ML24068A049

From kanterella
Jump to navigation Jump to search
ACMUI Fall 2023 Medical Events Subcommittee Report Final February 9 2024
ML24068A049
Person / Time
Issue date: 02/09/2024
From:
Advisory Committee on the Medical Uses of Isotopes
To:
References
Download: ML24068A049 (2)


Text

1 U.S. Nuclear Regulatory Commission Advisory Committee on the Medical Uses of Isotopes Subcommittee on Medical Events Final Report Submitted on February 9, 2024 Subcommittee Members:

Michael Folkert, MD, PhD (Brachytherapy Radiation Oncologist)

Richard Green, BS (Nuclear Pharmacist)

Richard Harvey, DrPH (Radiation Safety Officer; Chair)

Darlene Metter, MD (Diagnostic Radiologist)

Zoubir Ouhib, MS (Therapy Medical Physicist)

Harvey Wolkov, MD (Radiation Oncologist)

Consultant: John Angle, MD (Interventional Radiologist)

NRC Staff Resource: Daniel DiMarco, MS Subcommittee Charge: Review medical events (MEs) to advise the Advisory Committee on the Medical Uses of Isotopes (ACMUI) and U.S. Nuclear Regulatory Commission (NRC) about emerging trends that may need regulatory attention.

Background:

The subcommittee reviewed medical events from Fiscal Years 2021 and 2022 as part of its ongoing biannual review.

Findings: Medical events in Fiscal Years 2021 and 2022 were relatively low but there appears to be an increase in medical events related to 10 CFR 35.300 and 10 CFR 35.1000 uses. The volume of procedures involving 177Lu radiopharmaceuticals has increased during this period and this has resulted in an increased number of medical events in the category of 10 CFR 35.300.

Medical events related to 10 CFR 35.1000 uses have increased for treatments involving 90Y microspheres (for both TheraSpheres and SIR-Spheres). Reducing the number of medical events involving 90Y microspheres may be accomplished through improved education with emphasis on appropriate administration of the microspheres and proper set-up of the delivery device. In addition, a time-out may be a tool that can be implemented to reduce the number of medical events for procedures using 90Y microspheres.

The committee recognizes past effort and continues to discuss medical events that may benefit from a time-out or those that may be the result of infrequent/inexperience use. Improved education and use of a time-out may be beneficial in preventing medical events.

The subcommittee will continue to examine medical events involving radiopharmaceuticals approved for use under 10 CFR 35.300 as well as treatments performed with 90Y microsphere procedures (10 CFR 35.1000). Furthermore, the subcommittee will monitor trends that may continue and appear to be significant during the biannual review periods.

2 Concluding Remarks: The ACMUI Subcommittee on Medical Events appreciates the opportunity to continue reviewing these events. Emphasis will be placed on radiopharmaceuticals in 10 CFR 35.300 and 10 CFR 35.1000, specifically 90Y microsphere procedures, to discern if there are emerging trends in medical events involving these treatments. The subcommittee welcomes any comments and/or recommendations.

Respectfully Submitted on February 9, 2024 Subcommittee on Medical Events Advisory Committee on the Medical Uses of Isotopes (ACMUI)

U.S. Nuclear Regulatory Commission (NRC)

The ACMUI unanimously approved this report as presented during its public meeting on October 23, 2023.