ML20236S260

From kanterella
Jump to navigation Jump to search
Correction to Registry of Radioactive Sealed Sources & Devices Safety Evaluation of Device (Amended in Entirety) for Calibration device.Certificate:NR-121-D-103-S. Typographical Error in Third Bullet on Page 5
ML20236S260
Person / Time
Issue date: 07/15/1998
From:
NRC
To:
Shared Package
ML20236S253 List:
References
SSD, NUDOCS 9807240230
Download: ML20236S260 (1)


Text

- ____- -__-_-_. _ _ _ - _ _ _ _ _ _ _ _, _ - - _ _ _ _

i REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF. DEVICE l-(AMENDED IN ITS-ENTIRETY)

NO.:.NR-121-D-103-S' DATE:

pyL 15 jM PAGE 5 OF 6 DEVICE TYPE:

Calibration Device LIMITATIONS AND/OR OTHER' CONSIDERATIONS OF USE (Cont ' d) -

  • The device shall be labeled in accordance with Section 20.1901,.10 CFR.Part 20, or equivalent regulations of the Agreement State.
  • Handling, storage, use, transfer, and disposal:

To be determined by the licensing authority.

  • These devices shall.be leak tested at intervals not to exceed 6 months using techniques capable of detecting 0.005 microcurie (185 Bq) of removable contamination.
  • REVIEWER NOTE: The license reviewer should request procedures from the licensee as necessary to comply with the survey.and posting requirements of 10 CFR Part 20.

l

  • This registration sheet and the information contained within the references shall not be changed without the written consent of the NRC.

SAFETY ANALYSIS

SUMMARY

The device is locked in a closed position and only emita a maximum 0.0 mrem /hr (8 pSv) at one (1) foot (30.4 8 cm) from the device.when in this position.

The device will only be operated in the open position by an individual authorized by NRC.

Based on. review of calibration device, Model D-0062, and the

}

information cited below, we continue to conclude that this device l

.is acceptable for specific licensing purposes.

9907240230 990721 i

PDR RC e

i SSD PDR 5

l

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -