REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF SEALED SOURCE HQ21 NR-460-S-887-S DATE: January 26, 1996 PAGE 1 OF 6 SEALED SOURCE TYPE:

Sealed Source MODEL:

4D6D MANUFACTURER / DISTRIBUTOR:

3M Health Physics Services 3M Center, Building 224-2E-06 St. Paul, MN 55144-1000 ISOTOPE:

MAXIMUM ACTIVITY:

Cesium-137 10.00 curies (370.0 GBq)

LEAK TEST FREOUENCY:

6 Months l

PRINCIPAL USE:

(D) Gamma Gauges l

CUSTOM SOURCE:

YES X

NO

~

9905290329 960126 PDR RC SSD PDR e L M%LW6M2*D

l I

l

~

REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF SEALED SOURCE l

l NO.: NR-460-S-887-S DATE: January 26, 1996 PAGE 2 OF 6 SEALED SOURCE TYPE:

Sealed Source j

l DESCRIPTION:

The Model 4D6D is a doubly-encapsulated point source.

Cesium-137 is absorbed into small ceramic particles ranging from 0.001-0.008 in. (30-200 microns) in diameter.

The isotope is permanently affixed to the particles with a heat treatment to create 3M Brand Radiating Microspheres.

The microspheres are loaded into a 0.180 in. (0.457 cm) long inner capsule with a diameter of 0.250 in. (0.635 cm).

The cavity of this inner capsule is 0.140 in. (0.356 cm) deep and is nominally 0.171 in.

(0.434 cm) in diameter.

There is an opening on one end of this capsule.

The inner capsule and inner plug are made from type 304 stainless steel.

After loading the microspheres, a nominally 0.171 in. (0.434 cm) l diameter, 0.040 in. (0.102 cm) thick inner welding plug is inserted into the inner capsule and silver brazed irto place, sealing the radioactive material inside the inner capsule.

The entire assembly is then placed, plug end facing in, inside an outer capsule.

The outer capsule is 0.437 in. (1.110 cm) long and 0.375 in. (0.953 cm) in diameter.

There are two hollow cavities in the outer capsule.

One cavity is made with a #21 bit and a #10 tap and extends 0.159 in. (0.404 cm) into the capsule.

The other cavity, where the inner capsule assembly is inserted, l

is 0.230 in. (0.584 cm) deep and nominally 0.255 in. (0.648 cm) i in diameter.

An outer welding plug is inserted after the inner capsule and is heliarc welded into place, sealing in the inner capsule.

The outer capsule and outer plug are made from type 304 l

stainless steel.

l l

The completely assembled 4D6D source is 0.437 in. (1.110 cm) long and 0.375 in. (0.953 cm) in diameter.

DIAGRAM:

L See Attachment 1 F'

l l

l

REGISTRY OF RADIOACTIVE SdALED SOURCES AND DEVICES SAFEfY EVALUATION OF SEALED SOURCE NO.: NR-460-S-887-S DATE: January 26, 1996 PAGE 3 OF 6 SEALED SOURCE TYPE:

Sealed Source LABELING:

The following information was engraved on the exterior of the outer capsule:

3M Company X mc Cs-137 Serial XXX Model 4D6D Date

. CONDITIONS OF NORMAL USE:

The 4D6D sealed source was intended for use in applications where a high-integrity gamma source was needed, particularly where extreme environmental conditions were involved.

A cypical example of this type of environment is industrial gauging.

PROTOTYPE TESTING:

A prototype of the Model -iD6D source was tested according to the specifications of ANSI N542-1977 and achieved a classification of 77C66546.

A prototype model of this source loaded with 100.0 millicuries (3.700 GBq) of cesium-137 was subjected to the following tests:

1. Heat Test - Heat the source at 932*F (500*C) for 30 minutes.
2. Quench Test - Quench the source from 932*F (500*C) to cold water temperature.
3. Molten Lead Test - Immerse the source in molten lead at 932*F (500*C) for 30 minutes.
4. Impact Test - Drop a weight generating 2.0 ft-lb (2.7 N) on the sides of the source at two points 180* apart, performing the test 20 times at each point.

I' REGISTPY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF SEALED SOURCE l

NO,: NR-460-S-887-S DATE: January 26, 1996 PAGE 4 OF 6 SEALED SOURCE TYPE:

Sealed Source l

PROTOTYPE TESTING: (Continued)

After each test the source was wipe tested and leak tested.

The wipe tests were counted in a detection instrument capable of detecting less than 0.0001 yCi (3.700 Bq) of cesium-137.

All wipes showed less than 0.0001 4Ci (3.700 Bq) of removable cesium-137.

The leak tests were performed by immersing the source for a minimum of 10 seconds in water at 200'F (93.3'C) and observi.1g any bubbles generated.

All leak tests were negative.

EXTERNAL RADIATION LEVELS:

External radiation dose rates were calculated for a 10.00 Ci (370.0 GBq) Cs-137 source as follows:

Distance Maximum Radiation Level (from surface)

(cm)

(in.)

(R/hr)

(Sv/hr) 5 1.97 1320 13.2 30 11.81 36.6 0.37 100 39.37 3.3 0.033 OUALITY ASSURANCE AND CONTROL:

The following quality control procedures were followed during production of this source:

1. After the inner plug was silver brazed in place, the inner capsule was cleaned until wipe tests showed less than 0.0005 Ci.(18.50 Bq) of removable cesium-137.

i k

i

)

J

l l

REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF SEALED SOURCE i

NO.: NR-460-S-887-S DATE: January 26, 1996 PAGE 5 OF 6_

l l

SEALED SOURCE TYPE:

Sealed Source OUALITY ASSURANCE AND CONTROL: (Continued) i

2. After the outer plug was heliarc welded into place, the source was wipe tested.

If the wipe yielded more than 0.0001 Ci (3.700 Bq) of removable cesium-137 the source was cleaned until the 0.0001 Ci (3.700 Bq) wipe test limit was l

reached.

l

3. The source was stored for 7 days and wipe tested again.

If I

the wipe test yielded more than 0.0001 Ci (3.700 Bq) of removable cesium-137, the source was rejected.

'4.

Before shipment, the source was again wipe tested.

This wipe also had to show less than 0.0001 Ci (3.700 Bq) of i

removable cesium-137 or the source was rejected.

I LIMITATIONS AND/OR OTHER CONSIDERATIONS OF USE:

1 This source may be used only by persons specifically licensed by the NRC or an Agreement State.

j Handling, storage, use, transfer, and disposal: to be determined by the licensing authority.

This source shall be leak tested at intervals not to exceed 6 n.onths using techniques capable of detecting 0.005 Ci l

(185.0 Bq) of removable contamination.

i This source shall not be subjected to environmental or other conditions of use which exceed ANSI Classification 77C66546 (ANSI N542-1977).

This registration sheet and the information contained within l

the references shall not be changed without the written consent of the NRC.

1 l

i i

l:

1 i

i

L L

L REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES j

SAFETY EVALUATION OF SEALED SOURCE NO.: NR-460-S-887-S DATE: January 26, 1996 PAGE 6 OF 6 l

l SEALED SOURCE TYPE:

Sealed Source H

1 l

SAFETY ANALYSIS

SUMMARY

The Model 4D6D source is not a current product manufactured or distributed by 3M Health Physics Services.

However, 3M Health i

Physics Services will continue to receive Model 4D6D sources for i

disposal.

1 Based on our review of the Model 4D6D source, and the information and test data cited below, we continue to conclude that this sealed source is acceptable for specific licensing purposes.

Furthermore, we continue to conclude that the Model 4D6D source would be expected to maintain its containment integrity for normal conditions of use and accidental conditions which might occur during uses specified in this certificate.

REFERENCES:

The following supporting documents for the Model 4D6D source are hereby incorporated by reference and are made a part of this registry document:

3M Health Physics Services' letters dated August 6,

1991, and January 13, 1965, with enclosures thereto ANSI Tests - Reclassification of Radioactive Sources - Test e

Report dated August 31, 1982 I

ISSUING AGENCY:

U.S.

Nuclear Regulatory Commission Date:

January 26, 1996 Reviewer:

a

/bou las A. Broaddus Date:

January 26, 1996 Concurrence:

b

  • > 4

"' Steven L7*Baggett 1

I

^

i i

l l

l REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF SEALED SOURCE NO.: NR-460-S-887-S DATE: January 26, 1996 ATTACHMENT 1 l

WAW/l t,/

1

,7 INNE.R C APSULE.

N

/g/

OUTER CAPSULE h

s M h' (

- L A B E.L DATA 3 M BRAND R ADI ATING M IC RO S P H E.RES 3 M MODE.L No,

'4DCD (C s - 13 7) g I

g I

e a

~

t f*"%9 0-k UNITED STATES

{

j NUCLEAR REGULATORY COMMISSION t

WASHINGTON, D.C. 20555 4001 r*t% * * * *

  • p$

February 12, 1996 NOTE T0:

MINNESOTA MINING AND MANUFACTURING (3M) INACTIVE FILES hf FROM:

SUBJECT:

INACTIVATION OF REGISTRATIONS BASED ON AUGUST 6, 1991, LETTER.

A number of the background files for the registration certificates requested (by 3M letter dated August 6, 1991) to be made inactive contain limited information on the products. A search of 3M registration, archive, and license files, and information in NUDOCS was performed to obtain additional information. Also, 3M was contacted to obtain any additional information that may have been in their files.

3M provided limited additional information concerning these products and indicated that the information provided represented all additional information in their files.

Additional information obtained from these sources has been added to the applicable registration files.

At the present time, the information contained in the background registration files represents all available information for these products.

In addition, a number of these products were originally reviewed and approved by the Atomic Energy Commission (AEC).

The reviews performed for these products to make the certificates inactive were based on the information available, the previous reviews and approvals, and the historical use of the products (i.e., reports of failures or design problems, or the lack thereof).

In several instances, insufficient information was available to perform an adequate safety review.

In these instances, the certificates have been reissued based solely on the original AEC review and approval, and no additional safety review was performed.

/

l i

I

..b

February 29, 1996 Robert G. Wissink 3M Health Physics Services 3M Center Bldg. 224-2E-06 St. Paul, MN 55144-1000

Dear Mr. Wissink:

This letter is in response to your request dated August 6,1991, and subsequent letters, requesting to transfer to inactive status all registration i

sheets issued to 3M as listed in the first column of Enclosure 1.

We have 1

transferred the requested certificates to inactive status. The certificate numbers have been changed as listed in Enclosure 1.

Copies of the inactive certificates are enclosed for your records.

Please read over the registration certificates ailo notify us immediately if there are any errors or inconsistencies.

If you have any questions, please contact me at (301) 415-5847 or Mr. Steven Baggett at (301) 415-7273.

Sincerely,

/S/

{

- Douglas A. Broaddus, Mechanical Engineer Sealed Source Safety Section Source Containment and

)

Devices Branch i

Division of Industrial 'and Medical Nuclear Safety, NMSS

Enclosures:

As stated i

I l

)

Distribution:

SSSS R/F NE01 SSD-91-71 Registration certificate files listed in Enclosure 1 lDOCUMENTNAME:

n h,.o n.em.4

\\W.n, DOCS \\SSDS\\ INACTIVE \\NRC\\460lNACT. CMP C:

P m m. wn:

l i.

c. con we

.u.chen.nu.nei w.

r - con we.a.chen.nu.nei r - N py 0FFICE SCDB lD l

l l

l NAME DABroaddus DATE 02/27/96 OFFICIAL RECORD COPY

~~

axxfo@8 wavnme NsWnn

ENCLOSURE I REGISTRATIONS TRANSFERRED TO INACTIVE STATUS Old Certificate No.

New Certificate No.

Model Number (s)

NR-459-S-101-S NR-8008-S-801-S 4F6Y NR-459-S-102-S NR-8008-S-802-S 3 FIG NR-450-S-101-S NR-0460-S-875-S 4F60 NR-460-S-102-S COMBINED WITH 4F6H, 4F6B NR-0460-S-874-S NR-460-S-103-S NR-0460-S-876-S 4F6G NR-460-S-104-S COMBINED WITH.

6D6A NR-0460-S-865-S NR-460-S-105-S NR-0460-S-877-S 6B6F NR-460-S-106-S NR-0460-S-878-S 4F6P NR-460-S-107-S NR-0460-S-879-S IC2A, IC2B NR-460-S-108-S NR-0460-S-880-S 4F3B NR-460-S-109-S NR-0460-S-881-S 4F3C NR-460-S-110-S NR-0460-S-882-S 3MIC

{

NR-460-S-Ill-S NR-0460-S-883-S 3MlB NR-460-S-ll2-S NR-0460-S-884-S 3E4G NR-460-S-ll3-S NR-0460-S-885-S 403A NR-460-S-Il4-S NR-0460-S-886-S 4D3B NR-460-S-115-S NR-0460-S-887-S 4D60 NR-460-S-116-S NR-0460-S-888-S 4D6F NR-460-S-117-S NR-0460-S-889-S 5FID NR-460-S-Il8-S COMBINED INTO 5FIE NR-460-S-Il9-S

-~

~

5FIF NR-460-S-120-S NR-0460-S-891-S EFIG l

NR-460-S-121-S NR-0460-S-892-S SL2B l

NR-460-S-122-S NR-0460-S-893-S 3L2A NR-460-S-123-S NR-0460-S-894-S 3MIF l

NR-460-S-124-S NR-0460-S-895-S 3L2C

l l

l ENCLOSURE 1 l

REGISTRATIONS TRANSFERRED TO INACTIVE STATUS Old Certificate No.

New certificate No.

Model Number (s) l NR-460-S-125-S NR-0460-S-896-S 1E2J NR-460-S-126-S NR-0460-S-897-S 3 FIG NR-460-S-127-S NONE - NO INFORMATION 4FIE AVAILABLE NR-460-S-128-S NR-0460-S-899-S 5F1H l

NR-460-S-129-S NR-0460-S-900-S 3E40 NR-460-S-130-S COMBINED INTO 5 FIN NR-460-S-131-S NR-0460-S-901-S 5 FIN (MODIFIED)

NR-460-S-132-S NR-0460-S-902-S 788L NR-460-S-133-S COMBINED INTO 6H6A NR-460-S-134-S

~

~

6H6B NR-460-S-135-S NR-0460-S-904-S 4D6M NR-460-S-136-S NR-0460-S-905-S 3L2E NR-460-S-137-S NR-0460-S-906-S 6500, 6520 (FORM. 6D6C)

NR-460-S-138-S NR-0460-S-907-S 3L2D 1

NR-460-S-139-S NR-0460-S-908-S 3Q10 j

NR-460-S-140-S NR-0460-S-909-S 3E4L, 3E4S NR-460-S-141-S NR-0460-S-910-S 3 FIR NR-460-S-142-S NR-0460-S-911-S 902, 902F, 903 NR-460-S-143-S COMBINED WITH 4P6E NR-0460-S-871-S NR-460-S-144-S COMBINED WITH 4P6M NR-0460-S-872-S NR-460-S-145-S NR-0460-S-912-S 3Mll NR-460-S-146-S NR-0460-S-913-S 3G9A NR-460-S-147-S NR-0460-S-914-S 3B4G NR-460-S-148-S COMBINED WITH 4F3F NR-0460-S-871-S l

NR-460-S-149-S NR-0460-S-915-S 4F3G i

I ENCLOSURE I REGISTRATIONS TRANSFERRED TO INACTIVE STATUS l

Old Certificate No.

New Certificate No.

Model Number (s)

NR-460-S-150-S NR-0460-S-916-S 6DIA NR-460-S-151-S COMBINED WITH 6530, 6540 (FORM. 6B6G)

NR-0460-S-869-S NR-460-S-152-S NR-0460-S-917-S 6H6D NR-460-S-153-S NOT REQUIRED ALBUMIN MICR0 SPHERES NR-460-S-154-S NR-0460-S-919-S 6H6E, 8C9T NR-460-S-155-S COMBINED INTO 4D6L NR-460-S-156-S

~

~~

~

4D6P NR-460-S-158-S NR-0460-S-921-S 3F11, 3F1J, 3FIL NR-460-S-159-S NR-0460-S-922-S 3FIV NR-460-S-160-S COMBINED WITH 4P6T NR-0460-S-873-S NR-460-S-161-S COMBINED INTO d'

D NR-0460-S-920-S NR-460-S-162-S NR-0460-S-923-S 4F3H NR-460-S-163-S COMBINED INTO 4F6S NR-0460-S-920-S NR-460-S-164-S COMBINED WITH 3E40 NR-0460-S-900-S NR-460-S-165-S NR-0460-S-924-S 6701 NR-460-S-166-S NR-0460-S-925-S 6702 NR-460-S-167-S NR-0460-S-926-S 6711 NR-460-S-168-G NR-0460-S-927-G 702, 703, 704 NR-460-S-169-S NR-0460-S-928-S 6510, 6550, 6570 (FORMERLY 6B6G)

NR-460-S-170-S NR-0460-S-929-S 4P6V NR-460-S-171-S NR-0460-S-930-S 4F6ST