The following report was received via facsimile from the
PA Department of Radiation Protection.
Notifications: The department's western regional office received a phone call on January 17, 2012 and then referred the matter to the central office via email and telephone on January 18, 2012. This event is reportable within 24 hours2.777778e-4 days <br />0.00667 hours <br />3.968254e-5 weeks <br />9.132e-6 months <br /> under 10 CFR 30.50(b)(2)(i).
Event Description: A collimator failure was discovered during a monthly routine maintenance check on January 17, 2012. The collimator had broken off of the shutter mechanism.
The device is identified as - Manufacturer: Thermo Measure Tech; Model 5204; Isotope: Cs-137; Activity: 4 Ci.
CAUSE OF THE EVENT: Equipment Failure
ACTIONS: Upon discovery, a service provider was contacted and repairs were made by replacing the collimator with a spare one. Surveys were performed to verify appropriate conditions. No employees were exposed to excess levels of radiation as a result of this failure. A 30 day licensee report is expected. The department plans to conduct a reactive inspection.
Event Report ID No.: PA120002