ML20235P265

From kanterella
Jump to navigation Jump to search
Requests Addl Info for 870208 Application for License, Including Name of Mfg & Model Number of Gd-153 Source & Device & Procedures for Source Exchange.Info Requested within 30 Days
ML20235P265
Person / Time
Issue date: 03/02/1987
From: Graden K
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To: Passalaqua A
OSTEOPOROSIS CENTER OF CANTON STARK COUNTY WOMEN'S
Shared Package
ML20235P084 List:
References
82958, NUDOCS 8710070225
Download: ML20235P265 (2)


Text

_ _ -

g l

l' MAR 2 1967 Osteoporosis Center of Canton Stark County Women's Clinic ATTN: Anthony M. Passalaqua, M.D.

Treasurer 5000 Higbee, N.W.

Canton, OH 44718 Gentlemen:

We have reviewed your new license application dated February 8,1987 requesting a byproduct material license and find that we will need additional information as follows:

1.

Please submit the name of the manufacturer and the model number of both the gadolinium-153 source and the device in which it will be contained.

2.

With regard to source exchange of the gadolinium-153, please address the foliowing:

a.

Please submit the step-by-step procedures you will follow for source exchange.

You may either submit a copy of the manufacturer's procedures or reference them by date.

b.

Please submit the name of the individual (s) that will perform the source exchange procedures, c.

Please indicate that film badges will be worn during source exchange procedu res.

3.

Please submit the name and address of your film badge supplier.

4.

Please provide a sample calculation to demonstrate how leak test results will be converted into microcuries.

5.

Please indicate that you will maintain a copy of the manufacturer's instructions regarding the use of the device on file at your institution.

If you have any questions or require clarification on any of the information l

stated above, you may contact us at (312) 790-5625.

l i

8710070225 870420 i

REG 3 LIC30 34-24864-01 PDR

\\

i

_..__._.__.___________._.__m

Osteoporosis Center of Canton 2

We will continue our review of your application upon receipt of this information.

i l

Please reply in duplicate, within 30 days, and refer to Control Number 82958.

l l

Sincerely, 1

l Original Signed By Kathy J. Graden Materials Licensing Section l

l l

l RIII Gra n/pd J/;t /87