1165 S. Linden Road Flint, MI 48532 Report Number(s) 2025001 NRC/Regional Office Region III U. S. Nuclear Regulatory Commission 2056 Westings Avenue, Suite 400 Naperville, IL 60563-2657 Docket Number(s)
030-33371 License Number(s)
21-26540-01 Date(s) of Inspection 4/24/2025 LICENSEE:
The inspection was an examination of the activities conducted under your license as they relate to radiation safety and to compliance with the Nuclear Regulatory Commission (NRC) rules and regulations and the conditions of your license. The inspection consisted of selective examinations of procedures and representative records, interviews with personnel, and observations by the inspector. The inspection findings are as follows: Based on the inspection findings, no violations were identifie. Previous violation(s) closed. During this inspection, certain of your activities, as described below and/or attached, were in violation of NRC requirements, and were assessed at Severity Level IV, in accordance with the NRC Enforcement Policy. The following violation(s) is/are being cited in accordance with NRC Enforcement Policy. This form is a NOTICE OF VIOLATION, which may be subject to posting in accordance with 10 CFR 19.1 (Violations and Corrective Actions)
Statement of Corrective Actions I hereby state that, within 30 days, the actions described by me to the Inspector will be taken to correct the violations identifie This statement of corrective actions is made in accordance with the requirements of 10 CFR 2.201 (corrective steps already taken, corrective steps which will be taken, date when full compliance will be achieved). I understand that no further written response to NRC will be required, unless specifically requested. The violation(s), specifically described to you by the inspector as non-cited violations, are not being cited because they were self-identified, non-repetitive, corrective action was or is being taken, and the remaining criteria in the NRC Enforcement Policy were satisfie (Non-cited violation(s) was/were discussed involving the following requirement(s)
TITLE NRC INSPECTOR PRINTED NAME SIGNATURE AND DATE LICENSEE'S REPRESENTATIVE Zahid Sulaiman, Health Physicist BRANCH CHIEF Rhex A. Edwards, Chief, MIB ZAHID SULAIMAN Digitally signed by ZAHID SULAIMAN Date: 2025.05.09 13:14:32 -05'00'
Digitally signed by RHEX EDWARDS Date: 2025.05.16 07:26:37 -05'00'
NRC FORM 592M (10-04-2022)
U.S. NUCLEAR REGULATORY COMMISSION Materials Inspection Record Page 1 of 2 NRC Form 592M (10-04-2022) Licensee Name:
Premier Medical Clinics Docket Number(s):
030-33371 License Number(s)
21-26540-01 Report Number(s):
2025001 Date(s) of Inspection:
4/24/2025 Inspector(s):
Zahid Sulaiman, Health Physicist Program Code(s):
02201 Priority:
5 Inspection Guidance Used:
87130 1 Licensee Contact Name(s):
Mary Teffner, CNMT 1 Licensee E-mail Address:
mteffner@premiermedclinics.com 1 Licensee Telephone Number(s):
(810) 732-5400, Ext 247 1 Inspection Type:
Initial Routine
Announced Non-Routine Unannounced
1 Locations Inspected:
Hybrid Main Office
Field Office Temporary Job Site Remote 1 Next Inspection Date (MM/DD/YYYY):
4/24/2030 Normal
Reduced Extended No change 1 Location(s) Inspected List:
1165 S. Linden Road, Flint, MI 48532 1 Scope and Observations:
This was a routine, unannounced inspection of a medical clinic located in Flint Michigan. The licensee was authorized to use byproduct material for diagnostic uptake, imaging, and localization studies. The nuclear medicine department was staffed with a full-time nuclear medicine technologist (NMT) who performed approximately 4-5 cardiac stress tests on Mondays, Tuesdays, and Thursdays. The NMT works from 7 am - 5 pm, Monday -Thursda The licensee retained the services of a medical physics consultant to conduct audits of the program on a quarterly basi PERFORMANCE OBSERVATIONS This inspection consisted of a tour of the nuclear medicine department; interviews with select licensee personnel; a review of select records; an observation of security of the materials; and independent measurements. The inspector observed an administration of a technetium-99m (Tc-99m) unit dose stress test. The inspector verified the location of sealed sources from the licensee's most recent inventory, and all sources were accounted for. The inspector had the NMT demonstrate the dose calibrator constancy check, package receipt procedures, the end of the day daily and weekly area surveys, proper handling of radioactive waste, and disposal procedures. Through these demonstrations and discussions, the inspector found that the licensee personnel were knowledgeable of radiation protection principles, licensee procedures, and regulatory requirement The inspector reviewed the following records: quarterly physicist audits, package receipts, waste disposal records, DOT Hazmat training, linearity and accuracy of the dose calibrator, instrument calibration, spill report, sealed source leak tests and inventory, daily area surveys, weekly wipe tests, and dosimetry records. The inspectors performed independent and confirmatory radiation measurements which indicated results consistent with licensee survey records and posting No violations of NRC requirements were identified as a result of this inspectio NRC FORM 592M (10-04-2022)
U.S. NUCLEAR REGULATORY COMMISSION Materials Inspection Record (Continued)
Page 2 of 2 NRC Form 592M (10-04-2022)
Signature and Date - Branch Chief Digitally signed by RHEX EDWARDS Date: 2025.05.16 07:26:20 -05'00'