ENS 54461: Difference between revisions
Jump to navigation
Jump to search
StriderTol (talk | contribs) (StriderTol Bot change) |
StriderTol (talk | contribs) (StriderTol Bot change) |
||
Line 10: | Line 10: | ||
| utype = | | utype = | ||
| cfr = 10 CFR 30.50(b)(1) | | cfr = 10 CFR 30.50(b)(1) | ||
| emergency class = | | emergency class = Non Emergency | ||
| notification date = 01/02/2020 13:47 | | notification date = 01/02/2020 13:47 | ||
| notification by = Kay Kassel | | notification by = Kay Kassel | ||
Line 18: | Line 18: | ||
| title = Unplanned Contamination on Pet Ct Scanner Mobile Unit | | title = Unplanned Contamination on Pet Ct Scanner Mobile Unit | ||
| event text = The following is a summary of information received from Alliance Health Services (Alliance) via the phone: | | event text = The following is a summary of information received from Alliance Health Services (Alliance) via the phone: | ||
At approximately 0800 CST on January 2, 2020, a mobile PET CT scanner unit was received by Alliance from CardioNavix at the Henry Ford Medical Center in West Bloomfield, MI with rubidium-82 contamination on the top exterior surface of the generator cart. The initial wipes on the surface of the generator cart were 17,697 dpm and 112,368 dpm (2 different areas of the top of the cart). No other contamination was found. The contaminated generator cart is located inside the mobile unit with limited access. The unit has not been used since the discovery of contamination. No personnel were contaminated. The licensee notified CardioNavix and the unit will be picked up by CardioNavix later today. | |||
At approximately 0800 CST on January 2, 2020, a mobile PET CT scanner unit was received by Alliance from CardioNavix at the Henry Ford Medical Center in West Bloomfield, MI with rubidium-82 contamination on the top exterior surface of the generator cart. The initial wipes on the surface of the generator cart were 17,697 dpm and 112,368 dpm (2 different areas of the top of the cart). No other contamination was found. The contaminated generator cart is located inside the mobile unit with limited access. The unit has not been used since the discovery of contamination. No personnel were contaminated. The licensee notified CardioNavix and the unit will be picked up by CardioNavix later today. | |||
| URL = https://www.nrc.gov/reading-rm/doc-collections/event-status/event/2020/20200113en.html#en54461 | | URL = https://www.nrc.gov/reading-rm/doc-collections/event-status/event/2020/20200113en.html#en54461 | ||
}} | }} | ||
{{ENS-Nav}} | {{ENS-Nav}} |
Latest revision as of 12:00, 15 January 2021
Where | |
---|---|
Alliance Health Services Beckley, West Virginia (NRC Region 1) | |
License number: | 47-25570 |
Organization: | Alliance Health Services |
Reporting | |
10 CFR 30.50(b)(1) | |
Time - Person (Reporting Time:+1.78 h0.0742 days <br />0.0106 weeks <br />0.00244 months <br />) | |
Opened: | Kay Kassel 13:47 Jan 2, 2020 |
NRC Officer: | Caty Nolan |
Last Updated: | Jan 2, 2020 |
54461 - NRC Website | |