-l REGISTRY'OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY' EVALUATION OF DEVICE.
NO : NR-8095-D-809-S DATE:
OCT 0 71998 PAGE 1 OF 5 l
. DEVICE TYPE:
Instrument Calibrator l
l 4
'MODEL:
CDV-790 Model 1 l
l l
MANUFACTURER / DISTRIBUTOR:
Federal' Emergency Management Agency 1
(FEMA) 500 C Street S.W.
Washington, DC 20472 SEALED SOURCE DESIGNATION:
Nuclear Chicago Corp.
Model No.
OCD-S-104 ISOTOPE:
MAXIMUM ACTIVITY:
Cesium-137 16 millicuries (592 MBq)
LEAK TEST FREOUENCY:
6 months i
PRINCIPAL USE: (Y) Calibrator for Geiger-Mueller Instruments i
l CUSTOM DEVICE:
YES X
NO b
/
j 9810080296 981007 t
PDR RC s
SSD PDR l
l i
~
REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF DEVICE NO.: NR-8095-D-809-S DATE:
OCT 0 71998 PAGE 2 OF 5 DEVICE TYPE:
Instrument Calibrator DESCRIPTION:
The Model CDV-790 calibrator consists of a CDV-792 shipping container into which a collimated Nuclear Chicago 16 mci (592 MBq) Cs-137 source is placed.
During transportation the lid of the CDV-792 is locked to prevent unauthorized access.
In the calibration laboratory the lid is removed from the shipping container and the calibration jig (as shown in the attachments) is-then mounted onto the container and a lead storage plug is used to prevent unnecessary exposure to personnel while the unit is not in use.
When in use, the probe is placed on two V-blocks at a predetermined height.
The lead storage plug is unlocked and removed.
The radiation beam can be attenuated by sliding one or any combination of three tungsten attenuators in the beam path to approximate gamma exposure rates of 45, 30, 22.5, 15, 4.5, 3,
2.25, and 1.5 mrem /hr (450, 300, 225, 150, 45, 30, 22.5, and 15 pSv/hr) i 5%.
j DIAGRAM:
See attachments 1 and 2.
LABELING:
When not in their shipping containers, the source has an eyelet to which a cable and tag are attached.
The labeling on the tag meets the requirements of 10 CFR 20.1901 and 10 CFR 20.1904.
The shipping container has three adhesive backed labels affixed to it.
One label contains the trefoil symbol, the words
" Caution Radioactive Material," the words "DCPA CDV-790 Model 1 GM Instrument Calibrator," the isotope, physical form, activity and the assay date.
The second label contains the words " USA DOT 7A Type A,"
" Radioactive Material," " Federal Emergency Management Agency, Washington D.C."
and the gross weight of the container 210 lbs (95 kg).
The third label.is affixed to the lid and meets the requirements of 10 CFR 20.1901 and 10 CFR 20.1904.
l i
I I
l
REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF DEVICE NO.: NR-8095-D-809-S DATE: OCT 0 71998 PAGE 3 OF 5 DEVICE TYPE:
Instrument Calibrator CONDITIONS OF NORMAL USE:
The calibrators are used under 100% federally funded contracts with the States for inspection, maintenance, and calibration of radiation detection equipment.
The calibrators will be used by only trained technicians.
The expected useful life of the source in the calibrator is 48 years.
PROTOTYPE TESTING:
Nuclear Chicago manufactured and tested the sources to the requirements of their contractual agreement.
The following tests were performed on the source:
Leakage and surface contamination Atmospheric corrosion test
- Sea water immersion test Temperature resistant test to 1000 C (1832*F) for 30 minutes
- Impact resistance test, dropped 1/2 lb. (0.23 kg) weight 20 times from a height of four feet (1.2 m).
- Abrasion resistance test Hydrochloric acid immersion test, placed in 0.1N hcl for a
24 hours2.777778e-4 days <br />0.00667 hours <br />3.968254e-5 weeks <br />9.132e-6 months <br /> Salt Spray test, placed in fog chamber for 20 hours2.314815e-4 days <br />0.00556 hours <br />3.306878e-5 weeks <br />7.61e-6 months <br />
- Physical inspections Drop test The shipping container Model CDV-792 was tested and meets the requirements of Department of Transportation's Specifications for a 7A shipping container.
EXTERNAL RADIATION LEVELS:
FEMA reported an unshielded dose rate of 110 mR/hr (1100 pSv/hr) i 10% at 9 inches (228.6 mm) frnm the source.
They also reported an exposure rate of < 0.1 mR/hr (1 pSv/hr) at 2 inches (50.8 mm) from the surface of the shipping container Model CDV-792.
I
. ~
_.._ - - -. _.-- _ _ _ ~._-...
.~.. -.
REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES
)
SAFETY EVALUATION OF DEVICE j
NO.: NR-8095-D-809-S DATE: 00'O71998 PAGE 4 OF 5
- DEVICE TYPE:
Instrument! Calibrator
_ QUALITY ASSURANCE AND CONTROL:
I FEMA contracted out all fabrication of the source and shipping containers.
The contractors wereisubjected to various onsite-
' inspections to ensure they were meeting their contractual agreement.
This QA procedure was previously deemed acceptable by the NRC.
- LIMITATIONS AND/OR OTHER CONSIDERATIONS OF USE:
- The device shall be used by persons specifically licensed by the NRC or an Agreement State.
- These devices will no longer.be commercially distributed, but may still be approved for licensing purposes.
- Handling, storage, use, transfer,-and disposal: To lx3 determined by the licensing authority.
- The device shall be leak tested at intervals not to exceed 1
I 6 months using. techniques capable of detecting 0.005 microcurie (185 Bq) or removable contamination.
- This registration sheet and the information contained within the references shall not be changed without the written consent of the NRC.
SAFETY ANALYSIS
SUMMARY
The CDV-790 Model 1 instrument calibrator consists.of a Model CDV-792 shipping container that was previously approved by the AEC for licensing purposes in'1959; and a Model OCD-S-104 sealed
)
source which was also approved for licensing purposes in 1967.
- The Model CDV-790 Model 1 calibrator was' constructed using the previously approved component and jig that allows an instrument probe to.be held in one position while'the radiation dose rate is
- varied by the use of sliding attenuators.
This modified component device.was'put.into use by FEMA around 1976.
1 4
~n w
n-
.. ~ _ _. - _ _ ~
REGISTRY OF' RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF DEVICE-OCT 0 71998 NO.: NR-8095-D-809-S DATE:
PAGE 5 OF 5
- DEVICE TYPE:
Instrument Calibrator SAFETY ANALYSIS
SUMMARY
- (Cont.)
Based on the fact that prototype (s) of this device have been used since 1976 without reports of any problems and that the major components of the CDV-790 Model 1 were previously approved, we continue to conclude that the device is acceptable for licensing purposes when used in accordance with the Operations and Maintenance manual provided with the device and the Limitations and/or other Considerations of Use stated above.
REFERENCES:
The following supporting documents for the CDV-790 Model 1 instrument calibrator are hereby incorporated by reference and are made a part of this registry document:
- Federal Emergency Management Agency letter's dated March 17, 1997, and February 17, 1983.
ISSUING AGENCY:
U.S. Nuclear Regulatory Commission Date:
Reviewer:
rIA[
ows Michele
.' Burgess OLT 0 71998
/
Date:
Concurrence:
//
k n W.
Lubinski r
i l
i 3
REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF DEVICE NO.- NR-8095-D-809-S DATE:
ATTACMENT 1 OCT 0 71998
<?. -
3 4
-eg 1.'h.
- .s 4.eW
+
'hbf'i-Sf 5
c.
x
.-.., -f( -.i
. +. -
t 1
I of j
~
j
." V e
1 t
]
M-..ia,.
J g
gg e
.. t:x.
g -sw
\\ \\, ':.
.J '
i f.
. 7'..
. ' ,'?.....
g'
' iy.'-
j
?
- cts
.: r
%$.h. -&
< a '
$hW&.:.',...$;0. ;..~ -
k g
. g g.
n V*
=,-
Q
- 1. 3...-
l
- QQ,,h
' v ~s '. ~
j y M' %
- y..
2 i
~
sdh 1
'T'
. gq;
.,+ ~. -..
i p.
a a,
3+
Calibrator with Attenuator Slides Opsn and Storage Plug Removed
t p '
l, f :
1-t
[
^'
REGISTRY-OF RADIOACTIVE SEALED SOURCES AND DEVICES l -.
SAFETY EVALUATION OF DEVICE I
HQ,_t - NR-8 0 9 5 -D-8 0 9 -S '
DATE:
OCT o 71993 --
ATTAcnMENT 2 l
W-700 GM lube.,ffebe L.
l Probe Positioning "V" O
f Block.
j s
l i.
I l
r-7
_.I I
l l-l { ll l
- )
I Positionths f'
-(
l Table-M g
F ;.I p-I
}--
l l
l i.
i I
I
~
. Threaded ' Rod.---, k I
I I
I 1
Tungsten l
l,,__M C_ _. f j' Attenuatcrs-I At tenu'a to r l
1 9
j -- - j rM
. and Slides i
III Guide Block
)
I.
---d'
(- --*
1 o
lStoragePi Top g
Mounting %
i L _ _n l
4 "i I.,
- y
}\\
f t
- Plate f
\\
,l
. /<
CDV-792
/
[Contaise:.
11 =at'or /
\\
\\
l
\\
gt;,,,,._
g N
{7,;_
mm n
e x e
,_e 1
^g;= v,f,h
/
-1/2" Salt 1/2" Solt -
. Bottom
/
d.
l l
[,
Mounting \\ -
Plate T.
/
i gn l
gl o L3 il
's l._!
~ CROSS SECTIONAL VIEW OF THE CDV-S CXLIBRATOR
-l l
i
ce i
n UNITED STATES Y
j NUCLEAR RE2ULATORY COMMISSION i
.M 4
WASHINGTON, D.C. NJo01
\\...../
I NOTE TO:
SSD97-23/28 Federal Emergency Management Agency (FEMA) j 4
FROM:
Eric B. Compton, Engineering Aide Materials Safety Branch
[p ^
Division of Industrial and Medical F
Nuclear Safety Office of Material Nuclear Safety and Safeguards DATE:
September 14,1998 i
RE:
REFERENCES NOT FOUND IN FILE In the process of transferring several of FEMA's registration certificates to inactive status it was discovered that a number of items which had been referenced in the original registration j
certificates were not in the background file. These included the following:
1.
Federal Civil Defense Administration Standard item Specification dated May 23,1956, for the Model CDV-786.
2.
Office of Civil and Defense Mobilization letter dated April 6,1959, stating that the CDV-784 was to replace the CDV-786.
3.
Manual of Procedure for the Use and Control of the CDV-778 Radiation training Source Set dated August 1975.
4.
Equipment Manual /Use and Control of the CDV-798 Model I Radiological Survey Meter Calibrator dated August 1975.
5.
Federal Emergency Management Agency letter dated February 17,1983, with enclosures thereto.
6.
Federal Emergency Management Agency renewal application for Materials License No. 08-01297-06, dated August 10,1992 with enclosures thereto.
i i v w b t/ A
~.
l April 2,1997 -
~'
(.,
, ASSIGNMENT NUMBER: 97-23 to 97-28 i
.Mr. Vemon Wingert L~
i Federal Emergency Management' Agency l
500 C Street SYI.
Washington, DC 20472 l
SUBJECT:
ACKNOWLEDGEMENT OF. REQUEST FOR SAFETY EVALUATION '
i
Dear Mr. Wingert:
L
'This I,etter. acknowledges the receipt of your March 17,1997, application that requested a safety evaluation and registration of various Models. We have performed a cursory review of f
your application and determined that enough information has been provided to allow a technical reviewer to initiate the evaluation process. Applications are assigned to technical L
' reviewers on a first-in basis. Therefore, your application will be assigned in tum. Please note
(
that the technical reviewer may contact you to request information that was omitted from your application or to obtain clarification of technicalissues conceming your application. If you J
- have any questions conceming the status of your application, please contact me at (301) 415-7857. Please reference the assignment number listed above in your questions or
/
/
correspondence.
./
Please be aware that your request may be subject to the NRC's application fees in '
accordance with 10 CFR Part 170. Therefore, a copy of your application has been forwarded to the License Fee and Debt Collection Branch for approval of the fee category and amount.
If you have any questions conceming the fees associated with pur application, please contact the License Fee and Debt Collection Branch at (301) 415-7554.
Sincerely, ISI KimBerly Randall, Registration Assistant l
Sealed Source Safety Section i
Medical,- Academic, and Commercial Use Safety Branch Division of Industrial and ~
Medical Nuclear Safety Office of Nuclear Material Safety I
and Safeguards Distribution:
SBaggett SSSS r/f SSD-97-23 to 97-28 NE01 SSD File #
NR-283-D-101-S NR-283-S-102-S NR-283-D-105-S NR-283-D-106-S NR-283-D-110-S
...,.J3 DOCUMENT NAME:P:97-23.wpd n
h,..
,. w. e uni.nt. woe.t. m in. n.: c - copy without ett.cnm.ntienciosur. r - copy wita.it.cnm.ntiencio.or.
m - wo copy l'
OFFICE IMAB l
l l
l l
g l-NAME KRandafi F DATE 4/.2J97
u_....-
PAGE1 NRC FORM 567 U. S. NUCLEAR REOULATORY COMMISSION 14L L
REQUEST FOR A SEALED SOURCE OR DEVICE EVALUATION INSTRUCTIONS: Send this request AND a copy of aN reisted " ^^ : _.,u,. and drawirige to: The Sealed Source Safety Secten, ATTN: CNef, OWFN Mail Stop 6 H3. Chan9e the License Tracking System musetone to 19 and seeign to reviewer code I-5.
NOTE: Retam a copy of tNs request with the appilcation and back0round flies.
REQUESTER g
REGION / LOCATION:
/T /2/M
]l O Il 111 IV UV C HQ LFDCB TELEPHONE NUMBER DATE TYPE OF ACTION REQUESTED (Check as -
,6)
APPUCANTS NAME SOURCE REVIEW AMENDMENT OF l.
[/ f f y' g
/dy;qg/
REGISTRATION SHEET i
i MAL CONTROL NUMBER (S)
DEVICE REVIEW BER(S)
AIC438
~ M~b LETTER /APPUCA6TONATE/97 UCENSE NUMBER (S)
CUSTOM REVIEW kesg sse-ey w~m-+ ewy y
a SA,eeef Sub NASknel-utD C-c%V2v FOR 8888 USE ONLY REVIEWER NUMBERS -
NUMBER ASolGNED I
Tfl~ 990 made/l 9/mW
" ' ' * ~
4hlc7 4/&/'q7 y
TYPE OF ACTION (Indicate the number of each type)
/
/
l l COMMERCIAL DISTRIBUTION (FORMAL) l USE BY A SINGLE APPLICANT (CUSTOM) j.
SOURCE (9C)
DEVICE (BA)
SOURCE (9D)
DEVICE (99)
NEW NEW NEW
~
AMiiNDMENT NEW l-AMENDMENT dMENDMENT AMENDMENT
/
O V
^IN I
UCENSING ACTION REQUIRED IF KNOWF, EQ ED
) OTHER (Specoy)
TOTAL NUMBER OF NOTES g'
REVIEW HOURS gh NUMBER OF DEFICIENCY LETTERS NUMBER OF DEFICIENCY CALLS FOR BILLING PURPOSES ONLY l
j NEW REGISTRATION -
_j PRODUCT INACTIVE -
NAME CHANGE ADDRESS CHANGE ADD TO BILLING REMOVE FROM BILLING FOR FEE USE ONLY TYPE OF FEE FEE CATEGORY.
[
]M MM E
E AMOUNT RECENED CHECK NUMBER MATANN UPDATED Q
hap
~ AS REQUIRED L
DATE OF CHECK (f 7 LOG I MPOj TSYS PDATED R_,,
u~
n Nq
.,E RE,u.
-,E
/b I
?
COMMENTS i
i NRC FORM 587 (S.e3)
NE
~ r.
,ew..
. --=,
I a
x.;
' fi lf Federal Emergency' Management Agency
.. ~
Washington, D.C. 20472 i
March 17,~1997 Mr. Steven'Baggett-o U.S. Nuclear Regulatory Commission-
. Sealed Source Safety Section
~ Source Container and.Davices' Branch Division of-Industrial and Medical l Nuclear Safety, NMSS Washington, D.C. 20555-0001
Reference:
'NRC License No. 08-01297 '
Docket No. 030-7130 r
Dear Mr. Baggett,
The Director of.the Federal Emergency Management Agency. (FEMA) made
- l the decision'to close the National Instrumentation Center (NIC) at the Mt. Weather Emergency Assistance Center (MWEAC),.Bluemont, Virginia, effective August 15, 1996.
in Radioactive Sources in the possession of FEMA at the NIC are temporary storage, awaiting either the transfer to - a properly.
- licensed recipient or disposed of as radioactive waste.
The licensed recipient who possess radioactive. sources owned by FEMA will'either retain' possession of the sources, or dispose of them as radioactive waste.
Due.to'the closure of the NIC, it is requested that the following
-registry numbers.in our License be given inactive status:
NR-2 83-D-101-S NR-283-S-102-S NR-283-D-105-S NR-283-D-106-S NR-283-D-110-S NR-283-D-111-S
-The following registry numbers are inactive:
NR-283-D-801-S NR-283-D-802-S NR-283-D-803-S It is anticipated.that the annual fees for the devices covered by I
each registry number will be frozen for FY 1998.
We are' currently in'.the process of reconciling and updating our
. computerized inventory Radioactive Materials (RAM) Data-Base to and what radioactive
. reflect..what we currently 1 own and possess material and devices have been transferred to. licensed recipients or disposed of as radioactive waste.
When completed, we.will provide you with'an. updated' inventory.
~
I4 o
\\
1 I would be happy to discuss any questions or concerns that you or your staff have regarding this letter.
I can be reached at (202) 646-2872.
Sincerely, S
u.
Vernon L.
Wingert Chairperson, Radiation Control Committee (RCC) cc: