PNO-III-97-011, on 970225,radiopharmaceutical Package Exceeds Contamination Limits Occurred.Licensee Notified Medi-Physics Who Performed Surveys of Delivery Vehicle & Driver.Info in Preliminary Notification Reviewed W/Licensee Mgt

From kanterella
Jump to navigation Jump to search
PNO-III-97-011:on 970225,radiopharmaceutical Package Exceeds Contamination Limits Occurred.Licensee Notified Medi-Physics Who Performed Surveys of Delivery Vehicle & Driver.Info in Preliminary Notification Reviewed W/Licensee Mgt
ML20138F943
Person / Time
Site: 03018558
Issue date: 02/26/1997
From: Hayes B, Phillips M
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
References
PNO-III-97-011, PNO-III-97-11, NUDOCS 9705060106
Download: ML20138F943 (2)


- - - - - - . . - - . - _ - . - - - - - - - . - . . _ . . --

.s. + ..

t .

RECElVED 1997 FEB r 2gbgag,1997 PRELIMINARY NOTIFICATI0fWLID/DBEUMEWl#6% OCCURRENCE PNO-III-97-011 This preliminary notification constitutes EARLY notice of events of POSSIBLE safety or public interest significance. The information is as initially received without verification or evaluation, and is basically all that is known by Region III staff (Lisle, Illinois) on this date.

Facility Licensee Emergency Classification 1 MICHAEL LALA, M.D., P.C. Notification of Unusual Event Michael Lala, M.D.. P.C. Alert Allen Park Michigan Site Area Emergency License No: 21-24380-01 General Emergency X Not Applicable

Subject:

RADI0 PHARMACEUTICAL PACKAGE EXCEEDS CONTAMINATION LIMITS On February 25, 1997, the licensee's consultant notified NRC Region III '

(Chicago) that a package (ammo box) with radioactive contamination exceeding Department of Transportation limits was received earlier that day at the licensee's Livonia Internal Medicine Facility Livonia.

Michigan. The package was shipped by the Medi-Physics nuclear pharmacy.

Upon receipt of the package, the licensee's technician performed an initial wipe test on the outside of the package which indicated 27.000 disintegrations per minute (dpm). (The DOT limit is 2200 dpm/100 sq cm).  ;

Additional wipe tests indicated 6.270 dpm on the right side of the package, 4,681 dpm on top. 3,059 on the left side, 1,573 dpm on back. l 1.249 dpm on bottom, and 3.546 on front of the package. A survey of the I package indicated exposure rates between 1 to 1.5 millirem per hour. i The package contained 3 unit doses, each with 25 millicuries (925 MBq)  ;

of technetium-99m. Internal wipes of the package and contents were negative. The ammo box was placed in storage.  ;

i The licensee notified Medi-Physics who performed surveys of the delivery vehicle and driver. Survey results were negative. Two NRC inspectors were performing an inspection of the licensee during the time the .

package was delivered. The inspectors will review the event during an inspection at Medi-Physics on February 26, 1997.

~

9705060106 970226 fNO-I 97-011 PDR; s lE3Y l ch

L 4

4 i

~

The State of Michigan and NRC Office of Nuclear Material Safety and Safeguards have been notified. The information in this preliminary >

notification has been reviewed with licensee management.

NRC Region III was notified of the event at 11:15 a.m. (CST) on February

, 25, 1997. This information is current as of 8:00 a.m. (CST) on February 26, 1997.

-i

Contact:

MONTE PHILLIPS BOB HAYES (630)829-9806 (630)829-9819 E

?